TAKING A READING
Connect the probe to the external connector on the top of the meter.
Insert the probe into the bale.
Press the read button (#1).
Within 1 or 2 seconds, the meter will display the %MC between 8.0 and 37.0. The meter
will automatically correct the reading for the sample temperature. The reading will be
displayed for 2 seconds and if the “READ” key is still pressed, then the cycle will be
repeated. The buzzer will sound if the displayed reading is above the set point value
entered in the meter (default is 20.0).
The meter will “store” the reading, provided that the user releases the “READ” key,
otherwise the meter will initiate another “READ” cycle without “storing” the previous
reading. This feature lets the user take a number of readings in sequence and only
“keep” the readings of his choice.
When the “READ” key is released, the meter will add the last displayed value to the sum
of the previous readings (up to 100), provided the value is within the range. When all
the 100 readings are taken, the meter will automatically display their average before
turning off. No new value will be added to the sum unless the user erases the readings.
The reading below 8.0% will be displayed as “.0” and above 37.0% as “99.9”. These are
not considered valid readings and will not be stored.
NOTES
The prod is electrically insulated, except at the metal points near the tip. The moisture
content measured represents the tobacco in contact with the tip of the prod only.
Partially cured tobacco may have wide variations in moisture content throughout the
bale. Readings should be taken in several different parts of the bale and the highest
readings used as a guideline. The arrangement and compaction of tobacco fibers in a
bale may have an effect on meter readings.
If you are testing high density bales, we recommend using the H-4 handle with the
830-2 25cm prod, 830-3 45cm prod, or the 830-4 90cm prod. Using the handle/prod
combination eliminates excess stress on the instrument case that may occur when trying
to insert the prod into a high density or large bale.
When using the 90cm prod, be sure to guide the prod into the bale with one hand while
pushing on the H-4 handle.
CHECKING THE AVERAGE AND MAXIMUM VALUE OF THE ACCUMLATED READINGS
(And erasing them)
As mentioned before, the meter adds the value of the new reading (provided that the
“READ” key is released) to the sum of the previous readings. The meter holds this sum, as
well as the number of readings added so far and the HIGHEST reading. Up to 100 readings
can be accumulated. After that, the meter will stop adding the new reading and display the
average every time a new reading is taken to remind the user to erase the readings to start
accumulating the new ones.