▪Procedure:
1. Push the centre of the navigation pad.
2. On the toolbar, select Settings .
3. In the dialog box, select Measurement parameters.
4. In the dialog box, select Reflected apparent temperature and then set a value.
J. Changing the distance between the object and the camera
▪To measure temperatures accurately, the camera requires the distance between the
camera and the object.
▪Procedure to set the distance:
1. Push the centre of the navigation pad.
2. On the toolbar, select Settings .
3. In the dialog box, select Measurement parameters.
4. In the dialog box, select Distance - where you can select a distance.
K.Saving an image
▪Pressing the trigger button captures and saves the IR image into the memory card.
▪Image capacity - Approximately 500 images can be saved to the internal camera
memory.
▪Naming convention - the naming convention for images is FLIRxxxx.jpg, where xxxx
is a unique counter.
L. Cleaning the camera
▪Use a soft cloth and one of these liquids: (a) warm water, or (b) a weak detergent
solution
▪Procedure to clean the camera:
1. Soak the cloth in the liquid.
2. Twist the cloth to remove excess liquid.
3. Clean the part with the cloth. CAUTION: Do not apply solvents or similar liquids
to the camera, the cables, or other items. This can cause damage.
M. Cleaning the IR lens
▪Use cotton wool and one of these liquids: (a) commercial lens cleaning liquid with
more than 30% isopropyl alcohol, or (b) 96% ethyl alcohol (C2H5OH).
CAUTION: If you use a lens cleaning cloth it must be dry. Do not use a lens cleaning
cloth with the water or detergents. These liquids can cause material on the lens
cleaning cloth to become loose.