ACTIVE P30 Plus Series Owner’s Manual
EN-6
artery in the moments between beats when the heart is resting.
1. Place a stethoscope on the arm where there is a pulse. Wrap
the cu around the upper arm and hold in place with Velcro.
2. Press and hold until“ ”symbol appears on the display
with a beep. Release the button. Then cu begins to inate
automatically. (j)
3. After reaching the cu pressure, the deation begins. You can
adjust the ination pressure by keep pressing to inate
if necessary. If you release , the ination will stop and
deation begins.
4. The systolic pressure is measured when the operator rst hears
the pulse. Write down the value on the display. This value
indicates the systolic pressure. You have to record the values as
the result will not be stored in the memory.
5. The diastolic pressure is measured from the moment the
operator is unable to hear the sound of the pulse. Write down
the value on the display. This value indicates the patient’s
diastolic pressure. You have to record the values as the result
will not be stored in the memory.
6. The device will return to 0 mmHg after the measurement is
completed. To turn o the device, press and hold or it will
switch o automatically after idle for 3 minutes. (k)
Reviewing Test Results
Your device stores the 200 (P30 Plus BT only) most recent blood
pressure test results along with respective dates and times in the
memory. To recall the memory, start with the device o.
For P30 Plus
To review all test results, do the following:
1. Press and release . The“ ”icon appears on the screen and
the rst reading you see is the last blood pressure result along
with date and time. (l)
2. Press repeatedly to review other test results stored in the
device. (m)
To review the day-average test results, do the following:
1. Press and hold for 3 seconds until the“ ”icon
appears on the screen.
2. Press and the Day average result appears on the display. (n)
3. Press again to review the Day-time average result. (o)
4. Press again to review the Night-time average result. (p)
For P30 Plus and P30 Plus BT
Press to review the average and general memory recall mode.
Note:
• Day-time average ( ) is the average of the measurements taken
during 4:00 A.M. to 11:59 A.M.
• Night-time average ( ) is the average of the measurements taken
during 6:00 P.M. to 11:59 P.M.
• Press to exit the memory mode or leave it without any action for
3 minutes. The device will turn o automatically.
• If using the device for the rst time, the“---” icon appears when you
recall the test results or review the average result. It indicates that
there is no test result stored in the memory.