First Medica Medflex Premium UC-03L User manual

Medflex Premium
2nd Generation
Ultrasonic Cleaning Unit
Owner’s Manual
For model: UC-03L & UC-05L
UC03L210420

Introduction
Thank you for purchasing a 2nd Generation Medflextm Premium ultrasonic
cleaning unit. The machine is manufactured from the highest quality materials
and features both a stainless steel cleaning tank as well as stainless exterior
surfaces. When used as directed in this owner’s manual your cleaner will
provide years of service.
This cleaning unit features innovative functions such as a “degas” cycle that
conditions the water for optimal cleaning prior to the cleaning cycle. It also
offers two power settings for cleaning, a high power setting for quick cleaning
of dirty instruments and a quieter, lower power setting.
Ultrasonic cleaning is far more effective than soaking instruments or even
scrubbing them. Very high frequency sound waves are propagated through
the cleaning medium and permeate every nook and cranny of the item to
be cleaned. This ensures that even the hardest to reach areas are cleaned
without damaging the surface through hard or abrasive scrubbing. The process
of ultrasonic cleaning is called “cavitation”. Cavitation generates millions of
microscopic bubbles in the cleaning solution. As these bubbles move outward
from the sonic elements they grow in size and eventually implode. When these
bubbles collapse the cleaning fluid surrounding the bubbles collapses with a
great force, generating shock wave pressure in excess of 20,000 pounds per
square inch.
The Medflex Premium ultrasonic unit is designed to remove bioburden and
debris from medical and dental instruments. Jewelry and other small items can
also be effectively cleaned. An ultrasonic unit physically removes debris but
does not disinfect or sterilize.

Packing list
Ultrasonic Unit
Basket
Power cord
Owner’s manual
Specifications:
UC-03L UC-05L
120v, 50/60hz
100 watts 300 watts
33-40kHz cleaning frequency
External dimensions:
H = 9.4” H = 12.2”
W = 6.7” W = 7.1”
L = 10.8” L = 13.0”
Tank dimensions:
H = 3.9” H = 5.9”
W = 5.5” W = 6.1”
L = 9.4” L = 11.8”
Basket dimensions:
H = 3.0” H = 5.0”
W = 4.8” W = 5.5”
L = 8.7” L = 11.5”

Cautions
•Do not use flammable cleaning solutions.
•Use caution when inserting or removing cleaning baskets if the heat
function has been activated as the solution may be quite hot.
•Do not immerse the unit in any liquid.
•Do not use around flammable liquids or gasses.
•Ultrasonic cleaning does NOT disinfect or sterilize. Always use
proper disinfection/sterilizing practices on instruments after cleaning
with the ultrasonic unit.
•Always rinse instruments and items to be cleaned prior to putting in
ultrasonic unit.
•Check owner’s manual of item to be cleaned to be sure that it is
suitable to be cleaned by an ultrasonic unit.
•Unplug the unit when changing cleaning solutions if not using the
drain valve.
•Do not turn unit on unless the tank is at least 1/3 full.
•Always follow manufacturer’s cleaning instructions for any
instruments or items cleaned.
•Do not use acetone or phenol based cleaners such as Lysol to clean
the front control panel. Phenol may damage the plastic.
•If enzymatic cleaners are used they need to be replaced daily
•Be sure to follow all EPA and local regulations when draining
cleaning solutions.
•Typical cleaning times are from 3 to 30 minutes at temperatures
about 110oF. Follow cleaning instructions of each item cleaned and
of cleaning solutions.

Use of Ultrasonic Cleaning Unit
Before using:
1. Make sure that the drain valve is in the closed position
(5L model only – the 3L model does not have a drain)
Valve closed position:
2. The U2-5l drain fixture is threaded with standard ½” plumbing
fitting that can be found at any hardware store. These fittings
are inexpensive and allow the unit to be directly connected to the
plumbing or by using a barb and some hose it can drain into a sink.
ONLY USE PLASTIC FITTINGS ON THE STAINLESS STEEL THREADED
VALVE.
3. Fill the cleaning tank with solution to at least 1/3 full but no more than
1” from the top of the tank.
4. Plug the power cord to the back of the unit and then into a properly
grounded wall outlet.
Power line in here
5. For maximum cleaning power run the ultrasonic unit on the “degas”
setting with cleaning solution in it for 15 minutes to remove any air
bubbles in the solution.

Control Panel
Temp Cleaning
oF time
Temperature Heat Degas Regular High Time
Adjust On/Off On/Off Power Power Adjust
To Use:
1. Select desired cleaning time
2. If heating is needed select proper temperature (oF)
3. If heating is desired the cleaning solution will need to be preheated for
about an hour, turn heat on. The temperature can be changed while
the ultrasonic function is running. . The heat light will not illuminate if
the solution is at or above the set temperature.
4. For optimal cleaning select the degas function for about 15 minutes to
condition the cleaning liquid.
5. Place items to be cleaned in cleaning basket and lower into solution
(always use a cleaning basket, do not rest items on bottom of
ultrasonic unit). Rinse items to be cleaned prior to placing in
ultrasonic unit.
6. Place lid over unit to prevent splashing.
7. Select either regular or high power and press the appropriate button
to initiate the ultrasonic cleaning cycle.
8. When cycle is complete remove clean items. Always make sure that
ultrasonic cleaner is turned off before reaching in the tank.
9. Change the cleaning solution if it becomes discolored or has
noticeable contaminants

Cleaning & Maintenance
•After unplugging the ultrasonic unit the inner tank and cleaning trays
can be cleaned with hot soapy water.
•Use Perimotm or commercial spray cleaners such as Windex to clean
the exterior of the ultrasonic unit. Do not use phenol or acetone
based cleaners as they may damage the plastic. Do not immerse the
unit in any liquid.
Warranty
First Medica is pleased to offer a 1 year limited warranty on all Medflex
Premium ultrasonic cleaning units.
Warranty is valid for 1 year from date of initial purchase by an end user.
Warranty is transferable but warranty period is dated from original purchase.
Prior to sending in for warranty service please call First Medica at
(800) 777-7072 or e-mail: [email protected]
Warranty covers defects in manufacture. Warranty does not cover use
contrary to intended use. Warranty does not cover physical damage caused to
a unit by dropping or other physical abuse.
For warranty service send your Medflex Premium ultrasonic cleaner, securely
packaged along with a description of problem and your return address to:
First Medica Repairs
13030 Homer Smith Rd.
Piedmont, SD 57769
___________________ ____________________
Purchase Date Purchased From
_____________________
Serial Number
Manufactured in China exclusively for First Medica Inc.
www.firstmedica.com
This manual suits for next models