3
Display Features
sec: Shows when you look at the setting for the backlight
timer. The setting is in seconds.
dB: Shows when you look at the setting for the reflection
limit. The setting is in decibels.
dB: Shows when the display shows the reflectance of an
incident on the fiber. The measurement is in decibels.
: Shows when the display shows the power loss of an
incident on the fiber. The measurement is in decibels.
RL: Result loss.
MORE : Shows when the troubleshooter finds more
than one incident. Press to see more incidents. The
numbers show the number of the incident and the total
number of incidents. The troubleshooter shows up to 9
incidents. If there are more than 9 incidents on the fiber,
the last blinks when you look at the ninth incident.
SAVED: Shows after a test. The troubleshooter
automatically saves the test results.
TESTING: Shows as the troubleshooter does a test.
BREAK OR END: Shows when the troubleshooter shows
the distance to a break or the end of the fiber.
: The measurement is above or below the range that
the troubleshooter can show, or the troubleshooter
cannot give a more accurate measurement.
RESULT: Shows when you look at saved results.
Settings for the troubleshooter. LOSS LIMITWand
REFLECTION LIMITWblink if a measurement is equal to or
greater than the limit you select.