dynasonics TFXL series User manual

Transit Time Ultrasonic Flow Meters
TFXL Meter
TTM-UM-00369-EN-03 (April 2016) User Manual

Transit Time Ultrasonic Flow Meters, TFXL Meter
Page ii April 2016TTM-UM-00369-EN-03

CONTENTS
Scope of This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Terminology and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Quick-Start Operating Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Items Required for Basic Installation and Conguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Transducer Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Transducer Connections for Remote Mount Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Initial Settings and Powerup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Pipe Preparation and Transducer Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Application Versatility. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Temperature Ratings for Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
User Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Data Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Product Identication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Transmitter Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Transmitter Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Enclosures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Transducer Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
DC Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Transducer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mounting Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Select a Mounting Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Select a Mounting Conguration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Enter the Pipe and Liquid Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Mount the Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Transducer Mounting Congurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Inputs/Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Standard 4…20 mA Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Totalizer Output Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
User Manual
Page iiiApril 2016 TTM-UM-00369-EN-03

Frequency Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Parameter Conguration Using UltraLink Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
System Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Additional Parts Required for Conguration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Conguration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Basic Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Filtering Tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Remove the Zero Oset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Select Flow Rate Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Set Multiple Flow Rates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Target Dbg Data Screen Denitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Saving the Conguration on a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Printing a Conguration Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
UltraLink Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
KFactors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Calculating KFactors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
North American Pipe Schedules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Fluid Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Brad Harrison® Connector Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Control Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Software Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Part Number Construction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Transit Time Ultrasonic Flow Meters, TFXL Meter
Page iv April 2016TTM-UM-00369-EN-03

SCOPE OF THIS MANUAL
This manual is divided into two main sections:
• “Quick-Start Operating Overview” on page 6 is intended to help you get the TFXL meter up and running quickly. Refer to
the detailed instructions if you require additional information.
• The remaining chapters provide a detailed description of all software settings and hardware installation guidance.
Read this manual carefully before attempting any installation or operation. Keep the manual accessible for future reference.
UNPACKING AND INSPECTION
Upon opening the shipping container, visually inspect the product and applicable accessories for any physical damage such
as scratches, loose or broken parts, or any other sign of damage that may have occurred during shipment.
OTE:NIf damage is found, request an inspection by the carrier’s agent within 48 hours of delivery and file a claim with the
carrier. A claim for equipment damage in transit is the sole responsibility of the purchaser.
SAFETY
Terminology and Symbols
Indicates a hazardous situation, which, if not avoided, is estimated to be capable of causing death or serious
personal injury.
Indicates a hazardous situation, which, if not avoided, could result in severe personal injury or death.
Indicates a hazardous situation, which, if not avoided, is estimated to be capable of causing minor or moderate
personal injury or damage to property.
Considerations
The installation of the TFXL meter must comply with all applicable federal, state, and local rules, regulations, and codes.
EXPLOSION HAZARD - SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 2.
AVERTISSMENT
RISQUE D’EXPLOSION - LA SUBSTITUTION DE COMPOSANTS PEUT RENDRE CEMATÉRIEL INACCCEPTABLE POUR LES
EMPLACEMENTS DE CLASSE I, DIVISION 2.
DO NOT CONNECT OR DISCONNECT EITHER POWER OR OUTPUTS UNLESS THE AREA IS KNOWN TO BE NON-
HAZARDOUS.
AVERTISSMENT
RISQUE D’EXPLOSION. NE PAS DÉBRANCHER TANT QUE LE CIRCUIT EST SOUSTENSION, À MOINS QU’LL NE S’AGISSE
D’UN EMPLACEMENT NON DANGEREUX.
MPORTANTI
Not following instructions properly may impair safety of equipment and/or personnel.
MPORTANTI
Must be operated by a Class 2 supply suitable for the location.
Scope of This Manual
Page 5April 2016 TTM-UM-00369-EN-03

QUICKSTART OPERATING OVERVIEW
Items Required for Basic Installation and Conguration
• TFXL meter
• Transducers (remote-mount versions sold separately)
• Mounting straps (for remote-mount versions)
• Acoustic couplant
• Power source
• UltraLink Software
• Configuration cable kit P/N D010-0204-001
• USB-to-DB9 converter P/N D005-2116-004 (required if PC does not have a serial port)
• Phillips screwdriver
• Flathead screwdrivers (large and small)
• 5/16 in. nut driver (optional, for remote-mount versions)
Follow these instructions to get the system up and running quickly. Refer to the detailed instructions if you require
additional information.
OTE:NThe following steps require information supplied by the transmitter itself so it will be necessary to supply power
to the transmitter, at least temporarily, and connect to a computer using the UltraLink® software utility to obtain
setup information.
Transducer Location
1. Select a mounting location on the piping system with a minimum of ten pipe diameters (10 × the pipe inside diameter) of
straight pipe upstream and ve straight diameters downstream. See Table 1 on page 16 for additional congurations.
2. If the application requires DTTR, DTTN or DTTH transducers, select a mounting method for the transducers based on pipe
size and liquid characteristics. See Table 2 on page 17. The three transducer mounting congurations are shown in
Figure 4. See “Transducer Mounting Configurations” on page 20 for mounting procedures.
OTE:NAll DTTS and DTTC transducers use V–Mount configuration.
Power Connections
Power for the TFXL meter flow meter is obtained from a direct current (DC) power source.
1. Verify that the power source is capable of supplying 12…28V DC at a minimum of 250 milliamps.
2. With the power from the DC power source disabled or disconnected, connect the positive supply wire and ground to the
appropriate eld wiring terminals in the ow meter. See Figure 1. A wiring diagram decal is on the inside cover of the ow
meter enclosure.
O
N
1
PIC16F628
DC Ground
12 . . . 28V DC
DC Ground
12 . . . 28V DC
Figure 1: Power connections
Transducer Connections for Remote Mount Transducers
1. Guide the transducer terminations through the transmitter conduit hole on the bottom-left of the enclosure using a
sealed cord grip or NEMA 4 conduit connection. Secure the transducer cable with the supplied conduit nut (if exible
conduit was ordered with the transducer).
Quick-Start Operating Overview
Page 6 April 2016TTM-UM-00369-EN-03

2. The remote mount transducers use an add-in connection board on the left side of the meter below the LCD (TFXL meter
2 version). The terminals within the TFXL meter are of a screw-down barrier terminal type. Connect the wires at the
corresponding screw terminals in the transmitter. Observe upstream and downstream orientation and wire polarity. See
Figure 2 on page 7.
Upstream
Transducer
Downstream
Transducer
Up Blue/Red
Up White/Black
Down Blue/Red
Down White/Black
Figure 2: Remote mount connections
Initial Settings and Powerup
1. Apply power to the transmitter.
2. Enter the following data into the TFXL meter via the UltraLink software utility.
1Transducer mounting method 7Pipe liner thickness
2Pipe O.D. (Outside Diameter) 8Pipe liner material
3Pipe wall thickness 9Fluid type
4Pipe material 10 Fluid sound speed*
5Pipe sound speed* 11 Fluid viscosity*
6Pipe relative roughness* 12 Fluid specific gravity*
OTE:N* Nominal values for these parameters are included within the transmitter operating system. The nominal values may
be used as they appear or may be modified if the exact system values are known.
3. Record the value calculated and displayed as transducer spacing.
Quick-Start Operating Overview
Page 7April 2016 TTM-UM-00369-EN-03

Pipe Preparation and Transducer Mounting
DTTR, DTTN and DTTH Transducers
1. Place the transmitter in signal strength measuring mode.
This value is available on the data display of the UltraLink
software utility.
2. The pipe surface, where the transducers are to be mounted,
must be clean and dry. Remove scale, rust or loose paint to
ensure satisfactory acoustic conduction. Wire brushing the
rough surfaces of pipes to smooth bare metal may also be
useful. Plastic pipes do not require preparation other than
cleaning. On horizontal pipe, choose a mounting location within
approximately 45 degrees of the side of the pipe.
See Figure 5 on page 9. Locate the ow meter so that the
pipe will be completely full of liquid when ow is occurring in
the pipe. Avoid mounting on vertical pipes where the ow is
moving in a downward direction.
3. Apply a single 1/2 inch (12 mm) bead of acoustic couplant
grease to the upstream transducer and secure it to the pipe with
a mounting strap.
Downstream+
Downstream-
Upstream-
Upstream+
Figure 3: Transducer connections
4. Apply acoustic couplant grease to the downstream transducer and press it onto the pipe using hand pressure at the lineal
distance calculated in “Transducer Location” on page 6.
5. Space the transducers according to the recommended values from the UltraLink software utility. Secure the transducers
with the mounting straps at these locations.
TOP VIEW
OF PIPE
W-Mount V-Mount Z-Mount
TOP VIEW
OF PIPE
TOP VIEW
OF PIPE
Figure 4: Transducer mounting configurations
Quick-Start Operating Overview
Page 8 April 2016TTM-UM-00369-EN-03

45°
45°
YES
YES
45°
45°
FLOW METER
MOUNTING ORIENTATION
DTTS and DTTC TRANSDUCERS
TOP OF
PIPE
45°
45°
YES
YES
45°
45°
FLOW METER
MOUNTING ORIENTATION
DTTR, DTTN and DTTH TRANSDUCERS
TOP OF
PIPE
45°
45°
YES
YES
45°
45°
FLOW METER
MOUNTING ORIENTATION
2” DTTS and DTTC TRANSDUCERS
TOP OF
PIPE
Figure 5: Transducer mounting orientations
DTTS and DTTC Transducers
1. Place the transmitter in signal strength measuring mode. This value is available on the transmitter’s display Service Menu or
in the data display of the UltraLink software utility.
2. The pipe surface, where the transducers are to be mounted, must be clean and dry. Remove scale, rust or loose paint
to provide satisfactory acoustic conduction. Wire brushing the rough surfaces of pipes to smooth bare metal may also
be useful. Plastic pipes do not require preparation other than cleaning. On horizontal pipe, choose a mounting location
within approximately 45 degrees of the side of the pipe. See Figure 5 on page 9.Locate the ow meter so that the pipe
will be completely full of liquid when ow is occurring in the pipe. Avoid mounting on vertical pipes where the ow is
moving in a downward direction.
3. Apply a single 1/2 inch (12 mm) bead of acoustic couplant grease to the top half of the transducer and secure it to the pipe
with the bottom half or with U-bolts.
4. Tighten the nuts so the acoustic coupling grease begins to ow out from the edges of the transducer and from the gap
between the transducer and the pipe.
MPORTANTI
Do not overtighten. Overtightening will not improve performance and may damage the transducer.
5. Verify that signal strength is greater than 5.0.
6. Input the units of measure and the I/O data.
Quick-Start Operating Overview
Page 9April 2016 TTM-UM-00369-EN-03

INTRODUCTION
The TFXL meter is designed to measure the fluid velocity of liquid within a closed conduit. The transducers are a non-
contacting, clamp-on or clamp-around type, which provide the benefits of non-fouling operation and ease of installation.
The TFXL meter family of transit time transmitters uses two transducers that function as both ultrasonic transmitters and
receivers. The transducers are clamped on the outside of a closed pipe at a specific distance from each other.
The transducers can be mounted in V-Mount where the sound transverses the pipe two times, W-Mount where the sound
transverses the pipe four times, or in Z-Mount where the sound crosses the pipe once. The selection of how transducers are
mounted on opposite sides of the pipe and method is based on pipe and liquid characteristics, which both have an effect
on how much signal is generated. The flow meter operates by alternately transmitting and receiving a frequency modulated
burst of sound energy between the two transducers and measuring the time interval that it takes for sound to travel between
the two transducers. The difference in the time interval measured is directly related to the velocity of the liquid in the pipe.
Application Versatility
The TFXL meter can be successfully applied on a wide range of metering applications. The simple-to-program transmitter
allows the standard product to be used on pipe sizes ranging from 1/2 …100 inches (12…2540 mm)*. A variety of liquid
applications can be accommodated:
ultrapure liquids cooling water potable water river water chemicals
plant effluent sewage reclaimed water others
Because the transducers are non-contacting and have no moving parts, the transmitter is not affected by system pressure,
fouling or wear.
Temperature Ratings for Transducers
Because the transducers are non-contacting and have no moving parts, the flow meter is not affected by system pressure,
fouling or wear. Temperature ratings for each transducer are listed below.
Transducer Temperature Rating
DTTR –40…250° F (–40…121° C)
DTTC –40…194° F (–40…90° C)
DTTN –40…194° F (–40…90° C)
DTTH –40…350° F (–40…177° C)
DTTS –40…140° F (–40…60° C)
User Safety
The TFXL meter uses a low voltage DC power source that provides electrical safety for the user. Remove the cover to access to
the meter connections and the computer interface connection.
DANGER
THE POWER SUPPLY BOARD CAN HAVE LINE VOLTAGES APPLIED TO IT, SO DISCONNECT ELECTRICAL POWER BEFORE
OPENING THE INSTRUMENT ENCLOSURE. WIRING SHOULD ALWAYS CONFORM TO LOCAL CODES AND THE NATIONAL
ELECTRICAL CODE.
Data Integrity
Non-volatile flash memory retains all user-entered configuration values in memory indefinitely, even if power is lost or
turned off.
Product Identication
The serial number and complete model number of the transmitter are located on the top outside surface of the transmitter
body. Should technical assistance be required, please provide our customer service department with this information.
Introduction
Page 10 April 2016TTM-UM-00369-EN-03

TRANSMITTER INSTALLATION
Transmitter Location
Mount the enclosure in an area that is convenient for servicing, calibration and observation of the LCD readout (if equipped).
1. Locate the transmitter within the length of transducer cables supplied. If this is not possible, exchange the cable for one
that is of proper length.
2. Mount the transmitter in a location:
◊ Where little vibration exists.
◊ That is protected from corrosive fluids.
◊ That is within the transmitters ambient temperature limits –40…185° F (–40…85° C).
◊ That is out of direct sunlight. Direct sunlight may increase transmitter temperature to above the maximum limit.
3. See Figure 6 for enclosure and mounting dimension details. Allow enough room for door swing, maintenance and conduit
entrances.
4. Secure the enclosure to a at surface with two fasteners.
5. Feed the cables through the conduit holes in the enclosure. Use plugs to seal any unused holes.
OTE:NUse NEMA 4 (IP-65) rated fittings/plugs to maintain the watertight integrity of the enclosure. Generally, the right side
conduit hole (viewed from front) is used for power, the bottom conduit hole(s) for transducer connections.
Enclosures
Integral Enclosure Remote Enclosure
A
B
C
D
Conduit hole
E
F
C
D
Conduit hole
F
A
B
K
E
G
H
I
J
Two mounting holes
A
in. (mm)
B
in. (mm)
C
in. (mm)
D DIA
in. (mm)
E
in. (mm)
F
in. (mm)
G
in. (mm)
H
in. (mm)
I
in. (mm)
J DIA
in. (mm)
K
in. (mm)
6.72 (170.7) 3.17 (80.5) 2.57 (65.3) 0.87 (22.2) 1.33 (33.8) 0.85 (21.6) 0.77 (19.6) 1.78 (45.2) 3.74 (95) 0.22 (5.6) 7.01 (178)
Figure 6: Transmitter enclosure dimensions
The remote mount transmitter has three conduit holes in the flow meter enclosure that should be suitable for most
installations. Use a sealed cord grip or NEMA 4 conduit connection to retain the NEMA 3 integrity of the flow meter enclosure.
Failure to do so will void the manufacturer’s warranty and can lead to product failure.
The TFXL meter is housed in an insulating plastic enclosure that does not provide continuity of bonding between field wiring
conduit and the TFXL meter chassis or other conduits connected to the enclosure.
Transmitter Installation
Page 11April 2016 TTM-UM-00369-EN-03

Wiring methods and practices are to be made in accordance with the NEC (National Electrical Code®) and/or other local
ordinances that may be in effect. Consult the local electrical inspector for information regarding wiring regulations.
When making connections to the field wiring terminals inside the flow meter, strip back the wire insulation approximately
0.25 inch (6 mm). Stripping back too little may cause the terminals to clamp on the insulation and not make good contact.
Stripping back too much insulation may lead to a situation where the wires could short together between adjacent terminals.
Wires should be secured in the field wiring terminals using a screw torque of 0.5…0.6 Nm.
If using the DC ground terminal as a protective conductor terminal, apply the protective conductor first and secure it
independently of other connections. Connect the protective conductor so it is unlikely to be removed by servicing that does
not involve the protective conductor or post a warning requiring the replacement of the protective conductor after removal.
Power the TFXL meter with a Class 2 direct current (DC) power source. The power source should be capable of supplying
12…28V DC at a minimum of 250 milliamps. With the power from the DC power source disabled or disconnected, connect the
positive supply wire and ground to the field wiring terminals in the flow meter. See Figure 8 on page 14. A wiring diagram is
on the inside cover of the meter enclosure.
MPORTANTI
• FOLLOW INSTRUCTIONS TO PROVIDE SAFETY OF EQUIPMENT AND/OR PERSONNEL.
• MUST BE OPERATED BY A POWER SUPPLY SUITABLE FOR THE LOCATION.
• DO NOT CONNECT OR DISCONNECT EITHER POWER OR OUTPUTS UNLESS THE AREA IS KNOWN
TO BE NONHAZARDOUS.
• DO NOT CONNECT THE INTERFACE CABLE BETWEEN A TFXL METER AND A PERSONAL COMPUTER UNLESS THE AREA IS
KNOWN TO BE NONHAZARDOUS.
Transmitter Installation
Page 12 April 2016TTM-UM-00369-EN-03

TRANSDUCER CONNECTIONS
Figure 7: Transducer connections
To access terminal strips for wiring, first loosen the four screws holding the top of the case to the bottom.
OTE:NThe four screws are“captive”screws and cannot be removed from the top of the case.
If the unit has a display, remove the four Phillips head screws that hold the display to the main circuit board and carefully
move it out of the way. Do not over stress the ribbon cable located between the display and the microprocessor
circuit boards.
Guide the transducer terminations through the transmitter conduit hole located in the bottom-left of the enclosure. Secure
the transducer cable with the supplied conduit nut (if flexible conduit was ordered with the transducer).
OTE:NTFXL meter with integral transducers have the transducers connected at the factory and the transducer connections
section can be skipped.
The terminals within the TFXL meter are of a screw-down barrier terminal type. Depending on the type of transducers being
used there are two terminal strip arrangements possible.
Remote mount small pipe transducers are connected to the terminals on the main circuit board.
Remote mount transducers are connected to a daughter board on the left side of the meter.
Connect the appropriate wires at the corresponding screw terminals in the transmitter. Observe upstream and downstream
orientation and wire polarity. See Figure 7.
OTE:NHigh temperature transducer cables come with red and black wire colors. For the red and black combination, the red
wire is positive (+) and the black wire is negative (–).
OTE:NThe transducer cable carries low level, high frequency signals. In general, it is not recommended to add additional
length to the cable supplied with the transducers. If additional cable is required, contact the factory to arrange an
exchange for a transducer with the appropriate length of cable. Cables 100…990 feet (30…300 meters) are available
with RG59 75 Ohm coaxial cable.
Transducer Connections
Page 13April 2016 TTM-UM-00369-EN-03

DC POWER CONNECTIONS
The TFXL meter should be operated from an 12…28V DC Class 2 power source capable of supplying a minimum of 250 mA
of current.
1. Feed the power source through the conduit hole on the right side of the enclosure. Connect power to the screw terminal
block in the TFXL meter. Use wiring practices that conform to local and national codes.
2. Connect the DC power to 12…28V DC In and DC Gnd., as in Figure 8.
DC Ground
11 - 28 VDC
DC Ground
11 - 28 VDC
Figure 8: DC power connections
3. Connect an 12…28V DC Class 2 power source as illustrated in Figure 8. TFXL meter terminal blocks can accommodate wire
up to 14 AWG.
4. Connect a switch or circuit breaker in close proximity of the TFXL meter and within easy reach of the operator.
5. Mark the switch or circuit breaker as the disconnect device for the TFXL meter.
DC Power Connections
Page 14 April 2016TTM-UM-00369-EN-03

TRANSDUCER INSTALLATION
The transducers for the TFXL meter contain piezoelectric crystals that transmit and receive ultrasonic signals through the
walls of liquid piping systems.
DTTR, DTTN and DTTH transducers are relatively simple and straightforward to install, but spacing and alignment of the
transducers is critical to the system’s accuracy and performance. CAREFULLY EXECUTE THESE INSTRUCTIONS.
DTTS and DTTC small pipe transducers have integrated transmitter and receiver elements that eliminate the requirement for
spacing measurement and alignment.
Mounting the DTTR, DTTN and DTTH clamp-on ultrasonic transit time transducers takes four steps:
1. Select the optimum location on a piping system.
2. Select a mounting conguration.
3. Enter the pipe and liquid parameters into the UltraLink software utility or key them into the transmitter. The UltraLink
software utility or the transmitter’s rmware calculates proper transducer spacing based on these entries.
4. Prepare the pipe and mount the transducers.
Mounting Location
Select a Mounting Location
The first step in the installation process is the selection of an optimum location for the flow measurement to be made. For this
to be done effectively, a basic knowledge of the piping system and its plumbing are required.
An optimum location is defined as:
• A piping system that is completely full of liquid when measurements are being taken. If the pipe may become completely
empty during a process cycle an error code 0010 (Low Signal Strength) will be displayed on the transmitter while the
pipe is empty. This error code will clear automatically once the pipe refills with liquid. Do not mount the transducers in an
area where the pipe may become partially filled, such as the highest point in a flow loop. Partially filled pipes will cause
erroneous and unpredictable operation of the transmitter.
• A piping system that contains lengths of straight pipe such as those described in Table 1 on page 16. The optimum
straight pipe diameter recommendations apply to pipes in both horizontal and vertical orientation. The straight runs in
Table 1 apply to liquid velocities that are nominally 7 fps (2.2 mps). As liquid velocity increases above this nominal rate, the
requirement for straight pipe increases proportionally.
• An area where the transducers will not be inadvertently bumped or disturbed during normal operation.
• NOT on downward flowing pipes unless adequate downstream head pressure is present to overcome partial filling of or
cavitation in the pipe.
Transducer Installation
Page 15April 2016 TTM-UM-00369-EN-03

***
Flow
***
Flow
***
Flow
***
Flow
Flow
***
Flow
***
24
24
14
10
10
10
5
5
5
5
5
5
* **
Upstream
Pipe
Diameters
Downstream
Pipe
Diameters
Piping Conguration
and Transducer Positioning
Table 1: Piping configuration and transducer positioning
The TFXL meter will provide repeatable measurements on piping systems that do not meet these pipe diameter requirements,
but the accuracy of the readings may be influenced.
Transducer Installation
Page 16 April 2016TTM-UM-00369-EN-03

Select a Mounting Conguration
The transmitter can be used with five transducer types: DTTR, DTTN, DTTH, DTTS and DTTC. Meters that use the DTTR, DTTN
or DTTH transducer sets consist of two separate sensors that function as both ultrasonic transmitters and receivers. These
transducers are clamped on the outside of a closed pipe at a specific distance from each other. DTTS and DTTC transducers
integrate both the transmitter and receiver into one assembly that fixes the separation of the piezoelectric crystals.
The DTTR, DTTN and DTTH transducers can be mounted in:
• W-Mount where the sound traverses the pipe four times. This mounting method produces the best relative travel time
values but the weakest signal strength.
• V-Mount where the sound traverses the pipe twice. V-Mount is a compromise between travel time and signal strength.
• Z-Mount where the transducers are mounted on opposite sides of the pipe and the sound crosses the pipe once. Z-Mount
will yield the best signal strength but the smallest relative travel time.
Transducer Mounting
Configuration Pipe Material Pipe Size Liquid Composition
W-Mount
Plastic (all types)
2…4 in. (50…100 mm)
Low TSS (Total Suspended
Solids); non-aerated
Carbon Steel
Stainless Steel
Copper
Ductile Iron Not recommended
Cast Iron
V-Mount
Plastic (all types)
4…12 in. (100…300 mm)
Carbon Steel
Stainless Steel
Copper 4…30 in. (100…750 mm)
Ductile Iron 2…12 in. (50…300 mm)
Cast Iron
Z-Mount
Plastic (all types) > 30 in. (> 750 mm)
Carbon Steel > 12 in. (> 300 mm)
Stainless Steel
Copper > 30 in. (> 750 mm)
Ductile Iron > 12 in. (> 300 mm)
Cast Iron
Table 2: Transducer mounting modes for DTTR, DTTN and DTTH
The transducers can be mounted in V-Mount where the sound transverses the pipe two times, W-Mount where the sound
transverses the pipe four times, or in Z-Mount where the transducers are mounted on opposite sides of the pipe and the
sound crosses the pipe once. The selection of mounting method is based on pipe and liquid characteristics which both have
an effect on how much signal is generated. The transmitter operates by alternately transmitting and receiving a frequency
modulated burst of sound energy between the two transducers and measuring the time interval that it takes for sound to
travel between the two transducers. The difference in the time interval measured is directly related to the velocity of the liquid
in the pipe.
The appropriate mounting configuration is based on pipe and liquid characteristics. Selecting the proper transducer
mounting method is an iterative process. Table 2 contains recommended mounting configurations for common applications.
These recommended configurations may need to be modified for specific applications if such things as aeration, suspended
solids, out-of-round piping or poor piping conditions are present.
TOP VIEW
OF PIPE
W-Mount V-Mount Z-Mount
TOP VIEW
OF PIPE
TOP VIEW
OF PIPE
Figure 9: Transducer mounting modes for DTTR, DTTN and DTTH
Transducer Installation
Page 17April 2016 TTM-UM-00369-EN-03

45°
45°
YES
YES
45°
45°
Z-Mount
Flow Meter
Mounting Orientation
DTTR, DTTN and DTTH Transducers
Top of
Pipe
YES
45°
45°
YES
45°
45°
Flow Meter
Mounting Orientation
2” DTTS and DTTC Transducers
Top of
Pipe
45°
45°
YES
YES
45°
45°
Flow Meter
Mounting Orientation
DTTS and DTTC Transducers
Top of
Pipe
45°
45°
YES
W and V Mount
Top of
Pipe
Figure 10: Transducer orientation for horizontal pipes
For DTTS and DTTC transducers, the transducers are V-mount. The frequency setting depends on the pipe material.
Pipe Size Frequency
Setting Transducer Pipe Mounting
Configuration
1/2 in. 2 MHz
DTTSnP ANSI
V
DTTSnC Copper
DTTSnT Stainless Steel
3/4 in. 2 MHz
DTTSnP ANSI
DTTSnC Copper
DTTSnT Stainless Steel
1 in. 2 MHz
DTTSnP ANSI
DTTSnC Copper
DTTSnT Stainless Steel
1-1/4 in. 2 MHz
DTTSnP ANSI
DTTSnC Copper
DTTSnT Stainless Steel
1-1/2 in. 2 MHz
DTTSnP ANSI
DTTSnC Copper
DTTSnT Stainless Steel
2 in. 1 MHz DTTSnP ANSI
DTTSnC Copper
2 MHz DTTSnT Stainless Steel
DTTS transducer designation refers to both DTTS and DTTC transducer types.
Table 3: Transducer mounting modes for DTTS / DTTC
Transducer Installation
Page 18 April 2016TTM-UM-00369-EN-03

Enter the Pipe and Liquid Parameters
The TFXL meter calculates proper transducer spacing based on the piping and liquid information you enter into the
transmitter via the UltraLink software utility. See "Parameter Configuration Using UltraLink Software" on page 29.
The most accuracy is achieved when the transducer spacing is exactly what the transmitter calculates, so use the calculated
spacing if the signal strength is satisfactory. If the pipe is not round, the wall thickness not correct or the actual liquid being
measured has a different sound speed than the liquid programmed into the transmitter, the spacing can vary from the
calculated value. In that case, place the transducers at the highest signal level observed when moving the transducers slowly
around the mount area.
OTE:NTransducer spacing is calculated on“ideal” pipe. Ideal pipe almost never exists, so you may need to alter the
transducer spacing. An effective way to maximize signal strength is to configure the display to show signal strength,
fix one transducer on the pipe and then—starting at the calculated spacing—move the remaining transducer small
distances forward and back to find the maximum signal strength point.
MPORTANTI
Enter all of the data on this list, save the data and reset the transmitter before mounting the transducers.
The following information is required before programming the instrument:
Transducer mounting configuration Pipe liner thickness (if present) Pipe O.D. (outside diameter) Pipe liner material (if present)
Pipe wall thickness Fluid type Pipe material Fluid sound speed1
Pipe sound speed1Fluid viscosity1Pipe relative roughness1Fluid specific gravity1
Table 4: Parameters required
1Nominal values for these parameters are included within the transmitter’s operating system. The nominal values may be used as they appear or may be modied if exact system
values are known.
OTE:NMuch of the data relating to material sound speed, viscosity and specific gravity is pre-programmed into the
transmitter. You need to modify this data only if you know that a particular application’s data varies from the
reference values. See "Parameter Configuration Using UltraLink Software" on page 29 for instructions on entering
configuration data into the transmitter via the software utility.
After entering the data listed above, the transmitter will calculate proper transducer spacing for the particular data set. The
distance will be in inches if the transmitter is configured in English units, or millimeters if configured in metric units.
Mount the Transducer
After selecting an optimal mounting location and determining the proper transducer spacing, mount the transducers onto
the pipe.
1. Clean the surface of the pipe. If the pipe has external corrosion or dirt, wire brush, sand or grind the mounting location
until it is smooth and clean. Paint and other coatings, if not flaked or bubbled, need not be removed. Plastic pipes typically
do not require surface preparation other than soap and water cleaning.
2. Orient and space the DTTR, DTTN and DTTH transducers on the pipe to provide optimum reliability and performance. On
horizontal pipes, when Z-Mount is required, mount the transducers 180 radial degrees from one another and at least 45
degrees from the top-dead-center and bottom-dead-center of the pipe. See Figure 10 on page 18. On vertical pipes, the
orientation is not critical.
The spacing between the transducers is measured between the two spacing marks on the sides of the transducers. These
marks are approximately 0.75 inches (19 mm) back from the nose of the DTTR, DTTN and DTTH transducers. See Figure 11.
Mount DTTS and DTTC transducers with the cable exiting within ±45 degrees of the side of a horizontal pipe. On vertical
pipes, the orientation does not apply.
Alignment
Marks
Figure 11: Transducer alignment marks
Transducer Installation
Page 19April 2016 TTM-UM-00369-EN-03

Transducer Mounting Congurations
V-Mount and W-Mount Configurations
Apply the Couplant
For DTTR, DTTN and DTTH transducers, place a single bead of couplant, approximately 1/2 inch (12 mm) thick, on the flat
face of the transducer. See Figure 12. Generally, a silicone-based grease is used as an acoustic couplant, but any good quality
grease-like substance that is rated to not ow at the operating temperature of the pipe is acceptable. For pipe surface
temperature over 150° F (65° C), use Sonotemp® (P.N. D002-2011-011).
½ in.
(12 mm)
Figure 12: Application of couplant
Position and Secure the Transducer
1. Place the upstream transducer in position and secure with a mounting strap. Place the straps in the arched groove on the
end of the transducer. Use the screw provided to help hold the transducer onto the strap. Verify that the transducer is true
to the pipe and adjust as necessary. Tighten the transducer strap securely.
2. Place the downstream transducer on the pipe at the calculated transducer spacing. See Figure 13 on page 20. Apply rm
hand pressure. If signal strength is greater than ve, secure the transducer at this location. If the signal strength is not ve
or greater, using rm hand pressure slowly move the transducer both towards and away from the upstream transducer
while observing signal strength.
Signal strength can be displayed on the transmitter’s display or on the main data screen in the UltraLink software utility.
See "Parameter Configuration Using UltraLink Software" on page 29. Clamp the transducer at the position where the
highest signal strength is observed. The factory default signal strength setting is ve. However, there are many application-
specic conditions that may prevent the signal strength from attaining this level. Signal levels less than ve will probably
not be acceptable for reliable readings.
OTE:NSignal strength readings update only every few second. Move the transducer 1/8 inch then wait to see if the signal is
increasing or decreasing. Repeat until the highest level is achieved.
3. If, after adjusting the transducers, the signal strength does not rise to above ve, use an alternate transducer mounting
conguration. If the mounting conguration was W-Mount, re-congure the transmitter for V-Mount, move the
downstream transducer to the new spacing distance and repeat the procedure "Mount the Transducer" on page 19.
Transducer
Spacing
OTE:NMounting the high temperature transducers is similar
to mounting the DTTR/DTTN transducers. High
temperature installations require acoustic couplant
that is rated not to flow at the operating temperature
of the pipe surface.
Figure 13: Transducer positioning
Transducer Installation
Page 20 April 2016TTM-UM-00369-EN-03
Table of contents
Other dynasonics Measuring Instrument manuals

dynasonics
dynasonics UFX User manual

dynasonics
dynasonics TFXL User manual

dynasonics
dynasonics UFX User manual

dynasonics
dynasonics Fusion User manual

dynasonics
dynasonics 901 Series Manual

dynasonics
dynasonics DXN Series User manual

dynasonics
dynasonics TFXD Series Manual

dynasonics
dynasonics TFX Ultra User manual

dynasonics
dynasonics Badger Meter IS-6000 Doppler User manual

dynasonics
dynasonics 902 Series User manual