The small traffic lights graph, at the top of the computer screen during the signal capture, is updated each
cardiac cycle. If it doesn’t occur, it means:
•The R wave of the ECG is not detected (in the case of Ecg + Tonometry).
•The foot of Tonometer - sensor1 curve is not detected (in the case of 2 x Tonometry, ETT model only).
The ECG amplitude is too low
Re-position one or both the ECG electrodes in order to
obtain signals with higher amplitude.
The R wave is reversed (downwards)
Exchange the Ecg cables or electrodes.
The R wave has pathological shape
Proceed with the capture for at least 10 seconds, save
the signals and manually add markers on the R waves.
Tonometer1 (red curve) is off and/or no signal is
applied
Switch on the Tonometer1 and position it on the artery
under investigation (Carotid).
Captured signal’s amplitude on Tonometer1 is too low
Re-position the Tonometer1 in order to obtain signals
with higher amplitude.