
1157 - Wireless Spirometer Sensor
4
/
23
| Revision: 0 | DS155
Disclaimer
The Smart Wireless Spirometer is sold for the purpose of teaching and educational instruction. It is not
and has not been designed to be a diagnostic tool. The results are illustrative only; any information
gained from the use of this apparatus must not be used as the basis for any medical or clinical
diagnosis.
Read the instructions before use. Students should be supervised at all times.
Follow your local safety guidelines. It is the responsibility of the tutor to ensure
that individuals using this apparatus are healthy and have no history of
respiratory or cardiovascular problems. The students must be reassured about
the medical validity of the results. Respiratory tests can be challenging; limit the
number of tests in a session and take care that an individual is not overstressed
by the tests.
WE DO NOT RECOMMEND THE USE OF THIS SENSOR WITH INDIVIDUALS UNDER THE AGE
OF 10.
Overview
The Smart Wireless Spirometer is USB and Bluetooth compatible. Using Bluetooth, a sensor can
connect to mobile devices, tablets, laptops, and desktops.
The sensor has several ranges (flow (default), volume, adjusted volume, cyclic volume, breathing rate,
differential pressure). Any range can be selected on its own or selected alongside any other range.
The range required is set in the Devices icon of the EasySense2 software.
The Spirometer is used to measure changes in the volume of the inhaled and exhaled air. The total
volume of the lung can be calculated from various breathing processes and calculation of the
anatomical dead space.
The spirometer integrates the flow to produce volume (i.e. a flow of 1 l/s must produce 1 litre of volume
in one second and 2 litres of volume in 2 seconds etc.). Any value of the flow that differs from zero at
the start will become “integrated” into the volume and will be seen in the final data as a slope in the
data. A calibration function and Tare function are present to allow the no flow through of the Spirometer
become the zero base line to reduce this effect.
Additionally, the adjusted volume range uses mathematics to bring the data back to effective zero after
each breathing cycle.