Doran 1200 Series User manual

1200 Series
Digital Indicator
Technical Manual
MAN350 –Rev 2 Doran Scales, Inc. www.doranscales.com

Table of Contents
Introduction.......................................................................................................................... 1
Unpacking Your Scale.....................................................................................................................1
Specifications.................................................................................................................................2
Scale Function................................................................................................................................3
Battery Operation ................................................................................................................. 7
Power Off......................................................................................................................................7
Low Battery ...................................................................................................................................7
Recharging Battery.........................................................................................................................7
Load Cell Connection ............................................................................................................. 8
Load Cell Connector Pinout ............................................................................................................8
Scale Setup Guide.................................................................................................................. 9
Entering Setup Mode .....................................................................................................................9
Setup Mode Navigation..................................................................................................................9
Data Communications ......................................................................................................... 17
Troubleshooting.................................................................................................................. 18

1
Introduction
Thank you for purchasing a Doran Scales product. Please read this manual to ensure
obtaining all the benefits that this indicator can provide. If any questions arise, please
contact the Doran Scales Technical Support Department at 630-879-1288.
Unpacking Your Scale
Before unpacking your Doran scale, please read the instructions in this section. Your
new scale is a durable industrial product, but it is also a sensitive weighing instrument.
Normal care should be taken when handling and using this product. Improper handling
or abuse can damage the scale and result in costly repairs that will not be covered by
the warranty. If you notice any shipping damage, notify the shipper immediately. Please
observe the following precautions to insure years of trouble free service from your new
scale.
•DO NOT drop the indicator
•DO NOT immerse the indicator
•DO NOT drop objects on the platform
•DO NOT pick up the scale by the top of the weighing platform
•Carefully remove the scale from the shipping carton

2
Specifications
NTEP Certificate
Class III –10,000d; Cert. 19-013
Enclosure
ABS Plastic
Product Dimensions
9” W x 7.5” H x 2.75” D (includes u-bracket)
Environmental Protection
IP5X
Legal for Trade
Temperature Range
14 F to 104F (-10 C to +40 C)
Capacity Range
1 to 999,000 lb
Resolution Range
1,000 to 30,000d
Internal Counts
1,000,000
Analog Signal Range
0 mV/V to 10 mV/V
Excitation Voltage
5 VDC
Number of Load Cells
Up to 4 350 Ohm with 4 wire input
Scale Inputs
One
Calibration Range
Calibrate between 10% and 100% of capacity
Power Input
100 –240VAC 50/60Hz
Internal Rechargeable
Battery
35 hours of continuous use with four load cells
7.4 V, 4 AH, 1000 recharge cycles
Display
1.6" high, 6 digit Backlit LCD
Displayed Units
lb, kg
Serial Interface
Bi-directional RS-232 port standard
Controls
Total, Count, Units, Gross, Tare, Zero, Print (On/Off)

3
Fig. 1: Front Panel
Scale Function
~
Motion indicator. This symbol represents motion or instability of the
weight. The annunciator will illuminate when motion is sensed on the
platform. Changes in weight, vibration or air currents can cause the scale
to go into motion.
Center of zero indicator. The annunciator will illuminate while the scale is
displaying a zero weight.
NET
Indicates a net weight is displayed. Net weight is displayed when a tare
value is present.
lb / kg
Indicates current unit of measure
TOTAL
Lit when accumulator function is active. F1.10 parameter controls the
current function. Checkweighing is the default function.

4
Under, Accept and Over in checkweigh mode. F1.10 parameter controls
the current function. Checkweighing is the default function.
Indicates the count function is active
Indicates the accumulation function is active
Battery indicator
Power Up
Connect the cord to a compatible power source. Briefly press PRINT (On/Off) until the
scale beeps.
Basic Weighing Operation
1) Remove all items from the scale platform
2) Press the ZERO button to zero the scale
3) The weight display now reads zero
4) Place an item on the scale platform and wait for the motion annunciator to turn
off, indicating an accurate, stable weight
ZERO
ZERO is used to zero the scale display. To zero the scale, wait until the scale is stable
and press the ZERO button. The scale will not zero if the scale is in motion. The zero
function will operate over the entire capacity of the scale. ZERO will not function in the
net mode.
UNITS
UNITS selects the current unit of measure. Press UNITS to change the current unit. Lb
and kg are available.
PRINT
PRINT transmits weight data to a printer or other external devices.
GROSS
When in net weighing mode, press to return to gross mode and clear the tare value.
TARE
Press to tare the current displayed weight. The indicator enters net mode. Note that
tare only functions with positive weights. When in net mode, the ZERO function is not
functional.

5
TOTAL (Accumulator)
The TOTAL annunciator is lit when the accumulation function is active. The accumulator
will track the number of weights and total amount of weight on the scale. Do not change
units during an accumulation process.
Accumulate
1. Place item on the scale
2. Press TOTAL to accumulate current weight
3. Add--- is displayed while saving to memory
4. Remove item from scale
5. Weight must return to zero before the next weight can be added to the
accumulated total.
Display Accumulator
1. Press and hold TOTAL until total is displayed
2. Cn is displayed showing the current count
3. Press GROSS to display the total accumulated weight
4. Press PRINT to exit to weighing
Clear Accumulator
1. Press and hold TOTAL until total is displayed
2. Cn is displayed showing the current count
3. Press GROSS to display the total accumulated weight
4. Press ZERO to clear the accumulator
5. Press PRINT to exit to weighing
COUNT (Piece Counting Function)
Press COUNT to switch between weight and count display. When displaying piece
count, a nis displayed in the leftmost digit. Parameter F2.4 controls the method of
piece count entry. F2.4 = 0 will use platform weight to calculate piece weight. F2.4 = 1
will use manual piece weight entry.
Set piece weight with platform weight (F2.4 = 0)
1. Place the container on the scale
2. Press ZERO
3. Place the piece sample into the container
4. Press and hold COUNT until SANPLE is displayed
5. Press PRINT
6. Enter the number of pieces on the scale
7. Press PRINT
8. Piece count is displayed

6
Set piece weight with manual weight entry (F2.4 = 1)
1. Place the container on the scale
2. Press ZERO
3. Place the piece sample into the container
4. Press and hold COUNT until SANPLE is displayed
5. Press PRINT
6. Enter the number of pieces on the scale (UNITS and GROSS select digit, TARE
and ZERO increment and decrement digit value)
7. Press PRINT
8. Input the total weight of the sample
9. Press PRINT
Checkwieghing Function
For checkweighing set the F2.1 to the lowest weight where the checkweigh
annunciators will be active.
Set F2.2 to the Under limit
Set F2.3 to the Over limit
Any weight between the Under and Over limit will activate the OK annunciator
Operation Error Codes
【 】
Invalid operation
【 】
Overload condition: The weight on the platform is greater than scale
capacity.
【 】
Underload condition: The weight is under the calibration zero.
【 】
Zero is above the allowed range (F1.8)
【 】
Zero is below the allowed range (F1.8)

7
Battery Operation
The indicator will run continuously for up to 38 hours with four 350 ohm load cells. To
maximize battery life, set the Automatic Shutoff Timer (F3.5) which will automatically
powers down the scale after a period of non-use.
Power Off
1) Manual - Press and hold the PRINT push button until the display turns off.
2) Automatic - At the end of the Automatic Shutoff Timer (F3.5) scale parameter setting.
Low Battery
The battery annunciator indicates that the battery is in need of recharging. Once it
indicates no remaining bars, there will be approximately one hour of battery life
remaining before the scale turns off.
Recharging Battery
To charge the battery, plug the wall mount transformer into a wall outlet. While the scale
is charging, light on the wall mount transformer is red. When completed charging, the
wall mount transformer light is green. The battery will fully charge in approximately eight
hours. The scale can be used while recharging the battery.
Leaving the scale plugged in will ensure a fully charged battery and will not affect the life
of the battery. The battery supports up to 1000 recharges before requiring replacement.

8
Load Cell Connection
Load Cell Connector Pinout
Description
PIN
+SIG
+ Signal
2
- SIG
- Signal
3
+EX
+ Excitation
1
- EX
- Excitation
4
GND
Ground
5

9
Scale Setup Guide
Entering Setup Mode
The calibration push button is located behind a door on the rear of the indicator.
Remove the screw and press and hold the button down until the scale beeps. Setup
will be displayed. All menus are accessible.
Press and hold GROSS to access menus F2-5. Calibration menu F1 is not accessible.
Setup Mode Navigation

10
Set Scale Capacity
To set the scale capacity, select kg or lb before entering setup to select the primary unit.
Once this is set, a calibration in F1.4 and F1.5 must be performed to set the primary unit
and capacity.
1. Display reads Setup
2. Press PRINT
3. F1 is displayed
4. Press PRINT
5. F1.1 is displayed
6. Press PRINT
7. Press ZERO and TARE to change the value of the flashing digit
8. Press UNITS and GROSS to select the current digit
9. Press PRINT once the correct capacity is displayed
10.Press COUNT twice to exit the setup mode if desired
Set Number of Decimal Places
1. Display reads Setup
2. Press PRINT
3. F1 is displayed
4. Press PRINT
5. F1.1 is displayed
6. Press ZERO
7. F1.2 is displayed
8. Press PRINT
9. The number of decimal places is displayed
10.Press ZERO to select the correct value
11.Press PRINT once the correct capacity is displayed
12.Press COUNT twice to exit the setup mode if desired
Set Scale Resolution
1. Display reads Setup
2. Press PRINT
3. F1 is displayed
4. Press PRINT
5. F1.1 is displayed
6. Press ZERO
7. F1.2 is displayed
8. Press ZERO
9. F1.3 is displayed
10.Press PRINT
11.The count by value is displayed
12.Press ZERO to select the correct value
13.Press PRINT once the correct capacity is displayed
14.Press COUNT twice to exit the setup mode if desired

11
Calibration Error Codes
【 】
The entered span calibration weigh is too light, the calibration input
weight should be ≥10% of scale capacity.
【 】
The span calibration weight is too light, the loading weight should be
≥10% of scale capacity
【 】
The scale is in motion and cannot calibrate
Parameter Groups
The scale parameters are divided up into five parameter groups. Each group contains
related parameters.
F1 Capacity and Calibration
F2 Piece Count Configuration
F3 Date / Time and Shutoff Timers
F4 Data Output
F5 Indicator Test
F1.0
Base Capacity and Calibration Unit
0, 1
0: kg
1: lb
F1.1
Capacity Adjustment Menu
1 - 999000
3 lb / kg to 200,000 lb / kg
Press GROSS and UNITS to select flashing digit
Press TARE and ZERO to change digit value
F1.2
Resolution Decimal Places
0-4
Select number of decimal places for scale resolution

12
F1.3
Resolution Selection
1, 2, 5
10, 20, 50
Select number of decimal places for scale resolution
10, 20 and 50 are selectable when F1.2 is set to 0
F1.4
Zero Calibration
E-SCL
Empty scale platform. Press PRINT. The display will
count down from 10 to 0 and display End when
complete.
F1.5
Span Calibration
LOad
Press PRINT to enter the span calibration weight. The
span calibration weight must be at least 10% of the
total scale capacity. Place weights on scale and press
PRINT to calibrate. The display will count down from
10 to 0 and display End when complete.
Press GROSS and UNITS to select flashing digit
Press TARE and ZERO to change digit value
F1.6
Automatic Zero
0.5,1,3
Number of divisions the scale will automatically zero to
maintain zero. This feature is disabled when in net
mode.
F1.7
Power On Zero
OFF,2,10
Percentage of capacity that can be automatically
zeroed when powering on. Off (disabled), 2% of
capacity and 10% of capacity are selectable.
F1.8
Zero Button Range
OFF,1.9,100
Amount of capacity that can be zeroed. Off (disabled),
1.9% and 100% of capacity are selectable. 100% is
the default

13
F1.9
Digital Filter
0-9
0 is slowest 9 is fastest setting. 5 is the default
setting.
F1.10
Scale Function
0,1,2
0: activates checkweighing function
1: activates accumulation function
2: activates counting function (See F2.4)
F1.11
Factory Default
0,1
Set to 1 and press PRINT to default scale parameters.
Scale calibration data is retained.
F2.1
Minimum weight for checkweigh indicators to
activate
0to capacity
F2.2
Checkweighing Lower Limit
0to capacity
F2.3
Checkweighing Upper Limit
0to capacity

14
F2.4
Sampling Method for Counting Function
0,1
0: piece weight sample from platform
1: manual piece weight entry
F3.1
Date Format
0,1, 2
0: year.month.day
1: month.day.year
2: day.month.year
F3.2
Date Entry
00.00.00
Enter the date per the format selected in F3.1
Press GROSS and UNITS to select flashing digit
Press TARE and ZERO to change digit value
F3.3
Time Entry
00.00.00
Enter in 24 hour format, hour:minute:second.
Press GROSS and UNITS to select flashing digit
Press TARE and ZERO to change digit value
F3.4
Backlight Auto Shutoff Timer
0-999
Defaults to 0 which disables this feature
When set to 1-999 sec the display enters sleep mode
Press GROSS and UNITS to select flashing digit
Press TARE and ZERO to change digit value

15
F3.5
Automatic Shutoff Timer
0-60
Defaults to 0 which disables this feature
When set to 1-60 min the indicator shuts off
Press GROSS and UNITS to select flashing digit
Press TARE and ZERO to change digit value
F4.1
Data Output Format
0,1, 2
0: No data is transmitted
1: When the scale is stable as indicated by the motion
annunciator, a data string is transmitted
2: Data output when PRINT is pressed (see F4.4)
See Data Communications section for formats
F4.2
Data Bits and Parity Settings
8 n 1 8 databits, no parity and 1 stop bit
7 e 1 7 databits, even parity and 1 stop bit
7 0 1 7 databits, odd parity and 1 stop bit
8 E 1 8 databits, even parity and 1 stop bit
8 0 1 8 databits, odd parity and 1 stop bit
F4.3
Baud Rate
2400 4800
9600 19200

16
F4.4
Additional Line Feed (F4.1 set to 2)
0-9
Additional line feeds at the end of the data output
F5.1
Button Test
PRESS
Press each button to verify functionality. Press
TOTAL to exit.
F5.2
Display Test
Press PRINT to exit
F5.3
Raw Counts
Displays A/D raw output

17
Data Communications
DB9 Male Bulkhead Connector
Pin
Function
2
TXD
3
RXD
5
GND
Print Formats
F4.1 set to 1 –Continuous Output. Only weight is transmitted in this mode.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
S
S
,
G
G
,
SP
W
W
W
W
W
W
W
U
U
CR
LF
S
T
,
G
S
,
5
0
0
.
0
l
b
SS
•OL (overload or underload)
•ST (stable)
•US (unstable)
GG
•GS (gross weight)
•NT (net weight)
SP –Space
WWWWWWW –6 weight digits with decimal point. Spaces before weight if not 6 digits.
UU –Current unit lb or kg
CR –Carriage Return
LF –Line Feed
F4.1 set to 1 –Print on demand. Press PRINT to send weight data. The scale must
return to zero before printing again. Parameter F4.4 adds line feeds to the end of the
data string for ease of use with tape printers.
Weight Count
LIST
Date
MMDDYY
Time
HH:MM:SS
---------------------------------
Gross
XXXXXXX lb
Tare
XXXXXXX lb
Net
XXXXXXX lb
LIST
Date
MMDDYY
Time
HH:MM:SS
---------------------------------
Count
XXXXXX PCS

18
Troubleshooting
If any problem persists, contact Doran Tech Support at 630-879-1288.
Problem
What to Do or Check
Weight reading will not
repeat or does not return to
zero when weight is
removed
Examine the weighing platform for any interferences. Be
sure that nothing is inside the platform, under the load cell
or the weigh bridge structure
Scale overloads before
reaching full capacity
Make sure all four corner overload stops are properly set, if
present. Take the platter off the scale, invert it and place it
on the platform. With 1/2 of the scale's capacity in test
weights concentrated over a corner of the platform, there
should be approximately 1/32" of clearance between the
stop and the bottom of the spider. Check all four corners
then recalibrate the scale.
Scale will not indicate full
capacity or go into overload
Make sure that there is nothing caught in the scale under
or around the load cell or spider, which would interfere with
their movement. If not, check the overload stops using the
above procedure.
Scale will not zero when the
ZERO button is pressed
Make sure that the scale is stable ( annunciator is
off) when ZERO is pressed.
Weight readings don't seem
to be correct
Check the scale's accuracy with a test weight.
Recalibrate if necessary.
Scale drifts off of zero
Check for air currents and/or vibration around the scale.
If that is the cause, it may be necessary to set the F1.6
parameter to a wider setting to compensate
Scale reading is bouncing
Check for air currents and/or vibration around the scale.
If that is the cause, it may be necessary to change the
Display Filter parameter.
Error Message
What to Do or Check
【 】
【 】
Unable to start up. Perform zero calibration (F1.4).
【 】
Eeprom Checksum error. Perform a factory default
(F1.11) and recalibrate scale. If this occurs again, return
to factory for repair.
【 】
The scale is in continuous motion. Check platform for
interference. The load cell may need to be replaced.
【 】
Date / Time data format entry error. Verify correct format
(F3.1) and enter again.
【 】
A/D initialization error. Return to factory for repair.
Table of contents