
South Bend, Indiana USA | networketi.com DEICING CONTROLLER INSTALLATION MANUAL | PART NO. 25123 REV 0.0
7
When the output is turned ON, the voltage from Blue
to Orange will be zero volts.
The sensor can also be tested with at least one of
the Blue and Orange leads disconnected, using the
continuity-test mode on the DVM. When the output is
OFF, Blue and Orange will be open; when the output
is ON, they will have continuity.
3. After the meter connection is made,initiate the self-test
mode as follows. A system power switch or temporary
connection for the red power lead is needed to allow
briey interrupting the supply voltage to the sensor.
See Figure 5.
a. Power is applied to the sensor.
b. Then between 0.1 second and 4 seconds later,
the power is very briey interrupted – for 50 to
100 milliseconds – and then reapplied. This can be
done with a power switch, using a clip cable, or by
temporarily wiring a normally-closed pushbutton
so that it can be very briey pressed.
c. The power is maintained after that power-up pulse,
and the self-test then begins.
The self-test mode can be identied by the initial
increase in current draw (part 5a, below). If that
does not happen, it may be that self-test mode was
not entered; this happens if the power-off pulse
time is too long. In that case, remove power for at
least two seconds, then repeat the off-pulse
sequence (a-c) described above (Figure 7).
4. After the self-test mode is initiated, the following
action will happen:
a. The heater turns on for three seconds. During this
time, the current draw will increase to 0.25 amps.
This can be monitored with the current meter as
described in part 2.
b. As shown in Figure 7, the output and heater
temperature, measured to the nearest degree F.
One pulse means below 38, 2 means at 38, 3
means above 38, 4 means there is a problem.
c. After that, the unit will go into moisture-sense
mode. At this point, when moisture is sensed, the
output will turn on. When moisture is not sensed,
the output will turn off. This is monitored with the
DVM as described in part 2.
d. Moisture can be simulated with water on the sensor
(at any temperature) or with a metallic object
placed on the sensor to short one of the electrode
disks to the outside rim of the sensor.
When moisture is sensed, in addition to the output
turning on, the heater will begin pulsing with one
second on and two seconds off. You can see that
happening with the current meter.
e. After one minute, the self-test will end, and the
sensor will begin normal operation.
5. As an alternative to the self-test mode, the sensor can
be tested with ice. To do this, place ten pounds (4.54
kg) of crushed ice over the sensor and wait about 20
minutes before observing that a low output voltage (2
volts or less) is indicated by the DVM.When the sensor
temperature goes below 38° F (3.3° C) the sensor’s
output will turn ON and change to a lower voltage. If
the sensor surface temperature is initially very warm
then this procedure may take longer, perhaps 40
minutes or more.
6. With satisfactory results, disconnect the DVM and
restore all sensor wiring connections.