Doran 8200IS Series User manual

8200IS Series
Intrinsically Safe Indicator
Technical Manual
MAN289 – Rev 7 (2.3) Doran Scales, Inc. www.doranscales.com

Table of Contents
Introduction ..................................................................................................................... 1
Specifications ................................................................................................................. 2
Scale Operation .............................................................................................................. 3
Fig. 1: Model 8200IS Front Panel Layout ............................................................................................ 3
Scale Annunciators ................................................................................................................... 3
Powering On and Off ................................................................................................................ 4
Basic Weighing Operation ........................................................................................................ 4
ZERO ........................................................................................................................................ 4
TARE ........................................................................................................................................ 4
Keyboard TARE entry ............................................................................................................... 4
Digital Tare Entry ...................................................................................................................... 4
Display TARE value .................................................................................................................. 5
Clear TARE value ..................................................................................................................... 5
GROSS NET ............................................................................................................................. 5
UNITS ....................................................................................................................................... 5
PRINT ....................................................................................................................................... 5
Password Protected Setpoint, Preact, Tare and Check Limit Values ....................................... 5
Setpoint and Output Operation for Standard Configuration ...................................... 6
Enter SETPOINT Target Weight ............................................................................................... 6
Display SETPOINT Target Weight ............................................................................................ 6
Setpoint Learning Preacts ......................................................................................................... 6
Enter Setpoint Preact Weight .................................................................................................... 7
Display Setpoint Preact Weight ................................................................................................ 7
Tank Level Maintenance Mode Standard Configuration ............................................. 8
Product ID Standard Configuration .............................................................................. 9
Recall PRODUCT ID from Memory .......................................................................................... 9
Display Current PRODUCT ID .................................................................................................. 9
Create New PRODUCT ID ........................................................................................................ 9
Delete PRODUCT ID from Memory ........................................................................................ 10
Product Fields Standard Operation ............................................................................ 10
Display Product Fields ............................................................................................................ 10
Product Field Entry ................................................................................................................. 10
Accumulator and Counter Standard Operation ......................................................... 11
Accumulator and Counter Operation ...................................................................................... 11
Display Accumulator and Counter Values .............................................................................. 11
Clear Accumulator and Counter .............................................................................................. 11
Accumulator and Counter Data String Output to Printer or Data Collection ........................... 11
Checkweighing Operation ........................................................................................... 12
Three Band Checkweighing ........................................................................................ 12
Three Band Checkweighing .................................................................................................... 12
Enter and Display of Checkweigh Limits ................................................................................. 12
Weight Reference and Digital Entry of Checkweigh Limits ..................................................... 12
Weight Reference Entry of Checkweigh Limits ....................................................................... 12
Five Band Checkweighing ........................................................................................... 13
Five Band Checkweighing ...................................................................................................... 13
Enter and Display of High and Low Limits .............................................................................. 13
Weight Reference and Digital Entry of High and Low Limits .................................................. 13

Weight Reference Entry of High and Low Limits .................................................................... 13
Zero Band Checkweighing .......................................................................................... 14
Zero Band Checkweighing ...................................................................................................... 14
Product ID Checkweigh Operation ............................................................................. 15
Recall PRODUCT ID from Memory ........................................................................................ 15
Display Current PRODUCT ID ................................................................................................ 15
Create New PRODUCT ID ...................................................................................................... 15
Delete PRODUCT ID from Memory ........................................................................................ 16
User ID Checkweigh Operation ................................................................................... 16
User ID login ........................................................................................................................... 16
User ID Logout ........................................................................................................................ 16
Product Fields Checkweigh Operation ....................................................................... 17
Display Product Fields ............................................................................................................ 17
Product Field Entry ................................................................................................................. 17
QC Weigh Operation .................................................................................................... 18
QC Weigh Mode ..................................................................................................................... 18
Entering a User ID .................................................................................................................. 18
Entering a Product ID .............................................................................................................. 18
Checkweigh Operation ............................................................................................................ 19
ionSuite Syncing of Product IDs and User IDs ....................................................................... 19
Accumulator and Counter Checkweigh Operation ................................................... 20
Accumulator and Counter Operation ...................................................................................... 20
Display Accumulator and Counter Values .............................................................................. 20
Clear Accumulator and Counter .............................................................................................. 20
Accumulator and Counter Data String Output to Printer or Data Collection ........................... 20
Calibration Guide .......................................................................................................... 21
Entering Calibration and Parameter Setup Mode ................................................................... 21
Exit Calibration and Parameter Setup Mode ........................................................................... 21
Set Scale Capacity .................................................................................................................. 21
Set Scale Count By ................................................................................................................. 22
Two Point Calibration .............................................................................................................. 22
Single Point Calibration – Zero Only ....................................................................................... 22
Single Point Calibration – Span Only ...................................................................................... 23
Calibration Error Codes .......................................................................................................... 23
Scale Calibration Troubleshooting .......................................................................................... 24
Scale Parameter Setup ................................................................................................ 25
Entering Calibration and Parameter Setup Mode ................................................................... 25
Exit Calibration and Parameter Setup Mode ........................................................................... 25
Navigating Parameter Menu with Keypad ............................................................................... 26
Navigating Parameter Menu with ZERO, UNITS, and PRINT ................................................ 26
Parameter Groups .................................................................................................................. 26
Legal for Trade Restrictions .................................................................................................... 27
Audit Counters ........................................................................................................................ 27
Software Part Number and Revision Level ............................................................................. 27
Capacity and Calibration - 1 CAL ............................................................................................ 28
General Settings - 2 CNFG ...................................................................................................... 31
Fiber Optic Port 1 - 3 FIB1 ..................................................................................................... 34
Fiber Optic Port 2 - 4 FIB2 ..................................................................................................... 36
Standard Operation Setpoints and Outputs – 9 OPER ............................................................ 38

Checkweigh Operation Setpoints and Outputs – 9OPEr ...................................................... 40
Exit - 99
DON
........................................................................................................................... 42
Data Communications .................................................................................................. 43
Data String Formatting ............................................................................................................ 45
Custom Data String Configuration .......................................................................................... 50
Remote Commands ................................................................................................................ 52
Installation Instructions ............................................................................................... 54
Removing the Rear Panel ....................................................................................................... 54
Cable Glands .......................................................................................................................... 54
Internal Power Connections .................................................................................................... 55
Load Cell Connections ............................................................................................................ 55
Scale Installation ..................................................................................................................... 56
Electrical Ratings .................................................................................................................... 57
Installation with the 8BIS Battery ................................................................................ 58
Battery Installation .................................................................................................................. 58
Battery Pack Operation ........................................................................................................... 59
Installation with the 8AIS AC Power Supply .............................................................. 60
AC Power Supply Installation in Hazardous Location ............................................................. 60
AC Power Supply Installation In Safe Areas ........................................................................... 61
Extension Cable ...................................................................................................................... 62
Remote Switch Connections ................................................................................................... 62
Hazardous Area Fiber Optic Communications .......................................................... 63
8FB Safe Area Fiber Optic Option ............................................................................... 64
Installation ............................................................................................................................... 64
Fig. 8: 8FB Fiber Optic Option Connections ...................................................................................... 64
Remote Display Option ................................................................................................ 65
Fig. 11: Remote Display Option ......................................................................................................... 65
Troubleshooting ........................................................................................................... 66
Scale Messages ...................................................................................................................... 67
Default to Factory Settings ...................................................................................................... 68
Scale Default Settings ............................................................................................................. 68

1
Introduction
Thank you for purchasing a Doran Scales Model 8200IS Intrinsically Safe indicator. The
Model 8200IS indicator for Hazardous Locations is designed to provide consistent
reliability in the most demanding environments.
This manual describes the installation, operation and functionality of the 8200IS
Intrinsically Safe Indicator. Please be sure to read the entire manual and control
drawings to ensure that you obtain all the benefits that the 8200IS series can provide. If
any questions arise, please feel free to contact the Doran Scales Technical Support
Department at [email protected].
Doran Scales, Inc.
883 Enterprise Ct.
St. Charles, IL 60174
E485121

2
Specifications
UL Certificate Number
20190304-E485121
Enclosure
304 Stainless Steel
Product Dimensions
10” W x 6.75” H x 3.5” D
Environmental Protection
IP6X
Temperature Range
14 F to 104F (-10 C to +40 C)
8CHG only: 14 F to 86 F (-10 C to +30 C)
Altitude
Up to 2000 meters
Resolution Range
200d to 100,000d
Humidity
Maximum relative humidity 95%, non-condensing
Analog Signal Sensitivity
0.16 µV/e minimum, 0.5 µV/e typical
System Linearity
0.01% full scale
Analog Signal Range
-0.5mV/V to 5 mV/V with 4 and 6 wire input
Excitation Voltage
5 VDC
Number of Load Cells
Up to (4) 350 Ohm, 4 or 6 wire
Load Cell Entity
Parameters
Load cells must be certified for appropriate hazardous area
and entity parameters. See note one on control drawing
900243
Uo 7.14 V, Io 0.7076 A, Po 0.895 W, Co 10.8 uF, Lo 71 uH
Scale Inputs
One
Calibration Range
Calibrate between 2% and 100% of capacity
Power Input
100 VAC 50/60Hz
Battery Option
Rechargeable Sealed Lead Acid Battery
Charging time 36 for 8 hours of continuous use, 1000
recharge cycles
Display
1" high, 6 digit backlit LCD
Displayed Units
lb, kg, oz, g, lb:oz
Capacity Range
1 to 999,000 lb
Communication
Two Fiber Optic Ports standard
Communication Options
Safe Area Fiber Optic to RS-232 Output Converter
(Part#: 8FB)
Remote Inputs
Two programmable remote switch inputs

3
Scale Operation
Fig. 1: Model 8200IS Front Panel Layout
Scale Annunciators
Unit of measure lb, oz, kg, or g. The units annunciator to the right of the
display will indicate the current unit of measure.
Net weighing mode is indicated by the NET annunciator. The annunciator
will illuminate when a net weight is displayed. When not illuminated, a
gross weight is displayed.
►0◄
Center of zero. The annunciator will illuminate while the scale is
displaying a zero weight.
MOTION
Motion indicator. This symbol represents motion or instability of the
weight. The annunciator will illuminate when motion is sensed on the
platform. Changes in weight, vibration or air currents can cause the scale
to go into motion.
Battery annunciator. Indicates that the battery is near low battery status
that automatically turns off the indicator. Recharge the battery or replace
with a charged battery.

4
1 to 8 setpoint output status indicators. Below the weight display are
annunciators that are illuminated when an output is active in weighing
mode or the current setpoint or preact is being edited.
Powering On and Off
Connect the indicator to a compatible AC or Battery power source.
To turn the indicator on, press ZERO.
To turn the indicator off, press and hold ZERO until
OFF
is displayed.
Basic Weighing Operation
1) Remove all items from the scale platform
2) Press the ZERO button to zero the scale
3) The weight display now reads zero
4) Place an item on the scale platform and wait for the motion annunciator to turn
off, indicating an accurate, stable weight
ZERO
ZERO is used to zero the scale. To zero the scale, press the ZERO button. The scale
will not zero if the scale is in motion. The zero function will operate over the entire
capacity of the scale.
If the scale is displaying a net weight, pressing ZERO will return the scale to gross mode
and display a zero weight. The stored tare will remain in memory.
The scale is equipped with a Zero on Demand parameter which zeros the scale upon
the next stable reading after ZERO is pressed.
TARE
Place the item you wish to tare on the scale platform and press TARE. The scale will
display a net weight and the NET annunciator will illuminate.
Tare weights will remain in memory even if the indicator is turned off.
Keyboard TARE entry
Enter a weight and press TARE to save or press CLEAR to cancel tare entry. The scale
will display a net weight and the NET annunciator will illuminate.
Digital Tare Entry
Enter a weight and press TARE to save or press CLEAR to cancel tare entry. The scale
will display a net weight and the NET annunciator will illuminate.
When utilizing the Product ID memory, a Tare weight is stored with the associated
Product ID number if desired.

5
Display TARE value
To display the current tare value, press and hold TARE for three seconds. The display
will briefly read
TARE
then flash the tare weight in the currently selected units. To exit
press CLEAR.
Clear TARE value
Enter 0 and press TARE. This will remove the tare weight from memory. The display
will read
CLRTAR
to confirm the tare value has been cleared.
GROSS NET
Press the GROSS NET button to switch between the gross and net weighing mode.
Switching to the net mode is possible only when a tare is entered. Net mode is
indicated when the NET annunciator is illuminated.
UNITS
UNITS selects the unit of measure. Press UNITS to change the current unit. The units
annunciator to the right of the display will indicate the current unit or measure: lb, oz, kg,
or g. Lb:oz is disabled by default. Each unit can be enabled or disabled in the scale
parameter setup.
PRINT
PRINT transmits data to a printer or other external devices. When the data is
transmitted, the leftmost display digit will momentarily display an "r" to confirm data
transmission.
There are many parameters that customize the control of manual and automatic
transmission of data. Data can be transmitted via standard fiber optics to Doran’s safe
area fiber optic converter. Contact Doran Tech Support at [email protected] for
support.
Password Protected Setpoint, Preact, Tare and Check Limit Values
All values can be reviewed, but cannot be changed unless the password is deactivated.
If the password protection is activated, the display will display
PASS
when the
SETPOINT, TARE, UNDER or OVER values are changed. Enter the password and
press ENTER, the display shows
PASS
then
OFF
. Press SETPOINT to change or review
weight values or press and hold SETPOINT to edit or review preacts.
After entering the new setpoint or preact values, press and hold the ENTER button for 2
seconds to activate the password protection.
NOTE: If a password number has been activated in parameter 2.10
PASS
, the password
protection will be activated upon power up.

6
Setpoint and Output Operation for Standard Configuration
The 8200IS is equipped with eight outputs and eight setpoints. The output must be
assigned by the Output Configuration (9.7
OUT
) parameter to any of the eight setpoints,
remote input, batch program control and threshold weight to activate. A setpoint is a
target weight that triggers an output. The method of triggering the assigned output is
controlled by the configuration of the Setpoint Operation (9.4
S.O.
) parameter. No
outputs are available on board due to intrinsic safety requirements. All output status will
be communicated via fiber optic to a safe area communications box such as Doran’s
optional 8FB fiber optic to RS232 converter.
Enter SETPOINT Target Weight
Press SETPOINT. The last viewed or edited setpoint will be displayed. Press UNITS or
PRINT to scroll through the eight available setpoints. The annunciators below the main
display indicate the current setpoint.
Enter the setpoint weight using the numeric keypad. Press SETPOINT to accept the
change and return to the weigh mode or press UNITS or PRINT to save and edit other
setpoints. Press SETPOINT to exit this mode.
The display will read
ABORT
to indicate no changes were made to the setpoint values or
the display will read
SAVED
to indicate the setpoint value is saved.
Display SETPOINT Target Weight
Press SETPOINT. The last viewed or edited setpoint will be displayed. Press UNITS or
PRINT to scroll through the eight available setpoints. The annunciators below the main
display indicate the current setpoint. Press SETPOINT to exit this mode.
Setpoint Learning Preacts
A preact works with setpoints to allow the user to enter setpoint target weights that are
the final desired weight. The preact automatically adjusts the setpoint target weight
required for material in transit variations or line pressure changes. The output assigned
to the setpoint will then transition before the setpoint target weight is achieved.
Note: Preacts are always adjusting the weight through the learning process. If the
process has not changed, it should not be necessary to change this value. If the
process is not reliably in control, change Preact Adjustment % Configuration (9.6
PRE
)
parameter to dial in the learning process. Use the password protection feature if
adjustment of the preact could cause a safety issue.
Output Transition = Setpoint target weight – preact weight
The preact value changes based upon the final weight using the following formula:
Preact = previous preact + Adjustment % x (final stable weight – setpoint target weight)
The Preact Adjustment % Configuration (9.6
PRE
) parameter affects how the learning
preact will react to changes. The default value is 50% when the learning preact is
turned on. The final stable weight sample will be collected within 3.5 seconds of the
output transition. If no stable weight can be achieved in this time, the preact will not be

7
adjusted for that measurement. The overall change will be limited to a maximum of 63%
of the setpoint value, regardless of the adjustment percentage.
For Example:
20 pounds of a material is desired and material in transit is observed and estimated at
0.5 lb.
Setpoint 1 is set to 20 lb
Preact 1 is set to 0.5 lb
Adjustment % is left at the default of 50%
After running the process, the final weight is observed to be 20.3 lb
Preact = 0.5 lb + 0.5 x (20.3 – 20)
Preact = 0.65 lb
Enter Setpoint Preact Weight
Press SETPOINT. The last viewed or edited setpoint will be displayed. Press ZERO to
display the preact weight. Press UNITS or PRINT to scroll through the eight available
preacts. The annunciators below the main display indicate the current setpoint. Input
the desired preact weight, then press ENTER. Press SETPOINT to exit this mode.
Display Setpoint Preact Weight
Press SETPOINT. The last viewed or edited setpoint will be displayed. Press ZERO to
display the preact weight. Press UNITS or PRINT to scroll through the eight available
preacts. The annunciators below the main display indicate the current setpoint. Press
SETPOINT to exit this mode.

8
Tank Level Maintenance Mode Standard Configuration
The F1L Setpoint Operation (9.4
S.O.
) maintains a level in a tank between two setpoint
target weights. This allows the tank to be drained to a desired amount before being
refilled to a maximum target weight. Setpoint 1 will be the low level of the tank, and
setpoint 2 will be the high level. When the tank is drained to setpoint 1, the output turns
ON and the tank will begin filling. When the tank fills up to setpoint 2, the output turns
OFF and until the tank is once again drained to setpoint 1. The output has been
represented graphically below. No outputs are available on board due to intrinsic safety
requirements. All output status will be communicated via fiber optic to a safe area
communications box such as Doran’s optional 8FB fiber optic to RS232 converter.
To configure this operation:
1. Setpoint 1 must be configured to oFF in the Setpoint Operation (9.4
S.O.
)
parameter.
2. Setpoint 2 must be configured to FIL in the Setpoint Operation (9.4
S.O.
)
parameter.
3. Assign setpoint 2 to an output in Output Operation (9.7
OUT
) parameter.
4. Enter setpoint 1 as the lowest weight desired, as described in the setpoint section
of the manual.
5. Enter setpoint 2 as the highest weight desired, as described in the setpoint
section of the manual.
6. Ensure Setpoint Weight Operation (9.5
SW
) is configured to
DSP
for displayed
weight.
Note: While
FIL
is set to setpoint 2, the ZERO button is disabled

9
Product ID Standard Configuration
800 product IDs are available. Deploying a large library of IDs with multiple scales can
be easy to manage with Doran’s data management programs.
Product IDs save information that includes:
• Setpoint values
• Preact values
• Unit of measure
• Accumulator and counter values
• Tare
• Two 40 alphanumeric character fields
• Motion Aperture (1.6
M.A.
)
• Threshold (2.5
THS
)
• Setpoint operation (9.4
S.O.
)
• Output configuration (9.7
OUT
)
Recall PRODUCT ID from Memory
When powered on, no product ID will be loaded. This is indicated when pressing PROD
ID and the display reads
OFF
. Once a product ID is loaded, the unit of measure is
locked in the unit of the product ID.
To select a stored product, press PROD ID, enter the ID number and press ENTER.
The display will read
SAVED
to indicate the fields associated with that Product ID number
are active. After selecting a product, the scale will measure and display in the units
saved for that product. The UNITS button will then be disabled. Selecting product
‘OFF’ will re-enable the UNITS button.
Another method to select a product is to press PROD ID, then use the UNITS or PRINT
buttons to scroll through the available products. Press ENTER to select the displayed
product. The display will read
SAVED
to indicate the fields associated with that Product ID
number are active.
Display Current PRODUCT ID
Press PROD ID, the display will show
ID
followed by the currently active product.
Create New PRODUCT ID
Select the desired unit that will be used to checkweigh the new product. Enter a product
ID up to 6 digits not currently in memory and press PROD ID. The display will
momentarily show
NEW
then
ID
. Then return to weighing mode. All fields associated with
the new Product ID number will be blank.
To enter and save values for all fields associated with the current Product ID, enter
values for each field. When changing products, the display will read
SAVED
to indicate
that all fields associated with the new Product ID number are saved and will be recalled
when that product is used again.

10
Delete PRODUCT ID from Memory
Enter the product ID to be deleted and press PROD ID. The display will show
PRD ID
,
followed by the Product ID number. Press and hold the CLEAR button for more than 2
seconds. The display will show
CLR ID
and then
DONE
. All fields associated with that
Product ID number will be cleared. The previously used Product ID number will become
active.
Product Fields Standard Operation
The indicator memory has eight 40-character alphanumeric fields that can be entered
and transmitted as desired using custom data strings. In addition, there is a ninth
product for serialization which increments from the five digit number entered. This is
useful for custom data labels and data collection.
Display Product Fields
To access Product Fields, press and hold PROD ID on the front panel for 3 seconds.
The display will show “PF 1” for a second, then display the first 6 characters of the
product field if they are numeric. The eight fields can be cycled through by pressing
ENTER. Press PROD ID to exit from the Product Field mode.
Product Field Entry
To access Product Fields, press and hold PROD ID on the front panel. After 3 seconds,
the scale display will change from the current platform weight to show which Product
Field is ready for entry. For the first field, the display will show “PF 1” for a second, then
display the current entry of this field.
Product Fields can be entered by external communications, or by the keypad on the
front panel. A keypad entry followed by pressing ENTER, will store the entered value
as that Product Field. Once entered, the scale will then display the next Product Field,
in this case “PF 2”. The scale will cycle through the 9 Product Fields unless the user
presses PROD ID again, which will exit from the Product Field mode.

11
Accumulator and Counter Standard Operation
Accumulator and Counter Operation
When a manual or automatic print function is executed, the accumulator has the
currently displayed weight added to its current value and the counter is incremented.
To accumulate automatically, select an auto print function in the parameter setup menu.
To accumulate manually, allow the scale to become stable and press PRINT.
The maximum value that can be shown for the accumulator and counter is 999,999.
When the maximum value is reached, the accumulator and counter will rollover to a zero
value. This feature can only be used in a non Legal For Trade application.
If using Product ID functions, the Accumulator and Counter values are stored with the
associated product.
Display Accumulator and Counter Values
Press the ACCUM button to enter the accumulator and counter recall mode. The display
will show
ACCUM
followed by the accumulated weight in the units currently selected in the
weigh mode. Then
COUNTR
will be displayed followed by the counter value.
Press ACCUM to exit the accumulator and counter recall mode without changing their
values.
Clear Accumulator and Counter
Press the ACCUM button to enter the accumulator and counter recall mode. The display
will show
ACCUM
followed by the accumulated weight in the units currently selected in the
weigh mode. Then
COUNTR
will be displayed followed by the counter value.
Press CLEAR to clear the accumulator and counter values. The display will show
CLR AC
and exit from the recall mode.
Changing the current display units will clear both the accumulator and counter values.
Accumulator and Counter Data String Output to Printer or Data Collection
Press ACCUM to enter the accumulator recall mode. Press PRINT to transmit the LB4
custom data string that contains the accumulator and counter values by default. Both
the accumulator and counter values are cleared after transmission.

12
Checkweighing Operation
The indicator can be converted from the standard configuration into checkweighing
operation. Checkweighing removes the functionality of setpoints and outputs, and
activates checkweighing. Some sections will be repeated to reflect the full functionality
of the checkweighing operation.
Three Band Checkweighing
Three band checkweighing classifies weighments into over, accept and under. The
default configuration is three band checkweighing.
Three Band Checkweighing (9.1
C.O.
set to operation starting with 3)
1. Remove all items from the scale platform
2. Press ZERO and the display will read zero weight
3. Place an item on the scale platform and wait for the scale to stabilize
4. Accept, Over or Under annunciators indicate checkweigh status
Enter and Display of Checkweigh Limits (9.2
C.E.
default value
SCR
)
1. Press F1 (OVER) or F2 (UNDER)
2. The display will briefly read
OVER
or
UNDER
followed by the current limit
3. Enter limit value using the keypad and decimal point
4. Press ENTER to save the value or press CLEAR to exit without saving
5.
SAVED
is displayed if saved or
ABORT
is shown if aborted
Weight Reference and Digital Entry of Checkweigh Limits (9.2
C.E.
set to
SCS
)
1. Press ZERO
2. Place a target item on the scale
3. Press F1 (OVER) or F2 (UNDER)
4. The display will briefly read
OVER
or
UNDER
followed by the weight on the platform
5. Press either F1 (OVER) to increase the weight value or press F2 (UNDER) to
decrease the weight value. Pressing and holding will accelerate the weight
scroll.
6. Press ZERO to save the value or press PRINT to exit without saving
7.
SAVED
is displayed if saved or
ABORT
is shown if aborted
Weight Reference Entry of Checkweigh Limits (9.2
C.E.
set to PB)
1. Press ZERO
2. Place a weight equal to the desired F1 (OVER) or F2 (UNDER) limit on the
platform
3. Press F1 (OVER) or F2 (UNDER)
4.
OVER
or
UNDER
is displayed and then
SAVED
to indicate the new Over limit was
saved.

13
Five Band Checkweighing
Five band checkweighing classifies weighments into high, over, accept, under and low.
Five Band Checkweighing (9.1
C.O.
set to operation starting with S)
1. Press ZERO
2. Place an item on the scale
3. Checkweigh status is indicated as follows
a. Flashing OVER = HIGH
b. Solid OVER = OVER
c. Solid ACCEPT = ACCEPT
d. Solid UNDER = UNDER
e. Flashing UNDER = LOW
Enter and Display of High and Low Limits (9.2
C.E.
default value
SCR
)
1. Press and hold the F1 (OVER) or F2 (UNDER) until the display reads
HIGH
or
LOW
respectively
2. The current weight value of the saved limit is displayed and checkweigh status
annunciators will flash
3. Enter limit value using the keypad and decimal point
4. Press ENTER to save the value or press CLEAR to exit without saving
5.
SAVED
is displayed if saved or
ABORT
is shown if aborted
Weight Reference and Digital Entry of High and Low Limits (9.2
C.E.
set to
SCS
)
1. Press ZERO
2. Place an item of the desired weight on the scale platform
3. Press and hold the F1 (OVER) or F2 (UNDER) until the display reads
HIGH
or
LOW
respectively
4. The current weight value of the saved limit is displayed and checkweigh status
annunciators will flash
5. Press either F1 (OVER) to increase the weight value or press F2 (UNDER) to
decrease the weight value. Pressing and holding will accelerate the weight
scroll.
6. Press ZERO to save the value or press PRINT to exit without saving
7.
SAVED
is displayed if saved or
ABORT
is shown if aborted
Weight Reference Entry of High and Low Limits (9.2
C.E.
set to
PB
)
1. Press ZERO
2. Place an item of the desired weight on the scale platform
3. Press and hold the F1 (OVER) or F2 (UNDER) until the display reads
HIGH
or
LOW
respectively
4. The display will briefly read
OVER
or
UNDER
followed by the weight on the platform
and checkweigh status annunciators will flash
5. Press either F1 (OVER) to increase the weight value or press F2 (UNDER) to
decrease the weight value. Pressing and holding will accelerate the weight
scroll.
6. Press ZERO to save the value or press PRINT to exit without saving
7.
SAVED
is displayed if saved or
ABORT
is shown if aborted

14
Zero Band Checkweighing
Basic checkweighing - simply set the desired weight on the platform, press zero and
checkweigh based upon the standard tolerances in the O.U. parameter (9.3
O.U.
).
Zero Band Checkweighing (9.1
C.O.
set to operation starting with 0)
1. Remove all items from the scale platform
2. Place the target weight on the scale platform
3. Press ZERO and the display will read zero weight
4. Remove the target weight
5. Place an item on the scale platform and wait for the scale to stabilize
6. A zero weight will indicate the item is exactly the target weight. Any weight above
or below zero indicates the amount of weight away from the target weight.
7. Accept, Over or Under will be displayed based upon the tolerance set in 9.3
O.U.

15
Product ID Checkweigh Operation
800 product IDs are available. Deploying a large library of IDs with multiple scales can
be easy to manage with Doran’s QC Weigh and CheckWay data management
programs.
Product IDs save information that includes:
• Checkweigh limits
• Unit of measure
• Accumulator and counter values
• Tare
• Two 40 alphanumeric character fields
• Number of samples and alarm timer for QC Weigh
• Motion Aperture (1.6
M.A.
)
• Threshold (2.5
THS
)
• Checkweigh operation (9.1
C.O.
)
• Checkweigh limit entry (9.2
C.E.
)
• Output configuration (9.7
OUT
)
Recall PRODUCT ID from Memory
When powered on, no product ID will be loaded. This is indicated when pressing PROD
ID and the display reads
OFF
. Once a product ID is loaded, the unit of measure is
locked in the unit of the product ID.
To select a stored product, press PROD ID, enter the ID number and press ENTER.
The display will read
SAVED
to indicate the fields associated with that Product ID number
are active. After selecting a product, the scale will measure and display in the units
saved for that product. The UNITS button will then be disabled. Selecting product
‘OFF’ will re-enable the UNITS button.
Another method to select a product is to press PROD ID, then use the UNITS or PRINT
buttons to scroll through the available products. Press ENTER to select the displayed
product. The display will read
SAVED
to indicate the fields associated with that Product ID
number are active.
Display Current PRODUCT ID
Press PROD ID, the display will show
PRD ID
followed by the currently active product.
Press ENTER to leave this mode.
Create New PRODUCT ID
Select the desired unit that will be used to checkweigh the new product. Enter a product
ID up to 6 digits not currently in memory and press PROD ID. The display will
momentarily show
NEW
. Then return to weighing mode. All fields associated with the new
Product ID number will be blank.
To enter and save values for all fields associated with the current Product ID, enter
values for each field. When changing products, the display will read
SAVED
to indicate
the all fields associated with the new Product ID number are saved and will be recalled
when that product is used again.

16
Delete PRODUCT ID from Memory
Enter the product ID to be deleted and press PROD ID. The display will show
PRD ID
,
followed by the product ID number. Press and hold the CLEAR button for more than 2
seconds. The display will show
CLR ID
. The product ID will be set to off until another
product ID is selected.
User ID Checkweigh Operation
User logins can be entered prior to weighing when using Product IDs while the 8200IS
is set to QC Weigh mode. This locks scale weighments behind a login and records the
User ID of samples in ionSuite.
User ID login
With the display showing
LOGIN
, enter in through the keypad up to 20 digits for a user ID
number. The User ID entered is compared with a list of up to 200 User IDs stored in the
scale’s memory. If a User ID entered does not match any of the stored IDs, the display
will show
ERROR NO USER
message.
User ID Logout
Press and hold the CLEAR button for more than 2 seconds. The display will show
CLRUSER
. Display will show
LOGIN
to indicate scale is disabled and requires a user id to
login.
This manual suits for next models
1
Table of contents
Other Doran Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Wuhan Guide
Wuhan Guide HTWS-V manual

Keysight Technologies
Keysight Technologies 34980A Service guide

Emerson
Emerson Rosemount CT5100 Ex quick start guide

Schaller Messtechnik
Schaller Messtechnik Humimeter AW3 user manual

Fluke
Fluke OptiFiber Pro Getting started guide

Southland
Southland H2S-625 instruction manual