HD2124 - 15 - V2.5
THE PROBES
The instrument measures absolute, relative and differential pressure using the PP471 SICRAM
module and the TP704 and TP705 series probes. It also measures the temperature with a Pt100
sensor fitted with SICRAM module or 4-wire Pt100 probes.
The SICRAM module acts as an interface between the sensor on the probe and the instrument.
There is a microprocessor circuit with a permanent memory inside the module that enables the
datalogger to recognize the type of probe connected and to read its calibration information.
The probes are detected during turn on, and this cannot be performed when the instrument is
already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it
off and on.
The manometers are fitted with two inputs dedicated to the probes: it is possible to connect one or
two pressure probes, one or two temperature probes or one pressure probe and one temperature
probe. See the details on page 8.
The probes are fitted with SICRAM modules are calibrated in the factory; no calibration is required
by the user.
PP471 MODULE FOR PRESSURE MEASUREMENT
The PP471 module acts as an interface between the TP704 and TP705 series Delta Ohm probes and
the instrument. The TP704 series pressure probes have a 1/4” BSP threaded male connection and
must be screwed in the joint where you wish to measure the pressure. The TP 705 series have two
Ø5 connections where the suitable tubes are inserted in order to perform the desired measurement.
Please pay careful attention to the joint pressure sealing; use suitable gaskets and joints. A plastic
cap protects the threaded connection. Put it back after use as it protects the pressure cell from
foreign bodies. VERY IMPORTANT: ensure the probe's full scale is higher that the pressure
that will be measured. In case this value is unknown, start by using higher-capacity probes.
For every pressure probe a range of overload pressure and a burst pressure are declared:
pressures in the range of overload pressure don’t cause the break of the sensor but the
declared accuracy can be overcome. Pressure over the stated burst pressure may break the
sensor. The application of higher pressures than the overload pressure limit, although lower
than the burst pressure, may produce permanent damage to the probe (e.g. offset shift). Never
exceed the stated burst pressure.
Please see the technical characteristics of the probes on page 31.
Upon turning on the instrument automatically detects the PP471 module. The probe's type
(absolute, relative or differential) and full scale value are detected even when the instrument is on.
If no logging or recording operations are in progress, it is possible to change the probe connected to
the module without turning the instrument off and on again.
The measurement provided can be the instantaneous or the peak value of the pressure: please see
the PEAK key description on page 13.
To change the instantaneous or peak value unit of measurement, press UNIT. The following units of
measurement are available:
Pa, hPa, kPa, mbar, bar, atm, mmHg, mmH2O, kgf/cm2, PSI, inchHg, inchH2O.
Some following units of measurement require a multiplication factor: the “-3” symbol indicates the
displayed value must be divided by 1,000; the "3" and "6" symbols indicate the displayed value
must be multiplied respectively by 1,000 or by 1,000,000.