Excelitas Technologies OmniCure LM2011 Series User manual

LED Light Meter System
User Guide
Excelitas Canada Inc.
2260 Argentia Road
Mississauga (ON)
L5N 6H7 Canada
+1 905.821.2600
www.excelitas.com
035-00434R Rev. 3

User Guide
http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved.
Page 1
USER GUIDE
035-00434R
Excelitas Canada Inc. 2023
All rights reserved
No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval
system or translated into any language in and form by any means without the prior written
consent of Excelitas Canada Inc. every effort has been made to ensure information in this
manual is accurate, however, information in this manual is subject to change without notice and
does not represent a commitment on the part of the authors.

User Guide
http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 2
Trademarks:
OmniCure® is a trademark of Excelitas Canada Inc. All rights reserved. All other product names
are trademarks of their respective owners. All product or software photos shown are for
reference only and are subjected to change without notice.
Made in Canada. Doc. No. 035-00434R

User Guide
http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved.
Page 3
Table of Contents
1Introduction.............................................................................................................1
2Safety.......................................................................................................................1
2.1 Glossary of Symbols.................................................................................................... 1
2.2 Safety Precautions....................................................................................................... 2
3System.....................................................................................................................4
4LED Light Meter Features ......................................................................................4
4.1 Power ON/OFF............................................................................................................ 6
4.2 Peak Hold.................................................................................................................... 6
4.3 Wavelength nm:........................................................................................................... 6
4.4 Power/Irradiance.......................................................................................................... 7
4.5 LED Light Sensor Connection...................................................................................... 7
4.6 USB Connection.......................................................................................................... 7
5LED Light Sensor: ..................................................................................................7
6Calibration...............................................................................................................9
7Installation and Basic Operation:........................................................................10
8GUI Operation .......................................................................................................11
8.1 Driver Installation........................................................................................................11
8.2 GUI Control Panel Operation......................................................................................11
9Care and Maintenance..........................................................................................12
9.1 General.......................................................................................................................12
9.2 Cleaning .....................................................................................................................12
9.3 Lithium Batteries.........................................................................................................13
10 Technical Specifications......................................................................................14
11 Electrical Specifications (LED Light Meter)........................................................14
12 Regulatory Compliance........................................................................................15
12.1 Safety and Electromagnetic Compatibility:..................................................................15
12.2 China RoHS................................................................................................................16
12.3 WEEE Directive..........................................................................................................16
13 Warranty................................................................................................................17
14 Returning Equipment to Excelitas Technologies for Service...........................18
15 Contact Information..............................................................................................19

User Guide
http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 4
16 Replacement Parts and Accessories..................................................................20
Figure 1 LED Light Meter........................................................................................................... 4
Figure 2 LED Light Sensor Connection...................................................................................... 5
Figure 3 USB Connection .......................................................................................................... 5
Figure 4 LED Light Sensor......................................................................................................... 8
Figure 5 LED Light Sensor Aperture .......................................................................................... 8
Figure 6 LED Light Meter GUI Control Panel ............................................................................11
Figure 7 Lithium Battery Compartment......................................................................................13

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 1
1 Introduction
The main purpose of this LED Light Meter System is to provide accurate absolute
power/irradiance measurements for various lens options and wavelengths of an
Excelitas LX series UV LED spot curing system or any other commercially available
UV LED spot curing system.
This product will complement the Excelitas LX series and will also cater to
commercially available LED UV spot curing systems.
We recommend that you read this manual to discover all its features, and how to use
them
2 Safety
2.1 Glossary of Symbols
CAUTION - Risk of danger: consult accompanying documents
WARNING -Eye damage may result from directly viewing ultraviolet light –
protective eye shielding and clothing must be used at all times.
Input/Output Signals
Battery
D.C. Current
Caution, Hot Surface

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 2
2.2 Safety Precautions
DIN Multi Pin Connection Port For External Sensors
(for use only with sensors supplied by Excelitas Technologies)
For using the Sensor, connect cable to this port.
WARNING
When using LED Heads do not stare directly at the LED Aperture(s). This may be
harmful, resulting in eye injury. Always use the protective eyewear provided with this
device. Additionally, protect any exposed skin with appropriate clothing or shielding
as required.
WARNING
To prevent accidental exposure to hazardous optical/ UV radiation, always ensure
that the LED Heads are properly secured in a mounting fixture. Hand-held use of the
LED Heads is not recommended and may expose the user to dangerous optical
radiation.
Caution, Hot Surface
Due to elevated operating temperatures; avoid contact of the LED Head(s) when
energized The LED Heads are designed to be mounted in a suitable fixture prior to
use. Clamp type heat sink assemblies are provided with each LED Head to provide
user safety and optimum thermal management. Prior to handling of the UV LED
Head(s), allow a cool down for a period of approximately 5 minutes after system
power has been removed.
Warning
The LED Light Meter is supplied with two lithium batteries. Lithium batteries present
a potential fire, explosion or severe burn hazard. DO NOT attempt to re-charge,
disassemble, incinerate, short circuit or expose battery to temperatures above 100°
C or expose contents to water.
WARNING!
Used batteries are not to be discarded. Return to the nearest authorized Excelitas
Technologies service center for disposal/ recycling.
Lithium batteries must have terminals taped with non-conductive material prior to
returning for disposal/re-cycling to prevent short-circuiting. External packaging
material must provide adequate protection to contents.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 3
The lithium battery supplied in this system DOES NOT contain: mercury, lead,
manganese or cadmium. Substitution of any other type of battery is not recommended
and may void warranty.
Caution
The detection window operational life can be significantly shortened if it is handled
incorrectly. Do not touch the window surface. Skin oils and scratches can cause the
window to fail prematurely.
Cleaning:
Clean the exterior of the Light Meter with a slightly dampened cloth and simple water/
detergent solution only. Refer to section 9.2 for detailed cleaning instructions.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 4
3 System
LED Light Meter (model# LM2011, LM2011-xx)
LED Light Sensor with removable Aperture (model# LS100, LS200)
Component
Part Number
LED Light Meter System
010-00281R,
019-00427R
Lithium Battery (qty. 2)
055-00007R
4 LED Light Meter Features
Figure 1 LED Light Meter
POWER/IRRADIANCE
PEAK HOLD
WAVELENGTH SELECT
ON/OFF
LCD DISPLAY
LED LIGHT DETECTOR
CONNECTION
USB PORT

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 6
4.1 Power ON/OFF
4.1.1 Press ‘ON/OFF’button to switch on the unit. The power on sequence will flash full
screen information for 5 seconds and then display current wavelength.
4.1.2 The Meter will auto power off after 5 minutes (system will remember the last mode
and wavelength on power up).
4.1.3 Hold ‘ON/OFF’ for 4 seconds to turn power off.
4.2 Peak Hold
4.2.1 ‘Peak Hold’button (Irradiance or Power Mode): in either modes the meter latches
on to the peak value and refreshes the display when the value is higher than the
current displayed value.
4.2.2 Upon pressing the button once, the meter will activate the ‘Peak Hold’mode and
the system will display “peak hold”(displays “peak’ over top of ‘hold’) to indicate the
peak hold mode. Pressing ‘Peak Hold’ again the meter will revert back to the
normal default mode which displays the current value.
4.3 Wavelength nm:
Default Wavelength Selections (LM2011);
365nm
385nm
400nm
460nm
Default Wavelength Selections (LM2011-xx);
365nm
385nm
395nm
405nm
4.3.1 Pressing the ‘Wavelength’ button once will display the current set wavelength.
Pressing the ‘Wavelength’ button repeatedly will cycle through the 4 preset
wavelengths. (if you have increased or decreased the number of wavelengths
using the control panel, the unit will cycle through the list provided)

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 7
4.4 Power/Irradiance
4.4.1 This button will toggle between the Power and Irradiance mode.
Sensor Aperture
Light Meter Select
Power
Removed
Power
Irradiance
Installed
Irradiance
4.4.2 The display will show “OL” if the output power exceeds 500mW or 25W/cm² (when
used with a LS100)
4.4.3 The display will show “OL” if the output power exceeds 2000mW or 40W/cm²
(when used with a LS200)
4.5 LED Light Sensor Connection
4.5.1 Connect the LED Light Sensor via DIN multi-pin connection.
4.5.2 If the sensor is not connected the display will show “LS”
4.6 USB Connection
4.6.1 Connect to a computer to access the GUI.
5 LED Light Sensor:
5.1.1 Removable threaded aperture will allow a choice of Power or Irradiance to be
measured.
5.1.2 To remove the aperture, twist and rotate the aperture in a counter-clock wise
direction. (do not use mechanical tooling). Refer to Figure 5.
5.1.3 To install the aperture, engage threads and hand tighten the aperture in a clock
wise direction until a positive force is felt. (do not over tighten or use mechanical
tooling). Refer to Figure 5.
Note: Ensure the supplied aperture is used with the sensor, exchanging other
apertures invalidates the irradiance calibration.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 8
Figure 4 LED Light Sensor
Figure 5 LED Light Sensor Aperture
DETECTION WINDOW WITH THREADED
APERTURE REMOVED.
DETECTION WINDOW WITH THREADED
APERTURE INSTALLED.
TIGHTEN APERTURE -
CLOCK WISE DIRECTION
REMOVE APERTURE –
COUNTER-CLOCK WISE
DIRECTION.
Y
X

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 9
6 Calibration
6.1.1 It is recommended that the LED Light Meter system be calibrated every 12
months to ensure valid measurements. The calibration is traceable to NRC and a
calibration certificate is included at each calibration cycle.
6.1.2 Irradiance calibration is only valid with the correct aperture supplied, Exchanging
other apertures invalidates the calibration.
6.1.3 Calibration is authorized only by a certified Excelitas service center. When
calibration is due contact Excelitas for a return authorization number. Refer to
Section 0.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 10
7 Installation and Basic Operation:
7.1.1 Connect the LED Light Sensor to the LED Light Meter via the DIN Multi-pin
Connection.
7.1.2 Place LED Light Sensor into production position
7.1.3 Select wavelength to match the LED Head wavelength, ref. section 4.3.
7.1.4 If measuring Irradiance, install Aperture and ensure “Irradiance”is selected on the
meter.
7.1.5 If measuring Power remove Aperture and ensure “Power” is selected on meter.
Sensor Aperture
Light Meter Select
Power
Removed
Power
Irradiance
Installed
Irradiance
7.1.6 If utilizing an XY alignment tool/fixture adjust XY positions over the detection
window/aperture to achieve maximum Power/Irradiance level. Secure LED Head
into desired position.
7.1.7 If not using an alignment tool/fixture manually move LED Head over the detection
window until maximum Power/Irradiance level is achieved and secure LED Head
into desired position.
XY Alignment Tool:
Provisions for an alignment tool/fixture is recommended to
eliminate the alignment challenges which may be encountered to
capture accurate peak power or irradiance measurement.
Note; Do not use mechanical device to remove or install Aperture.
Carefully install by hand until snug. Ensure the Aperture and the
detector optical window are clean before each use.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 11
8 GUI Operation
8.1 Driver Installation
Minimum Computer Specifications.
600+ MHz Intel Compatible processor.
Windows XP, Windows Vista, Windows 7.
256 Mb for RAM.
10 Mb for software installation.
20 Mb for data storage.
SVGA video 1024x768 resolution, 16 bit color.
USB port (1 port), USB 1.1 or 2.0 format.
Download the GUI installation file from Excelitas website.
8.1.1 Connect a USB cable to the LED Light Meter and the Computer USB port. Refer
to Figure 3 for connecting cable locations.
8.2 GUI Control Panel Operation
8.2.1 The GUI allows the user to change the wavelengths that may be accessed by the
radiometer. The radiometer by default has 4 wavelengths, this may be increased
to 5 wavelengths, customized using the GUI.
Figure 6 LED Light Meter GUI Control Panel
8.2.2 Allowable wavelength values are between 320nm and 750nm. Click Set Values to
update the selectable wavelengths in the power meter.
8.2.3 Enter 0 if you wish to use less than 5 wavelengths.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 12
8.2.4 Click Exit to exit out of the program.
9 Care and Maintenance
9.1 General
9.1.1 The LED Light Sensor can be stored in its original individual packaging, or in the
plastic case with the LED Light Meter –there is a special cut-out in the foam
insert for this purpose.
9.1.2 Do not touch the LED Light Sensor’s optical window with fingers or any abrasive
object. Large measurement errors may result from be generated through
scratches, digs and other coating damage on the Sensor optical window surface.
Refrain from touching the optical surface with any foreign object which could
scratch the optical window.
9.2 Cleaning
9.2.1 Cleaning of the Sensor is not generally required. However, if any visible
contamination or finger prints appear in the optical window, the contaminants may
cause inaccurate measurement. If required, gently clean the Sensor optical
window and the aluminum housing with reagent grade alcohol and soft cotton
swab, a minimum amount alcohol is preferred to avoid alcohol contamination in
the detector.
9.2.2 Large measurement errors may result from be generated through scratches, digs
and other coating damage on the detector optical window surface. Refrain from
touching the optical surface with any foreign object which could scratch the optical
window.
Warning:
Seals are not 100% water tight. Use of excessive liquid will flood and
damage the electronics.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 13
9.3 Lithium Batteries
9.3.1 The lithium batteries (qty.2) in the LED Light Meter are expected to last for 1 year
under normal operating conditions. Do not remove the batteries from the Meter
until the low battery code ‘BAT’ appears on the LCD screen.
9.3.2 The batteries are accessible through a cover located on the underside of the LED
Meter. When replacing the batteries ensure that both batteries are replaced at
the same time.
9.3.3 Use the batteries supplied by Excelitas Technologies, do not substitute with
another brand. The batteries in the Meter are not standard AA batteries. The
voltage of the standard AA batteries will not supply enough power to operate the
LED Light Meter.
9.3.4 To replace the Excelitas Technologies supplied lithium batteries:
Remove the protective rubber boot.
Remove qty.2 Philips Screws.
Open battery cover.
Remove and discard old batteries
Insert new batteries (refer to the + / - placement decal inside the battery cavity).
Figure 7 Lithium Battery Compartment

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 14
10 Technical Specifications
Wavelength Range: 5 wavelengths are user-selectable within the range of
320nm - 750nm
Irradiance (LS100): 50mW/cm² - 25W/cm²
Power Range (LS100): 1mW –500mW
Irradiance (LS200): 50mW/cm² - 40W/cm²
Power Range (LS200): 1mW –2000mW
Measurement Resolution: 1% of displayed reading (worst case)
Sensor Uncertainty:
The uncertainty is claimed to be ±10.0%. This means there is a 95% probability the
reported power measurement from the detector falls within ±10.0% of the NRC
(National Research Council of Canada) reference detector standard
11 Electrical Specifications (LED Light Meter)
Battery Type: 3.6-volt Lithium, non-rechargeable 2.2 Ah
I/O Ports:
External Detector Port: DIN Connector, only use with Excelitas external LED Light
Meter Sensor.
USB 2.0.
Electrical Specifications (Sensor)
DIN Multi-Pinned Connection Cable: use only connect to LED Light Meter.
General
LED Light Meter
Dimensions (L x W x H): 6.39 x 3.93 x 1.70 inches (163mm x 100mm x 44mm)
Weight: 1lb (450g).
LED Light Sensor
Dimensions: 3 x 1 x 0.35 inches (75mm x 25mm x 9mm) (without aperture).
Weight: 2.9oz (82g) without protective cover.

http://www.excelitas.com/omnicure
Excelitas Canada Inc. 2023
All rights reserved
Page 15
12 Regulatory Compliance
12.1 Safety and Electromagnetic Compatibility:
The LED Light Meter system (model # LM-2011/LM-2011-xx/LS100/LS200), has been tested
and found to comply with product safety and electromagnetic compatibility requirements. For a
complete list of tests and for certification details, please contact your OmniCure representative
or visit: https://www.excelitas.com/product/omnicure-lx500-led-uv-radiometer-and-calibration-kit
CE Marking
Council Directive 2014/35/EU
Low Voltage Directive
Council Directive 2014/30/EU
EMC Directive
Council Directive 2012/19/EU
WEEE Directive
Council Directive 2011/65/EU
as amended by (EU) 2015/863
RoHS
Information to User
FCC Class B Digital Device or Peripheral
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a residential
installation. This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and then the user is
encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
WARNING
Changes or modifications not expressly approved by Excelitas Canada Inc. could void the
user’s authority to operate the equipment.
This manual suits for next models
1
Table of contents
Other Excelitas Technologies Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

MULTISPAN
MULTISPAN AVF-19N product manual

Niigata seiki
Niigata seiki DEI-127 instruction manual

Hanna Instruments
Hanna Instruments TDS 1 instruction manual

PCB Piezotronics
PCB Piezotronics 357B69 Installation and operating manual

Active Tools
Active Tools CoxOrb Platinum user guide

Southwire
Southwire 14060S operating instructions