FLOWTECH KF700H Series User manual

ISO9001-2000
NINGBO KIO FLOW INSTRUMENTS CO.,LTD
KF700H Series
Electromagnetic Flowmeter
OPERATION INSTRUMENTS

CATALOGS OF SENSORS
1. Operation principle
2. Major feature and application
3. Structure
4. Appearance and install size
5. Flow Specifications
6. Installation
7.Installation position of sensor
8. Connecting cable
9. The preparation of pre-operational
10. Maintenance and common trouble shooting
11. Complete sets of products and Open Box
12. Quality Assurance
13. Transport and storage
14. Order Information
15. Associated instruments part number and type
16. The Corrosion resistance of Common electrode materials
17. Performance and application of Common lining materials
1
1
2
3
4
5
7
8
8
9
10
10
11
11
11
11
12

CATALOGS OF TRANSIMITTER
I.Installation
1.Working condition
2.Installation Notes
II.Power Line connectiong
1.Instrument Power supply
2.Power Line connectiong
3.Output signal
III.Instrument started
1.Working instrument showing
2.Interface show
3.Power supply fuse replacement
IV.Operations
1.Panel construction and ket definition
2.Menu construction
3.Select menu item meassurement mode
4.Operation guide for regular function of transmitter
1.)Zero Correction
2)Damping time
3)Instantaneous flow Resolution
4)Cumulative total flow resolution
5)Scale of Flow
6)Small flow termination%(low %)
7)Flow direction
8.) The flow of indication
9.) Frequency upper limit Hz
10.) Liter/pulse
11.) Cumulate Clear
12.) Empty Trim
13.) Full Trim and Tube region setup
14.) Unit of Flow
15.) Unit of Accumulation
16.) Direction of Total
Ⅴ.Technical Data
Ⅵ.Type selection
Ⅶ.Error message
Ⅷ.Common failures and how to deal with
1、Not Flow signal show on LCD display
2、Zero point Instability
3、Real flow data different with Instrument show
ⅧI. Transport, storage
13
13
14
14
15
17
17
18
18
19
20
21
21
22
23
24
25
26
27
29
30
31
33
34
35
37
38
38
39
39
40
41
42
43
44

01
Http: // www.flowmeter.cc
1. Operation Principle
<>The operation of a magnetic flowmeter or mag meter is based upon Faraday's Law, which
states that the voltage induced across any conductor as it moves at right angles through a
magnetic field is proportional to the velocity of that conductor.
Faraday's Formula:
E is proportional to V x B x D where:
E = The voltage generated in a conductor
V = The velocity of the conductor
B = The magnetic field strength
D = The length of the conductor
To apply this principle to flow measurement with a magnetic flowmeter, it is necessary first to state
that the fluid being measured must be electrically conductive for the Faraday principle to apply. As
applied to the design of magnetic flowmeters, Faraday's Law indicates that signal voltage (E) is
dependent on the average liquid velocity (V) the magnetic field strength (B) and the length of the
conductor (D) (which in this instance is the distance between the electrodes).In the case of
wafer-style magnetic flowmeters, a magnetic field is established throughout the entire
cross-section of the flow tube (Figure 1). If this magnetic field is considered as the measuring
element of the magnetic flowmeter, it can be seen that the measuring element is exposed to the
hydraulic conditions throughout the entire cross-section of the flowmeter. With insertion-style
flowmeters, the magnetic field radiates outward from the inserted probe (Figure 2).
2. Main Features andApplications
Electromagnetic flowmeter consists of two parts of the sensor and transmitter. KF700-type
electromagnetic flowmeter for measuring a variety of acid, alkali, salt solution, paper pulp, slurry
and other conductive liquid or liquid-solid two-phase medium volume flow. In the chemical, alloy,
water supply and drainage, sewage treatment, food, sugar, paper making, environmental protection
and other departments to be widely application.
Following characteristics of sensor:
(1) The whole welded structure,good sealing performance;
functional diagram
SENSORS PARTS
SENSORS

02
Http: // www.flowmeter.cc
3.Structure:
(2) Structure is simple and reliable,no moving parts inside;virtually no pressure loss
(3) low-frequency square wave excitation,anti-jamming performance,zero stability;
(4) The instrument measured medium has nothing to do with pressure, viscosity, temperature,
density and other physical parameters of the impact of changes
(5) The instrument reflects the sensitivity; the output signal has a linear relationship with flow. Width
of measurement;
(6) As measured medium only measuring tube lining and electrode contact, easy to satisfy
anti-corrosion, anti-wear requirements;
(7) Power consumption is small, complete set of instrument power consumption <10VA, has
nothing to do with the size of the sensor aperture;
(8) To install, usage, and easy maintenance.
Structure diagram
1.Junction box; 2.flange;
3.insulated liner; 4.electrode; 5.measuring tube;
6.excitation coil; 7.shell
Sensor structure shown in Figure 2, the following components:
(1) Measuring tube: Measuring tube flow measured medium, measuring tube by the non-magnetic
stainless steel and welded flange, lined with insulation lining.
(2) Lining: the inside of the measuring tube and the flange sealing surface of a complete corrosion
resistant electrical insulation material, to prevent the traffic signal is short-circuit.
(3) Excitation system: measurement of the outer tube from top to bottom with a group of coils to
generate magnetic fields of work.
(4) Electrode: with the magnetic field lines perpendicular to the direction of the measurement pipe
wall with a pair of electrodes to detect flow signals, the electrode materials can be corrosion
testing medium selected.
(5) Shell: protecting the instrument and steal up it.
SENSORS PARTS
SENSORS

4.Appearance and install size
03
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

DN Common choice of full scale flow range(m3/h)
10
15
20
25
32
40
50
65
80
100
125
150
200
250
300
350
400
450
500
600
700
800
900
1000
1200
1400
0.16, 0.2, 0.25, 0.3, 0.4,0.5,0.6,0.8,1.0,1.2,1.6,2.0,2.5
0.4,0.5,0.6,0.8,1.0,1.2,1.6,2.0,2.5,3.0,4.0,5.0,6.0
0.6,0.8,1.0,1.2,1.6,2.0,2.5,3.0,4.0,5.0,6.0,8.0,10.0,12.0
1.0,1.2,1.6,2.0,2.5,3.0,4.0,5.0,6.0,8.0,10.0,12.0,14.0,16.0
1.6,2.0,2.5,3.0,4.0,5.0,6.0,8.0,10.0,12.16,20,25
2.5, 3.0, 4.0, 5.0, 6.0, 8.0,10.0,12,16,20,25,30,40
4.0, 5.0, 6.0, 8.0, 10, 12,16,20,25,30,40,50,60,70
6.0,8.0,10,12,16,20,25,30,40,50,60,80,100,120
10,12,16,20,25,30,40,50,60,80,100,120,160
16,20,25,30,40,50,60,80,100,120,160,200,250
25,30,40,50,60,80,100,120,160,200,250,300,400
40,50,60,80,100,120,160,200,250,300,400,500,600
60,80,100,120,160,200,250,300,400,500,600,800,1000
100,120,160,200,250,300,400,500,600,800,1000,1200,1600
160,200,250,300,400,500,600,800,1000,1200,1600,2000,2500
200,250,300,400,500,600,800,1000,1200,1600,2000,2500,3000
250,300,400,500,600,800,1000,1200,1600,2000,2500,3000,4000
300,400,500,600,800,1000,1200,1600,2000,2500,3000,4000,5000
400,500,600,800,1000,1200,1600,2000,2500,3000,4000,5000,6000
600,800,1000,1200,1600,2000,2500,3000,4000,5000,6000,10000
800,1000,1200,1600,2000,2500,3000,4000,5000,6000,10000,12000
1000,1200,1600,2000,2500,3000,4000,5000,6000,10000,12000,16000
1200,1600,2000,2500,3000,4000,5000,6000,8000,10000,12000,16000
1600,2000,2500,3000,4000,5000,6000,10000,12000,16000,20000
2500,3000,4000,5000,6000,8000,10000,12000,16000,20000,25000,30000
3000,4000,5000,6000,8000,10000,12000,16000,20000,25000,30000,40000
5. Specifications
1). Max flow range for EMF (Choice of reference map) :
04
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

6. Installation
1.) Installation ways
3 .) The choice of installation location
2.) The choice of installation environment
2.) Flowmeter curve graph of the relationship between Diameter, flow rate and flow
Electromagnetic flow sensor if not installed properly, will significantly affect measurement
accuracy, even lead instrument not working properly, so Before installation instructions
carefully read the relevant sections.
Sensor installation method for the flange connection. Criteria and process pipe welding flange
screw holes in the inter-position, bolt can be passed smoothly to facilitate the integration
between the sensor and process piping.
Installation must ensure that sensors Center and the process pipe center line, and then a good
grounding line, otherwise it would cause measurement error.
According to the work of instrument characteristics and technical characteristics, the choice of
instrument installation environment should pay attention to:
(1) The instrument should be installed at the ventilation to dry, avoid installing at the place easy
accumulation water.
(2) The instrument should try to avoid just sun shine and rain. Open-air installation, should
keep out rain in places;
(3) Installation of places as much as possible to avoid strong vibration;
(4) as far as possible to avoid a strong electromagnetic field equipment; such as large motors,
large transformers.
(5) Select the ease of maintenance, activities, convenient place.
Installed on the sensor in the pipeline, we should note the following points
(1) sign the direction of the flow sensors and pipeline flows in the same direction within the
medium;
(2) The need to ensure that sensor tube is full filled with the measured medium;
(3) The sensors should be five times the upstream straight pipe section D above, the sensor
should be three times the downstream straight pipe section D above (available from the center
of sensor , D to measure the pipe diameter);
SENSORS PARTS
05
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

图3
picture 4 Install reducing pipe or gradually expanding tub to effect pressure loss
(d/D)
4. ) When the pipe diameter is inconsistent with the sensor, the sensor is installed at both ends tapered
or gradually expanding tube, and then with the pipe connection. Gradually expanding, tapered conical
tube should be no more than 15 degrees. When using 15 cone angle gradually expanding, tapered tube,
the pressure loss resulting from the curve in Figure 4 :
d.Sensor internal diameter; D.diameter, V.flow rate of
sensor(m/s)
5) sensor installation should be taken in the horizontal electrode as horizontal position, once the media
containing bubbles or sediment, the bubbles will not be adsorbed on the electrode in the vicinity, resulting
in conversion of the input terminal open circuit; sediment will not cover the electrode, caused by zero drift;
(6) liquid-solid two-phase medium, the vertical installation of a more favorable, a measured medium can
prevent phase separation, two lining wear sensor allows more uniform. Vertical installation, the medium
flow direction should be bottom-up, so as to ensure that the sensor tube is always filled with medium.
PRESSURE LOSES (Pa)
06
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

C/CE
7. Installation position of sensor
Pipe to the highest point (Gathered bubble
in tube to make error measurement)
Preferred position
(Sensor cannot be installed at highest level of pipeline)
Easily lead to non-full
pipes (wrong measurement way)
Pls put flow meter into low position,
exposure for Injection or discharge Pls install exhaust valve if
drop pipe longer than 5M.
5m
Exhaust valve
picture 5 recommend installation
In order to enable reliable flow meter, Pls note the following installation requirements:
1. As far as possible to avoid ferromagnetic objects and equipment with strong magnetic fields
(big electric machine, big transformer) is protecting of the magnetic field sensor.
2.Flow meter should be enough space around for easy installation and Maintenance.
In longer pipe system, pls install
control valve and Cut-off valve on
the flow meter Downstream side
Pump: Prohibited to install
flowmeter at suction
side(Vacuum)
07
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS
C/CE
C/CE
C/CE
C/CE
C/CE
C/CE
C/CE

8. Connecting cable
BOM 1 -Connecting cable
Items
Signal lines Accessory kit
Accessory kit
provided by users
8 or 15m
8 or 15m
1
2
3
PVC sheathed 2 core shielded cable 2X16/015 (SBWP)
PVC sheathed 2 core shielded cable 2X80/015 RWP
Ship sealed with rubber insulated cable 2X0.5
Two core plastic cable 2X1.0mm2(YHZ)
Marine soft ethylene-propylene rubber
insulated cable 2X1.0(CEFR-C)
General Color plastic
double-stranded copper wire
Excitation wire
Output signal lines
Name specifications Qua. Remarks
SENSORS PARTS
08
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

9. The preparation of pre-operational
Important Note: the whole instrumentation (including sensors and Transmitter) to run a rigorous
adjustment and flow calibration in the factory, one by one after passing the test
In Plants. To the end user, So, it can be put into operation without any adjustments. Therefore, the initial
operation of the problems encountered, should be in accordance with the specifications Reference points
one by one inspection, a careful analysis, troubleshooting.Avoid Blindly tamper to make a good set of
adjustments to the original instrument to confuse or even damage.
Instrument can be put into operation the following steps:
1) First open upstream and downstream the valve with Sensor, so that sensors measuring tube filled with
test medium;
2. ) Power for one minute, Transmitter showed an immediate increase to a certain numerical value, pls
wiring is correct or not, when the flow direction is wrong,pls
Check the direction of the sensor installation is correct.
3) zero Trim , instrument through the media 15 minutes after the first sensor tightly closed valve
downstream side, and then shut down the upstream side of the valve, so that tube
Stop the flow of fluid and no leakage, the flow shows zero. Zero if too high or too low, Pls do Zero Trim on
the Transmitter. Specific operations see Transmitter Manual p. 27
SENSORS PARTS
09
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

10. Common failures and how to deal with
11. Open box and product sets
Check packing box, and check the sensor model, the contract specifications in line with
the order randomized complete documents
KF700 series EMF manual instructions
1 pcs
1 pcs
1 pcs
1 pcs
certification
packing list
CD
Regular failure and obviate method
Troubles
Use a multimeter to check signal good or not
Use a multimeter to check sensor loop is good or not
According transimitter manual inspection,
troubleshooting
Grounding bad or electrode dirty, Zero
Trim after inspections
Adjusted in accordance with the parameters
set, and then zero trim
Inspection process to improve the
installation method
Clear scaling
Carried out using standard flow meter to compare
2.Signal circuit break
2.medium in tube not full
2.Transmitter configure be modified
3.Imperfect earth
3.Medium in tube not full
4.Electrode or wall scaling
3.Open-loop excitation
4.Transmitter failure
Possible Causes Check the troubleshooting
Failure to deal with reference to Table
Check signal line to ground resistance to use a
multimeter measuring electrode-to-ground
resistance, generally from thousands of OM to
some 10k om
Electromagnetic flow sensors generally do not require regular maintenance, but medium of measured
is easily adhesion in the electrode or dielectric tube wall measurements or scaling of the occasion, the
need for regular cleaning pipe wall of measurement and electrodes, attention do not damage liner,
electrode.
Liquid flow
through the
meter ithout
instructions
Changes in
the flow
apparatus
showed that
full-scale on
transmitter
Instrumented
inconsistent
with the actual
flow
1.Moisture sensor or signal-to-ground
short-circuit resulting in damage
1.A signal-to-ground short-circuit or
open circuit
1.Change the zero point to make
measurement error
5.error in actual determination to
check with Flowmeter
Use a multimeter to check insulator of
signal line whether good or not
SENSORS PARTS
Use a multimeter to check signal lines open
or not and to improve the installation method
Check the signal shielding layer and then
place resistors, re-installation of grounding
devices
10
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

12. Quality Assurance
13. Transport and storage
14. Order Information
15. Instrument serial number and model matching
(1) sensor ID
(2) sensor coefficient
(3) Converter Model
(4)other
Instruments and accessories from the factory from the date of 12 months, when a user products in
full compliance with the technical requirements and installation instructions provided in the
transport, installation and use of the provisions of instruments and accessories have found
products that do not meet technical standards, the instrument may be returned factory, this factory
is responsible for free repair.
Instruments to prevent damage in transit, on arrival at the scene after the installation, please keep
the state at the time of factory shipment, storage location of the indoor conditions on the follow:
a) ventilation, rain, moisture, indoor air should not contain the harmful effects of corrosive
substances;
b) a small mechanical vibrations and to avoid the impact;
c) temperature range in -20°C~ +60°C
d) do not dry 90% humidity
Orders must be measured in accordance with specific targets and measuring conditions of
access to the Company on the selection of technical information and ordering the right choice. E.
According to the actual situation, the order should be determined:
1) the scope of the model and flow measurement. Order to determine, based on sensor diameter
models. Full-scale instrumentation (ie, range) should not be less than the actual measured flow of
the largest pipeline, and the normal flow of more than 50% of the selected range in order to
obtain high precision.
2) pressure meter, temperature must meet the test medium pressure and temperature.
3) come into contact with the measured medium liner, electrode corrosion media should be
capability measured. Therefore, users must be ordered according to their own experience of
anti-corrosion,
Refer to the company's existing varieties of lining and electrode materials (see 12th page), the
correct selection of materials.
4) the need for installation with matching flanges, please specify when ordering.
11
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

16. Common electrode materials corrosion performance
17. Common properties of lining materials and application.
Lining material
PTFE -80~250°C
0~80°C
0~60°C
Chloroprene
rubber
Polyurethane
Rubber
The main performanceThe main performance Application
temperature
HastelloyB
HastelloyC
Ti
Acid-resistant
steel
1Cr18Ni9Ti
Corrosion performance
To nitric acid, phosphoric acid and other cold inorganic acid,a variety of salt and
alkali solutions, organic acids, water-resistant for a good corrosion resistance.
Formic acid of boiling, Oxalic acid , industrial acid-ming, as well as sodium
carbonate and chlorine, bromine, iodine, such as poor chemical stability of
medium, not corrosion.
Stainless steel
containing
molybdenum
0Cr18Ni12Mo2Ti
0Cr18Ni12Mo3Ti
In the reductive medium (such as hydrochloric acid) compare with 1Crl8N 9Ti
have stronger corrosion resistance.Less than 50% of the nitric acid, at room
temperature less than 50% of sulfuric acid and 20% of the hydrochloric acid,
alkali solution, boiling phosphoric acid, formic acid, under pressure from a certain
sub-sulfuric acid, water, acetic acid and other media have a strong corrosion
resistance , can be widely used in petrochemical, urea, vinylon industry.Intolerant
of hydrofluoric acid, chlorine, bromine, iodine and other medium.
Below Boiling point of all of the following concentrations of hydrochloric acid with
good corrosion resistance, is also resistant to sulfuric acid, phosphoric acid,
hydrofluoric acid, organic acids and other non-oxidizing acids, alkalis, salt
solution of non-oxidative corrosion.
Oxidation of acid-resistant, such as nitric acid, mixed acid or chromic acid and
sulfuric acid corrosion of mixed medium, but also resistant to oxidation of the
salts such as Fe ", Cu +2 corrosion or other oxidants. Such as higher than normal
temperature of the hypochlorite solution. Seawater corrosion resistance is very
good
Medium resistance of oxygen and nitric acid, chloride, hypochlorite and chlorine
the corrosion resistance of a good medium.
Material
It is the most stable plastic
material. Ability to boiling Of
hydrochloric acid, sulfuric acid,
nitric acid and aqua regia, but also
strong capability Alkali, organic
agent. Molten alkali metals and
intolerance ,Its ammonia
solution, abrasion resistance and
poor bonding
Have good flexibility, high tensile
strength, abrasion resistance,
good impact resistance. Acid,
alkali, salt and other corrosive
media. Intolerance oxidative
corrosion media
Has an excellent wear resistance,
which is equivalent to ten times
that of natural rubber.
Has an excellent wear
resistance, which is
equivalent to the pressure
of natural rubber pipe
measurement acids,
alkalis,Like salt, strong
corrosive medium or
medium health category.
A non-oxidizing
acids, alkalis,
salt solution.
Oil drilling, slurry, mud,
Mortar serious occasions,
such as wear and tear
12
SENSORS PARTS
Http: // www.flowmeter.cc
SENSORS

Ⅰ. Installation
2.) Installation environment
2)The working conditions::
a) ambient temperature:-20°C~55°C
b) relative humidity: 5% -90%
c) power supply
AC Power Supply:85VAC~265VAC,50Hz
DC Power Supply:18VDC~36VDC
Rated Power:<10VA(include sensor)
1. Working condition
2. Installation Notes
Transmitter’s configuration in accordance with the contract requirements, run the data set in
accordance with contract requirements. Users can check the meter nameplate to confirm
that the instrument provided by the model and instrument operating parameters set. It can
be put into use after connect power. In order to ensure your system run on normal
operation, make sure that the sensor measurement is full of medium channels
Our company electromagnetic flowmeter is dedicated to conductive fluid (conductivity ≥ 5μS/cm,
water ≥ 20μS / cm) the volume of flow measurement.
Flowmeter installation and usage should be in strict accordance with the specification, and to comply
with the relevant national standards, safety requirements and accident prevention requirements.
MF7200 electromagnetic flow converter directly connected with the electromagnetic flow sensor
compact into one body
It should avoid large temperature changes. If the meter installed by the thermal radiation Plateau,
please provide the thermal isolation or ventilation. Instrumentation installed in the switch box, there
should be appropriate measures of net and ventilation are examples such as the fan. Compact
structure flow meter should give full consideration to the transmitter working environment
temperature.
1.)The working conditions of the reference test:
a) operating temperature:20°C±2°C
b) relative humidity:45%~85%
c) the supply voltage:220VAC±2%
d) Power frequency:50Hz±5%
e) harmonic content:<5%
f) warm-up time:>15min
It should avoid in a strong corrosive atmospheric environment. Installation location should have
adequate ventilation. Protective properties of instruments should be maintained to prevent the
corrosive gases and moisture into the instrument cavity. It Should avoid direct sunlight, especially
the liquid crystal display part. Strong vibration should be avoided.
1. ) Environment temperature
3.) Installation methods
A compact body flowmeter transmitter and Sensor two parts is connected to the end before out
of factory, the user can be installed directly. Sub-constructer flowmeter for Transmitter and
sensors to be in two parts at field by the user through a dedicated cable (factory annex) to
connect. Part of the sensors installed on both glycosides, Transmitter are usually installed in the
indoor or meter box. The distance is shorter for better.
13
CONVERT PARTS
Http: // www.flowmeter.cc
CONVERT

Ⅱ.
Power line connecting
Operation principle and requirements:
This series of transmitter have the following power supply types.
1) AC Power supply range
85VAC-265VAC, 50Hz power ~ 10VA (including sensors)
2) DC power supply range
18VDC 36VDC; power ≤ 10W (including sensors)
2. Power line connecting
Operation steps:
Before connecting the power supply, please see the nameplate and part of the electrical terminals
and the types of instructions ,in order to avoid misoperation to make instrument error or damage
1. Instrument Power supply
Instrumentation and electrical installation work must be connected by a certain qualified staff to
implement. Be sure to operate before reading this operation manual carefully to avoid wrong operation
caused by solid work instrument error or damage. Converter must have a good grounding in order to
protect their personal safety. Use in hazardous area must be provided by columns according to special
"EX" in the operating manual instructions. Instrument should be in the electrical connection before
power run.
1.The lid of Instrument just open in the wiring, once opening the lid of instrument will affect protection
performance. It should be back to the factory sealed lid state after the wiring finished.
2, In order to ensure the insulation of instrument performance, to prevent the instrument as a result of wet
insulation caused by poor rains ,in the outdoor wiring should be avoided.
3, with a threading instrument connectors, cable connectors must be approved by
threading into the instrument cavity (see circle). After the wiring , waterproof joints
should be tightened to ensure that moisture and corrosive gases do not enter
the electronic part.
4, Input and output lines recommend the use of the installation of conduit,
conduit tube be recommended thick solid brass or soft metal pipes Road. Conduit should be noted that the order
to prevent flood water flowing into the conduit or conduit into the internal instrumentation. Input power line and
signal lines should be separated through their respective special perforated cavity into the instrument, prohibited
parallel and banding together.
5, The choice of input and output cables to reference the following description:
Excitation, a dedicated flow meter signal cable is provided by our company.
Power Line: owned by the user, customer can option YH Z-2xlmm2 two soft-core rubber insulated cable, cable
length should concern voltage drop.
Output line: owned by the user, customer can option RVVP2X16 / 0 15 PVC insulated sheathed wire shielding,
cable length will affect the load characteristics.
1, open the rear cover of the transmitter
2, a dedicated power supply input line threading through the hole (Waterproof connector) into the instrument
cavity.
3, the grounding wire connected to the transmitter side of the ground.
4, AC (AC) power lines L and N lines were connected to the converter side of the L and N-side.
Direct current (DC) power supply to power + lines and - receiving line side converter, respectively, the +
terminal and - terminal.
5, water-proof connector threading screw instrumentation and to return to the status of the factory sealed.
CONVERT PARTS
14
Http: // www.flowmeter.cc
CONVERT

L
N 85 VAC-265VAC 50Hz
+ 24 VDC Power, positive terminal
- 24 VDC Power, negative terminal 18 VDC -36VDC
3. Output signal
Diagram of output signal terminal
Terminal Description
Diagram of output signal terminal
Terminal Description
Description:
FUSE
L
NAC220V
4-20mA
0-2KHZ
RS DATE
+
-
+
-
+
-
FUSE
+
-DC24V
4-20mA
0-2KHZ
RS DATE
+
-
+
-
+
-
FUSE
L
NAC220V
4-20mA
0-2KHZ
RS DATE
+
-
+
-
+
-
FUSE
+
_DC24V
4-20mA
0-2KHZ
RS DATE
+
-
+
-
+
-
Identifier Functional Description
220VAC Power, L terminal
220VAC Power, Nterminal
Power Ground terminal
Power range:
Power range:
Grounding resistance <= 10 Ω
Pay attention to the following questions on connect the power supply :
1, the power cable of inside the sensor will not wound.
2, the power input line should separated with other input and output lines , threading through the
holes into their own dedicated instrument cavity.
3, noted that the positive and negative polarity DC power supply, if the reverse then the
instrument does not work.
4, power supply should have a good grounding in order to protect the operator's personal safety.
15
CONVERT PARTS
Http: // www.flowmeter.cc
CONVERT

1 4-20mA +
4-20mA -
0~2KHz+
0~2KHz-
0~2KHz+
0~2KHz-
2
3
4
5RS485date+
6RS485date -
4-20mA +
1
2 4-20mA -
3
4
1.)Electric current output
2.)Pulse, frequency output
3.) Function of communication
4-20mA Output positive terminal
Function Instructions
Terminal description
4-20mAOutput negative terminal
Frequency/pulse output
positive terminal
Frequency/pulse output
negative terminal
RS485 Communication
positiveterminal
RS485 Communication
negativeterminal
1, Load resistance:750Q (with link)
2, Hart Communications external 24VDC
power supply Active Output Mode
The output amplitude of 24V load
current <=50mA
RS485 communication function (n just
for transmitter has this function
Receive
Instruments
The converter current output electrical isolation has been achieved. Output to take an active
approach. Current output mode 4-20mA, 20mA current output value from the corresponding traffic
parameter item "scale flow of value" is determined (reference to factory nameplate on the
instrument measuring range a value). The maximum current output load resistance of 750Q, the
load resistance includes the cables used to connect the resistance. Current output cable is
recommended RWP2x16/015 PVC insulated sheathed cable shield. 2.
Counter
The Transmitter frequency, pulse output has been achieved electrical isolation, the output for the
active mode (see above chart). Transmitter frequency, pulse output with transistor output mode.
Maximum pulse output frequency 5KHZ, the output pulse amplitude of 24V. Active mode the
maximum load current 50mA, passive mode the maximum load current 0.2A. As the frequency
and pulse output terminals are shared, it is not to choice two output modes. Users can set the
parameter "frequency output" to select the work. Frequency output upper limit corresponding to
the measured flow value by the parameter item "scale flow value" decision (reference instruments
to measure the scope of a factory nameplate values). Pulse output equivalent by the parameter
'pulse equivalent L / P "decision.
Transmitter communication with RS485, MODBUSASC, MODBUSRTU communications
capabilities (requires a user specified when ordering). Available through the "485 output
communication protocol" parameter is set to specify. Instrument Communication Interface specific
technical note on "protocol."
Items Lable
TRANSIMITTER
16
CONVERT PARTS
Http: // www.flowmeter.cc
CONVERT
Remarks

TM
-8888.8
m/h
82345678.8
3
m3
TOTAL +
115.8%
OVER LIMIT !
REVISION V50.2
Ⅲ. Instrument started
1 .Working instrument showing
2 . Interface show
Flow direction
Instantaneous delivery
Unit of instantaneous delivery
Total accumulated
Unit of total accumulated
Warn of instrument
Edition number of program
Percentage of floe
Means of total accumulated
Specific alarm items
Before connect power, follow the manual of the first part of the note 1,2 and sensor part of
the installation manual to verify whether the system is properly installed and connections.
End customer can directly run instrument, the reason is including sensors and signal
transmitter of two parts, all of the data have been based on user requirements and technical
specifications of the company's manufacturing process in the factory setting.
After complete correct electric connection, run power on instrument. Instrument transmitter first
implementation of the initialization; display the company logo (see below). Wait 3 seconds after the
instrument into the own measurement mode, immediately began to flow measurement and
displays the current flow measurement value or other self-assemblies off information. If there is no
meter to power after the show (display without backlight), then the power supply and connectivity
in identifying ways to meet the requirements, may view the instrument power supply fuse is intact
(you can see the 8 common faults and processing).
If there is no meter to power after the show (display without backlight), then the power supply
and connectivity in identifying ways to meet the requirements, may view the instrument power
supply fuse is intact (you can see the 8 common faults and processing).
17
CONVERT PARTS
Http: // www.flowmeter.cc
CONVERT
Table of contents
Other FLOWTECH Measuring Instrument manuals