10 pfm Rotary 3006 EM Rev.2022-01-28
Please take special note of the following warnings:
Beware of explosive gases
The instrument may not be used if dangerous gas
concentrations could occur in the surroundings.
Risk of infection
When working with infectious materials, appropriate
safety measures must be implemented. The samples used
may potentially be infectious. The general laboratory rules
regarding protection against infections must therefore be
observed.
Risk of radioactivity
When working with radioactive materials, the radiation
protection regulations must be followed. Protective
clothing must be worn in accordance with the rules and
regulations that apply in the respective laboratory for
handling radioactively contaminated materials. For the
disposal of radioactive waste, the applicable regulations
will apply in each instance.
Risk of chemical substances
For better cutting properties, some of the component
alloys of this instrument contain small amounts of lead,
intended, the lead in the alloys is not released and does not
represent any risk to the user. As an additional precaution,
we recommend that you wear gloves when changing the
Declaration on the EU Regulation 1907/2006 (REACH) at:
http://www.pfmmedical.com/downolads
Avoiding malfunctions
To avoid malfunctions the instrument may not be used
close to radio transmitters, e.g. mobile phones. If the
your distributor.
Caution
please check at the installation location that the supply
voltage/frequency match the information given on the
nameplate.
Protective earthing of the instrument
In order to rule out the risk of electrocution, the
instrument must be connected to the earth conductor.
The instrument is equipped with a mains power plug
with earth conductor. It is very important to check that
the mains power outlet used is equipped with an earth
conductor, and that it complies with the regulations of the
International Electrotechnical Commission (IEC).
Mains voltage precautions
Never remove housing components. The replacement of
components and alignment must only be carried out by
trained service technicians.
Interference due to electromagnetic interactions
The electromagnetic environment should be assessed before
the microtome is operated. Do not use this instrument near
strong sources of electromagnetic radiation (for example,
unshielded, intentionally operated high-frequency sources)
because these can interfere with the proper operation
requirements of the standard IEC 61326-2-6 on immunity
and emissions.
1.6 General safety precautions
WARNING
Caution! Switch off the instrument and pull out
the mains plug before removing any part of the
housing.
Safety instructions