Dynamica HALO XB-10/BIOmaster/VIS-20/ENVmaster User manual

HALO XB-10/VIS-20
/BIOmaster
Instruction manual
Accessories for Model HALO XB-10/VIS-
20/BIOmaster UV-VIS Spectrophotometer
Revision : D01-201407

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
1
XB-10/BIOmaster/VIS-20/ENVmaster Spectrophotometer
Thank you for your purchase of Dynamica XB-10/BIOmaster/VIS-20/ENVmaster spectrophotometer. In
order to ensure proper use and your safety, please read this manual carefully and keep it well before using
the instrument.
Information contained in this manual is subject to change without notice for product appearance and
technical data. Enquiries are welcome.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
2
IMPORTANT
Precautions on Electromagnetic Wave Interference
(1)Possible Electromagnetic Wave Interference Caused by This Instrument
When this instrument is used in a residential area or an adjacent area thereto,
it may cause interference to radio and television reception.
To prevent this, use the specified system connection cables in strict
accordance with the instruction manual.
The instrument is designed to minimize possible electromagnetic wave
interference caused by it if the specified cables are connected properly.
However, there is no guarantee that electromagnetic wave interference will not
be caused by the instrument.
If the instrument does cause interference to radio or television reception, which
can be determined by turning off and on the instrument, the user is encouraged
to try to correct the interference by one or more of the following measures:
Increase separation between the instrument and radio/TV receiver.
Connect the instrument to an outlet on a circuit different from that to which
the radio/TV receiver is connected.
(2)Possible Electromagnetic Wave Interference Affecting This Instrument
If this instrument is used near an intense electromagnetic source, interference
noise may be given to the instrument to incur an adverse effect on its
performance or functionality.
To prevent this, use the specified system connection cables in strict
accordance with the instruction manual.
The instrument is designed to minimize possible electromagnetic wave
interference affecting it if the specified cables are connected properly.
However, there is no guarantee that electromagnetic wave interference will not
occur in this instrument.
If the instrument does incur electromagnetic wave interference, which can be
determined by turning on and off possible sources of electromagnetic wave
interference nearby, the user is encouraged to try to correct the interference by
one or more of the following measures:
Reorient the instrument.
Increase separation between the instrument and possible sources of
electromagnetic wave interference.
Increase separation between the power cable of the instrument and
possible sources of electromagnetic wave interference.
Connect the instrument to an outlet on a circuit different from that to which

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
3
possible sources of electromagnetic wave interference are connected.
Confirm that any other device connected with the instrument is not
affected by electromagnetic wave interference.
Warranty on Product
The Model XB-10/BIOmaster/VIS-20/ENVmaster spectrophotometer is
warranted to operate according to the specifications given in the instruction
manual, provided it is used in accordance with the instructions described in the
manual.
(1)Scope of Warranty
(a) Any parts which prove to be defective in design or workmanship during
the warranty period will be repaired without charge.
(b) A substitute part may be used for repair, or replacement with an
equivalent product may be made instead of repair.
(c) Such system components as a personal computer and printer to be
updated frequently for improvement may not be available in original versions at
the time of replacement.
(2)Warranty Period
One year from the date of initial installation.
(3)Availability of Technical Support Service
Technical support service for this instrument is available within regular working
hours on workdays predetermined by us.
(4)Limitations and Exclusions on Warranty
Note that this warranty is void in the following cases, even if they occur within
the warranty period.
(a) Failure due to operation at a place not meeting the installation
requirements specified by us
(b) Failure due to power supply voltage/frequency other than specified by us
or due to abnormality in power supply
(c) Corrosion or deterioration of the tubing due to impurities contained in
reagent, gas, air or cooling water supplied by the user
(d) Corrosion of the electric circuits or deterioration of the optical elements
due to highly corrosive atmospheric gas
(e) Failure due to use of hardware, software or spare parts other than
supplied by us
(f) Failure due to improper handling or maintenance by the user
(g) Failure due to maintenance or repair by a service agent not approved or
authorized by us

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
4
(h) After disposal of this instrument, or after its resale without our approval
(i) Failure due to relocation or transport after initial installation
(j) Failure due to disassembly, modification or relocation not approved by us
(k) Consumables, and failure of parts that have reached the end of specified
useful life
(l) Failure of parts excluded from the warranty in the instruction manual or
other documents
(m) Failure due to acts of God, including fire, earthquake, storm, lightning,
social disturbance, riot, crime, insurrection, war (declared or undeclared),
radioactive pollution, contamination with harmful substances, etc.
(n) Failure of the hardware, or damage to the system software, application
software, data or hard disk due to computer virus infection
(o) Failure of the personal computer connected with the instrument, or
damage to the system software, application software, data or hard disk due to
power interruption or momentary power voltage drop caused by lightning or the
like
(p) Failure of the personal computer connected with the instrument, or
damage to the system software, application software, data or hard disk due to
disconnection of main power to the personal computer without taking the
specified normal shutdown procedure
(5)Disclaimer of Warranty
(a) Any express warranties other than the explicit conditions indicated in (1)
are excluded from this warranty.
(b) Any other implied warranties of merchantability and fitness for a particular
purpose are not included in this warranty. No liability is assumed for direct or
indirect damages arising out of explicit or implied warranties.
(c) Oral or written information or advice given by our dealers, distributors,
agents or employees without our express permission shall not create a
warranty or in any way increase the scope of this warranty.
Installation, Relocation and After-sale Technical Service
Installation of this instrument shall be carried out by or under supervision of
qualified service personnel of Techcomp or its authorized service agent.
Before installation of the instrument, the user shall make preparations for
satisfying the installation requirements in accordance with the instruction
manual.
If relocation of the instrument becomes necessary after initial installation
(delivery), please notify your local sales representative or nearest service office
of Dynamica.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
5
Disposal this instrument
When you discard equipment, please check and discard a related statute etc.
or ask the service section of Dynamica.
Other Precautions
(1)Handling of Chemicals and Samples
(a)The user is responsible for following relevant legal standards and
regulations in handling, storage and discarding of chemicals and samples used
in analytical operations of this instrument.
(b) Reagents, standard solutions and accuracy-control samples shall be
handled, stored and discarded as instructed by the respective suppliers.
(2)Notice on Instruction Manuals
(a)Information contained in the instruction manuals furnished with the
instrument is subjected to change without notice for product improvement.
(b) This manual is copyrighted by Techcomp with all rights reserved.
(c) No part of this manual may be reproduced or transmitted in any form or by
any means without our express written permission.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
6
SAFETY SUMMARY
CAUTION
For your safety please read the following precautions carefully before
using the instrument.
General Safety Guidelines
For safe handling of this product, please follow the instruction procedure in
the manual for this product.
Pay special attention to follow all the hazard warnings on the product and
in the manual. Failure to do so can cause injury to you or damage to the
product.
After installation, please do not move the equipment. A vibration might
affect the adjustment of the product.
The hazard warnings, which appear on the warning labels on the product
or in the manual, have one of the following alert headings consisting of an
alert symbol and a signal word, DANGER, WARNING, or CAUTION.
DANGER: Indicates an imminently hazardous situation that, if not
avoided, will result in death or serious injury. (It does not
apply to this equipment.)
WARNING: Indicates a potentially hazardous situation that, if not
avoided, can result in death or series injury.
CAUTION: Indicates a hazardous situation that, if not avoided, will or
can result in minor to moderate injury, or serious damage to
the product.
: The alert symbol shown precedes every signal word for hazard
warnings, and appears in safety related descriptions in the manual.
In addition, the following “Attention” and “Note” are not directly related to the
safety of a person:

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
7
△ Attention: It is used to present warnings, which are not directly related to
personal injury hazards. It is used to indicate prevention against damage to
the equipment.
◇ Note: This is used to indicate instructions that enable you to operate the
equipment accurately and perform accurate measurements.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
8
SAFETY SUMMARY (Continued)
General Safety Guidelines (Continued)
Before using
Before using this product, please make sure you read and understand the
instructions.
Please keep this manual in a safe and easily accessible place so that you
can use it when necessary.
Please make sure to use this product properly and follow the instructions
as specified in this manual.
Please make sure to understand and follow the instructions regarding
safety in this manual.
If you do not follow the instructions in this manual, an inaccurate analysis
may result or bodily injury may occur.
Because of danger, please make sure not to modify or alter the product,
make sure not to use unspecified parts, and make sure not to operate the
equipment by removing/defeating the safety device(s).
When using chemicals, please make sure to ventilate the room well. If
there is not enough ventilation, it may be hazardous to your health.
Although we have carefully considered the instructions written on the
products and manuals, it is possible for an unexpected event to occur.
When operating the equipment, aside from following the instructions, be
very cautious.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
9
SAFETY SUMMARY (Continued)
General Safety Guidelines (Continued)
Precautions for Installation • Maintenance • Relocation and After Sale Technical
Service
Before installation, confirm that there are no missing items or standard
accessories. If there is something missing or damaged, or you have
noticed any problems, please contact our nearest representative.
■ Operating the equipment without a standard part can damage the
equipment and cause safety concerns. If that occurs, please follow the
instruction of the installer.
■ Installation of this instrument shall be carried out by or under supervisions
of qualified service personnel of Dynamica or its authorized service agent.
■ When relocation of this instrument becomes necessary after initial
installation (delivery), please notify your local Dynamica sales
representative or nearest Dynamica service office. Technical support
service for this instrument is available from service agent approved or
authorized by Dynamica within regular working hours or workdays.
■ Please do not perform any other operations that are not included in the
manual. If any problem occurs with the equipment, please contact the
agent from whom you purchased it or the service department of Dynamica.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
10
SAFETY SUMMARY (Continued)
WARNING: Poisoning from Organic Solvent Gas
Handling Organic Solvents
■ The organic solvent vapor may be harmful to your health.
WARNING: Eye Injury from Organic Solvents
Handling Organic Solvents
■ Please wear protective glasses when using organic solvents. If the
organic solvent should get into your eye, flush your eye immediately under
running water for at least 5 minutes while keeping your eyelids open. See
a physician for appropriate treatment.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
11
SAFETY SUMMARY (Continued)
WARNING: Electrical Shocks from Improper Grounding
■ When wiring the personal computer, power supply for the thermostatic cell
holder and the like, please make sure to use the 3-prong wire (with ground)
provided.
WARNING:Electrical Shock from Contact
■ High voltage is used inside of the equipment, be sure to turn off the unit
before connecting the power cord.
CAUTION: Burns from high Temperature
■ The lamp will become very hot during operation.
■ Make sure that the instrument is switch off, the power cord is pulled off and
the D2 lamp and the tungsten lamp is cooled off when replacing the lamp.
CAUTION: Fatigue due to Prolonged Work
■ Viewing the display in your work can cause eye and physical fatigue if you
continue to work in the same posture for extended periods.
■ When working with the display for a prolonged period, for your health,
make sure to take breaks for 10 to 15 minutes every hour in order to rest
your eyes and body.
CAUTION:Indoor ventilation
■ If using UV zero detection system in a small room for a long time, it may
cause indoor nitrogen concentration increases and the oxygen
concentration drops, which will affect human health.
Please install exhaust fan or open the windows frequently to maintain good
indoor ventilation.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
12
SAFETY SUMMARY (Continued)
Electricity
(a) The voltage for the Spectrophotometer system and personal computer
must be a single-phase AC 100V to 240V; Variations in the voltage and
noise generated in the power line will cause adverse effects on the
spectrophotometer and may also cause accidents.
(b) Please make sure that grounding is provided together with the power
supply wires, and make sure that it is connected with a grounding
resistance of less than 4. Defective grounding may not only cause
lower resistance against noise from the outside but it can also cause the
Mass Spectrometer and personal computer to generate static electricity,
which may involve the danger of electrical shocks.
(c) A high voltage circuit is used inside the Spectrophotometer. Do not open
the covers when this circuit is operational because of the danger from
electrical shock.
Fire Extinguishers
(a) Do not smoke or use fire within 3 meters of the Spectrophotometer
(b) Make sure to keep a fire extinguisher near the Spectrophotometer at all
times. Obtain an ABC Powder extinguisher that can be used for normal fires,
oil fires, and electrical fires.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
13
Functional characteristics
This series instrument includes 4 model number spectrophotometers. Model VIS-20 is traditonal visible
spectrophotometer, ENVmaster is enhancement type spectrophotometer with method invocation mode,
XB-10 is traditonal UV-VIS spectrophotometer, BIOmaster is UV-VIS spectrophotometer only for nucleic
acid and protein detection.
Feature
Small and beautiful
appearance
4.3 inches, 480×272 resolution, TFT colorful liquid crystal touch display screen
size: like A3 paper, 400(w)×280(d)×160(h) mm
weight: 4 kg
Easy operation One-button operation interface into the measurement function
High-speed wavelength wave, arrive to any specified wavelength within 1 second
For VIS-20/EVNmaster spectrophotometer, when sampling interval 为is 0.2nm, its
scanning speed is 2400nm/min
Unique application method manager changes the original function-oriented
operation method and user sample measurement method operation mode.
Energy saving and
environmental
protection
High conversion efficiency switching power supply, 100~240VAC voltage input,
and complete riddance of low grid voltage
VIS-20/EVNmaster spectrophotometer uses 2000 hours lifetime imported
halogen lamp
XB-10/BIOmaster spectrophotometer uses low-power, high brightness and 109
times lifetime pulses xenon lamp
More ports Serial printer port connects thermal printer
USB port connects PC
SD card port saves data and measurement methods
Options port connects and controls several options
More options Auto 5-cell holder
Auto sample sipper
Flow cuvette holder
Rectangular long-path cuvette holder, put 10,20, 30, 50, 100mm cuvette holder
100uL micro-cuvette holder
Tube holder(only for VIS-20/ENVmaster)
Electronic thermostat TC cuvette holder(only for XB-10/BIOmaster)

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
14
Appearance
Keypad
Though the instrument uses touch display screen, all operation will be acheived by touch screen. The
purpose of setting the two shortcut key is to make easy operation.
BACK. Return to the previous menu.
MEASURE. Begin to a new measurement.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
15
Operation environment
Power Voltage: AC 100V~240V, Frequency :50/60Hz±1Hz, Capacity: more than 200W
Grounding line resistance of 10Ω or less is required
The input power mutations, without interference from other large electrical equipment
Operating
temperature
5~35 Deg C
In order to perform a measurement under the most stable condition, we recommend
that the instrument is used in an air conditioned room of 20 ~ 25 Deg C.
Storage temperature:-20~55 Deg C
Operating
humidity
Less than 85%.
Because it is optical equipment, in order to ensure it in good working condition, we
recommend that the instrument is used under 30%~70% humidity.
Storage humidity: less than 85%。
Workbench Width: more than 550mm, Depth: more than 400mm, Load bearing: more than 15kg,
horizontal and reliable workbench.
If there are other auxiliary devices, it is necessary to enlarge workbench area.
More than 200mm space at the both sides of the instrument, avoid being close to the
wall.
Height of the instrument: about 160mm. In order to easy operation, please choose
suitable workbench.
Connect with
other equipment
Connect with PC(additional optional software and computer). Indirect specify type
thermal printer(options)and other accessories.
When main unit connects with above equipments, please swithch off the power of the
main unit and other equipments.
All grounding cords are in good condition and can be connected with the grounding
line of the main unit.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
16
Specification(I)
VIS-20/ENVmaster
Wavelength range 320~1100nm
Wavelength controlled
variable
0.2nm
Wavelength accuracy ±1nm
Wavelength repeatability ≤0.5nm
Transmittance accuracy ±0.5%T(NIST 930 Filter)
Transmittance repeatability 0.2%T
Baseline flatness ±0.002Abs(330~1090nm)
Noise level ≤0.0008Abs(500nm)
Baseline stability ≤0.0008Abs/h(500nm, 2 hours after power on)
Spectral bandwidth 6nm±1.2nm
Stray light ≤0.3%T
Wavelength scan speed 2400nm/min (0.2nm sampling interval, without filter
switchover)
Wavelength move speed To any specified position within 1 second
Light source WI lamp
Detector silicon photodiode
Display screen 4.3 inches, 480×272 colorful touch LCD screen
Printer specified 80-column thermal printer(series port)
Metering mode Single beam
Dimension 400(W)×280(D)×160(H)mm
Weight About 4 kg
Power requirement 100~240V,(50/60Hz)
Power consumption 100VA(100V~240V,50/60Hz)

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
17
Specification(II)
XB-10/BIOmaster
Wavelength range 190~1000nm
Wavelength controlled
variable
0.2nm
Wavelength accuracy ±2nm
Wavelength repeatability ≤1nm
Transmittance accuracy ±1%T(NIST 930 Filter)
Transmittance repeatability 0.5%T
Baseline flatness ±0.005Abs(200~990nm)
Noise level ≤0.005Abs(250nm 处)
Baseline stability ≤0.005Abs/h(250nm, 2 hours after light on)
Spectral bandwidth 5nm±1nm
Stray light ≤0.5%T
Wavelength scan speed 300nm/min (0.2nm sampling interval, without filter
switchover)
Wavelength move speed To any specified position within 1 second
Light source pulsed-xenon lamp
Detector silicon photodiode
Display screen 4.3 inches, 480×272 colorful touch LCD screen
Printer specified 80-column thermal printer(series port)
Metering mode Single beam
Dimension 400(W)×280(D)×160(H)mm
Weight About 4 kg
Power requirement 100~240V,(50/60Hz)
Power consumption 100VA(100V~240V,50/60Hz)

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
18
The operation interface
Function shortcut button These buttons appear on every interface and can quickly switch to different
measurement functions.
Measure button These buttons appear on every measure interface. Measure basic operation.
Processing button These buttons appear on scan interface. For the results of the scan data
processing and according to different scaning characteristic, automatically
varycorresponding processing ways.
Spectrogram button These buttons appear on scan interface. For the results of spectrogram
processing.
Status bar These buttons appear on every interface. Show current time, wavelength and
vaule(transmission or absorbance).
Click wavelength in the status bar, it will pop a digital input interface, then set
up wavelength to achevie GOTO λ function.
Click value in the status bar, it will pop a big font measure interface to achieve
simpletransmittance and absorbance read function.
The value in the status bar of VIS-20/ENVmaster spectrophotometer is
real-time reading value of current wavelength.
The value in the status bar of XB-10/BIOmaster spectrophotometer is the last
time measured value.

XB-10/BIOmaster/VIS-20/ENVmaster instruction manual
19
Word input keypad
This keypad appears at the area where can input word, for example, sample name, user and document
name.
BACK close word input keypad
CLEAR clear incorrect word
Word switchover majuscule, lowercase, digit and sign
Enter ensure input word
Digit input keypad
This keypad appears at the area where can input digit, for example, wavelength value and scanning time.
Table of contents