IMPORTANT NOTICE
•
This manual should be read and understood before operating.
•
This equipment can be wrong operated by which generates electromagnetic waves (mobile phone,
radiotelegraph, remote-controlled things etc). Please pull apart this equipment from those instruments.
•
When this manual is issued, all information in this manual are confirmed and concluded carefully. But EZER
is not responsible for the mistake or oversight of using this manual.
•
EZER can correct, improve and reform the product or its specification on EZER own sole authority without
any notice whenever have to do. And then, this manual may not contain those alteration sometimes.
•
This manual should be placed in a user
’
s sight.
EZER
ERK-5200

ERK-5200
EZER
9990 NW 14th Street Unit # 105
Doral, FL 33172
Phone: 786.621.0521
www.ezerusa.com
CONTENTS
IMPORTANT NOTICE
………………………………………
2
PRODUCT CONSTRUCTIONS
………………………………………
4
SPECIFICATION
………………………………………
4
SAFETY INFORMATION
………………………………………
5
1.
Special Ability
………………………………………
7
2.
Required Attentions
………………………………………
7
3.
The name and function of each parts
………………………………………
8
3-1
Body
………………………………………
8
3-2
Operating Board
………………………………………
10
4.
Practice with a model eye
………………………………………
12
5.
Measurement
………………………………………
14
5-1
R/K Mode
………………………………………
14
5-2
REF Mode
………………………………………
17
EZER
ERK-5200

ERK-5200
5-3
KER Mode
………………………………………
18
5-4
C/B Mode
………………………………………
20
6.
Other Functions
………………………………………
21
6-1
SIZE Mode
………………………………………
21
6-2
RET Mode
………………………………………
23
6-3
DISPLAY Mode
………………………………………
26
6-4
SETUP Mode
………………………………………
26
6-5
Power saving Mode
………………………………………
29
7.
Self diagnosis and repairing
………………………………………
30
7-1
Before calling to service engineer
………………………………………
30
7-2
Solving problem
………………………………………
31
7-3
Changing paper
………………………………………
31
7-4
Cleaning the equipment
………………………………………
32
7-5
Moving the equipment
………………………………………
32
7-6
Service information
………………………………………
32
Product Constructions
▣ERK-5200 Body
EZER
ERK-5200

ERK-5200
▣Model Eye
▣User Manual
▣Power Cable
▣Print Paper
▣Chin rest Paper
▣Dust Cover
Specifications
[Refractometry]
VD
0.0, 12.5, 13.5
SPH
-25.00 ~ +22.000D (VD= 12mm) (0.12 / 0.25D step)
CYL
0.00 ~ ±10.00D (0.12 / 0.25D step)
AX
1 ~ 180°
(1°step)
Astigmatism Indication
-, +, MIX
PD
10 ~ 85 mm
Minimal Pupil Size
Ø
2.0 mm
[Keratometry]
Cornea Curvature Radius
5.0 ~ 10.2 mm (0.1 mm step)
Cornea Refraction
33.00 ~ 67.50D (Cornea equivalence power=1.3375)(0.05 / 0.12 / 0.25D)
Cornea Cylinder Power
0.00 ~ -15.00D (0.05 / 0.12 / 0.25D step)
Cornea Cylinder Axis
1 ~ 180°(1°step)
Cornea Diameter
2.0 ~ 12.0 mm (0.1 mm step)
Memory
Can store 10times measurements result of both eyes
Printer
Built-in Printer
Monitor
6.4”TFT LCD Monitor
[Environmental Conditions]
Operating condition
Temp: +10℃~ + 40℃
Humidity: 30% ~ 85% RH
Keeping & moving
Temp: -10℃~ +55℃
Humidity: 10% ~ 95% RH
EZER
ERK-5200

ERK-5200
Power
AC100-240V, 50/60Hz
Power Consumption
Approximately 100VA
Size
300(W) ×500(D) ×430(H) mm
Weight
Approximately 20Kg
SAFETY INFORMATION
■
Symbols marked on the instrument
Type B EQUIPMENT
~
Alternating current
Protective earth (ground)
Attention, consult ACCOMPANYING DOCUMENTS
0
Off (power : disconnect to the mains)
1
On (power : connection to the mains)
Warning! When you see
‘
WARNING
’
or
‘
CAUTION
’
, please do as this manual says.
Please read and understand this manual enough before using and operating.
WARNING
The mark is that a danger could happen, unless be careful.
You or the other person could be injured seriously.
EZER
ERK-5200

ERK-5200
CAUTION
The mark is that a danger could happen, unless be careful.
You or the other person could be injured slightly and the equipment could
be damaged too.
NOTE
To emphasize the basic information.
To avoid using inappropriately, please read this information carefully.
■
General Safety Information
WARNING
•
Unless use this equipment in a standard power source, a fire or an electric shock
can happen.
•
Must connect or separate the electric wire after switching off the main power.
•
Never operate this equipment with a wet hand or can be injured by a serious
electric shock.
•
Never disassemble or try to modify this equipment yourself or a fire, an electric
shock can happen.
•
There are some dangerous parts inside of this equipment. Touching those parts in
carelessness, can make you to be injured or die.
•
Please switch off the main power and separate the electric wire from A/C socket
EZER
ERK-5200

ERK-5200
when the symptoms as occur belows.
- when it is smoky, smelling, noisy.
- when slop any liquid or drop any metal into the equipment.
- when drop this equipment or case
’
s broken.
CAUTION
•
To reduce the electric shock, please connect the cable into the grounded A/C
socket all the time.
•
Don
’
t put the object person
’
s hand or fingers on the lower part under the chin rest.
It
’
s dangerous and can be injured.
•
The forehead rest should be always cleaned with cleansing liquid like ethyl or
something to protect a every new patient from a contagion.
•
Chin rest paper should be replaced every time for a new patient for hygiene.
•
Do never put anyone
’
s hand or finger between the Stage and the Base. It could
be dangerous and injured.
EZER
ERK-5200


ERK-5200
•
Please keep clean the objective lens, or there can be a test error.
•
Do not use alcohol, thinner, benzene to clean this equipment. The equipment can be damaged.
•
Do use as the manufacturer’s lead when you use it with other machinery in connection.
•
If you find smokes, smells or noise, during you are using this equipment, pull the power out from the
socket and call to service engineer or distributor.
•
When you try to move this equipment, check the power off first. Then, fix the body with clamping screw
and hold the bottom of the body with both hands.
•During
you don't use it for a long time, disconnect the power cable from socket and put the dust cover
on it.
3-1 Body
3. Name and Function of each Parts
EZER
ERK-5200

ERK-5200
①Measurement Head
Performing measurement part.
②Height Indication
Set the height of eyes here with controlling the chin-rest.
③Chin Rest Control Button
Control the chin-rest.
④Monitor
Display current situation.
⑤Stage Clamping Screw
Fix the body.
⑥Measurement Button
Do the measurement.
⑦Control Lever
Control location and focus on.
⑧Print Button
Print the test result.
⑨Operating Board
All sorts of buttons here.
⑩Clamping Bolt
Fix all to move.
⑪Interface Device
Communication port to connect with other equipments.
③, ⑧Buttons
Front Side
EZER
ERK-5200

ERK-5200
①Power Switch
Power ON / OFF switch
②Fuse
Protect the equipment from overvoltage.
③Power Supply Socket
Connect with power cable.
④Product Label
Indicate the product name and some information.
⑤Forehead Rest
Forehead rest
⑥Measurement Window
Window where a patient should see
⑦Chin-rest
Chin-rest
Back Side
①, ②, ③Buttons
EZER
ERK-5200

ERK-5200
3-2 Operating Board
①MODE
Change the measurement mode
EZER
ERK-5200

ERK-5200
②AUTO
Set up Auto or Manual measurement
③SIZE
Measure the size of cornea
④ILLUM
Suspend the monitor and check the cornea or the contact lens
⑤IOL
Measure the eyesight of Cataract patients or artificial crystalline patients.
In this mode, PRK-9000 determines the condition of the eyes automatically
to detect even if there are IOL's or cataracts and measure them.
⑥DISPLAY
Show the result of the measurement
EZER
ERK-5200

ERK-5200
<Each Buttons and Modes>
Practice the measurement with a model eye before do it with a person.
(1) Release the stage clamping screw and bolt.
(2) Turn the switch on.
(3) Install the model eye
Install the model eye on the chin rest with the fixing pin.
4. Practice with a Model eye.
EZER
ERK-5200

ERK-5200
(4) Control the height of model eye
Press the chin rest control buttons and set the height of model eye like the picture above.
(5)
Converting to R/K Mode or REF Mode
Press the MODE button several time until the icon(R/K or REF) occur, unless there are one of them on the monitor.
Pupil Distance
Current setup
Frequency
EZER
ERK-5200

ERK-5200
<Picture 2>
(6) Setting focus and location of the model eye
①With watching monitor, control the lever to move and set the aiming point around the inside collimating mark.
②Move and set the aiming point on the middle of the inside collimating mark.
③Control the lever back and forward and make the mire ring to be seen clearly.
(7) Measurement
•
Manual Measurement
(a) Focus a point and location of the model eye in accordance with (6) above.
(b) Press the measurement button. Then if an error massage occur on the monitor, do (a) again and
press the button again.
(c) Check and compare the measurement result with the value of diopter written on lower part of the model eye.
Unless it is satisfied, measure and check gain as a same way.
•
Auto Measurement
Operate the auto measurement continuously.
(a) Press AUTO button on the operating board
(b) Focus a point and location of the model eye in accordance with (6) above.
(c) Measurement would start automatically when the Bright Dot occur in the inside collimating mark.
(d) Do the process (c) of Manual Measurement part.
Mire ring
Outside
Collimating Mark
Inside
Collimating Mark
Aiming Point
Left / Right
Mode
Measurement
Auto /Manual
Function of buttons
Test Frequency
Test result
EZER
ERK-5200

ERK-5200
5-1 Continuous Refraction/Cornea Measurement (R / K Mode)
In R / K Mode, you would do Cornea measurement first and Refraction measurement.
(1) Converting to R / K Mode
Press MODE button until 'R/K' icon displayed on the top left of the monitor.
<
Measurement Scene>
(2) Setting up the height of eye
Make a patient to sit on the chair and put his chin and forehead on the chin rest and forehead rest.
To set up the height of the eye, turn and control the chin rest control lever.
(3) Setting focus and measurement location.
(a) Control the lever and set the patient’s right eye to be seen on the monitor.
(b)
Make him to see the fixed target on the middle of inside
.
(c) With watching monitor, control the lever to make the ‘mire ring’to be seen clearly.
•
On the monitor, if the Mire ring is not seen clearly, request to the object man to open his eye widely until
when the measurement has finished. If it’s necessary, you raise his eyelid with your finger.
5. Measurement
EZER
ERK-5200

ERK-5200
(d) If you are not satisfied with controlling lever, you push the stage for all directions to focus.
•
When pupil and center of collimating mark do not fit, request to the object man to see the red roof of the
center of the picture.
(4) Measurement
(a) Press Measurement Button.
(b) After measuring, the result would be displayed on the monitor.
(d) When you measure continuously, the last result would be displayed.
(e) You can select an expression of Vertex Distance, the value of CYL and KER in USER's SET.
(Refer to USER's SET)
NOTES:
①If the pupil is out of the outside collimating mark, an error could happen.
②An error could happen by the modification of cornea after surgery or the lens pollution.
③During IOL switch is on, the trust-ability of measurement can be low.
④IOL button is only operated on 'R/K Mode' and 'REF Mode'.
⑤You can convert to IOL or normal with pressing IOL button.
(5) Continuous Measurement; You can measure continuously if it
'
s necessary.
(a)
Whenever you measure, the last result would be displayed
.
(b) Last 10 Measurement results but except 'error' can be remembered.
(c) Those results would be seen on the monitor in DISPLAY MODE.
(6) Measuring the eye other side.
(a) You measure a left eye to push the stage to right with the control lever.
(b) After the binocular measurement, the PD (pupil distance) would be displayed on the monitor.
•
If you push the stage to right eye without pressing the PRINT button, then the last measurement result of
right eye would be displayed on the monitor. Then you press the button, the new measurement result would
be added.
EZER
ERK-5200

ERK-5200
(7) Print
(a) Press PRINT button to print the result.
(b) It would be printed in accordance with SETUP MODE. (Refer to page. SETUP MODE)
(c) Pull
down
the printed paper to take off.
(d) Write the name of patient on 'NAME' if it is necessary.
NOTES:
①After Printing / Sending, all data would be removed.
②All printings are printed by heat. To keep the printing for a long time, make a copy of it.
<Controlling Location>
<After Measurement>
EZER ERK-5200
2007 / 05 / 28 09:33:45
NO: 00001
NAME:
------------------------------
[REF]
VD:
12. 0
Cyl.
Form:
(-)
<R>
SPH
CYL
AX
-5.00
+0.00
A
-5.00
+0.00
A
-5.00
+0.00
A
AVE
-5.00
+0.00
<L>
SPH
CYL
AX
-2.50
+0.00
A
-2.50
+0.00
A
-2.50
+0.00
A
AVE
-2.50
+0.00
A
PD = 63mm
[KER]
Index: 1.3375
<R>
mm
D
AX
R1
7.95
42.50
R2
7.95
42.50
AVE
7.95
42.50
CYL
<L>
mm
D
AX
R1
7.95
42.50
R2
7.95
42.50
AVE
7.95
42.50
CYL
EZER
786.621.0521
EZER
ERK-5200
Table of contents
Other EZER Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Nasco
Nasco Life/form LF01108U instruction manual

Carestream
Carestream DRX-Revolution Safety and regulatory information

Oximeter Plus
Oximeter Plus Oxi-Go Premier Operator's manual

MedPro
MedPro Defense AquaCare 770-300 manual

REMSTIM
REMSTIM 3000 operating instructions

Comau
Comau MATE Exoskeleton User handbook