manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Eastern Energy
  6. •
  7. Gas Detector
  8. •
  9. Eastern Energy 7721 User manual

Eastern Energy 7721 User manual

This manual suits for next models

3

Other Eastern Energy Gas Detector manuals

Eastern Energy JL269 LED User manual

Eastern Energy

Eastern Energy JL269 LED User manual

Eastern Energy 7201 User manual

Eastern Energy

Eastern Energy 7201 User manual

Popular Gas Detector manuals by other brands

Macurco GD-6 Operation manual

Macurco

Macurco GD-6 Operation manual

Evikon E2632 user manual

Evikon

Evikon E2632 user manual

Perry 1GA 50917MET/P Installation and operating instructions

Perry

Perry 1GA 50917MET/P Installation and operating instructions

MSA altair 5 operating manual

MSA

MSA altair 5 operating manual

DOD Technologies ChemLogic CL96 operating manual

DOD Technologies

DOD Technologies ChemLogic CL96 operating manual

DEGA NS III LCD Series instruction manual

DEGA

DEGA NS III LCD Series instruction manual

RKI Instruments 35-3001A-08-01 Operator's manual

RKI Instruments

RKI Instruments 35-3001A-08-01 Operator's manual

isweek AS8900 manual

isweek

isweek AS8900 manual

Beacon MEGA Installation, operation and maintenance instructions

Beacon

Beacon MEGA Installation, operation and maintenance instructions

MSA ULTIMA X5000 operating manual

MSA

MSA ULTIMA X5000 operating manual

Critical Environment Technologies CGAS-A Series Installation & operation manual

Critical Environment Technologies

Critical Environment Technologies CGAS-A Series Installation & operation manual

Sensidyne SensAlert Plus instruction sheet

Sensidyne

Sensidyne SensAlert Plus instruction sheet

Gas Sense GS-300.A Operation manual

Gas Sense

Gas Sense GS-300.A Operation manual

Evikon PluraSens E2638 Series user manual

Evikon

Evikon PluraSens E2638 Series user manual

Bosean Electronic Technology BH-4M user manual

Bosean Electronic Technology

Bosean Electronic Technology BH-4M user manual

Coster RGS 328 instructions

Coster

Coster RGS 328 instructions

Sygonix 2483186 operating instructions

Sygonix

Sygonix 2483186 operating instructions

Compur Monitors Tracer manual

Compur Monitors

Compur Monitors Tracer manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

OPERATION MANUAL
WALL-MOUNT CO MONITOR
2
Model:
7722
77232
www.ete.co.th
EASTERN ENERGY CO.,LTD. (HEAD OFFICE)
40/4 Vitoondumri Rd., Banbueng, Banbueng,
Contact :
Tel: 0-3844-6117 Fax: 0-3844-6200
Email: [email protected]
Banbueng, Chonburi 20170
INTRODUCTION
Features:
Triple displays of level, temp.
and humidity (7722/77232)
s
calibration
PC connect
CO2
Stable NDIR sensor for CO detection
2
Statistics of weighted average
TWA (8 hours weighted average)
STEL(15 minutes weighted average)
Visible and audile CO warning alarm
2
Alarm output for ventilation control
ABC(Automatic Baseline Calibration)
and manual CO
2
via RS232 interface
1
Thank you for purchasing this wall mo unt
CO monitor. It could me asure the CO
2 2
level, air temp., dew point, wet bulb temp.
and humi dity (DP, WB , RH are for mo dels
7722/77232) and is an ideal instrume nt
for indoor air quality (IAQ) diagnosis.
Poor indoor air quality is considered
unhealthy because it causes tiredness,
loss of ability to concentrate, and even
illness(ex. Sick Building Syndrome ). IAQ
monitoring and survey, especially on CO
2
level and air ventilation become widely
applied in public areas such as offices,
classrooms, factories, hospitals and
hotels. It is also suggested in regulations
of industrial hygiene in some countries.
(See appendix)
With NDIR (non-dispersive infrared)
sensor used, this CO mo nitor is stable
2
in long term mo nitoring. And the built-in
alarm output is especially helpful in
ventilation control and HVAC system
performance verification.
Meter
Adaptor
Operation manual
Plain white box
This package contains:
Optional accessory:
33% calibration salt (VZ0033AZ1)
75% calibration salt (VZ0075AZ1)
RS232 cable and software
The meter is powered by an AC adaptor
(12V/1A output).
2
TWA Time weighted average(
STEL hort-term exposure limit
(15 minutes weighted
DP Dew point temp. (7722,77232)
AIR Air temp
WBT Wet bulb temp. (7722,77232)
% Unit of relative humidity
oE (C/F) Celsius/Fahrenheit
AVG/CO/
ALM/HLD
MAX/MIN Maximun/Minimum readings
8 hours)
S
average)
.
/ Vain icons in these models
%%
OO
ALM
HLD
CO2
CO2 con-
centration
in ppm
Relative
humidity in %
Air temp.
Dew point
Wet bulb
temp. in
oC or F
o
Symbols
MAX MIN AVG TWA STEL PPM
CO
MATERIAL SUPPLIED
LCD DISPLY
POWER SUPPLY
3
Enter setup mode.
Save and finish settings.
Enter CO calibration with .
2
Enter RH calibration with .
Exit setup page/mode.
Reset the meter and clear MAX/MIN.
Terminate during calibration.
Select AIR, DP, WBT temp. display.
(7722/77232 only)
Select mode or increase value in
calibration and setup.
Activate MAX, MIN, STEL,TWA
function. Select mode or decrease
value in calibration and setup.
KEYPAD
SETSET
CAL
ESC
CAL
ESC
DP/WB
M /
X
N
AVG
RESE TRESE T
DP/WB
M /
X
N
AVG
OPERATION
Plug in the adaptor and the meter turns
on automatically with a short beep. It
performs 30 sec. countdown(Fig.1) for
meter warm up, then enters normal
mode with current CO , temp.and
2
humidity(7722/77232) readings
displayed (Fig.2).
Fig. 1
Fig. 2
Model 7721/77231 Model 7722/77232
POWER ON/OFF
OO
CO2
PPM
%%
OO
CO2
PPM
or
4
%%
OO
CO2
%%
OO
CO2
Fig.3 Fig.4
TAKING MEASUREMENT
Press to switch temp. display. The
lower left LCD will cycle through air
temp., Dew point temp. (Fig.3), and wet
bulb temp.(Fig.4).
The meter starts taking measurement
after power on and updates readings
every second. In the condition of opera
-ting environment change (ex. from high
to low temp.), it takes 30 sec to respond
for CO sensor and 30 minutes for RH.
2
AIR(all models),DP,WBT(7722/77232)
DP/WB
PPM PPM
Under normal mode, press key to see
the minimum, maximum, and weighted
average readings. Press key, it
displays MIN, MAX, STEL, TWA in
sequence and returns to normal mode.
Fig.5
M /
X
N
AVG
M /
X
N
AVG
%%
OO
CO2
PPM
MAX,MIN, STEL,TWA
NOTE: Do not hold the meter close to
face in case that exhalation affects
CO levels.
2
In MIN and MAX modes, it shows the
minimum and maximum reading of
on the upper LCD. The reading of
AIR or DP or WB and humidity
on the lower LCD.
(Fig.5)
CO
2
temp.
(7722/77232) shows
In STEL and TWA modes, the upper
LCD shows the
(STEL) and 8 hours(TWA).
weighted average of
CO reading for the past 15 minutes
2
The lower
LCD shows the current AIR, DP/WB
temp. and humidity (7722/77232).
(Fig.6)
5
%%
OO
STEL
CO2
Fig.6
PPM
NOTE:
1.
If the meter is turned on for shorter
than 15 minutes, the STEL value will
be the weighted average of readings
taken since power on. Same for the
TWA value.
2.It takes at least 5 minutes to calculate
the STEL and TWA value. The LCD
shows "----" (Fig.7) during the first 5
minutes from power on.
Fig.7
The meter features visible and audible
alarm to give warnings when CO con-
2
centration exceeds the limit. Users can
set up 2 limits:
1. Upper limit: the alarm threshold
requires air conditioning
2. Lower limit: to stop the alarm.
(See page 7 P1.0 to set the alarm limits).
ALARM & OUTPUT
ALARM
%%
OO
STEL
PPM
CO2
6
LED alarm keeps blinking when beeps
are manually shut. It stops only when
readings fall below the lower limit.
ALARM OUTPUT
The meter is designed with a relay
to send output for further connection.
When CO readings go over the upper
2
limit and cause to alarm. The relay picks
up automatically and send output.
It can be connected to a ventilation
system or activator for conditioning the
air quality. The relay will drop out when
CO readings fall below the lower limit.
2
If the beeper is temporarily shut, it will
sound again when readings fall below
the lower limit and then go over the
upper limit again, or users press key
for more than 1 sec to activate it.
RESE TRESET
Alarm
Output
RS232
Adaptor
It emits beeps (Abt.80dB) with blinking
LED when CO level goes over the upper
2
limit. Beeps can be stopped by pressing
any key or automatically stop when
CO reading falls below the lower limit.
2
DP/WB
M /
X
N
AVG
Press or key to increase the value
or key to decrease. Each press tunes
100ppm and the alarm range is from
100 to 9900ppm.
7
CO2
Fig. 8
SETSET
CO2
PPM
P1.0 CO ALARM: UPPER&LOWER LIMITS
2
Fig. 9
When entering setup mode, P1.0 and
"AL" (Fig.8) are displayed on the LCD.
Press key again to enter P1.1 to set
the CO upper limit. The current setting
2
value will be blinking on the LCD(Fig.9).
CAUTION:
It’s suggested to set up the alarm
value within specification range
that accuracy is ensured. The out
of spec readings are only for
reference and not suitable to use
as alarm limits.
SETUP
SETSET
CAL
ESC
CAL
ESC
Hold key under normal mode for
more than 1 sec to enter setup mode.
To exit setup, press key in P1.0, P3.0,
or P4.0 and it returns to normal mode.
Note:
P2.0 is not applicable in these models
but for future model with CO and CO
2
measurements.
Fig. 11 Fig. 12
OO
Press key in P1.0 to access P3.0 to
set up the temperature scale(Fig.11).
Press key to enter P3.1, current
setting of °C or °F would blinking on the
lower LCD (Fig.12). Press or key to
switch °C or °F. Then press key to
save the setting or without saving
and return to P3.0.
DP/WB
SETSET
DP/WB
M /
X
N
AVG
SETSET
CAL
ESC
CAL
ESC
P3.0 TEMPERATURE SCALE
8
P4.0 ABC SELECTION
Fig. 11
ABC (Automatic Baseline Calibration)
is to implement baseline calibration to
eliminate the zero drift of the infrared
sensor. The ABC function default is on
when turning on the meter. Users can
disable it by following the procedure.
Press in P3.0 or in P1.0 to access
P4.0 to select the ABC function.(Fig.13).
Fig. 10
SETSET
CO2
PPM
SETSET CAL
ESC
CAL
ESC
When the preferred value is set, press
key to enter P1.2 (Fig. 10) to set the
lower limit. After both settings are done,
press key to save or key without
saving and return to P1.0.
DP/WB
M /
X
N
AVG
Fig.13
CO2
9
CO CALIBRATION
2
CAUTION:
1. Do not calibrate the meter in the
air with unknown CO level.
2
Otherwise, it will be taken as 400
ppm and leads to inaccurate
measurements.
2. When operating the meter in the
closed indoor areas that are not
ventilating,
be sure
to disable the ABC function to
prevent incorrect calibration.
such as hospitals or
offices with windows shut,
The meter is calibrated at standard
400ppm CO concentration in factory.
2
It’s suggested to do either ABC or
manual calibration regularly to maintain
the good accuracy.
Fig. 14 Fig. 15
CO2CO2
Press key enter P4.1 with blinking
"en" icon (Fig.14) on the lower LCD. To
disable the ABC function, press or
and "dis" blinks for choice (Fig.15).
After the preferred selection is done,
press key to save the setting or
without saving and return to P4.0.
SETSET
DP/WB
M /
X
N
AVG
SETSET CAL
ESC
CAL
ESC
Note:
When the accuracy becomes a concern
after a long time usage or other special
conditions, return to dealers for standard
calibration.