7
2 digit
10s
t
Ο
2
C
1
3) SAL mg/l (Setting the salinity): the message "SAL mg/l – UNT=RSET" is scrolled in the
comment line. The display central line shows the current salinity value, and the secondary
line the corresponding dissolved Oxygen concentration value. Use the arrows to enter this
value, and confirm using ENTER. To disable the salinity compensation, set the value to
zero: pressing the °C/°F key rapidly sets the value to zero without the need to use the ar-
rows.
Note: the dissolved Oxygen concentration depends on the measured liquid salinity. In con-
trast, salinity has no effect on the saturation index.
4) Management of memorized data: the message "LOG_DUMP_or_ERAS" (Transfer data
or erase) is scrolled in the comment line. The center figure reports the number of free mem-
ory pages (FREE). All memory data are permanently erased by pressing SE-
RIAL/EraseLOG. By pressing LOG/DumpLOG, the data transfer of the logged data on the
serial port is started: the "BAUD-RATE" must have previously been set to the maximum
value (please see the menu items described below and the paragraph "STORING AND
TRANSFERRING DATA TO A PERSONAL COMPUTER" on page 25).
5) Sleep_Mode_LOG (Automatic turning off during recording): this function controls the
instrument's automatic turning off during logging, occurring between the capture of a sample
and the next one. When the interval is lower than 60 seconds, the instrument will always
remain on. With intervals greater than or equal to 60 seconds, it is possible to turn off the in-
strument between loggings: it will turn on at the moment of sampling and will turn off im-
mediately afterwards, thus increasing the battery life. Using the arrows select YES and con-
firm using ENTER in order to enable the automatic turning off, select NO and confirm to
disable it and keep the instrument on continuously.
Note: even if Sleep_Mode_LOG=YES is selected, the instrument does not turn off for less
than one minute intervals.
6) Print and log interval: sets the interval in seconds between two loggings or data transfers
to the serial port. The interval can be set at 0s, 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min),
300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1
hour). If the value 0 is set, SERIAL works on command: the sending of data to the se-
rial port is performed each time the key is pressed. Recording (LOG) is performed with
one second intervals even if the interval is set to 0. With an interval from 1 to 3600s, con-
tinuous data transfer is started when the SERIAL key is pressed. To end the recording
(LOG) and continuous data transfer operations (SERIAL with an interval greater than 0),
press the same key again.
7) RCD MODE (Record mode): the instrument captures the dissolved Oxygen concentration,
saturation index, barometric pressure and temperature values every second. The relationship
existing between these parameters is such that the maximum or minimum value of one gen-
erally does not correspond to the maximum or minimum value of the others. If the RCD
MODE parameter is set to "I Sat", the maximum (MAX) and minimum (MIN) values dis-