Model LI1X Universal Process Meter Instruction ManualModel LI1X Universal Process Meter Instruction Manual
Select Input Signal
The meter can be programmed to accept current or voltage input
using the Signal Input Selection Array.This jumper array is located at
the rear of the instrument, next to the screw terminal block. Remove
jumper J3 (if installed) to disable Lockout feature.
Table 1: Signal Input Selection Array
Initial Calibration with External Source (If Needed)
Initial calibration is required only when the I-CAL feature is to be used
with an input other than the default input of 4-20 mA. For best results,
allow the meter to warm up for at least 30 minutes.
1. Program the Signal Input Selection Array for desired input per
Table 1 above.
2. Apply power with ENTER and ACK buttons pushed. Release when
display stops flashing.
3. The meter scrolls through the choices 20, 5, and 10. Which
corresponds to input signals of 4-20 mA, 0-5V or 1-5V, and 0-10V.
Push ENTER when desired value appears.
4. The meter flashes InPt 1. Apply low end signal, (4 mA or 0 V)*,
push ENTER.
5. The meter flashes
InPt 2
. Apply high end signal, (20 mA, 5V,
10 V), push ENTER.
6. This completes the Initial Calibration and initializes INPUT 1 and
INPUT 2 scaling points to 4.000 & 20.000 mA, 0.000 & 5.000 V, or
0.000 & 10.000V.
7. Refer to I-CAL scaling below for further instructions.
*For 1-5 V input signals, calibrate the meter with input1=0 V,
input 2=5 V.
20
Input Signal Jumper J1 Jumper J2
1-5, 0-5V OFF OFF
0-10 V OFF ON
4-20,0-20 mA ON OFF
21
Select Calibration Method
The meter may be calibrated using an external signal source such as a
calibrator (E-CAL), or scaled using the internal source (II-CAL).With
I-CAL, a 4-20 mA input can be scaled for any display range without
applying a signal.To scale inputs such as 0-5V, 1-5V or 0-10 V without
applying a signal, it is necessary to first complete an Initial Calibration,
see Initial Calibration with External Source (If Needed) on page 20.
Scale or Calibrate the Meter
To select a Calibration Method, apply power with ACK button pressed
.
1. When display stops flashing, release ACK button. Display
alternates between E-CAL and I-CAL.
2. To calibrate meter with an external source, such as a calibrator,
press ENTER when E-CAL appears.
3. To scale meter with internal source, press ENTER when I-CAL
appears.
Select Decimal Point (dECPt)
To program the decimal point or light up the extra "0", push ENTER.
The meter displays SCALe or Calib, then dECPt. Push ENTER when
dECPt appears.The meter then reads 199990 with the decimal point
changing locations and the extra "0" illuminating. Push ENTER when
the decimal point is in the desired position; or if the extra "0" is
desired; when the extra "0" is displayed.
Select Linear Input (L or S)
The meter can be programmed to either process the input signal in a
linear fashion (L) or to take the square root of it (S).To program the
meter for a linear input, push ENTER.When (L or S) appears, push
ENTER again.The meter will flash (L or S) alternately. Press ENTER
when (L) is flashing. After (L) is selected the meter will switch to indi-
cation mode and is ready to be calibrated.
Select Number of Calibration Points (no. PtS )
The user may input up to 11 calibration points and corresponding
displays to provide custom linearization.To program the number of
calibration points, push ENTER, when (no. PtS) appears, push ENTER
again.The meter displays a number from 2 to 11 that represents the
number of calibration points. Push ENTER when the number equals
the desired number of calibration points. As soon as the number of
points is programmed, the meter flashes InPt 1 (Indicating it is ready
to be calibrated). Refer to Scale Using Internal Calibration (I-Cal),
page 22 or Calibrate Using an External Calibrator
(e-Cal) on page 23.