
GENERAL SM6000
Page 3 - 2DELTA ELEKTRONIKA B.V. rev. June 2018
fig. 3- 4
Programming inputs (internal circuit)
9) ANALOG PROGRAMMING
Theoutputvoltageandcurrentcanbeprogrammedbyanexternalanalog
voltage. This programming is very accurate and linear. The levels are all
standardized on 5 V.
Standard theSM300-20andSM600-10areequipped with the ISO-AMP,
see next paragraph.
Fortheotherunits,theinputshave a protection circuit formed by aseries
resistor andaparallelzener (seefig. 3-4).Thecapacitorlimits thespeed
to a safe value. Note that the analog inputs (and outputs) are not float-
ing, but the common is connected to the negative output terminal.
Wrong connection of Øcan causeearthloopswhich can trip thefuse.Af-
ter removing the fault, the fuse will reset (PTC-fuse). For isolated pro-
gramming see next paragraph.
10) ISOLATED ANALOG PROGRAMMING
To prevent earth loops which can cause programming errors, use an iso-
lated programming source. If this is not possible, use the optional ISO
AMP CARD (δ-product) which can be built inside the unit.
With the ISO AMP CARD earth loops between the unit and the program-
ming source are prevented.
11) ETH / IEEE488 / RS232 / CAN / PROFI PROGRAMMING
The Delta Elektronika PSC-ETH, PSC-488, PSC-232, PSC CAN and
PSC PROFI controllers can be factory installed inside the unit.
Voltage and current can easily be programmed and read back. Also all
the status outputs can be read by the computer.
12) MONITORING OUTPUTS
The monitor outputs give a voltage 0 - 5 V proportional to the output cur-
rent or voltage. The output current can easily be measured using the
I-monitor (see fig. 3- 6). The monitor outputs are buffered with op-amp’s
andprotectedwithseriesresistorsandparallelzeners(seefig. 3-7).The
tableinfig.3-5 showstheimpedancelevelsofthemonitoringoutputs.
For using Imon on a pulsating load, see paragraph 20) of this chapter.
13) +12 V ON PROGRAMMING CONNECTOR
The +12 V onthe programming connector can beused to supply external
circuits. The output is current limited, but should not be overloaded. The
fuseF27_3onP598couldblow.ThefuseF27_3alsoprotectstheinternal
circuit, in case an external high voltage is applied by accident. Note: this
fuse is a special 600 V type, always replace with the same type.
14) STATUS OUTPUTS
All the status outputs arelogic outputs. Logic "0" means the output is0 V,
logic "1" means the output is 5 V (Ro = 500 Ohm). This makes it possible
to drive directly: an opto-coupler, a TTL gate or a CMOS gate.
The Limit Status or LIM-statusis"1"incasetheoutput voltageorcurrent
reaches the limit setting. Which limit circuit is active can be seen on the
front panel LED’s.
The Over Temperature Status or OT status is "1" in case of an over tem-
perature, the OT LED will be on and the output shuts down. As a
pre-warningtheOT LEDstartstoblinkwhen theunitrunshot butthesitu-
ation of over temperature is not reached yet. The status will still be low
when the LED is blinking.
fig. 3- 5
Connections ANALOG PROG. CONNECTOR
pin Description, see par.12)...17) for details
1Ø, return of reference, prog. inputs and
monitor outputs (Ro = 1.2 Ohm).
2current monitor output 0 - 5 V
(Ro = 1.2 Ohm, Io max = 4 mA)
3current programming input (0 - 5 V),
Ri = 8 MOhm
4CC status output, logic 1 = CC mode
(5 V / 500 Ohm)
5Remote ShutDown (4 - 12 V),
Ri = 5 kOhm
6PSOL status output, logic 1 = PSOL
(5 V / 500 Ohm)
7+12 V output
(Ro = 3 Ohm, Io max = 0.2 A)
8Ø, return of status outputs, +12 V
and Remote ShutDown
9reference voltage 5.1 V
(Ro = 1.2 Ohm, Io max = 4 mA)
10 voltage monitor output 0 - 5 V
Ro = 1.2 Ohm, Io max = 4 mA)
11 voltage programming input (0 - 5 V)
Ri = 8 MOhm
12 OT - status output, logic 1 = OT
(5 V / 500 Ohm)
13 LIM - status output, logic 1 = LIM
(5 V / 500 Ohm)
14 DCF - status output, logic 1 = DCF
(5 V / 500 Ohm)
15 ACF - status output, logic 1 = ACF
(5 V / 500 Ohm)
V
A
Imon
0-5V
Vmon
0-5V
Return of
ref, prog, mon.
fig. 3- 6
External meters
using monitor outputs
fig. 3- 7
Buffered monitor outputs