5
Usage
2.2.1 Connect the power via suitable adaptor.
2.2.2 The PW lamp will light up and the sensor lamp will flash once from HS lamp
to LS lamp one by one. That mean is the sensor had readied.
2.2.3 When someone approaches the bed, the HS lamp will flash, but the sensor
still won’t be activated.
2.2.4 When the patient got up or anyone approaches the detective range, the LS /
CS / RS lamp(s) and HS lamp will light up, then the alarm will be activated.
2.2.5 The AL lamp will flash, and a pulse alarm signal will transmit to receiver.
Pause Function
When the nurse helps the patient to take medication or other needs, the alarm
sensor need to pause.
2.3.1 When press pause button, the main unit will be suspended the detect
function. All sensor lamps (HS, RS, CS, LS) will light up.
2.3.2 After 30 seconds, HS lamp will off;
2.3.3 After 60 seconds, RS lamp will off;
2.3.4 After 90 seconds, CS lamp will off;
2.3.5 After 120 seconds, all sensor lamps will flash.
2.3.6 After 150 seconds, all sensor lamps will off. And all sensors will resume
detection automatically.
2.3.7 If press pause button during suspend time, all sensor lamps will off and all
sensors will resume detection.
Type B & Type C:
The base function of type B and type C that is same. They just different in detection
time only. The detection time unit of type B is second; the detection time unit of type
C is minute. That is for different application.
These types require three distance sensors for detection.
Although the detective distance can adjust to 1.0m & 1.3m, it is recommended to use
0.6m for type B & type C.
When the patient is lying in bed, the distance sensor will monitor the head of patient.
If the head leave the monitor zone, the alarm signal will be activated.