- 9 -
11 LEAK TEST
This test may be performed at any time to check that the pipette does not leak,
especially after performing a maintenance or decontamination procedure. If a pi-
pette fails this test, replace the O-ring and seal. After making sure that the pipette
is correctly reassembled, repeat this test.
Symptom Possible Cause Action
Tips fall o or Low quality tips Use PIPETMAN DIAMOND tips
do not t correctly Tip-ejector damaged
Replace tip-ejector
Ejector spacer damaged Replace ejector spacer
Dirty tip-holder Clean them with ethanol or isopropanol
Damaged tip-holder Contact your local Gilson authorized service center
Pipette won’t aspirate Connecting nut is loose Tighten connecting nut
Pipette is inaccurate Connecting nut is loose Tighten connecting nut
Pipette is not precise
Connecting nut is loose Tighten connecting nut
Incorrect operator technique Operator training
For Multichannel models
no leak
For the P2L to P200L Single models,
and the F1L to F250L Fixed models:
Fit a PIPETMAN DIAMOND Tip.
For the variable volume,set thepi-
pettetothemaximumvolumegiven
in the specifications, and pre-rinse.
Aspiratethesetvolumefromabeaker
of distilled
water.
Maintain the pipette in the vertical
position and wait for 20 seconds.
If awater droplet appears atthe end
of the tip there is a leak.
If you see no droplet, re-immerse
the tip below the surface of water.
Thewater levelinside the tip should
remain constant; if the level goes
down there is a leak.
For the P1000L, P5000L and P10mlL
Singlemodels,fortheF300LtoF5000L
Fixed models:
Fit a PIPETMAN DIAMOND tip.
For the variable volume, set the
pipette to the maximum volume
given in the specifications.
Aspirate the set volume from a
beaker of distilled water.
Maintain the pipette in the vertical
position and wait for 20 seconds.
If awater droplet appears atthe end
of the tip, there is a leak.
FortheMultichannel models (8x-12x):
Fit the PIPETMAN DIAMOND tip.
Set the pipette to the maximum
volume given in the specifications,
and pre-rinse.
Aspirate the set volume from a
reagent reservoir of distilled water.
Maintain the pipette in the vertical
position and wait for 20 seconds;
uid level in tips shoulds remain
constant.
If awater droplet appears atthe end
of the tip, there is a leak.
If you see no droplet, for volumes
below 200 µL, re-immerse the tip
below the surface of water.
Thewater levelinside the tip should
remain constant; if the level goes
down there is a leak.
12A MAINTENANCE FOR THE SINGLE AND FIXED MODELS ONLY
Routine maintenance will help keep your pipette in good condition, ensuring a
continued high level of performance.
Maintenance is limited to:
Cleaning or decontamination (see Chapter 13 - Cleaning and Decontamination)
Replacing spare parts
Greasing the piston assembly.
PIPETMAN P2L and P10L, F1L, F2L, F5L and F10L should not be disassembled,
so you may only replace the push-button, tip-ejector, dual position tip-ejector and
its adapter. With these pipettes if the tip-holder is damaged, the piston may also
be damaged.
After replacing any parts you
should verify the performance
of your pipette following the
verification procedure available
on the Gilson website (www.
gilson.com). If the pipette needs
to be readjusted, please contact
your local Gilson authorized
Service Center.