
Instruction No. EH 45004 E
05/2004
8
When the handwheel (120) is turned clockwise, the
stop (122) moves up the threaded rod, compressing
the spring (134) and raising lever No. 2 (22) (VariPak
28002) or (113) (VariPak 28001). This closes the
valve if the valve is equipped with a direct-action
actuator (Air-to-Close) and opens the valve if it is
equipped with a reverse-action actuator (Air-to-
Open).
The valve can be returned to automatic mode
(neutral position), by turning the handwheel counter-
clockwise until the stop (122) comes into contact with
the bracket (108).
Note: During this operation, the compression is
released when the stop (122) moves away from the
lever (22) (VariPak 28002) or (113) (VariPak 28001).
Continue the operation until a slight tension
reappears, then tighten the handwheel lock (121).
VariPak actuators are equipped with a limit stop.
This consists of parts (180) and (181), and is
designed to prevent damage to the plug and seat
ring assembly and/or the plug stem in the event of
handwheel or actuator overstroke.
Note: The limit stop is not used with air-to-open
valves fitted with a handwheel, and must be screwed
down at the bottom of the piston.
2.5 REVERSING THE VALVE ACTION
(Figures 5, 6, 26, 27 and 29)
A. Loosen and remove cover (110). On valves
equipped with a handwheel, loosen cover screw
(109), back off handwheel lock (121), and turn
handwheel (120) counter-clockwise to release
cover (110).
B. Adjust the signal so that the valve closes.
Change the signal slightly so that the plug just
moves off the seat ring. On the VariPak 28002
adjustable-Cvvalve, set the adjustment knob
(24) to the minimum Cvsetting.
C. Loosen locknut (103) and, using a screwdriver,
turn the plug stem one and three-quarter turns
counter-clockwise. Shut off the signal and the
supply pressure. Slightly tighten locknut (103)
against the clevis.
D. Unlock nut (117) and completely loosen screw
(116). Unhook spring (114) from spring clamp
(115).
2.5.1 VariPak 28001 single-lever valve
E. Remove retainer clips (112b) from pin (18), and
remove the pin from lever (113) and plug clevis
(104a).
Note: This operation will be simplified by
relieving the load exerted on the plug clevis by
the conical compression spring (106). To do this
push against the plug stem end with a
screwdriver while driving out the pins.
F. Remove the two retainer clips (112a) from the
pin (105) and disengage it to uncouple the lever
(113) from the bracket (108).
G. Refit the levers (113) making sure that the pin
(105) is fitted into the correct holes in the lever
and bracket for the new action of the actuator
(see figure 6). Fit the spring clamp (115) into the
corresponding lever hole.
Note: Ensure that the clevis is correctly
positioned before fitting the lever on the bracket.
H. Couple the plug stem clevis (104a) to the lever
(113) following the normal reassembly
procedure (step I. in the "Reassembly" section
on page 17). Then follow the same procedure as
for the VariPak 28002 adjustable-Cvvalve as
from step L.
2.5.2 VariPak 28002 adjustable-Cvvalve
(2 levers)
E. Remove retainer clips (112b) from each of the
two pins (184) and remove pins from lever (113)
and plug clevis (104a).
Note: This operation will be simplified by
relieving the load exerted on the plug clevis by
the conical compression spring (106). To do this
push against the plug stem end with a
screwdriver while driving out the pins.
F. Remove the two retainer clips (112a) from pin
No. 1 (105) and disengage it to uncouple lever
No. 1 from bracket (108). Loosen adjustment
knob (24) and slide it to the maximum Cvsetting.
Disengage the smooth end of the adjustment pin
(23) from the groove in lever No. 2 while
removing lever No. 1.
G. Drive out pin No. 3 (18) and remove lever No. 2
(22) from the bracket.
H. Remove the two screws (26) and refit Cv
adjustment plate (25) after turning it round to
comply with the new action.
I. Refit levers No. 1 and 2 in sequence, making
sure that pins (105) and (108) are fitted into the
correct holes in the levers and bracket for the
new action of the actuator (see figure 6).
Caution: The valve must be isolated and
pressure vented before disassembly.