cool cloth may impact your reading. It is advised to wait 10 minutes before taking
a reading.
In the following situations it is recommended that 3-5 temperatures in the same
location be taken and the highest one taken as the reading:
• Newborn infants in the first 100 days.
• Children under three years of age with a compromised immune system and for
whom the presence or absence of fever is critical.
• When the user is learning how to use the thermometer for the first time until
he/she has familiarized himself/herself with the instrument and obtains
consistent readings.
Care and cleaning
Use an alcohol swab or cotton swab moistened
with alcohol to clean the thermometer casing
and the measuring probe. After the alcohol has
completely dried out, you can take a new
measurement. Ensure that no liquid enters the
interior of the thermometer. Never use abrasive
cleaning agents, thinners, or benzene for
cleaning and never immerse the instrument in
water or other cleaning liquids. Take care not to
scratch the surface of the LCD.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the instructions, periodic re-adjustment is
not required. If, at any time, you question the accuracy of the temperature
measurements, please contact seller.
This thermometer was designed and manufactured for a long service life,
however it is generally recommended to have it inspected once a year by the
Authorized Service Center located in your country to ensure correct function and
accuracy.
Errors and Troubleshooting
Symptom Possible Cause Description & Solution
Fail to
turn on
The battery level is too
low.
Replace batteries