Elbur PB 325 User manual

Instruction for Use
ELBUR: PB 325, PB 326,
PB 331, PB 337
ELBUR PB 325
ELBUR PB 326
ELBUR PB 331
ELBUR PB 337


Dear Customer,
Congratulations on having chosen a quality ELBUR Care Bed which will
certainly fully meet your expectations. We would like to take this opportunity to
thank you for showing such confidence and trust in our company and purchasing
one of our products.
Due to designing and manufacturing in accordance to the latest technological
development at the time of delivery, our nursing beds, which are used to alleviate
and compensate for injuries and disabilities, are very durable and ensure essential
performance. They combine modern design and technical accuracy with user-
friendly handling, also meet the varying needs of patients and professional
caregivers.
Based on great experience of Elbur company, we can guarantee high quality
and reliability of our products. We also care about attractive design, as well as
matching furniture systems and accessories, which can be chosen depending on
the individual needs.
This instruction manual provides important information regarding safe
operation for the Operators, Users and Patients. It describes: assembling, care
and maintenance of the nursing bed and operating the adjustment functions.
It will help you to get familiar with the assembly actions, operation, capabilities
and limitations of your bed. Read these instructions carefully and follow them
exactly in order to provide reliable service and avoid damage or incorrect
operation.
If you have any questions regarding setting up, using, maintaining this
product or you would like to receive other customer service information, please
do not hesitate to contact your local distributor or if in doubt our company
directly.
We always meet customer expectations and we will do our best to fulfil your
needs in future.
Team Elbur

Table of contents
General information .......................................................................... 6
Contact details ................................................................................. 6
Design policy ................................................................................... 6
Safety symbols used ......................................................................... 6
1.0. Product description..................................................................... 7
1.1. Intended purpose and range of application .................................. 7
1.2. Technical specification.............................................................10
1.3. Labelling of ELBUR Care Beds...................................................12
1.4. Explanation of symbols used on labelling ...................................13
1.5. Definition of the groups of persons involved ...............................15
2.0. Safety information and general warnings ......................................15
3.0. Scope of delivery, transport and storage ......................................19
4.0. Electric components ..................................................................23
5.0. Accessories and spare parts........................................................24
6.0. Mattress selection .....................................................................25
7.0. Assembly and use .....................................................................26
7.1. Preparation ...........................................................................26
7.2. Mattress Support Platform .......................................................26
7.3. Assembly of Control Box and MSP Actuators ...............................27
7.4. Actuator cable clip protection ...................................................28
7.4.1. New type of actuator cable clip ...........................................28
7.4.2. Old type of actuator cable clip.............................................28
7.5. Connecting electric components and cable routing.......................29
7.5.1. Standard version –Linak CB6 Control Box ............................29
7.5.2. Optional version –Linak CA40 Control Box ...........................30
7.6. Castors operating ...................................................................31
7.7. Assembly of the Bed Ends with the MSP End Sections ..................32
7.8. Height and angle adjustment of MSP and its segments................32
7.8.1. Standard Handset .............................................................33
7.8.2. Comfort Handset...............................................................34

7.8.3. Trendelenburg Handset......................................................35
7.8.4. Locking individual functions on the Handset ..........................35
7.9. Additional adjustment of the Lower Leg Rest ..............................36
7.9.1. Electric adjustment of the Lower Leg Rest ............................36
7.9.2. Manual adjustment of the Lower Leg Rest.............................36
7.10. Emergency lowering of the Back Rest ......................................37
7.11. Side Rails ............................................................................38
7.11.1. Assembly of Side Rails .....................................................39
7.11.2. Lowering of Side Rails......................................................40
7.12. Lifting Pole ..........................................................................41
8.0. Movements of ELBUR Nursing Beds..............................................43
9.0. Functionality Check ...................................................................44
11.0. Cleaning and Disinfection .........................................................45
12.0. Disassembly ...........................................................................46
12.0. Troubleshooting ......................................................................47
13.0. Maintenance ...........................................................................48
14.0. Electromagnetic Compatibility statement ....................................55
15.0. Disposal.................................................................................58
16.0. Warranty ...............................................................................58

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 6 of 60
General information
This instruction for use is part and parcel of the product and must accompany
every product sold.
Article name, [number]
Instruction for use of ELBUR:
PB 325, PB 326, PB 331, PB 337 [35131]
Document number,
version, date of issue
TF-03[6.7.8].5
ver.02, 26.04.2022
All rights reserved - no part of this manual may be reproduced in any form
without written permission of the manufacturer.
Contact details
When assistance is needed regarding setting up, using, maintaining this
product or when is necessary to receive other customer service information,
please do not hesitate to contact your local distributor or if in doubt our company
directly at the following address:
Address
Elbur sp. z o.o. sp.k.
Działosza 34
56-500 Syców, POLAND (PL)
NIP: 9112035149 | REGON: 386426270
BDO: 000467132
Telephone
+48 62 786 97 80 (Secretariat)
+48 62 786 97 90 (Sales Deparmtent)
E-mail / Website
Design policy
This manual reflects the latest product development. However as our policy is
one of continuous improvement, we reserve the right to modify design or
introduce technical changes without prior notice or without being imposed to any
obligation to adapt or replace similar products previously delivered.
Safety symbols used
In this Instruction for use, safety information is displayed in the following way:
WARNING
Indication of a potentially hazardous situation that, if disregarded could
lead to death or injury or could result in damage / failure to the
products or something around it.
ATTENTION
General information or tips regarding the safe use and helpful courses
of action.
It is very important to read all written safety information included in this
manual and follow the instructions exactly!
Pictures are used to clarify instructions in this manual. Details of the depicted
product may deviate from your product.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 7 of 60
1.0. Product description
1.1. Intended purpose and range of application
ELBUR Care Beds: PB 325, PB 326, PB 331, PB 337 are classified as
Medical Devices Class 1 in accordance with the Medical Device Regulation
2017/745 on medical devices. They are developed, manufactured and tested
according to the current state of technical knowledge and all relevant safety
standards for medically used beds. Most importantly, they ensure basic safety
and essential performance, which are provided by the fulfilment of requirements
presented in EN 60601-2-52 standard.
Those products are beds with electrically operated functions, intended to be
used in a long-term, continuous care of elderly or infirm, frail and handicapped
adult patients. They provide a high degree of lying comfort. Subject to risk
assessment, ELBUR Care Beds are used to and may help to maintain, improve,
compensate or alleviate for injury, disability, disease and the overall condition
of the patient.
WARNING
A risk assessment must be carried out before the bed is used by a patient
by a clinically qualified person (age, size, condition should be assessed).
ELBUR: PB 325, PB 326, PB 331 and PB 337 may be used in the following
application groups according to EN 60601-2-52:
•3 –Long-term care in a medical facility in which medical supervision is
required and monitoring is provided if required. A medical electrical device
used in medical procedures can be provided to help maintain or improve the
condition of the resident (e.g. retirement and nursing homes, rehabilitation
facilities and geriatric institutions).
•4 –Care in the home. A medical electrical device is used to alleviate or
compensate for injuries, disabilities or illnesses.
ELBUR Care Beds type PB described in this manual are approved by
manufacturer for use by patients having a physical size equal to or greater than
146 cm, a mass equal to or greater than 40 kg and a body max index (BMI) equal
to or greater than 17.
It is recommended to use the original ELBUR Mattress Platform Extension for
patients taller than 190 cm.
Special dimensions are an essential part of the Elbur production. We enable
our customers to have their nursing bed tailored to fit the individually physical
requirements of the patient, which allow to achieve everybody the optimal lying
comfort while maintaining the same level of functionality.
The ELBUR Nursing Beds described in this manual are not suitable for the
patient transport, but they can be moved within the patient’s room for cleaning
purposes or to allow access to the patient for care activities.
ELBUR Care Beds are designed and manufactured solely as resting place to
support one patient with maximum weight of 140 kg. The safe working load of
the beds presented in this manual is calculated as specified by EN 60601-2-52:

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 8 of 60
Safe Working Load:
175 kg
200 kg
Maximum Patient Weight:
140 kg
165 kg
Mattress:
20 kg
Accessories:
15 kg
WARNING
The indicated values of Maximum Patient Weight and Safe Working Load
cannot be exceeded.
Only use suitable Side Rails and Mattress on those beds.
Only use original accessories and spare parts approved by Elbur.
ELBUR Care Beds: PB 325, PB 326, PB 331 and PB 337 are not suitable
for child patients or for use in hospitals.
ATTENTION
Any other used than described above is considered improper and is
excluded from a possible liability claim.
ELBUR Care Beds described in this manual are not licensed for use on
the North American market, especially in the United States of America.
The distribution and use of ELBUR Care Beds in these markets, including
through third parties, is prohibited by the manufacturer.
The ELBUR Care Beds are developed and constructed in such a way that they
can function long and safely. If the bed is used and operated properly, the
estimated product life is 4-8 years –it all depends on the conditions within and
frequency at which is used.
ELBUR Care Beds are appropriate for situations where assistance is required
in positioning the patient, or in reducing lifting and manual handling stresses on
the carer –it makes the nursing staff work easier. Electric regulation simplifies
patients getting into or out of the bed, comfortable setting in chosen position and
provides easy access to the patient during hygienic actions given by caregivers.
They are classified as a Class II electrical products and are suitable for indoor
use only. The electrical building components comply with safety standard
EN 60601-1 for Medical Electrical Equipment.
ATTENTION
ELBUR Care Beds have no special connectors for potential equalisation.
Please be aware of that before connecting additional medical electrical
equipment. The operator of the medical products has to ensure that the
combination of the equipment meets the requirements of EN 60601-1
standard.
The indivdual ELBUR Care Bed components are made from the best quality
materials, which biocompatibility is described in the technical documentation.
Our Beds are made predominantly of steel sections coated with a durable
polyester-powder coating. All wooden or wood-based parts are protected with an
ecologically coating that is low on harmful substances. The indicated surfaces are
safe for contact with the skin.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 9 of 60
A wide range of wooden finishes and a variety of colours allow for the
harmonious integration of each ELBUR Care Bed with any kind of furniture. Also,
a wide product range of auxiliary furniture complements the various bed models
up to the complete interior design of your home. The combination creates a care
and living comfort resulting in perfect harmony.
The Mattress Support Platform consists wooden slats as a base standard, but
each model can be also ordered with metal slats as a base. It is divided into four
functional sections: Back Rest, Stationary Seat, Upper Leg Rest, Lower Leg Rest.
The various parts of the bed are moved with protective-low voltage by Linak
Actuators. Two of them (Main Lift Actuators) are positioned at either end of the
bed and control the bed height. Two other Actuators are positioned underneath
the Mattress Support Platform and control the Back Rest and the Leg Rest. Each
individual position can be controlled adjusted continuoulsy with a touch of
a button on the Handset. The Actuators, Handset and Power Cable are connected
to a Control Box, which is fixed onto the Back Rest Actuator.
The main difference between beds described in this manual are Bed Ends,
which type defines each model. The standard height adjustment is 40 ÷ 80 cm,
except ELBUR PB 337, where is 23 ÷ 63 cm.
The other and last differences are values of Safe Working Load and Maximum
Patient Weight, which are respectively 200 kg and 165 kg for ELBUR PB 325
where for the rest beds are as follows: 175 kg and 140 kg.
The Mattress Support Platform is exactly the same for all models. The manual
adjustment of Leg Rest with two rastomats assembled in that section of the bed
is an option for ELBUR PB 325, where it is as a standard for all other ELBUR Care
Beds described in this Instruction for Use.
All beds are equipped with Full-length Side Rails and Lifting Pole in the
standard configuration. Instead of ussually used side rails, ELBUR Care Beds
described in this manual can be equipped with Split Side Rails, Folding Side Rails
and Metal Full-length Side Rails.
Basic UDI
ELBUR nursing
bed Model
GS1 Company
Prefix
Internal Number
/ Model Ref.
Valid-check
character pair
ELBUR PB 325
590426134
ELBURPB325
X7
590426134ELBURPB325X7
ELBUR PB 326
590426134
ELBURPB326
X9
590426134ELBURPB326X9
ELBUR PB 331
590426134
ELBURPB331
X2
590426134ELBURPB331X2
ELBUR PB 337
590426134
ELBURPB337
XE
590426134ELBURPB337XE
ELBUR Care Beds are designed to be dismantled into sections for ease of
transportation, but they can be easily assembled with a minimum of tools.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 10 of 60
No.
Description
No.
Description
1
MSP (Head Section)
2
Back Rest Section
3
MSP (Leg Section)
4
Leg Rest Section
5
Bed End (Head End)*
6
Bed End (Foot End)*
7
Lower Side Rail
(with top holes drilling)
8
Upper Side Rail
(with bottom holes drilling)
9
Castor with brake pedal lever
10
Handset
11
Lifting Pole
12
Triangle Handle with Strap
13
Drive system (concealed in picture):
Control Box with Back Rest Actuator, Leg Rest Actuator
*On the picture below is ELBUR PB 331. Only difference with other models are
Bed Ends, which are presented in the Chapter 4.0. of this manual.
1.2. Technical specification
ATTENTION
All indications of dimensions and weights in this manual are approximate
(measurement tolerance: ±10 mm / ± 0,5kg / ± 1,5°). All parts and
data are subject to a constant further development and therefore may
differ from mentioned data.
The technical details on the next page are only valid for the standard
configuration and dimensions and optimal ambient conditions.
The values are no longer applicable if your Care Bed has been modified,
damaged or suffers from severe wear.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 11 of 60
TECHNICAL SPECIFICATION
MODEL: ELBUR
PB 325
PB 326
PB 331
PB 337
Load
[kg]
Maximum Patient Weight
165
140
Safe Working Load (SWL)
200
175
Lifting Pole
80
Dimensions
[cm]
External [WxL]
102x218
104x216
102x215
102x219
Mattress Support Platform
(MSP) [WxL]
90x200
Under bed clearance (in the
lowest position of the MSP)
approx.
24
approx.
8
Adjustment
Height –Mattress Support
Platform [cm]
40 ÷ 80
23 ÷ 63
Angle –Back Rest [o]
0 ÷70
Angle –Upper Leg Rest [o]
0 ÷35
Mattress
Dimensions [Width x Length]
90 x 200 cm
Acceptable thickness
15 cm
Volume density of foamed
material
35 - 50 kg/m3
Drive
system
Control Box
LINAK CB6
Lifting Hi-Lo Actuator
LINAK LA27
Back Rest / Leg Rest
Actuator
LINAK LA27
Handset
HL 74
Comfort
HL74 Standard
Input voltage*
230V AC, 50Hz
Output voltage
24V DC
Power input*
Max. 1A
Operating time*
Duty cycle: 10%, 2 min / off 18 min
Ingress Protection*
IP X4
Electrical appliance Class*
II
Noise level
Typically < 65 dB (A)
Weight
[kg]
Total
87,0
91,2
101,0
102,6
MSP (Head End / Leg End)
With wooden slats
17,4 / 16,6
Bed End (1)
15,8
17,9
22,8
23,6
Side rail (1 lath / set)
2,8 / 11,2
Lifting Pole with Triangle
Handle
5,3
Back Rest / Leg Rest
Actuator
1,5 / 1,5
Handset / Control Box
0,3 / 1,6
Operating
conditions
Temperature [oC]
+10 ÷ +40
Humidity [%]
30 ÷ 75
Storage
conditions
Temperature [oC]
-20 ÷ +50
Humidity [%]
30 ÷ 75
* Electrical data values visible on the Name Plate of the ELBUR Nursing Bed.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 12 of 60
1.3. Labelling of ELBUR Care Beds
Manufacturer’s Name Plate (including serial number*)
[located on the outside of the Bed End lower section]
*Please quote SN on all correspondence regarding sales and customer service
information or to report unexpected operation.
Warning / Information label no. 35350
-for ELBUR: PB 326, PB 331, PB 337
Warning / Information label no. 35352
-for ELBUR: PB 325
[located on the inside of the MSP Head End frame]
Warning
ELBUR Care Beds are designed for patient having a physical size equal to
or more than 146 cm, a mass equal to or more than 40 kg and a body max
index (BMI) equal to or more than 17.
Warning / Information label no. 35360
[located on the inside of the MSP Head End frame]
Warning
Incompatible Side Rails and Mattresses can create entrapment hazards.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 13 of 60
Warning / Information label no. 35365
(Mattress Support Platform –Head End Section)
[located on the top of the Back Rest Section]
Warning / Information label no. 35370
(Indication of the appropriate assembly of drive units)
[located underneath the bed]
Warning / Information label no. Linak CB6
(Indication of the appropriate cables connection to Control Box channels)
[located on the CB6 Control Box]
Warning / Information label no. Linak CA40
(if the bed is equipped with this type of the control box)
(Indication of the appropriate cables connection to Control Box channels)
[located inside the CA40 Control Box]
zzzz
1.4. Explanation of symbols used on labelling
Symbol
Description
Catalogue number
Serial number
Date of manufacture (month/year)
Manufacturer

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 14 of 60
Symbol
Description
Maximum Patient Weight
Safe Working Load (SWL)
Medical Device
Caution, read instructions before use
Consult the Instruction for Use
Temperature limit
For indoor use only
Do not dispose of in household or commercial waste.
This product must be disposed via the separated
municipal waste.
Class II electrical device, double insulated,
degree of protection against electric shock
Type B applied part according to EN 60601-1
(symbol indicates this product has a degree of
protection against electric shock for type B equipment)
Conformity mark according to the Medical Device
Regulation (EU) 2017/745
Minimum Patient weight = 40 kg;
Minimum Patient Height = 146 cm;
Minimum Patient BMI = 17.
Detachable side rail suitability
Incompatible Side Rails can create hazards!
Mattress suitability -
Incompatible Mattresses can create hazards!
Product with a thermofuse
Safety isolating transformer, general
Pollution control mark (China)
Symbol regulatory compliance mark of The Australian
Safety/EMC Regulations

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 15 of 60
1.5. Definition of the groups of persons involved
WARNING
All persons working with the ELBUR nursing beds must be acquainted
with the safety instructions contained in this manual. Improper
operation can result in personal injuries or product damage.
Operator –every natural and legal person (e.g. care home operator, medical
supply distributors) responsible for safe operation and supervising of the bed.
User –person who is qualified how to operate the bed or handle its functions,
based on performed trainings, experience or briefing. The user is able to
recognize and avoid possible risks and to assess the health condition of occupant.
Patient –person who is injured, disabled or handicapped and who occupies
the ELBUR nursing bed in need of care.
For homecare application area Operator, User and Patient can be perceive as
the same person.
2.0. Safety information and general warnings
WARNING
Before assembling and operating the Care Bed, you must read,
familiarize, understand and follow all instructions in this manual and with
rated values/symbols provided on labelling carefully. It helps to prevent
any damage and risk occurring due to incorrect operation or use. All
safety features instructions mentioned in this document must be strictly
observed.
Use the Care Bed in accordance to its intended purpose and in
accordance to all instructions detailed in this manual, describing the
normal operating procedure, to ensure convenient handling, safe
operation and effective use of this bed and the safety of users and
carers.
ELBUR Care Beds: PB 325, PB 326, PB 331 and PB 337 are not suitable
for child patients or for use in hospitals.
Remember to keep this instruction manual for future reference. Please
ensure that the manual is available to users and carers at any time
throughout the bed’s service life.
All actions regarding assembling, handling, positioning and using of the
bed must be performed in accordance with the provided information to
ensure safe operation.
Only use technically perfect bed, in proper and faultless condition -
it must not be used if faults have been detected on it that may injure
the patient, staff or a third person, the bed or surroundings.
If damage or malfunction is suspected, take the bed out of service
immediately, label it as non-conforming product and disconnect it from
the power until exchange or repair of broken components.
Before using the Bed, the operator should understand the Bed and its
functionality.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 16 of 60
WARNING
Any actions that are inconsistent with the manual are performed at your
own risk and can result in serious injury or death. Elbur sp. z o.o. sp.k.
shall not be held liable for any damage, injuries or accidents arising from
unauthorised modifications, non-genuine spare parts, negligence or use
that is at variance with this manual.
Any serious incident that occurs in relation to this medical device,
affecting the user/operator or the patient, should be reported to
manufacturer - Elbur sp. z o.o. sp.k. and the Competent Authority of the
Member State in which the device is used.
The bed must only be operated by persons who are able to operate it in
accordance with the manual. Operators and patients must be risk
assessed to ensure they are able to operate the ELBUR Care Bed safely,
without risk to themselves or others.
Patients should used the bed adjustment function on their own only if
their physical and mental condition allows it and if they have been
positively verified (assessed as competent to do so) in this regard by
qualified medical personnel. They must then understand any safety
instructions in this manual. Before start using the Handset, the operating
staff must make the patient aware of its control functions.
It is recommended (but not required) to perform the assembly process
with another person, qualified in ELBUR’s Care Bed operating.
This Bed must not be exposed to smoke, naked flame, extreme
temperature, flammable gases or other hazardous substances or
situations.
The Bed must not be used where there is a danger of explosion or in the
presence of uncontained flammable liquids.
Electrical installations must meet local requirements.
Electrical equipment can be hazardous if misused or abused. Do not
open any electric components under no circumstances.
ELBUR Care Beds do not have emergency-off switch. Remove the power
plug in emergency situation.
Ensure that all cables (Power Mains Cable, Handset cable, Actuators
cables) are not damaged by crushing, that there is no loop in the cables
and that they are not get entrapped between the moving parts of the
bed.
Only the Power Supply Lead supplied with the bed may be used. Connect
it directly to an electrical outlet –do not use extension cords or multiple
outlet extension cords.
ELBUR Care Bed should be located in place which allow to easily
removing mains plug from the socket.
The electrical supply cable could create a trip hazard. Safely position and
route the cable underneath the bed using its holders.
Make sure that the Power supply cord does not get squished under the
castors.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 17 of 60
WARNING
The Handset should be positioned to avoid strangulation risk. Make sure
it is placed properly, that it is not trapped between Side Rails and
Mattress Support Platform/other furniture.
At the same time only one person can operate the bed functions. It is
strictly prohibited to press all buttons simultaneously on the Handset.
The bed must not be used if there is any visible damage to the any
electrical part of the drive system.
When routing cable for other electrical equipment used with the bed,
ensure that those cables cannot be squeezed, crushed or damaged by
the moving parts on the bed. Inappropriate routing of accessory cables
could lead to dangerous electric hazards. The bed must not be used if
there is any visible damage to any cables.
If using the electrical functions adversely affects the health of the bed
user, disconnect the power supply and only use the bed in static mode.
Do not exceed the maximum duty cycle of 2 minutes. Observe
a subsequent break of at least 18 minutes by all means.
For reasons of safety and electromagnetic compatibility, use only
original electric components from Elbur that have been released for the
Care Bed model in order. Otherwise it can caused danger for patients
and loss of bed functionality.
It is still possible that the operation performance is influenced by
electromagnetic fields like those of cell phones, power generators of high
power energy sources, despite sample representing products family of
ELBUR Care Bed type PB was tested on electromagnetic combability and
complied to the EN 60601-1-2 standard.
On the other hand, it is also possible that the electronics of the bed can
influence other electronic devices.
The special attention should be given to children and pets - keep them
away from this bed unless supervised by an adult as there is a risk of
injury and/or choking on small parts.
Maximum Patient Weight and Safe Working Load must never be
exceeded.
Only one person should occupy the bed at any time.
Use the bed on the flat, level, solid floors - remember that all four castors
must touch the ground and must be locked if there is a patient on the
Bed and location is not going to move, as an unlocked bed castor can
cause injury to a patient who leaves the bed or changes position.
This bed is not adapted for patient’s transport, but it can be moved
within patient room for cleaning or patient access. Remember to unplug
the Power Supply Lead before relocating the bed and make sure it will
not be rolled over or damaged in other way.
ELBUR Care Bed with a patient may only be moved in an emergency
situation. This activity should be then performed by two people due to
force the propulsion, but above all to ensure necessary safety. Patient
has to stay in flat position and Care Bed should be adjusted in the lowest
height.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 18 of 60
WARNING
When moving ELBUR Care Bed on the TLSU Transport Bracket, take
special care to prevent it from tipping over or unexpectedly moving.
Do not use the bed as a means of transport, ladder, or storage for heavy
or hot objects. Never stand on the bed.
Do not perform medical operation on the ELBUR Care Bed.
Before operating this bed, ensure that its user is safely positioned to
reduce the risks of bed fall or entrapment.
There must be enough space for bed height adjustment in the full range.
It cannot be blocked by any obstacles such as bedside lockers or window
sills etc.
When operating the moving parts of the bed, care must be taken to
ensure that the patient, operator or any other people, especially children
or pets, and objects do not become trapped, jammed or injured.
The height of the Mattress Support Platform must be adjusted to the
correct level for the patient’s health condition. It should also facilitate
the provision of appropriate care to the staff.
The Bed should be left in its lowest position and with side rails in the
upper and locked position when the patient is unattended to reduce risk
of injury due to falls.
Always check for any entrapment risks under the bed before lowering to
its lowest position.
Never use area under the bed as a storage space.
Do not lower the bed while a Patient Hoist that extends beneath the bed
is being used. Hoist access is obtained when the bed is event in its lowest
position, but for ELBUR PB 337 is necessary to raise its Mattress Support
Platform to at least 30 cm measured from the floor.
Use only original Elbur accessories which are compatible with this bed –
any modification is strictly forbidden without authorization of our
company.
Contact your local distributor or manufacturer directly if the exchange
or repair of faulty parts have to be made. Only original materials and
components may be used in those kind of situations.
Only use Side Rails that are compatible with this bed as supplied by
Elbur. Incompatible Side Rails can create hazards and entrapment risks.
There must be a minimum height of 22 cm without compression
measured from the top of the mattress to the top edge of the top rail.
Where a speciality mattress (thicker than 15 cm) or mattress overlay is
used and the distance from the top of uncompressed mattress to the top
of the side rails, if fitted, is less than 220 mm a risk assessment must
be performed to assure equivalent safety, in order to prevent trapping
or person completely falling out in accordance to EN 60601-2-52.
If a Lifting Pole is fitted on the Bed, increased attention must be taken
during movement or lifting, to the space around the lifting pole, so that
the equipment is not damaged.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 19 of 60
WARNING
Make sure that other attached equipment, which is compatible with the
ELBUR Care Bed, can function in a secure and safe manner. Contact your
local distributor or manufacturer, if you have any doubts.
ELBUR Care Bed must be cleaned and washed down with a mild
household disinfectant in case of a patient change or in other situation
required it. Before start cleaning procedure, the bed’s power supply
lead must be disconnected from the socket.
Clean / disinfect the bed by wiping all rigid parts with a soft, damp cloth
(not drenched) with a mild household disinfectant, suitable for varnishes
and synthetics (only gentle, not aggressive agents should be used in
order to maintain the material resistance).
ELBUR Care Bed must not be cleaned in automatic bed washer or with
high-pressure or steam cleaner, as liquid penetrates into the electrical
components could result malfunctions and danger. Protect those
components from moisture.
Only specially authorized technicians are allowed to carry out
troubleshooting activities and replace individual electric components in
the event of their failure.
Maintenance must be carried out only by or under supervision of suitably
trained and qualified personnel or professional persons such as
electricians or electro-technically instructed persons who have
knowledge of the relevant provisions and are able to recognize possible
risks and hazards. Failure to do so my result in injury or an unsafe
product.
It is obligated to perform periodic inspections and electrical tests
according to EN 62353 performed by technically trained person at least
once a year, with every re-use for a new patient or with every
maintenance, to ensure safe condition of the Nursing Bed.
The subsequent technical safety check has to be documented.
During performing service work or functionality check patient cannot lay
on the bed.
The manufacturer is not liable for damage caused as a result of not
following instructions from this manual or by the lack or improper
service.
3.0. Scope of delivery, transport and storage
ELBUR Care Beds are delivered unassembled and mounted on the TLSU
Transport Bracket, what simplify transport and relocating on the place of delivery
(at the customer’s special request, they can also be packed in cardboard boxes).
Reliable package protects beds from external damages and ensures that
patient or user get a bed in the perfect condition.

ELBUR: PB 325, PB 326, PB 331, PB 337
Elbur sp. z o.o. sp.k. 20 of 60
WARNING
Extreme caution must be taken when moving the bed on its TLSU
Transport Bracket to prevent from the bed tipping over or moving
unexpectedly.
Before start assembling or operating ELBUR Care Bed, you must read,
understand and follow all instructions written in this manual carefully.
If the bed has become soiled or contaminated during transport refer to
cleaning and disinfection instructions.
If the Bed is stored in conditions outside the normal operating range, it
should be allowed time to stabilise, in normal operating conditions,
before use.
ATTENTION
Always verify shipment contents with your order (items described below
may not be included in the ordered model configuration) to ensure that
delivery is complete and undamaged. Check whether any visible damage
has occurred to the Bed during transport.
Any missing parts, faults or damage must be reported immediately to
the carrier and your local distributor or Elbur directly in writing.
This manual suits for next models
3
Table of contents
Other Elbur Medical Equipment manuals