
GB
INSTRUMENT SAFETY MSA
8 ALTAIR 5 - Operating Manual
•Do not use silicone tubing or sampling lines.
•Wait sufficient time for the reading; response times vary, based on gas and
length of sampling line.
•Do not use the instrument with integral pump for prolonged periods in an
atmosphere containing a concentration of fuel or solvent vapours that may be
greater than 10%LEL.
It is imperative that this operating manual be read and observed when using the
product. In particular, the safety instructions, as well as the information for the use
and operation of the product, must be carefully read and observed. Furthermore,
the national regulations applicable in the user's country must be taken into
account for a safe use.
Danger!
This product is supporting life and health. Inappropriate use,
maintenance or servicing may affect the function of the device and
thereby seriously compromise the user's life.
Before use the product operability must be verified. The product must
not be used if the function test is unsuccessful, it is damaged, a
competent servicing/maintenance has not been made, genuine MSA
spare parts have not been used.
Alternative use, or use outside this specification will be considered as non-
compliance. This also applies especially to unauthorised alterations to the product
and to commissioning work that has not been carried out by MSA or authorised
persons.
1.2. Liability Information
MSA accepts no liability in cases where the product has been used
inappropriately or not as intended. The selection and use of the product are the
exclusive responsibility of the individual operator.
Product liability claims, warranties also as guarantees made by MSA with respect
to the product are voided, if it is not used, serviced or maintained in accordance
with the instructions in this manual.
1.3. Safety and Precautionary Measures to be Adopted
Perform a blocked flow test (see chapter 4.2) before each day's use if equipped
with an integral pump.
MSA recommends to check the instrument's functionality (see chapter 3.6) and
carry out a routine inspection each day before use.
Perform calibration check (see chapter 3.8) before each day's use to verify proper
instrument operation. Adjust calibration if the readings are not within the specified
limits.