(1) What’s in the box
Rolling Stand
IR42b Forehead Thermometer
an iPad
(2) Hardware overview
Auto Temperature Measurement Station (ATMS)
User Guide
(3) How to assemble the stand?
- Also see the assembly video (https://foracare.com/fora-auto-temperature-measurement-station-atms-assembly-video/)
IR42 Non-Contact
Forehead Thermometer
iPad
(Clamp with lock)
Table desk
with holding handler
MCS stand
Clamps
Flexible Coils
Bayonet Mount
Basket
Wheels
Step 1.
Pull up the center stand.
Adjust the position of the basket, and tighten up the bolts
Step 2.
Adjust the table desk and x it.
Step 3.
Fix the iPad holder
Step 4.
Insert the exible conduit and tighten up it.
Step 5.
Insert the iPad in the holder. Lock and secure the iPad
with the key.
Step 6.
Place the thermometer. Make the sensor direction
parallel to the ground.
(4) How to use the application?
1. iFora TM should be already installed in the iPad, and the thermometer has been paired with the APP. If not, refer to the section
(7) How to install the application.
2. Open the app (iFORA TM) and turn on the thermometer. The color of the Bluetooth icon on the app will turn to blue (Figure 2).
When you see the blue icon, it’s ready to go. You can take the rst measurement.
3. If the unit of temperature is not for your region, please change it under Setting →Temperature Unit (Figure 3).
Figure 3Figure 2
2.
1.
3.
TM
(5) Setup notes
Adjust the height of the thermometer to suit
the needs of your facility. The recommended
height is a little lower than the average height.
Once the height is set, please do not adjust
the conduit, instead use the adjustable rod to
adjust the height.
Keep the measuring direction of the forehead
thermometer horizontal to the ground.
Secure your iPad into the iPad holder by
locking up the security lock in the back.
1.
2.
3.
12.
Be careful not to touch the buttons on both
side of the thermometer.
4.
Adjust the angle and height for the forehead
thermometer so a clear portrait will display in the
lens zone on the top left of the iFORA TM APP.
5.
The ATMS measurement is designed to be used
indoors. Avoid direct sunlight and windy areas
as dust will cause inaccurate results.
6.
9.
10.
11.
After placing the ATMS stand in a proper
location and locking the latches of the wheel,
we recommend marking the measurement
and waiting areas in order to clearly guide the
users. Make sure the thermometer sensor is
vertical to the ground.
When visitors step up to the ATMS, they
need to follow the on-screen instructions
and stay 1.97” to 2.75” (5 to 7 cm) away from
the thermometer sensor, stand still, and wait
for 1 second. The measurement is completed
after the beep. The temperature will be
displayed on the iPad within 4 seconds.
If the user has just come from either a
freezing or a hot environment, a break is
advised before measuring your body
temperature.
Please switch your forehead thermometer
and iPad o when not in use. Charge your
iPad and replace the batteries when neces-
sary.
Please do not use the assigned iPad for any
other use including changing settings or
other software installation that may void
original pairing and operation.
Wi-fi is required for active cloud users to
export and upload data. If you do not want
to upload data, you can switch the option
o in the “Setting -> Export -> Auto Upload
Data” conguration.
7.
8.
(6) FAQ
What do I do if cannot get a reading?
If you cannot get the reading on the screen, please make sure:
•The Bluetooth icon is a blue color. If not, check the connection with the
thermometer. The thermometer might be o or have a low battery.
•The environmental temperature cannot be too high. Avoid direct heat and
sunlight projecting onto the thermometer. Additionally, a windy entrance
might cause the sensor to fail.
•If you’ve just come from either a freezing or a hot environment, a break is
advised before measuring your body temperature.
•If the iPad cannot display any reading after 6 seconds, please step back and
move forward again to initiate another measurement.
Why does the app beep when no one is standing in front of it?
The sensor of the thermometer could be activated from the sunlight or other
heat sources. Please move the station to another location or adjust the direction
of the sensor.
(7)
How to install the application
1. In Apple Store, search for iFORA TM, download and
install the mobile application.
2. When opening the app for the rst time, you will be
asked to turn on your tablet’s Bluetooth (Figure 4).
3. Turn on the thermometer by pressing the power button
4. In the app, under Available Meters, click the ⊕button (Figure 5).
5. The thermometer will show up in the list, representing it’s been successfully
paired (Figure 6).
Figure 4 Figure 5 Figure 6
If you need more information, please refer to:
https://foracare.com/fora-atms/

Other manuals for ATMS

1

Other FOR-A Medical Equipment manuals

FOR-A VSM100 Pro User manual

FOR-A

FOR-A VSM100 Pro User manual

FOR-A TN'G SpO2 Reference guide

FOR-A

FOR-A TN'G SpO2 Reference guide

FOR-A TN'G SpO2 User manual

FOR-A

FOR-A TN'G SpO2 User manual

FOR-A FORA O2 User manual

FOR-A

FOR-A FORA O2 User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.