EZee Life CH5075 User manual

Model # CH5075
EZee Life™ Products
34 Futurity Gate, Unit 15
Concord, Ontario L4K 1S6
Folding Electric Patient Lift
User Guide
…...............................................

2
Upon receipt of your patient, lift please verify that all the parts are
included, and it is free of shipping damage. If there seems to be
damage due to shipping, please contact the shipping company to begin
a claim.
Contents
Components 2
Safety Warnings 3
Assembly 5
Maintenance & Use 7
Transferring Instructions 10
Troubleshooting 13
Warranty 14
Components
a. Boom
b. Mast
c. Base
d. Spreader Bar
e. Pendant
f. Actuator
g. Control Box
h. Caster w/brake
i. Front Caster
j. Emergency Lowering
Device

3
Safety
The CH5075 patient lift reflects an innnovtive state of the art design to
enable the lifting and movement of the user. When properly maintained
as required, this lift will provide years of trouble free service.
Please pay attention to the information found in this manual regarding
care, maintenance and use of the lift. Carefully read these instructions
before asssembling the lifter, or attempting to lift anyone.
If you have any questions, or conccerns, or don’t understand the
inforamtion in this manual please contact your dealer.
1) Special care must be taken with users/patients who are unable to
assist while being lifted. (eg. Patients who are comatose, spastic,
agitated or otherwise severly handicapped).
2) The Ch5075 should be used solely for transferring form on utility (bed,
wheelchair, toilet, etc.) to another. The lift ashould not be used as a
transporting device to move users from one location to another.
3) Keep the base legs spread as wide as possible during use.
4) The base should be closed while being transported between
locations.
5) Do not roll the lift over any object or change of level while there is a
user in the sling.
6) Do not lock the casters while lifting or lowering.
7) The user should always be centered on ot he lift and facing the
caregiver operting the lift.
8) The maximum user weight ofr this lift is 400 pounds.
9) Never leav a user unattended in the sling or during lifting.

4
The CH5075 lift allows the user to be lifted and transferred safely with
minimum physical effort by the caregiver. Before attempting to lift
anyone, practice operating the lift. Explain the procedure for
transferring to the patient prior to actually lifting them.
WARNINGS
Do not attempt to lift a patient without prior approval of the primary
caregiver.
Practice using the lift before trying to transfer a patient.
If the patient hasn’t been lifted before consult a medical professional
of attendant to select the proper practical and comfortable sling. The
sling selected should be one that serves the needs of the patient and
provides optimal safety.
Never interfere with the lift unless instructed by the attendant.
Do not use the lift unsupervised during the first few transfers by the
caregiver.
Our sings are soecially designed for this lift. Do not use slings
manufacturered by other companies with our equipment. Using slings
not designed for this model of lift is unsafe and may result in serious
injury to users and caregivers.
Do not put anything (cushion, pad, etc.) between the user and the
sling. This may cause the patient to slide out of the sling.
Avoid personal injury by ensuring there are no children or pets near
the lift while being used.
Avoid persoan injury by ensuring extremities (legs, arms, hands fingers,
etc.) are clear of the lift mechanism.
Avoid personal injury by ensuring the lift is not used on a patient who
exceeds the lifts 400 pound capacity.
Replace any worn/damaged parts with genuine parts only. Parts from
other manufacturers or models are not interchangeable with the parts
of this lift and may create an unsafe condition that can result in serious
injury.
Avoid injury by removing the battery while servicing electrical
components. Consult a dealer for electrical repairs.
Service and repairs should only be performed by a dealer.

5
Assembly Instructions
Step 1. Engage the brakes, remove the black hand nut from the bottom of the
mast and remove the quick release pin from the bottom of the mast and base.
Step 2. Pull the mast, boom and spreader bar to an upright position. Re-
insert the quick release pin into the holes at the bottom of the mast.
Step 3. Attach the control box to the mast and tighten the two screws.

6
Step 4. Insert the actuator plug into the control box.
Step 5. Insert the pendant plug into the control box.
Note: Charge the batteries for 10 to 12 hours before the first use.
Step 6. Press on the pendant together to move the
spreader bar up to the desired height.

7
Opearating Instructions
Inspection Check List
The operator of the lift shall:
•Inspect the lift before each use.
•Check all bolts for tightness.
•Ensure the lift can be widened properly.
•Ensure all parts are in place.
•Ensure the casters are turning freely and the brakes can be
engaged.
•Ensure all the area of the transfer is ready for the lifting process.
Once a month the lift should be inspected by a person qualified to
recognize any signs of wear, loose parts or bolts.
Lubbricate the top of mast, spreader hinge and caster axles monthly with
a drop of oil.
Charging the battery – The lift can be charged by plugging the charger
directly into the control box while mounted onthe lift
or
the battery pack can be removed and charged in the optional
wall/desk charging caddy.
Using the Control Box
Pressing either of the control
buttons for less than ½ a second
will activate the view screen so
the battery charge indicator can
be seen.

8
Pressing the left button for more
than ½ a second will cause the
lift boom to rise.
Pressing the right button for
more than ½ a second will cause
the lift boom to lower.
Using the Pendant
Pressing the “UP” button for more than
½ a second will cause the lift boom to
rise.
Pressing the “DOWN” button for more
than ½ a second will cause the lift
boom to lower.

9
Maintenance & Inspection Schedules
Item
Each Use
Monthly
Check for loose or worn
electrical connections
x
x
Check the boom for
irregularity
X
Check the mast for
irregularity
X
Check the steering bar for
bending or looseness
X
X
Ensure all nuts and bolts
are tight
X
X
Ensure control box is
securely attached
X
X
Ensure actuator is securely
attached to boom and
mast
X
X
Ensure actuator
connecting cable is secure
X
X
Check battery capacity
X
Check sling for wear
X
X
Check sling hardware for
damage
X
X
Cleaning (as needed)
X
Item
Quarterly
Check axle bolts casters
X
Check rubber parts of
casters for wear/distortion
X
Apply grease to caster
bearings (if needed)
X
Cleaning (as needed)
X

10
Transferring Instructions
Turn the lift on by rotating the red “RESET” button clockwise.
Turn the lift off by pressing the “RESET” button down.
Transfer from a Bed
Note: Explain each step to the patient to reduce any anxiety.
1. Ensure the patient is in the center of the bed.
2. Roll the patient onto their side.
3. Roll the sling about halfway from side to side. The handle on the
back should face outward.
4. Position the rolled section of the sling against the patient’s back
with commode opening at the bottom of the spine and the top
near the patient’s neck.
5. Roll the patient onto the sling and up onto their opposite side so
the rolled section can be unrolled.
6. Roll the patient onto their back.
7. Feed the leg sections under the thighs and pull up between the
thighs.
8. Raise the head of the bed if this function is available.
9. Move the lift slowly toward the patient and position the
spreader bar above the patient’s chest.
10.Attach loop “a” of the sling to hook “A”, loop “b” to hook “B”,
loop “c” to hook “C” and loop “d” to hook “D” pf the spreader
bar.
11. Lift the patient using the hand pendant. (ensure the lift brakes
are off)
12. Pull the lift away from the bed to the wheelchair or chair and
lower the patient onto the surface.
Note: Reverse procedure to return the patient to their bed.
Warning: Only use slings supplied by the lift manufacturer. Using
mismatched slings can lead to accident and injury.

11
Transfer from Wheelchair
Note: Explain each step to the patient to reduce any anxiety.
1. Grasp the sling at each corner of the “U” shape of the commode
aperture.
2. Help the user lean forward slightly and slide the sling down
between the patient’s back and the wheelchair upholstery with
the grab handles on the back facing outward.
3. Position the commode aperture where the buttocks meet the
wheelchair seat
4. Position the leg sections of the sling along the outer thighs of
the patient.
5. Ensure the sling is centered on the patient by checking the
lengths of the leg sections when pulled forward.
6. Draw the leg sections under the patient’s thighs ensuring they
aren’t folded over or bunched.
7. Reposition the sling if leg sections aren’t equal length.
8. Ensure the leg sections are midway under the thighs to
provide
good support and patient comfort.
9. Move the lift towards the patient centering the spreader
above
the patient’s chest.
10. Attach loop “a” of the sling to hook “A”, loop “b” to hook “B”,
loop “c” to hook “C” and loop “d” to hook “D” pf the spreader
bar.
11. Lift the patient using the hand pendant.
12. Pull the wheelchair away and move the lift to the bed to finish
the transfer.
Note: Reverse procedure to return the patient to their bed.
Warning: Only use slings supplied by the lift manufacturer. Using
mismatched slings can lead to accident and injury.

12
Emergency Lowering Mechanism
In case of lift failure during use, please follow the following
procedure to safely lower the patient.
The emergency lowering device is located on top of the actuator
shaft. It is intended for use if the actuator stops working while
the patient is lifted. The device consists of a plastic collar ring
that should be turned clockwise continually until the patient is
lowered.
Contact your dealer immediately if the standard troubleshooting
techniques fail to correct the issues. Do not attempt to use the
lift until all failure issues have been safely resolved.
Note: The emergency lowering devices is intended for use
during lift failure only. This device will allow lowering of patients
only. Please contact your dealer immediately in case of failure.

13
Troubleshooting Chart
The following list of encountered problems and their solutions will assist
you in determining what may be causing your lift not to function as
designed.
If you have a problem occurring that is not listed below contact your
dealer or supplier for further information. Do not attempt to repair
components or parts of your lift as this may result in patient injury. Stop
using your lift immediately if it is not functioning correctly or any
warning beeps are heard.
Maintenance Notification:
Check the state of the lift and actuators.
Low Battery Capacity:
Recharge the battery. Replace the
battery if it will not hold a sufficient
charge.
Overload Alarm:
Weight has exceeded the set maximum
limit. Remove weight from lift and retry.
Force Interference:
If any force interferes when the lift is
lowered the lift will stop working to
prevent patient injury. Remove
interference immediately.
This image will also appear if the
actuator connectors are installed in the
wrong positions in the control box.

14
Warranty:
EZee Life™ products carry a one-year warranty against manufacturing
defects, faulty materials and workmanship.
In the event of a defect covered by this warranty, we will, at our option,
repair or replace the item. Please contact your supplier/dealer for assistance
should a defect become apparent.
The warranty does not extend to consequential costs resulting from the fault
or defect in the product, in particular freight and travel costs, loss of
earnings, expenses or other expenses which may occur.
The warranty does not cover normal wear and tear resulting from normal
use or damage caused by accident, misuse, negligence or improper
installation or assembly.
The vendor/manufacturer will not accept responsibility for any damages or
injury caused by misuse, negligence or non-observance of the instructions
supplied.
Imported & Distributed by:
HPU Medical Wholesale
34 Futurity Gate, Unit # 15
Concord Ontario L4K 1S6
Local: 416-739-1267
Toll Free: 1-888-634-5808