Gima 28085 ultrasound aerosol Parts list manual

Aerosol ad Ultrasuoni Professionale
Professional Ultrasonic Nebulizer
M28085-I-G-Rev.1.02.11
Gima S.p.A. - Via Marconi, 1 - 20060 Gessate (MI) Italy
Italia: tel. 199 400 401 - fax 199 400 403
Export: tel. +39 02 953854209/221/225 fax +39 02 95380056
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
ATTENZIONE: Gli operatori devono leggere e capire completamente questo manuale
prima di utilizzare il prodotto.
ATTENTION: The operators must carefully read and completely understand the present
manual before using the product.
MANUALE D’USO E MANUTENZIONE
USE AND MAINTENANCE BOOK
0476

8
CHARACTERISTICS
- Easy to use: it uses water as a conductor of ultrasounds
- Simply functions: touch controls for operation of the aerosol
- Silent: thanks to the technology of ultrasound vibrations
- 3 speed aerosol: it allows for excellent treatment for all types of medicines of different densities
- Multifunction: it acts as a dehumidier if lled with water and as an aerosol if tted with medicinal products
- Special design: with an adjustable air nozzle for all types of treatment
- Easy to clean: air nozzle, cover and container for the medicines are easy to remove and clean
PREPARATION
Warnings for preparation
- Only use the device for the purposes indicated in this manual. Do not use any other components not
recommended by the manufacturer.
- Only use the device with medicines if recommended by your doctor.
- Do not use the device if damages have been caused or if the product is dropped in water. Take the
device to the assistance centre for control and repairs.
- Always remove the plug before lling it and after use.
- When used for the rst time or if a long time has passed from the rst use, make sure that all parts are
cleaned and sterilized, as indicated in the cleaning instructions.
- The medicine container, the mouthpiece and the air lter should be checked before any use. Replace
them with new ones if they are dirty or worn.
- The device should not be used close to ammable gas, spray products or oxygen.
- When the container for medicine or water is empty, do not try to turn the device back on
- In the case of faults, interrupt use until the device has been checked and repaired
- Do not leave the device unprotected when connected to electricity
- Do not use solutions other than water in the tank, they may cause damage to the transducer
- Only use the device on solid and at surfaces
- Opening of the cover during operation may cause serious damage
OPERATION
Warnings for operation
- If used by children or the disabled, the supervision of an adult is recommended
- Keep eyes far from the emission of medicinal products
- In the case of defects or inconvenient situations, interrupt use of the device immediately
- Comply with the general safety norms while using the UT-12L ultrasound aerosol
- Do not use during driving
STORAGE
Warnings for storage
- Keep out of direct sunlight, far from high temperatures and humidity
- Keep out of reach of children
- Always remove the plug when not in use
CLEANING
Warnings for cleaning
- Before every use check the medicine container, the air nozzle, the masks, the extended tube and the
air lter. The dirty or worn areas must be replaced.
- Do not plunge the device in water, it may get damaged
- Before carrying out the cleaning operations of the device, remove the plug
- Clean every part after use as indicated in the manual
ENGLISH

9
TECHNICAL CHARACTERISTCS
Name: 28085 ultrasound aerosol
Power: 230 VAC 50 Hz
Energy consumption: 30 VA
Ultrasound frequency: 2.5 MHz
Fuse characteristics: 250V, 315 mA
Nebulisation capacity: 3 adjustable speeds: 0.3~0.6mL/min, 0.6~0.9mL/min, 0.9~1.2mL/min
Diameter of the particle: 3-5 um typical
Capacity of the medicine container: 10ml
Environmental conditions
Temperature: 10°C – 40°C (50-104°F)
Humidity: 30% - 75% RH no condensation
Atmospheric pressure: 700 hPa 1 060 hPa
Storage temperature
Temperature: -10°C – 50°C (14-122°f)
Relative humidity: 10% - 95% RH no condensation
Atmospheric pressure: 700 hPa 1 060 hPa
Dimensions: 220 x 110 x 205mm
Weight: approximately 1300 g
DESCRIPTION OF THE PARTS
NB: The water level indicator indicates the water level recommended for the ultrasound
aerosol. When the water level is below this point, the water may not come into contact with
the medicine container and nebulisation may be poor. However, when the water level is
above this mark, the water may overow when the medicine container is inserted.
Air nozzle
Cover
Medicine
container
Controls
Plastic
structure
Switch
AC current
cable
Indication of water level +/-70,
until reaching the base of the black
plastic, the metal thread
ENGLISH

10
DESCRIPTION OF COMMANDS
Operation light: it indicates the state of the 28085 ultrasound aerosol
- No light means that the device is not on
- A green light indicates that the device is in a stand-by mode
Level 1 - Level 3: it indicates the level of nebulisation
- Level 1: low nebulisation
- Level 2: average nebulisation
- Level 3: high nebulisation
Stand-by button: nebulisation begins
Nebulisation control: it regulates the quantity of nebulisation. By pressing these buttons Level 1, Level
2 or Level 3 may be selected.
ACCESSORIES
Before using the 28085 ultrasound aerosol, check that the packaging contains the following:
User manual: 1 piece
28085 ultrasound aerosol: 1 piece
Medicine container: 5 pieces
Air nozzle: 1 piece
Optional: adult and children’s mask, extension tube, container for transportation
Only use the specic parts indicated in the list above, the use of any other parts may be
dangerous.
The medicine container may only be used once. After each use, the medicine container
should be disposed of according to legal requirements.
Air nozzle
Medicine
container
Mask
Extension tube
Level 3 (0.9 ~1.2 ml/min)
Level 2 (0.6 ~0.9 ml/min)
Level 1 (0.3 ~0.6 ml/min)
Operation light
Stand by button
Nebulisation
controller
ENGLISH

11
USE
The 28085 ultrasound aerosol is used to treat asthma, chronic obstructive pulmonary disorders and other
breathing problems in which a therapy with aerosol is required. It is suitable for use in one single patient.
CLEANING AND ORDINARY MAINTENANCE
It is important to clean the Aerosol before use and after every use, drying it well in every part.
Remember: remove the plug and do not plunge in water
1. Remove the cover
2. Remove any medicine left in the container
3. Remove the water from the tank
4. Rinse the air nozzle, the extension tube and the mask in warm running water for 30 seconds
5. Clean the main element, the cover and the lower air cover as well as the water tank using a damp
cloth and a normal detergent
6. Dry using a clean cloth
Once or twice a week
1. Clean the air nozzle, the extension tube and the mask in a much more in-depth manner. Plunge them
in a solution containing 1 glass of white vinegar and 2 glasses of warm water for 30 minutes.
2. Remove and rinse them.
3. Clean the main unit, the cover, the small air cover and the water tank using a soft cloth with water
and a normal detergent.
4. Dry using a clean cloth.
OPERATION OF THE AEROSOL
Following the instructions provided above for cleaning before using the ultrasound Aerosol for the rst
time or after a long period in which the device is not used.
Remember: Always remove the plug before carrying out cleaning operations and lling with
every use.
1. The operator should sit in a comfortable and erect position.
2. Connect the aerosol to the plug
3. Press the ignition button at the back of the device in order to activate the stand-by mode. The green
light will turn on.
First step: ll the water tank
1. Remove the cover by lifting it, pull out the medicine container if available
2. Use water at a temperature of between 20°C and 30°C (68°F and 86°F), otherwise nebulisation will
be affected. The use of distilled water or a physiological solution is preferred as these liquids prevent
the formation of limestone inside the device that may be ruined if limestone is formed at the base of the
device, a base that allows for the activation of ultrasounds.
3. Fill the tank with 70 ml of water as described above, until reaching the black plastic base.
NB: -The use of cold water (below 20°C or 68°F) would generate a low level of nebulisation
-The use of hot water (more than 30°C or 86°F) would generate greater nebulisation
Second step: Fill the medicine container
1. Place the medicine container inside the water tank
2. Pour the medicine directly into the relative container
3. Put the cover back on as well as the air cover and press the stand-by button to start nebulisation
Remember:
- The maximum capacity of the medicine container is 10 ml and this level should not be
exceeded
- Only use medicinal products recommended by your doctor
ENGLISH

12
The aerosol is programmed initially for level 1 (low level of nebulisation).
According to indications given by a therapist, the operator may select level 2 or level 3 (the maximum level
of nebulisation) by pressing the respective nebulisation commands. By pressing the button repeatedly,
level 2 or 3 may be selected or level 1.
After use press the stand-by button in order to stop nebulisation. The switch on the side is used to turn
the aerosol off.
Warning: Do not put the air nozzle directly in your mouth. Use the accessories recommended
with the air nozzle.
Third step: after each use
1. Remove the plug of the aerosol
2. Carefully remove the medicine container and dispose of the medicine left over
3. Empty the water tank
4. Dry the product in the best possible way
PROBLEM SOLVING
In the event of a fault, check and correct as follows:
Conditions
When I press the ignition button of the
device it does not turn on
The device is on but it does not nebulise
or nebulisation is poor
The device is on but the fan does not
work correctly
Solutions
- Make sure that the plug has been correctly inserted into
the wall
- Check that the water level in the tank is as indicated by
the relative marker
- Check that the medicinal product is in the container.
- Check that the medicine container is not damaged.
- Check that the transducer is clean
Check that the fan has been correctly tted and xed to
the motor shaft
Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it to a specic recycling point for electric and
electronic equipment.
For further information on recycling points contact the local authorities, the local recycling
center or the shop where the product was purchased. If the equipment is not disposed of
correctly, nes or penalties may be applied in accordance with the national legislation and
regulations.
GIMA WARRANTY CONDITIONS
Congratulations for purchasing a GIMA product.
This product meets high qualitative standards both as regards the material and the production. The
warranty is valid for 12 months from the date of supply of GIMA.
During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defected
parts due to production reasons. Labor costs and personnel traveling expenses and packaging not
included. All components subject to wear are not included in the warranty.
The repair or replacement performed during the warranty period shall not extend the warranty.
The warranty is void in the following cases: repairs performed by unauthorized personnel or with non-
original spare parts, defects caused by negligence or incorrect use.
GIMA cannot be held responsible for malfunctioning on electronic devices or software due to outside
agents such as: voltage changes, electro-magnetic elds, radio interferences, etc.
The warranty is void if the above regulations are not observed and if the serial code (if available) has
been removed, cancelled or changed. The defected products must be returned only to the dealer the
product was purchased from. Products sent to GIMA will be rejected.
ENGLISH
Table of contents
Other Gima Respiratory Product manuals
Popular Respiratory Product manuals by other brands

Nebu-tec
Nebu-tec SaHoMa II quick start guide

Trudell Medical International
Trudell Medical International AeroChamber Plus Flow-Vu quick start guide

Philips
Philips OptiLife Instructions for use

INTERSPIRO
INTERSPIRO INCURVE user manual

DeVilbiss
DeVilbiss EasyFit SilkGel Instructions for use

Sundstrom
Sundstrom SR 100 APR user manual

Omron
Omron compA-I-R NE-C28P instruction manual

ResMed
ResMed Ultra Mirage user guide

Suzhou Fangtian Industries
Suzhou Fangtian Industries FT-N059 User instructions

S.E.A. Group
S.E.A. Group SmallTalk ST2-F User instructions

CELIMED
CELIMED U-615 instruction manual

3M
3M Aura 9300+ Series Technical data sheet