Fluke PV350 User manual

PV350 English Instruction Sheet
Page 1
®
PV350
PV350
PRESSURE / VACUUM MODULE
cmHg
inHg
kPa
psi
OFF
METRIC
ENGLISH ZE
Pressure/ Vacuum Transducer Module
Instruction Sheet
Introduction
The PV350 Pressure/Vacuum Transducer Module (the module)
measures gauge pressure; the difference between atmospheric
(barometric) pressure and the pressure (or vacuum) applied to the
pressure port on the transducer. It then converts that measurement to
1 mV dc per unit. A module and a high-performance digital multimeter
(DMM) with min/max function become a datalogger capable of measuring
peak high and peak low.
Box Contents
•Pressure/Vacuum Transducer Module
•1/4 in. male flare to 1/4 in. female pipe thread connector
•9 V alkaline battery
•Instruction Sheet
•Quick Reference Guide
Safety Information
Warning
To avoid eye injury, always wear approved safety glasses
when working with this instrument.
To avoid injury from pressure explosion:
•Be sure that pressure is removed before
disconnecting any hoses or tubing from the PV350
transducer.
•Follow accepted safety procedures for systems on
which pressure or vacuum measurements are made.
•Make sure the PV350 is serviced by a qualified
technician.

PV350 English Instruction Sheet
Page 2
•To ensure that the module is used safely, follow all
safety and operating instructions in this Instruction
Sheet. If the module is not used as described in this
Instruction Sheet, the safety features of the module
might be impaired.
•Do not use the module if the module or fittings look
damaged, or if you suspect that the module is not
operating properly.
• This symbol on the module indicates that an
operator should refer to an explanation in the
instruction sheet.
•Follow accepted safety procedures for the systems on
which pressure or vacuum measurements are being
made.
Materials Compatibility
The PV350 is compatible with industrial gases and liquids that are
compatible with 316 stainless steel. To determine the compatibility of a
liquid or gas, refer to the manufacturer's specifications. When making
measurements on more than one type of pressure system, make sure the
fluids or gases in those systems are compatible. If they are not,
contamination or damage to the system being measured may result.
Cleaning
Clean the module using a soft cloth and mild soap and water. Do not
immerse the PV350 in liquids.
Clean the transducer and connectors after each use with a cleaner
appropriate for the liquid or gas connected during measurement. Refer to
the manufacturer’s instructions for the proper cleaning material.
Measuring Pressure and Vacuum
Caution
The sensor diaphragm can be damaged by solid or sharp
objects. Never insert any object into the inlet port.
Clean the sensor immediately after use with appropriate
solvents (refer to the solvent manufacturer's instructions).
Note
Fluke recommends that Teflon sealing tape (or its equivalent)
be used in fitting threads to ensure good seals between the
PV350 and any hoses or fittings.
1. Plug the PV350 into the DMM (red polarity dot to volts input) and set
the multimeter to mV dc.
2. Check the battery voltage (see Testing the Battery).
3. Press the English/Metric button to select the desired units.
4. Slide the switch to the desired setting.
5. Before applying pressure, zero the module by turning the ZERO
potentiometer until the DMM displays zero. (Zero on the range you
will be using and rezero the DMM whenever the scale is changed.)

PV350 English Instruction Sheet
Page 3
mV
87
TRUE RMS MULTIMETER
OFF
mA
A
A
µ
mV
V
V
400mA
FUSED
!
ACOM V
mA A
10A MAX
FUSED
MIN MAX
RANGE
HOLD
Hz
PEAK MIN MAX
REL
H
PV350
P
R
E
S
S
U
R
E
V
A
C
U
M
N
M
O
D
U
L
E
M
E
T
R
I
C
M
E
T
R
I
C
OFF
kPa.psi
cmHg.inHg
FLUKE
To
pressure/vacuum
input
Sensor
Module
Figure 1. Operation
Note
The Relative or Zero function on your DMM (if available) can
be used instead of the ZERO potentiometer to zero the meter.
Refer to the multimeter's user manual for instructions on using
this function. Zeroing the meter does not affect the module's
specified accuracy.
6. Apply pressure/vacuum to the PV350 and read the DMM display.
•When measuring pressure, if the DMM displays OL,
switch the multimeter range from millivolts to dc volts.
Increasing the range to dc volts moves the decimal
position three places to the left (500 psi displays as
0.500 V).
•When measuring vacuum the DMM displays a minus sign.
Testing the Battery
1. Turn the DMM and the module OFF and connect the PV350 to the
DMM.
2. Turn the DMM on and set the DMM to the mV dc range.
3. If the DMM displays less than 100 mV, replace the battery.
How to Replace the Battery
Warning
To avoid electrical shock, disconnect the module from all
devices before opening the case. Be sure to completely
reassemble the module before attempting to use it.

PV350 English Instruction Sheet
Page 4
Battery
Figure 2. Battery Replacement
1. Turn the module OFF and unplug it from the DMM.
2. Remove the screw from the back of the module, separate the case,
and replace the battery.
3. To reassemble, line up the holes in the case with the
METRIC/ENGLISH switch and zero knob. Then snap the two halves
together. Reinstall the screw.
Typical Applications
Refer to Figures 3 and 4 for typical applications.
88
AUTOMOTIVE TESTER
MIN MAX RANGE HOLD
H
mA
A
mV
V
V
OFF
A
mA A
10A MAX
FUSED
ZERO
SMOOTH Hz
RPM
% DUTY
mA
A
RPM
ALERT
TRIGGER
_
+
ms-PULSE
!
COM
1000V MAX
400mA MAX
FUSED
DC
mV
4000
mV
20 2 46 3824684
V
RPM
METRIC
ENGLISH Z
PV500
PRESSURE / VACUUM MODULE
cmHg
inHg
kPa
psi
OFF
Figure 3. Pressure for HVAC/R Super Heat Measurements
88
AUTOMOTIVE TESTER
MIN MAX RANGE HOLD
H
mA
A
mV
V
V
OFF
AmA A
10A MAX
FUSED
ZERO
SMOOTH Hz
RPM
% DUTY
mA
A
RPM
ALERT
TRIGGER
_
+
ms-PULSE
!
COM
1000V MAX
400mA MAX
FUSED
DC
mV
4000
mV
20 2 46 3824684
RANGE
V
RPM
METRIC
ENGLISH Z
PV500
PRESSURE / VACUUM MODULE
cmHg
inHg
kPa
psi
OFF
Figure 4. Measure Pneumatic Pressures

PV350 English Instruction Sheet
Page 5
Conversion Factors
psi = inches of H2O x (3.6127 x 10-2) inches of H2O = psi x 27.68
psi = mm of H2O x (1.4223 x 10-3) mm of H2O = psi x 703.1
psi = cm of H2O x (14.223 x 10-3) cm of H2O = psi x 70.3
psi = bar x (14.503) bar = psi x 0.0689
psi = lbs/ft2 x (6.9444 x 10-3) lbs/ft2 = psi x 144
psi = mbar x (1.4503 x 10-2) mbar = psi x (68.9513)
psi = Pascals x (1.4503 x 10-4) Pascals = psi x (6.895 x 103)
All H2O conversion factors are @ 4°C
Specifications
The following specifications apply at 23 ±5 °C for 1 year after calibration
when the module is zeroed prior to each measurement.
Pressure
Range Accuracy
0 psi to 500 psi:
(0 kPa to 3447 kPa)
±0.25 % of reading ±0.3 psi
±0.25 % of reading ±2.1 kPa)
Vacuum
Range Accuracy
0 in-Hg to -29 in-Hg
(0 cm-Hg to -76 cm-Hg)
±0.25 % of reading ±0.5 in-Hg
(±0.25 % of reading ±1.3 cm-Hg)
Temperature coefficient: Add 0.05 % of reading per °C for temperature
ranges -10 °C to 18 °C and 28 °C to 55 °C.
Output: 1 mV per unit (psi, kPa, cm-Hg or in-Hg)
Resolution: Determined by the resolution of the measuring device.
RFI Specification: Typically <1 % error from 80 to 200 MHz at 3 V/m.
Electromagnetic Compatibility (EMC)
International IEC 61326-1: Portable; IEC 61326-2-2; CISPR 11: Group 1,
Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power supply
network that supplies buildings used for domestic purposes.
Korea (KCC) Class A Equipment (Industrial Broadcasting &
Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and not to be
used in homes.
USA (FCC) 47 CFR 15 subpart B. This product is considered an exempt
device per clause 15.103.
Meter Input Impedance: ≥1 MΩ
Maximum working pressure: 500 psi
Burst Pressure: 1000 psi

PV350 English Instruction Sheet
Page 6
Storage Temperature: -40 °C to 71 °C (Battery removed)
Operating Environment
Temperature Humidity
-10 °C to 10 °C
10 °C to 30 °C
30 °C to 40 °C
40 °C to 50 °C
50 °C to 55 °C
Non-condensing
0 °C to 95 % Relative Humidity
0 °C to 75 % Relative Humidity
0 °C to 45 % Relative Humidity
0 °C to 35 % Relative Humidity
General
Weight: 12 oz, (336 grams)
Battery: Standard 9 V Alkaline battery (NEDA #1604, IEC 6LR61)
Battery Life: 300 hours (typical)
Electrical Cable: Black PVC, will withstand up to 105 °C
Cable Length: 8 ft
Pressure Port Connection: 1/4 in. NPT
Connector: Brass 1/4 in. male flare to 1/4 in. female pipe thread
List of Replaceable Parts
Part Number Description
614487
2813197
2816172
913207
926873
926881
Battery, Primary, 9 V, 0-15 mA
Case Bottom, PV350
Case Top, PV350
Fitting, BR, SAE, 45, 1/4 FP
PV350 Instruction Sheet
PV350 Quick Reference Guide
Fluke Electronics Multimeters and Testers
Table of contents
Other Fluke Control Unit manuals

Fluke
Fluke DTX-MFM2 User manual

Fluke
Fluke DTX-GFM2 User manual

Fluke
Fluke DSP-PM10B User manual

Fluke
Fluke 80TK User manual

Fluke
Fluke DSP-PM11B User manual

Fluke
Fluke DSP-PM12B User manual

Fluke
Fluke 700PRV User manual

Fluke
Fluke 700PCK User manual

Fluke
Fluke PM Series User manual

Fluke
Fluke DSP-PM14A User manual
Popular Control Unit manuals by other brands

Texas Instruments
Texas Instruments INA190EVM user guide

DFI
DFI SH960MD-CM236 user manual

urmet domus
urmet domus 2 VOICE 1083/69 manual

WURM
WURM HVL-G3 Product information

oventrop
oventrop Multidis SF Installation and operating instructions

Parker
Parker 71215 INSTALLATION, OPERATING & MAINTENANGE INSTRUCTIONS