GE SimpliNano User manual

Spectrophotometer Health &
Safety Document including
General Operating Instructions
Original Instructions

29-0047-78 AC 04-2014 2
1. Health and Safety 3
1.1. Scope 3
1.2. Intended use 3
1.3. Safety Notices 3
1.4. General Safety 4
1.5. General Hazards 5
1.6. Unpacking and Installation 5
2. Operating and Maintenance 7
2.1. Preparation before starting 7
2.2. Performing a run - cuvettes 7
2.3. Performing a run – Micro Volume Instruments 8
2.4. Post run procedures 9
2.5. User Maintenance 9
2.6. Novaspec III/Plus Lamp Change 10
2.7. Ultrospec 8000/9000 Lamp Change 10
3. Troubleshooting 11
3.1. Emergency Procedures 12
3.2. Recycling Procedures 12
3.3. Recycling of hazardous substances 12
3.4. Disposal of electrical components 12
3.5. Specifications 13
3.6. Manufacturing information 13
4. Legal 14
Page finder

29-0047-78 AC 04-2014 3
1. HEALTH AND SAFETY
1.1. Scope
This booklet covers the following ranges of instruments:
• Ultrospec™10celldensitymeter
• Novaspec™IIIandNovaspecPlusVisiblespectrophotometers
• GeneQuant™100&1300UV/Visiblespectrophotometers
• Ultrospec2100/7000/8000/9000 UV/Visible spectrophotometers
• SimpliNano™andNanoVue™Plusmicrovolumespectrophotometer
Translations of these instructions into 22 European languages are available on the
User Manual CD supplied with the instrument and on the internet at GE home page
www.gelifesciences.com
1.2. Intended use
Visible and UV/Visible spectrophotometers shine light through a liquid sample and
measure its Absorbance. This sample is typically held in a cuvette but when sample
volume is limited, instruments such as the SimpliNano or NanoVue Plus can be used
where a 2 µl sample can be pipetted directly onto the measurement area. Typical
applications of spectrophotometers include DNA/RNA/Oligo concentration and purity
measurements as well as protein concentrations.
This manual addresses the needs of scientists and technicians who operate UV/
Visible spectrophotometers. The level of information presented in this manual
assumes the user possesses basic laboratory and technical skills, and has the
knowledge and documentation to safely operate. If you need assistance with the
instructions in this manual, contact GE for more information.
1.3. Safety Notices
These operating instructions contain Warnings and Cautions concerning the products
with meanings as defined below:
WARNING
Warning indicates a hazardous situation which, if not avoided, could result in death or serious injury. It is important
not to proceed until all stated conditions are met and clearly understood.
CAUTION
CAUTION indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. It is
important not to proceed until all stated conditions are met and clearly understood.

29-0047-78 AC 04-2014 4
WARNING
• Therearenobio-hazardousmaterialswithintheunit;howeverthisunitcouldbeusedwithbio-hazardoussamples.
Before using the instrument, the customer should have in place decontamination procedures designed to protect
laboratory workers from occupationally acquired infections. The sample chamber cell holders are removable and
may be decontaminated using the appropriate disinfectant for the bio hazard in question, rinsed with distilled
water and then allowed to dry. The sample chamber and exterior may be wiped with a suitable disinfectant
cleaning wipe. It is the responsibility of the customer to ensure that the user of the equipment is provided with a
safe working environment.
• AnychemicalsusedinAnalysesshouldbeused,storedanddisposedofinaccordancewithmanufacturer’s
guidelines and local safety regulations.
• ToxicFumes.Efficientlaboratoryventilationmustbeprovidedwhenworkingwithvolatilesolventsortoxic
substances.
• Wastedisposal.Disposalofsomesolventsandchemicalsmaybeclassedashazardouswasteandmustbedealt
with in accordance with local regulatory practice.
• Personalprotectiveequipment.ThisisnotrequiredtooperatetheunitbutthesamplesmeasuredmayrequirePPE.
A local risk assessment should be carried out.
• Decontamination.Equipmentreturnedforrepairshouldincludeanappropriatedecontaminationcertificate.
1.4. General Safety
This equipment has been designed to conform to the following directives
2006/95/EC Low voltage directive (LVD)
2004/108/EC Electromagnetic Compatibility (EMC) directive
2012/19/EU Waste Electrical and Electronic Equipment directive recast (WEEE Recast)
2011/65/EU Restriction on the use of certain hazardous substances (ROHS) directive
2006/42/EC Machinery directive
1999/5/EC Radio and Telecommunications terminal equipment directive (Applicable
only to products fitted with Bluetooth communication devices)
Standards, to which conformity is declared, are as follows:
EN61010 - 1:2010 Safety requirements for electrical equipment for measurement,
control and laboratory use. General requirements
EN61326 - 1:2006 Electrical equipment for measurement, control and laboratory
use -EMC Requirements*
EN50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous
substances.
EN ISO 12100:2010 Safety of machinery-General principles for design, risk
assessment and risk reduction
EN 301 489-1/-17 Electromagnetic Compatibility (EMC) standard for radio
equipmentandservices;Part1:Commontechnical
requirements (Applicable only to products fitted with Bluetooth
communication devices)
EN 300 328 Data transmission equipment operating in the 2,4 GHz ISM
bandandusingwidebandmodulationtechniques;essential
requirements under article 3.2 of the R&TTE Directive.
(Applicable only to products fitted with Bluetooth
communication devices)

29-0047-78 AC 04-2014 5
1.5. General Hazards
There a number of warning labels and symbols that may be present on your
instrument. These are there to inform you where a potential danger exists or
particular caution is required. Before commencing installation, please take time to
familiarise yourself with these symbols and their meaning. This instrument is subject
to the following hazards:
CAUTION
Some of the above instruments contain a UV source which generates a light beam that traverses the sample
chamber and is accessible in the lamp chamber. Under normal use the lamp beam is confined within the instrument.
The unit should not be operated with the sample chamber lid open or the lamp housing lid removed as prolonged
exposure to the beam may cause permanent eye damage.(Not applicable to Ultrospec 10, Novaspec III/Plus)
WARNING
HighvoltagesexistinsidetheUltrospec2100/7000/8000/9000,GeneQuant100/1300,SimpliNanoandNanoVuePlus
instruments. Repair and maintenance should only be carried out by individuals trained specifically to work on these
instruments.
CAUTION
These instruments may be connected to a separate PC. To preserve the integrity of the measuring equipment it is
essential that the attached PC itself conforms to basic safety and EMC standards and is set up in accordance with
themanufacturers’instructions.IfindoubtconsulttheinformationthatcamewithyourPC.Incommonwithall
computer operation the following safety precautions are advised.
•To reduce the chance of eye strain, set up the PC display with the correct viewing position, free from glare and with
appropriate brightness and contrast settings.
• To reduce the chance of cross contamination from biological samples, use appropriate personnel protection
measures and disinfectant wipes on keyboard and mouse
CAUTION, HOT SURFACE
• These instruments may be fitted with heated cell holder accessories that depending on operation may become hot
to touch. Care should be taken to avoid touching the heated accessory when running at elevated temperatures
1.6. Unpacking and installation
NanoVue Plus instrument External power supply
• Inspecttheinstrumentforanysignsofdamagecausedduringtransit.Ifany
damage is discovered, do not use the instrument and report the problem to your
supplier.
• Ensureyourproposedinstallationsiteconformstotheenvironmentconditionsfor
safe operation
• Indooruse
• 5to40°C

29-0047-78 AC 04-2014 6
• Maximumrelativehumidity90%upto31°Cdecreasinglinearlyto50%at40°C
• Extremesoftemperaturemayrequirerecalibrationoftheunitforoptimum
performance
• Iftheinstrumenthasbeenstoredinacoldenvironmentthenitshouldbeallowed
to come to thermal equilibrium for 2 to 3 hours before operation so that start up
calibration is not compromised by condensation.
• Theinstrumentmustbeplacedonastable,levelbenchortablecapableoftaking
its weight with sufficient space around the instrument for ventilation to circulate
freely.
• Theequipmentmustbeconnectedtothelocalsupplyoutletusingtheprovided
power cables, compatible voltages are shown in the table below. Replace power
inlet fuses only with the same type and rating as follows
Instrument Type Power requirements Replacement Fuse
Ultrospec 10 units 100–240 V AC 50–60 Hz 3 A (UK plug only)
Novaspec III and Novaspec Plus 100–240 V AC 50–60 Hz 3 A (UK plug only)
GeneQuant100&1300units 100–240 V AC 50–60 Hz 3 A (UK plug only)
Ultrospec 2100 units 85–264 V AC, 50–60 Hz T 2 A H 250 V AC (Anti-Surge,
High breaking capacity)
Ultrospec 7000/8000/9000 units 85–264 V AC, 50–60 Hz T 1.6 A H 250 V AC (Anti-Surge,
High breaking capacity)
SimpliNano units 100–240 V AC 50–60 Hz 3 A (UK plug only)
NanoVue Plus units 90–250v, 50/60Hz, Max 30 VA 3 A (UK plug only)
• Theinstrumentshouldbepositionedsothatthepowercablemaybereadily
removed in the event of a hazard or malfunction occurring.
• Sitetheinstrumentinanatmospherefreefromdustandcorrosivefumes.
• Ultrospec2100instrumentshavetheon/offswitchonthebackoftheinstrument
on the right hand side. Ultrospec 7000/8000/9000 instruments have the on/off
switch on the left hand side of the instrument. Ultrospec 10, Novaspec III/Plus,
GeneQuant100/1300,SimpliNanoandNanoVueinstrumentshaveanon/off
button on the keypad – to disconnect power from these units remove the power
lead from the instrument. The instruments will automatically perform some start
up self diagnostic checks, please wait for these to finish before attempting to use
the equipment.
• Thespectrophotometermaybefittedwitharangeofaccessoriesforspecific
applications.FortheGeneQuant100/1300andNanoVuePlustheseincludeabuilt
in printer, bluetooth or SD card accessory. The SimpliNano has an optional built-
in printer. The Ultrospec 2100/7000/8000/9000 have a range of accessories that
can be fitted in the sample compartment including a sipper, long path length cell
holder and thermostatted cell holder. Full instructions on fitting accessories are
within the instrument or accessory user manual.

29-0047-78 AC 04-2014 7
All spectrophotometers are operated in the same generic way by comparing the
amount of light Absorbed by a sample when compared to a reference solution:
2.1. Preparation before starting
• Theinstrumentisswitchedonandallowedtoinitialise,Ultrospec10,
NovaspecIII/Plus,GeneQuant100/1300,SimpliNanoandNanoVuePlusunitshave
the on/off switch on the keypad.
• Ultrospec2100unitshaveanon/offswitchonthebackoftheinstrumentonthe
right hand side.
• Ultrospec7000/8000/9000instrumentshavetheon/offswitchonthelefthand
side of the instrument.
• Therelevantsoftwaremodulefortheapplicationbeingcarriedoutisselectedfrom
within the instrument software e.g. fixed wavelength measurement, concentration
measurement, wavelength scan, kinetics measurement or a dedicated method
such as DNA concentration & purity. The options within the application set up the
instrument parameters such as wavelength and define any calculations taking
place on the measured results.
2.2. Performing a run - cuvettes
• OntheUltrospec10,NovaspecIII/Plus,GeneQuant100/1300andUltrospec
2100/7000/8000/9000 units, samples are measured in a cuvette. A suitable
cuvette must be used for the application to be carried out – it must be of the
correct material to allow the selected wavelength of light to pass through and
must be of the correct physical size to fit the relevant instrument and the sample
volume available. The cuvette is filled with a solution of reference material which
is typically the solution that the sample is dissolved in, the cuvette is placed in the
cell holder in the correct orientation as per the diagrams below:
Ultrospec 10, front to back All other units, right to left
• TheinstrumentisthenzeroedbypressingtheZerokeyonthekeypad:
2. OPERATING AND
MAINTENANCE
Ultrospec10NovaspecIII/PlusGeneQuant100/1300Ultrospec2100

29-0047-78 AC 04-2014 8
• Thereferencesolutionisthenremoved,thecuvettefilledwiththesampleandthesamplemeasuredbypressingtherun
button on the keypad. The measured results will be displayed on the instrument display.
• TheUltrospec7000/8000/9000instrumentshaveacoloredTouchscreenforoperationandadoublebeamoptical
system. In normal operation a cuvette filled with reference solution is put in the rear cell holder , a cuvette filled with
sample in the front cell holder, the relevant application method is loaded and the Run Icon pressed. If a single cell is being
usede.g.amicrocellitshouldbefilledwithreferencesolution,placedinthefrontcellholder,theZeroiconpressed,the
cuvette filled with sample, placed in the front cell holder and the Run Icon pressed
• Typicalresultsscreen(DNAapplicationGeneQuant)
2.3. Performing a run - micro volume instruments
• OntheSimpliNanounitssamplesaremeasuredinasampleport.Usingapipette
a solution of reference material, which is typically the solution that the sample is
dissolved in, is dropped onto the measurement position in the sample port and the
instrument zeroed by pressing the Reference key on the keypad .
Zero Run
Ultrospec10NovaspecIII/PlusGeneQuant100/1300Ultrospec2100

29-0047-78 AC 04-2014 9
Thereferencesolutionisthenwipedawaywithatissue;thesampleispipettedinto
place and measured by pressing the run button on the keypad.
•Themeasuredresultswillbedisplayedontheinstrumentdisplay.
• OntheNanoVuePlusunitssamplesaremeasuredonasampleplate.Usinga
pipette a solution of reference material, which is typically the solution that the
sample is dissolved in, is dropped onto the measurement position on the sample
plate and the instrument zeroed by pressing the Reference key on the keypad.
• Thereferencesolutionisthenwipedawaywithatissue;thesampleispipettedinto
place and measured by pressing the run button on the keypad.
• Themeasuredresultswillbedisplayedontheinstrumentdisplay.
2.4. Post run procedures
Cuvettes – empty cuvette of sample and rinse with deionised water. Cuvettes may be
periodically cleaned by using commercially available cell cleaning solutions or dilute
detergent solutions but should always be rinsed in deionised water.
SimpliNano – wipe the sample port with a dry, lint-free tissue between samples and
at the end of the run, this is sufficient to remove sample. If more thorough cleaning is
desired, the port can also be cleaned with dilute (2%) detergent solution, followed by
isopropyl alcohol.
NanoVue Plus – wipe sample with tissue. At end of sample run wipe plates with tissue
wetting with deionised water. The plates can also be cleaned with isopropyl alcohol
or a dilute detergent solution but the plates should always be rinsed with deionised
water after cleaning.
Sample disposal
•
Waste disposal. Disposal of some solvents, chemicals or samples may be classed as hazardous or bio hazardous waste
and must be dealt with in accordance with local regulatory practice.
2.5. User Maintenance
Instrument Type User Maintenance
Ultrospec 10 units None
Novaspec III and Novaspec Plus Changing Tungsten lamp, removal of cell
holder for cleaning
GeneQuant100&1300units
Ultrospec 7000 unit Removal of cell holder for cleaning
Ultrospec 2100 Removal of cell changer for cleaning
Reference Key

29-0047-78 AC 04-2014 10
Ultrospec 8000 & 9000 units Changing Deuterium/Tungsten lamps,
removal of cell holders for cleaning
NanoVue Plus units Cleaning of sample plates, recalibration of
pathlength
SimpliNano units Cleaning of sample port
2.6. Novaspec III/Plus Lamp Change
1. Switch off the instrument, remove the sample from the cell holder and disconnect
the power supply cord
2. Remove the protective layers at the lamp access and plug in points on the
underneath of the instrument
3. Remove the lamp wires from the groove by gently unclipping it
4. Remove the lamp by twisting the lamp assembly counter-clockwise
5. Remove the lamp connection end by gently pulling with your fingers
6. Replace with new lamp using the reverse of these actions
7. Dispose of the faulty lamp in accordance with local regulatory practice.
2.7. Ultrospec 8000/9000 Lamp Change
1. Switch the instrument off, disconnect the power supply cord and allow lamps to
cool.
2. Locate the lamp access cover at the rear of unit and remove the two top screws
and slide top cover back and lift off. Never operate the unit with the lamp
housing cover removed.
3.
4. Remove the lamp and dispose of in accordance with local regulatory practice.
5. Follow the handling instructions supplied with the lamp. Do not touch the glass
envelopes of the replacement bulbs directly.
6. Replace the lamp cover and screw securely.
7. Attach power cord, switch on and wait at least 30 minutes for the unit to warm
up.
8. Perform a new instrument baseline (on all bandwidth settings if variable
bandwidth unit) and save this as the permanent baseline (see the online help or
user manual).
9. Reset the lamp hours after replacing the lamp (see the online help or user
manual).

29-0047-78 AC 04-2014 11
3. TROUBLESHOOTING
Detailed in the table below are the most common faults found whilst using our
spectrophotometers
Instrument Problem Possible causes
Ultrospec 10 A flashing Absorbance reading of 2.00 Abs is
obtained.
This indicates an Absorbance of more than
1.99 and is therefore out of range. The
sample needs to be diluted.
Ultrospec 10 REF is displayed when T is pressed. The instrument has not been zeroed
before a sample measurement has been
attempted.
All instruments A negative reading is obtained. In normal measurements the test sample
has a positive Absorbance compared to that
of the Reference. Negative readings will be
obtained if the Reference and Test cuvettes
are mixed up.
All instruments Unexpected results are obtained. Any bubbles in solution will produce
considerable error.
All instruments Absorbance values higher than expected. Possible issue with optical alignment.
Contact service support.
All instruments Absorbance values lower than expected. Check sample cuvette filled to 20 mm from
base of cuvette.
Possible stray light issue. Contact service
support.
All cuvette models Instrument will not zero. Cuvette placed in wrong orientation, Wrong
cuvette material used for wavelength.
Novaspec III Error code 001 symbol . Lamp needs changing.
Novaspec III
Error code 006 symbol .
Pixel clock too high. Instrument needs
optical alignment. Contact service support.
Novaspec III Error code Flashing number! Wavelength accuracy failure. Instrument
needs optical alignment. Contact service
support.
Novaspec plus X against Lamp on initialisation screen. Lamp needs changing.
Novaspec plus X against Wavelength calibration on
initialisation screen.
Check sample beam is clear when
instrument switched on.
Novaspec plus - against all calibration items on
initialisation screen.
Instrument needs optical realignment or
reprogramming. Contact service support.
GeneQuant100/1300,
Ultrospec 2100,
Ultrospec 7000/8000/9000
Poor reproducibility on DNA concentration
and purity measurements.
Wrong cuvette being used – not a 15 mm
beam height or wrong material – not letting
UV light pass through.
Background correction mode not switched
on.
Measured Absorbance too low, should be
>0.1 A for accurate results.
GeneQuant100/1300,
Ultrospec 2100,
Ultrospec 7000/8000/9000
Instrument fails start up calibration check. Check sample and reference beams are
clear when instrument switched on.
GeneQuant100/1300,Ultrospec
2100/7000/8000/9000 report failure
message to service support.

29-0047-78 AC 04-2014 12
Ultrospec 10, Novaspec III/Plus,
GeneQuant100/1300,NanoVue
Plus, SimpliNano
Instrument switches off after calibration. You may be keeping you finger on the ON/
OFF button too long, so that the instrument
receives both ON and OFF signals and
switches off after the calibration.
Try adjusting the timing of your finger press
at switch on.
SimpliNano Instrument fails to initialise – “cell holder
obstructed UV” error message.
Ensure sample port is clean and there is no
dried on sample
SimpliNano Absorbance readings unstable EnsureReferenceQAswitchedon.
Ensure Background correction is switched
on.
Ensure 2 µl sample being used
NanoVue Plus Instrument fails to initialise – “cell holder
obstructed UV” error message.
Ensure sample plates are clean.
NanoVue Plus Absorbance readings unstable. EnsureReferenceQAswitchedon.
Ensure Background correction is switched
on.
Ensure 2 µl sample being used.
NanoVue Plus Absorbance readings stable but incorrect. Check pathlength calibration.
Detailed instructions on how to carry out the above user maintenance and trouble
shooting for the different models may be found in the relevant user manual supplied
on a CD with the instrument.
3.1. Emergency Procedures
In the event of a malfunction or hazard occurring, the user responsible shall
disconnect the unit from power and isolate the instrument for decontamination by an
appropriate means if bio hazardous material is spilt on, in or around the instrument.
Ultrospec 2100 instruments have the on/off switch on the back of the instrument on
the right hand side. Ultrospec 7000/8000/9000 instruments have the on/off switch on
the left hand side of the instrument.
Ultrospec10,NovaspecIII/Plus,GeneQuant100/1300,SimpliNanoandNanoVueand
instruments have an on/off button on the keypad – to disconnect power from these
units remove the power lead from the instrument.
3.2. Recycling Procedures
This equipment should be decontaminated before decommissioning and all local
regulations shall be followed with regard to scrapping of the equipment.
3.3. Recycling of hazardous substances
Spectrophotometers contain hazardous substances. Detailed information is available
from your local GE representative.
3.4. Disposal of electrical components
Waste of electrical and electronic equipment must not be disposed of as unsorted
municipal waste and must be collected separately. Please contact an authorized
representative of the manufacturer for information concerning the decommissioning
of your equipment.

29-0047-78 AC 04-2014 13
3.5. Specifications
Specification SimpliNano NanoVue Plus Ultrospec
9000
Ultrospec
8000
Ultrospec
7000
Ultrospec
2100
GeneQuant
1300
GeneQuant
100
Novaspec
Plus
Novaspec III Novaspec 10
Wavelength
Range 190–1100 nm 200–1100 nm 190–1100 nm 190–110 190–900 nm 190–900 nm 330–800 nm 600 nm dedicated
Light Source Xenon Xenon Deuterium/Tungsten Xenon Xenon Xenon Tungsten 600 nm LED
Bandwidth 5 nm 5 nm
0.5 nm,
1.0 nm,
2 nm,
4 nm
1.0 nm 2.0 nm < 3 nm 5 nm < 7 nm 40 nm
Absorbance
Range (A) -0.3 to 50A
0 to 125 A
(10 mm
pathlength
equivalence)
-4 A to 4 A -3 A to 3 A -0.3 A to 2.5 A - 0.3 A to 1.99 A
Pathlength 0.5 mm 0.5 mm,
0.2 mm 1 mm to 100 mm 10 mm
Wavelength
Accuracy
± 2 nm across
range,
± 1 nm from
240 nm to
330nm
± 2 nm across
range,
± 1 nm from
240 nm to
330nm
± 0.3 nm ± 0.5 nm ± 1 nm ± 2 nm ± 2 nm N/A
Absorbance
Accuracy
+/-1%at
259 nm at
0.7–0.8 A using
uraci
+/-1%at
259 nm at
0.7–0.8 A
using uraci
± 0.002 A at 0.5 A
± 0.004 A at 1 A
± 0.006 A at 2 A
at 440 nm, 465 nm, 546.1 nm,
590 nm, 635 nm
± 0.005 A using 60 mg/l K2Cr2O7
within0.5%of
Absorbance
value to 3.000 A
at 546 nm
±0.005Absor1%ofthe
reading, whichever is the
greater at 546 nm
±2.0%or±0.010Ato
1.000A at 546nm, whichever
is the greater
<± 0.05 A at 1
A using Neutral
Density Filters
Full specification and operating instructions are contained within the relevant user manual.
3.6. Manufacturing information
Requirement Content
Name and address of manufacturer
Biochrom Ltd, 22 Cambridge Science
Park, Milton Road, Cambridge,
CB4 0FJ, UK
Name and ID notified body dB Technologies (Cambridge) Ltd. FCC
reg. no 90528
Place and date of declaration
NanoVue Plus Family
GeneQuantFamily
Novaspec Family
Ultrospec 10, 2100Family
Ultrospec 7000, 8000, 9000 Family
SimpliNano Family
Cambridge, September 2013
Cambridge, September 2013
Cambridge, September 2013
Cambridge, September 2013
Cambridge, September 2013
Cambridge October 2013
Identity of person authorized to sign DoC Howie Pether, Managing Director
Full specification and operating instructions are contained within the relevant user manual.

29-0047-78 AC 04-2014 14
4. LEGAL
GE, GE monogram and imagination at work are trademarks of General Electric
Company.
Nanovue and SimpliNano are trademarks of GE Healthcare Company or one of its
subsidiaries.
Ultrospec,NovaspecandGeneQuantaretrademarksofBiochromLtd.
All other third party trademarks are the property of their respective owner.
© 2011–2014 General Electric Company – All rights reserved.
First published September 2011
All goods and services are sold subject to the terms and conditions of sale of the
company within GE Healthcare which supplies them. A copy of these terms and
conditions is available on request. Contact your local GE Healthcare representative
for the most current information.
http://www.gelifesciences.com
GE Healthcare UK Limited
Amersham Place, Little Chalfont,
Buckinghamshire, HP7 9NA UK
GE Healthcare Bio-Sciences AB
Björkgaton, 30 751 84,
Uppsala, Sweden
GE Healthcare Europe GmbH
Munzinger Strasse, 5 D-79111,
Freiburg, Germany
GE Healthcare Bio-Sciences Corp
800 Centennial Avenue, PO Box 1327,
Piscataway, NJ 08855-1327, USA
GE Healthcare Japan Corporation
Sanken Bldg 3-25-1, Hyakunincho Shinjuku-ku,
Tokyo, 169-0073, Japan

29-0047-78 AC 04-2014 15
Page intentionally left blank

GE Healthcare offices:
GE Healthcare Bio-Sciences AB
Björkgatan 30, 751 84 Uppsala,
Sweden
GE Healthcare Europe GmbH
Munzinger Strasse 5, D-79111 Freiburg,
Germany
GE Healthcare Bio-Sciences Corp.
800 Centennial Avenue, P.O. Box 1327,
Piscataway, NJ 08855-1327,
USA
GE Healthcare Japan Corporation
Sanken Bldg. 3-25-1, Hyakunincho,
Shinjuku-ku, Tokyo 169-0073,
Japan
For contact information for your local office,
please visit: www.gelifesciences.com/contact
GE Healthcare Limited
Amersham Place
Little Chalfont, Buckinghamshire,
HP7 9NA, UK
http://www.gelifesciences.com
imagination at work
29-0047-78 AC 04-2014
imagination at work
This manual suits for next models
10
Table of contents
Other GE Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Vega
Vega VEGAWELL S 51 operating instructions

Rohde & Schwarz
Rohde & Schwarz FPC1000 user manual

Indu-Sol
Indu-Sol EmCheck ISMZ I user manual

Tascam
Tascam Portacapture X8 owner's manual

GrayWolf
GrayWolf IAQ SURVEYOR PRO operating instructions

Velleman Instruments
Velleman Instruments HPS40 Technical documentation

Testboy
Testboy TV 340 operating instructions

PCB Piezotronics
PCB Piezotronics IMI SENSORS M645B01 Installation and operating manual

Radiation Solutions
Radiation Solutions RS-330 Operator's manual

Red Seal Measurement
Red Seal Measurement neptune S Operating and maintenance manual

BOSSLASER
BOSSLASER HP Series Operator's manual

T&D
T&D TR-73U manual