GREISINGER GIR 300 User manual

E44.X.01.6C-03
GHM Messtechnik GmbH • Standort Greisinger
Hans-Sachs-Str. 26 • D-93128 Regenstauf
+49 (0) 9402 / 9383-0 +49 (0) 9402 / 9383-33 [email protected]
Universal displaying and controlling device
as of Version 1.1 Operating Manual GIR 300
WEEE-Reg.-Nr. DE 93889386

E44.X.00.6C-03 Connection and operating manual GIR 300 page 2 of 28
Index
1GENERAL NOTE ......................................................................................................................................................3
2INTENDED USE ........................................................................................................................................................3
3SAFETY.....................................................................................................................................................................4
3.1 SAFETY SIGNS AND SYMBOLS.................................................................................................................................4
3.2 SAFETY GUIDELINES..............................................................................................................................................4
3.3 SKILLED PERSONNEL.............................................................................................................................................4
4ELECTRIC CONNECTION........................................................................................................................................5
4.1 TERMINAL ASSIGNMENT.........................................................................................................................................5
4.2 CONNECTION DATA ...............................................................................................................................................5
4.3 CONNECTION AN INPUT SIGNAL ..............................................................................................................................6
4.3.1 Connection a Pt 100 or Pt 1000 RTD probe or a thermocouple probe.......................................................6
4.3.2 Connecting a 4 - 20 mA transmitter in 2-wire-technology...........................................................................6
4.3.3 Connecting a (0)4 - 20 mA transmitter in 3-wire-technology.......................................................................6
4.3.4 Connecting a 0-1V, 0-2V or 0-10V transmitter in 3-wire-technology...........................................................7
4.3.5 Connecting a 0-1/2/10V or 0-50 mV transmitter in 4-wire-technology ........................................................7
4.3.6 Connecting a frequency- or rotation-signal .................................................................................................7
4.3.7 Connecting a counter signal........................................................................................................................9
4.4 CONNECTING SWITCHING OUTPUTS......................................................................................................................10
4.5 COMMON WIRING OF SEVERAL DEVICES................................................................................................................10
5CONFIGURATION...................................................................................................................................................11
5.1 SELECTING AN INPUT SIGNAL TYPE.......................................................................................................................11
5.2 MEASURING VOLTAGE OR CURRENT (0-50MV, 0-1V, 0-2V, 0-10V, 0-20MA, 4-20MA)............................................12
5.3 MEASURING TEMPERATURE (PT 100, PT 1000 RTD PROBES AND THERMOCOUPLES TYPE J, K, N, SOR T) .................................14
5.4 MEASURING OF FREQUENCY (TTL, SWITCHING-CONTACT).....................................................................................15
5.5 MEASURING OF ROTATION SPEED (TTL, SWITCHING-CONTACT) .............................................................................16
5.6 UP-/DOWNWARDS COUNTER................................................................................................................................17
5.7 INTERFACE MODE................................................................................................................................................19
5.8 SELECTION OF THE OUTPUT FUNCTION.................................................................................................................19
6SWITCHING POINTS AND ALARM-BOUNDARIES .............................................................................................21
6.1 2-POINT-CONTROLLER,3-POINT-CONTROLLER......................................................................................................22
6.2 2-POINT-CONTROLLER WITH ALARM FUCTION ........................................................................................................22
6.3 MINIMUM/MAXIMUM ALARM (INDIVIDUAL OR COMMON)............................................................................................23
7OFFSET- AND SLOPE-ADJUSTMENT..................................................................................................................24
8MIN-/MAX-VALUE MEMORY..................................................................................................................................25
9SERIAL INTERFACE..............................................................................................................................................25
10 ERROR CODES ......................................................................................................................................................25
11 SPECIFICATION .....................................................................................................................................................27
12 RESHIPMENT AND DISPOSAL.............................................................................................................................28
12.1 RESHIPMENT......................................................................................................................................................28
12.2 DISPOSAL INSTRUCTION......................................................................................................................................28

E44.X.00.6C-03 Connection and operating manual GIR 300 page 3 of 28
1 General Note
Read this document carefully and get used to the operation of the device before you use it.
Keep this document within easy reach near the device for consulting in case of doubt.
Mounting, start-up, operating, maintenance and removing from operation must be done by qualified, specially trained
staff that have carefully read and understood this manual before starting any work.
The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended one, ig-
noring this manual, operating by unqualified staff as well as unauthorized modifications to the device.
The manufacturer is not liable for any costs or damages incurred at the user or third parties because of the usage or
application of this device, in particular in case of improper use of the device, misuse or malfunction of the connection
or of the device.
The manufacturer is not liable for misprints.
2 Intended Use
The GIR 300 is a microprocessor controlled displaying, monitoring and controlling device..
The device is supporting one universal interface for the connection of:
- Standard transmitter signals (0-20 mA, 4-20 mA, 0-50 mV, 0-1 V, 0-2 V
and 0-10 V),
- RTD (Pt 100 and Pt 1000),
- Thermocouple probes (type K, J, N, T and S)
- Frequency (TTL and switching contact)
As well as rotation measuring, counting, etc...
The device features two switching outputs, which can be configured as 2-point-controller, 3-point-controller, 2-point-
controller with min./max. alarm, common or individual min./max. alarm.
The state of the switching contacts is displayed with 2 LED´s. LED “1” displays the state of the contact of relay 1 and
LED “2” displays the state of contact of relay 2. (LED illuminate = contact is closed)
Furthermore the device supports one EASYBus -interface for communicating with a host computer that makes the
device to a full functions EASYBus-module.
When leaving our factory the GIR 300 has been subjected to various inspection tests and is completely calibrated.
Before the GIR 300 can be used, it has to be configured for the customer’s application.
Hint: In order to avoid undefined input states and unwanted or wrong switching processes, we suggest
to connect the device’s switching outputs after You have configured the device properly.
Hint: By calling a configuration menu (configuration of the measuring input, configuration of the output
function, offset- and slope-adjustment) the measurement and regulation of the device will be deac-
tivated.
By leaving the menu the device will be reinitialised and the measuring/regulation will be started
again.
At the input function “counter” the counter state will be reset by leaving the menu.

E44.X.00.6C-03 Connection and operating manual GIR 300 page 4 of 28
3 Safety
3.1 Safety signs and symbols
Warnings are labeled in this document with the followings signs:
Caution! This symbol warns of imminent danger, death, serious injuries and
significant damage to property at non-observance.
Attention! This symbol warns of possible dangers or dangerous situations
which can provoke damage to the device or environment at non-observance.
Note! This symbol point out processes which can indirectly influence operation
or provoke unforeseen reactions at non-observance.
3.2 Safety guidelines
This device was designed and tested considering the safety regulations for electronic measuring devices.
Faultless operation and reliability in operation of the measuring device can only be assured if the General Safety
Measures and the devices specific safety regulations mentioned in this users manual are considered.
1. Faultless operation and reliability in operation of the measuring device can only be assured if the device is used
within the climatic conditions specified in the chapter “Specifications“.
2. Standard regulations for operation and safety for electrical, light and heavy current equipment have to be ob-
served, with particular attention paid to the national safety regulations (e.g. VDE 0100).
3. When connecting the device to other devices (e.g. the PC) the interconnection has to be designed most thor-
oughly, as internal connections in third-party devices (e.g. connection of ground with protective earth) may lead to
undesired voltage potentials.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked
accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device.
- the device is not working as prescribed.
- storing the device under inappropriate conditions for longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Do not use these products as safety or emergency stop devices or in any other application where failure
of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.
6. Modifications or repairs of the device may not be performed by the customer.
For maintenance or repair the device must be sent to the manufacturer.
3.3 Skilled personnel
Are persons familiar with installation, connection, commissioning and operation of the product and have professional
qualification relating to their job.
For example:
•Training or instruction and qualifications to switch on or off, isolate ground and mark electric circuits and devices or
systems.
•Training or instruction according to the state.
•First-aid training.

E44.X.00.6C-03 Connection and operating manual GIR 300 page 5 of 28
4 Electric connection
Wiring and commissioning of the device must be carried out by trained and skilled personnel only.
Wrong connection may lead to the destruction of the display device, in which case we cannot assume any
warranty!
! Mind for the maximum input conditions !
4.1 Terminal assignment
1
EASYBus-Interface
2
EASYBus-Interface
3
Input: 0-10V
4
Input: 0-50mV, thermocouples, Pt100
5
Input: 0-1V, mA, frequency, Pt100, Pt1000
6
Input: GND, Pt100, Pt1000
7
Power supply GND
8
Power supply +Uv
9
Output 2: relay, break contact
10
Output 2: relay, input
11
Output 1: relay, input
12
Output 1; relay, make contact
4.2 Connection data
between
terminals
typical
limitations
Note
min.
max.
min.
max.
Power supply voltage
7 and 8
9
28
0
30
or corresponding designation
on the type plate
Output 1:
relay: make contact
11 and 12
253 VAC
5A
ohmic load
or corresponding designation
on the type plate
Output 2
relay: break contact
9 and 10
253 VAC
5A
ohmic load
or corresponding designation
on the type plate
Input 0-50mV, TC, ...
4 and 6
0 V
3.3 V
-1 V
10 V,
I<10mA
Input mA
5 and 6
0 mA
20 mA
0 mA
30 mA
Input 0-1(2)V, freq., ...
0 V
3.3 V
-1 V
30 V,
I<5mA
Input 0-10V
3 and 6
0 V
10 V
-1 V
20 V
Input Pt100 (Pt1000)
4, 5 and 6
0 Ω
∞Ω
active signal not allowed
EASYBus-Interface
1 and 2
12 V
36 V
0 V
42 V
These limits must not be exceeded (not even for a short time)!
12 10

E44.X.00.6C-03 Connection and operating manual GIR 300 page 6 of 28
4.3 Connection an input signal
Please take care not to exceed the limitations of the inputs when connecting the device as this may lead to
destruction of the device.
4.3.1 Connection a Pt 100 or Pt 1000 RTD probe or a thermocouple probe
Pt100
6
5
4
Pt100
6
5
4
Pt100
6
5
4
Pt100-RTD probe (3-wire)
Pt100- RTD probe (2-wire)
Pt100- RTD probe (4-wire)
Pt1000
6
5
-
+TC
6
4
Pt1000- RTD probe (2-wire)
Thermocouple probe
4.3.2 Connecting a 4 - 20 mA transmitter in 2-wire-technology
+
_
9-28VDC
Supply:
-Uv
+Uv Transmitter
4-20mA
_
+for transmitter
Supply:
8
7
6
5
+
_
9-28VDC
Supply:
-Uv
+Uv Transmitter
4-20mA
8
7
5
with individual transmitter supply
without separate transmitter supply
4.3.3 Connecting a (0)4 - 20 mA transmitter in 3-wire-technology
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv
Transmitter
0(4)-20mA
_
+for transmitter
Supply:
8
7
6
5
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv
Transmitter
0(4)-20mA
8
7
5
with individual transmitter supply
without separate transmitter supply
set link
non-connected

E44.X.00.6C-03 Connection and operating manual GIR 300 page 7 of 28
4.3.4 Connecting a 0-1V, 0-2V or 0-10V transmitter in 3-wire-technology
+
_
9-28VDC
Supply:
10V
1V / 2V
-Uv
Sig.
+Uv
Transmitter
0-1(10)V
_
+for transmitter
Supply:
8
7
6
5
3
+
_
9-28VDC
Supply:
10V
1V / 2V
-Uv
Sig.
+Uv
Transmitter
0-1(10)V
8
7
5
3
with individual transmitter supply
without separate transmitter supply
4.3.5 Connecting a 0-1/2/10V or 0-50 mV transmitter in 4-wire-technology
+
_
9-28VDC
Supply:
10V
50mV
1V / 2V
-Uv
Sig-
Sig+
+Uv
Transmitter
_
+for transmitter
Supply:
8
7
6
5
4
3
+
_
9-28VDC
Supply:
10V
50mV
1V / 2V -Uv
Sig-
Sig+
+Uv
Transmitter
8
7
6
5
4
3
with individual transmitter supply
without separate transmitter supply
(Note: Sig- and -Uv of the Transmitter must be the same potential)
4.3.6 Connecting a frequency- or rotation-signal
When measuring frequency or rotation three different input signals can be selected in the device’s configuration.
There is the possibility of connecting an active signal (= TTL, ...), a passive sensor-signal with NPN
(= NPN-output, push-button, relay, ...) or PNP (= a PNP output switching to +Us, high-side push-button, ...)
When configuring the device with a NPN switching output, a pull-up-resistor (~7 kΩreferring to +3.3V) is con-
nected internally. So when you use a device with NPN output you don‘t need to connect a resistor externally.
When configuring the device with a PNP switching output, a pull-down resistor (~7 kΩreferring to GND) is connected
internally. So when you use a device with PNP output You don‘t need a resistor externally.
It may be that your measuring-signal source needs the connection of an external resistor e.g. the pull-up-voltage
of 3.3V is not enough for the signal source, or you want to measure in the top level frequency range. In this case
the input signal has to be treated like an active signal and you have to configure the device as „TTL“.
Hint: when connecting the device You have to take care not to exceed the limits of the input volt-
age respective the input current of the frequency-input.

E44.X.00.6C-03 Connection and operating manual GIR 300 page 8 of 28
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv
Transducer
_
+for transducer
Supply:
8
7
6
5
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv
Transducer
8
7
5
Connection of a transducer (with separate power
supply) with TTL or PNP output
Connection of a transducer (without separate power
supply) with TTL or PNP output
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv Transducer
_
+for transducer
Supply:
8
7
6
5
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv Transducer
8
7
5
Connection of a transducer (with separate power
supply) with NPN output
Connection of a transducer (without separate power
supply) with NPN output
+
_
9-28VDC
Supply:
Rv
-Uv
Sig.
+Uv Transducer
_
+for transducer
Supply:
8
7
6
5
+
_
9-28VDC
Supply:
Rv -Uv
Sig.
+Uv Transducer
8
7
5
Connection of a transducer (with separate power
supply) with NPN output and necessary external
resistor
Connection of a transducer (without separate power
supply) with NPN output and necessary external
resistor
Connection note: Rv = 3 kΩ(with power supply voltage =12 V) or 7 kΩ(at 24 V), device configuration: Sens = TTL
+
_
9-28VDC
Supply:
Rv2
Rv1
-Uv
Sig.
+Uv
Transducer
_
+for transducer
Supply:
8
7
6
5
+
_
9-28VDC
Supply:
Rv2
Rv1
-Uv
Sig.
+Uv
Transducer
8
7
6
5
Connection of a transducer (with separate power
supply) with PNP output
and necessary external resistor wiring.
Connection of a transducer (without separate power
supply) with PNP output
and necessary external resistor wiring.
Connection note: Rv2 = 600Ω, Rv1 = 1.8kΩ(with power supply voltage =12V) or 4.2kΩ(at 24V), device configuration: Sens = TTL
(Rv1 is a current limiting resistor and may be shorted if necessary. It should never exceed the mentioned value)

E44.X.00.6C-03 Connection and operating manual GIR 300 page 9 of 28
4.3.7 Connecting a counter signal
When configuring the device you can select 3 different input signal modes similar to the connection of frequency-
and rotation-signals.
The connection of a sensor-signal for a counter-signal is the same used for the frequency- and rotation-signal.
Please use the wiring diagram given in this chapter.
There is the possibility to reset the counter. When connecting contact 4 with GND (e.g. contact 6) the counter will
be reset. You can do this manually (e.g. with the help of a push-button) or automatically (with one switching out-
put of the device).
Hint: When connecting the device, take care not to exceed the limits of the input-voltage or the input-
current of the frequency input.
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv
transducer
Impulse
8
7
6
5
4Reset-
button
-Uv
Sig.
+Uv
transducer
Impulse
_
+9-28 VDC
Supply:
no.2
device
8
7
5
no.1
device
12
11
8
7
5
4
Manually reset the device
with the help of an push-button
+
_
9-28VDC
Supply:
-Uv
Sig.
+Uv
transducer
Impulse
12
11
8
7
6
5
4Reset-
button
automatically resetting with the help of output 1 and
additional resetting the device via push-button
Connection note: output 2 can only be used to
switching of low voltage potential!
Cascading of GIR 300
(Configuration note for the GIR 300:
device 1 –input signal like impulse transducer
device 2 –input signal = switching-contact

E44.X.00.6C-03 Connection and operating manual GIR 300 page 10 of 28
4.4 Connecting switching outputs
Hint: In order to avoid unwanted or wrong switching processes, we suggest to connect the device’s
switching outputs after you have configured the device’s switching outputs properly.
The device features 2 switching outputs by default
• output 1: relay, make contact
• output 2: relay, break contact
Please take care that you must not exceed the limits of the voltage and of the maximum current of
the switching outputs (not even for a short period of time).
Please take extreme care when switching inductive loads (like coils or relays, etc.). Because of
their high voltage peaks, protective measures (e.g. RC-element) to limit these peaks have to be
taken.
Note: You will find detailed information to the switching states of the different output functions in chapter 6
(“Switching points and alarm-boundaries“)
Note: In case of configuring one output as an alarm output, the relay contact will be closed at the idle state (no
alarm is present). If an alarm condition will occurred the relay contact will be opened.
(for detailed information please refer the notes in chapter 6.2 and 6.3)
4.5 Common wiring of several devices
The input of the device are not electrically isolated to the power supply.
When interconnecting several device`s you have to make sure that there is no potential displacement
Make sure to observe the following points:
- When several devices are connected to the same power supply unit it is highly recommended to isolate the sensors,
measuring transducers etc.
- When the sensors, measuring transducers etc. are electrically connected, and you can’t manage to isolate them, you
should use separate electrically isolated power supply units for each devices.
Please note, that an electric connection may also be created via the medium to be measured (e.g. pH-electrodes
and conductivity-electrodes in fluids).

E44.X.00.6C-03 Connection and operating manual GIR 300 page 11 of 28
5 Configuration
Please note: When you are configuring the device and don’t press any button for more than 60 sec.
the configuration of the device will be cancelled. The changes you made will not be
saved and will be lost!
Hint: The buttons 2 and 3 are featured with a ‘roll-function‘. When pressing the button once the value will be
raised (button 2) by one or lowered (button 3) by one.
When holding the button pressed for longer than 1 sec. the value starts counting up or down, the counting
speed will be raised after a short period of time.
The device also features a ‘overflow-function‘, when reaching the upper limit of the range, the device
switches to the lower limit, vice versa.
5.1 Selecting an input signal type
- Turn the device on and wait until it completed its built-in segment test.
- Press button 2 for >2 sec. (e.g. with a small screw driver)
The device displays “InP“ ('INPUT').
- Use button 2 or button 3 (middle or right button) to select the input
signal (see table below).
- Validate the selection with button 1 (the left button). The display will
show “InP“ again
Depending on the selected input signal, additional configurations will be needed.
Input type
Input Signal
To select as
input
proceed in chapter
Voltage signal
0 –10 V
U
5.2
0 –2 V
0 –1 V
0 –50 mV
Current signal
4 –20 mA
I
5.2
0 –20 mA
RTD
Pt 100
T.RES
5.3
Pt 1000
Thermocouples
NiCr-Ni (type K)
T.TC
5.3
Pt10Rh-Pt (type S)
NiCrSi-NiSi (type N)
Fe-CuNi (type J)
Cu-CuNi (type T)
Frequency
TTL-signal
FREQ
5.4
Switch-contact NPN, PNP
Rotation
TTL-signal
RPN
5.5
Switch-contact NPN, PNP
Counter up
TTL-signal
[O.VP
5.6
Switch-contact NPN, PNP
Counter down
TTL-signal
[O.DN
5.6
Switch-contact NPN, PNP
Interface mode
Serial interface
SER,
5.7
Please note: When changing the measuring mode “InP“, the input signal “SEnS“ and the display-unit “Unit“
all settings will be changed to factory default. You have to set all the other settings.
This also regards the settings for offset and slope-adjustment as well as the switching points!
button 1 button 2 button 3

E44.X.00.6C-03 Connection and operating manual GIR 300 page 12 of 28
Parameter Overflow for the available measuring types:
Measuring type
U
I
T.RES
T.TC
FREQ
RPN
[O.VP
[O.DN
SER,
Input signal / sensor
X
X
X
X
X
X
X
X
-
Signal edge
-
-
-
-
-
-
X
X
-
Devisor
-
-
-
-
-
X
X
X
-
Input signal low
-
-
-
-
X
-
-
-
-
Input signal high
-
-
-
-
X
-
X
X
-
Decimal point
X
X
X
-
X
X
X
X
-
Display low
X
X
-
-
X
-
-
-
-
Display high
X
X
-
-
X
-
X
X
-
Temperature unit
-
-
X
X
-
-
-
-
-
Limit
X
X
-
-
X
-
X
X
-
Filter
X
X
X
X
X
-
-
-
-
Address
-
-
-
-
-
-
-
-
X
Output
X
X
X
X
X
X
X
X
X
5.2 Measuring voltage or current (0-50mV, 0-1V, 0-2V, 0-10V, 0-20mA, 4-20mA)
This chapter describes how you configure the device for measuring voltage- or current-signals from an external
transmitter. This instruction demands that you selected “U“ or “I“ as your desired input type like it is explained in chap-
ter 5.1. The display has to show “InP“.
- Press Button 1. The display shows “SEnS“.
- Select the desired input signal using button 2 or button 3 (middle or right button).
Display
Input signal
(voltage measuring)
Notes
10.00
0 –10 V
2.00
0 –2 V
1.00
0 –1 V
0.050
0 –50 mV
Display
Input signal
(current measuring)
Notes
4-20
4 –20 mA
0-20
0 –20 mA
- Validate the selected input signal by pressing button 1. The display shows “SEnS“ again.
- Press button 1 again, The display will show “dP“ (decimal point).
- Select the desired decimal point place by pressing button 2 or button 3.
- Validate the selected decimal position by pressing button 1. The display shows “dP“ again.
- Press button 1 again, the display will show “di.Lo“ (Display Low = low display value).
- Use button 2 and button 3 to select the desired value the device should display when a 0mA, 4mA or
0V input signal is attached.
- Validate the selected value by pressing button 1. The display shows “di.Lo“ again.
- Press button 1 again, the display will show “di.Hi“ (Display High = high display value).
- Use button 2 and button 3 to select the desired value the device should display when a 20mA, 50mV,
1V, 2V or 10V input signal is attached.
- Validate the selected value by pressing button 1. The display shows “di.Hi“ again.
SENS
DP
D,.K,
D,.LO

E44.X.00.6C-03 Connection and operating manual GIR 300 page 13 of 28
- Press button 1 again. The display will show “Li“ (Limit = Measuring range limit).
- Use button 2 and button 3 to select the desired measuring range limit.
Display
Measuring range limit
Notes
off
Deactivated
Exceeding of the measuring range limit is
tolerable for about 10% of the selected input
signal.
on.Er
Active, (displays error)
The measuring range limit is exactly bounded
by the input signal. When exceeding or short-
falling the input signal the device will display
an error message.
on.rG
Active, (displays the selected limit)
The measuring range limit is exactly bounded
by the input signal. When exceeding or short-
falling the input signal the device will display
the selected lower/upper display value.
[e.g. humidity: when shortfalling or exceeding,
the device will display 0% or 100%]
Hint: When exceeding the measuring range limit > 10% independently from the setting, the device
will always display an error message (“Err.1“ or “Err.2“).
- Press button 1 to validate the selection, the display shows “Li“ again.
- When pressing button 1 again, the display will show “FiLt“ (Filter = digital filter).
- Use button 2 and button 3 to select the desired filter [in sec.].
Selectable values: 0.01 ... 2.00 sec.
Explanation: this digital filter is a digital replica of a low pass filter.
Note: when using the input signal 0-50 mV a filter value of at least 0.2 is recommended
- Press button 1 to validate your value, the display shows “FiLt“ again.
Now your device is adjusted to your signal source. Now the only thing left to do is to adjust the outputs of the device.
- When pressing button 1 again, the display shows “outP“. (output)
For configuring the outputs of the device, please follow the instructions given in chapter 5.8.
L,
F,LT
OVTP

E44.X.00.6C-03 Connection and operating manual GIR 300 page 14 of 28
5.3 Measuring temperature (Pt 100, Pt 1000 RTD probes and thermocouples type J, K, N, S or T)
This chapter describes how to configure the device for temperature measuring with the help of external platinum RTD
probes or thermocouple probes. This instruction demands that you selected “t.res“ or “t.tc“ as your desired input type
like it is explained in chapter 5.1. The device has to display “InP“.
- When pressing button 1 the display shows “SEnS “.
- Use button 2 or button 3 (middle or right button) to select your desired input signal.
Display
Input signal
(thermocouples)
Notes
niCr
NiCr-Ni (type K)
Meas. range: -270 ... +1350 °C (-454 ... + 2462 °F)
S
Pt10Rh-Pt (type S)
Meas. range: -50 ... +1750 °C (- 58 ... + 3182 °F)
n
NiCrSi-NiSi (type N)
Meas. range: -270 ... +1300 °C (-454 ... + 2372 °F)
J
Fe-CuNi (type J)
Meas. range: -170 ... + 950 °C (-274 ... + 1742 °F)
t
Cu-CuNi (type T)
Meas. range: -270 ... + 400 °C (-454 ... + 752 °F)
- Validate the selected input signal by pressing button 1. The display shows “SEnS“ again.
- Press button 1 again, the display will show “dP“(decimal point, for the resolution).
This menu parameter is only available at input signal Pt 100 and Pt 1000!
- Use button 2 and button 3 to select whether the temperature is displayed with 0.1° or 1°.
- Validate the selected decimal position by pressing button 1. The display shows “dP“ again.
- Press button 1 again, the display will show “Unit“ (the unit you want to display).
- Use button 2 and button 3 to select whether you want to display °C or °F.
- Press button 1 to validate the selected unit, the display shows “Unit“ again.
- Press button 1 again, the display will be showing “FiLt“ (Filter = digital filter).
- Use button 2 and button 3 for setting the desired filter-value [in sec.].
Selectable values: 0.01 ... 2.00 sec.
Explanation: this digital filter is a digital replica of a low pass filter.
- Press button 1 to validate your selection, the display shows “FiLt“ again.
Now your device is adjusted to your signal source. Now the only thing left to do is to adjust the outputs of the device.
- When pressing button 1 again, the display shows “outP“. (output)
For configuring the outputs of the device, please follow the instructions shown in chapter 5.8.
For setting the offset and for setting the slope-adjustment, please follow the instructions given in chapter 7.
Display
Input signal
(RTD)
Notes
P100
Pt100 (3-wire)
Meas. range: -50.0 ... +200.0 °C (-58.0 ... + 392.0 °F)
Meas. range: -200 ... + 850 °C (-328 ... + 1562 °F)
1000
Pt1000 (2-wire)
Meas. range: -100.0 ... +200.0 °C (-148.0 ... + 392.0 °F)
Meas. range: -200 ... + 850 °C (-328 ... + 1562 °F)
SENS
DP
UN,T
F,LT
OVTP

E44.X.00.6C-03 Connection and operating manual GIR 300 page 15 of 28
5.4 Measuring of frequency (TTL, switching-contact)
This chapter describes how to configure the device for measuring frequency.
This instruction demands that you selected “FrEq“ as your desired input type like it is explained in chapter 5.1.
The device has to display “InP“.
- When pressing button 1 the display will show “SEnS“.
- Use button 2 or button 3 (middle or right button) to select the desired input signal.
Display
Input signal
Notes
ttL
TTL-signal
nPn
Switching contact, NPN
For direct connection of a passive switching
contact (e.g. push button, relay) or Transmitter
with NPN output.
A pull-up-resistor is internally connected.
Hint: when using push-buttons or relays, they
must be bounce-free!
PnP
Switching contact, PNP
For direct connection of a transmitter with
PNP output.
A pull-down-resistor is internally connected.
Hint: For the connection of a frequency-transmitter, please follow the instructions given in chapter.
When connecting a switching-contact-transmitter with increased frequency range (= with external
circuitry) you have to select TTL as your desired input signal.
- Validate your selected input signal by pressing button 1. The display shows “SEnS“ again.
- When pressing button 1 again, the display will show “Fr.Lo“ (frequency low = lower frequency range limit).
- Use button 2 and button 3 to select the lowest frequency that may occur when measuring.
- Press button 1 to validate your selection. The display shows “Fr.Lo“ again.
- When pressing button 1 again, the display will show “Fr.Hi“ (frequency high = upper frequency range limit).
- Use button 2 and button 3 to select the highest frequency that may occur when measuring.
- Press button 1 to validate your selection. The display shows “Fr.Hi“ again.
- When pressing button 1 again, the display will show “dP“ (decimal point).
- Use button 2 and button 3 to select the desired decimal point position.
- Press button 1 to validate your selection. The display shows “dP“ again.
- When pressing button 1 again, the display will show “di.Lo“ (display low = display at lower frequency range limit).
- Set the value the device shall display at the lower frequency range limit by pressing button 2 or button 3.
- Press button 1 to validate your selection. The display shows “di.Lo“ again.
- When pressing button 1 again, the display will show “di.Hi“ (display high = display at upper frequency range limit).
- Set the value the device shall display at the upper frequency range limit by pressing button 2 or button 3.
- Press button 1 to validate your selection. The display shows “di.Hi“ again.
- When pressing button 1 again, the display will show “Li“ (limit = measuring range limitation).
- Use button 2 and button 3 to select the desired measuring range limitation.
Display
Measuring range limit
Notes
off
Inactive
Exceeding of the measuring-frequency is tol-
erable until you reach the maximum measur-
ing range limit.
on.Er
Active, (displays error)
The measuring range is exactly bounded by
the selected frequency-measuring-range-limit.
When exceeding or shortfalling of the limit the
device will display an error message.
SENS
DP
FR.LO
FR.K,
D,.LO
D,.K,
L,

E44.X.00.6C-03 Connection and operating manual GIR 300 page 16 of 28
on.rG
Active, (displays the selected limit)
The measuring range is exactly bounded by
the selected frequency-measuring-range-limit.
When exceeding or shortfalling of the limit the
device will display the lower or upper display-
range-limit.
[e.g. for humidity: when shortfalling or exceed-
ing the device will display 0% or 100%]]
Hint: When exceeding the maximum range limit (10 kHz) independently from the limit setting an error message
will be displayed (“Err.1“).
- Press button 1 to validate your selection. The display shows “Li“ again.
- When pressing button 1 again, the display will show “FiLt“ (Filter = digital filter).
- Use button 2 and button 3 to select the desired filter value [in sec.].
Usable values: 0.01 ... 2.00 sec.
Explanation: this digital filter is a digital replica of a low pass filter.
- Press button 1 to validate your selection. The display shows “FiLt“ again.
Now your device is adjusted to your signal source. The only thing you left do is to adjust the outputs of the device.
- When pressing button 1 again, the display will show “outP“. (Output)
For configuring the outputs of the device, please follow the instructions shown in chapter 5.8.
5.5 Measuring of rotation speed (TTL, switching-contact)
This chapter describes how to configure the device for measuring rotation speed.
This instruction demands that you selected “rPn“ as your desired input type like it is explained in chapter 5.1.
The device has to display “InP“.
- When pressing button 1 the display will show “SEnS“.
- Use button 2 or button 3 (middle or right button) to select the desired input signal.
Display
Input signal
Notes
ttL
TTL-signal
nPn
Switching contact, NPN
For direct connection of a passive switching
contact (e.g. push button, relay) or Transmitter
with NPN output.
A pull-up-resistor is internally connected.
Hint: when using push-buttons or relays, they
must be bounce-free!
PnP
Switching contact, PNP
For direct connection of a transmitter with
PNP output.
A pull-down-resistor is internally connected.
Hint: For the connection of a frequency-transmitter, please follow the instructions given in chapter 4.3.6.
When connecting a switching-contact-transmitter with increased frequency range (= with external
circuitry) you have to select TTL as your desired input signal.
- Validate your selected input signal by pressing button 1. The display shows “SEnS“ again.
- When pressing button 1 again, the display will show “diu“ (divisor).
- Use button 2 and 3 to select your desired divisor.
Set the divisor to the pulses per rotation the transmitter supplies.
- Press button 1 to validate your selection. The display shows “diu“ again.
- When pressing button 1 again, the display will show “dP“ (decimal point).
- Use button 2 and button 3 to select the desired decimal point position.
Use the decimal point position to change the resolution of your measurement.
The more the decimal point position is on the left, the finer the resolution will become. Please note
that you lower the maximum value that can be displayed, either.
Example: your engine runs with 50 rotations per minute.
With no decimal point the device will display something like 49 –50 –51, the maximum value that can
be displayed is 9999 rotations per minute.
F,LT
SENS
OVTP
D,V
DP

E44.X.00.6C-03 Connection and operating manual GIR 300 page 17 of 28
With the decimal point position on --.-- the device will display something like 49.99 –50.00 –50.01,
but the maximum value that can be displayed is 99.99 rotations per minute.
- Press button 1 to validate your selection. The display shows “dP“ again.
Now your device is adjusted to your signal source. The only thing left to do is to adjust the outputs of the device.
- When pressing button 1 again, the display will show “outP“. (Output)
For configuring the outputs of the device, please follow the instructions shown in chapter 5.8.
5.6 Up-/Downwards counter
The upwards counter starts counting upwards from 0 according to its settings.
The downwards counter starts counting downwards from the upper value that had been selected.
The current value of the counter can be reset anytime by connecting terminal 4 to GND (e.g. terminal 6)
The counter starts from its beginning as you disconnect the pin connection.
Feature: The current counter value won‘t be lost if the voltage supply is disconnected.
After restarting the counter starts from this value.
This chapter describes how to configure the device as a counter.
This instruction demands that you selected “Co.up“ or “Co.dn“ as your desired input type like it is explained in chapter
5.1. The device has to display “InP“.
- When pressing button 1 the display will show “SEnS“.
- Use button 2 or button 3 (middle or right button) to select the desired input signal.
Display
Input signal
Notes
ttL
TTL-signal
nPn
Switching contact, NPN
For direct connection of a passive switching
contact (e.g. push button, relay) or Transmitter
with NPN output.
A pull-up-resistor is internally connected.
Hint: when using push-buttons or relays, they
must be bounce-free!
PnP
Switching contact, PNP
For direct connection of a transmitter with
PNP output.
A pull-down-resistor is internally connected.
Hint: For the connection of a frequency-transmitter, please follow the instructions given in chapter 4.3.6 and 4.3.7.
When connecting a switching-contact-transmitter with increased frequency range (= with external circuitry)
you have to select TTL as your desired input signal.
- Validate your selected input signal by pressing button 1. The display shows “SEnS“ again.
- When pressing button 1 again, the device will be displaying “EdGE“ (signal edge).
- Use button 2 or button3 (middle or right button) to select the desired signal edge.
Display
Signal edge
Notes
PoS
positive
The counter is triggered on the positive (ris-
ing) edge.
nEG
negative
The counter is triggered on the negative (fal-
ling) edge.
- Press button 1 to validate your selection, the display shows “EdGE“ again.
- When pressing button 1 again, the display will show “diu“ (divisor = pre-scaling factor).
- Use button 2 and button 3 to select the desired pre-scaling factor.
The incoming pulses will be divided with the selected pre-scaling factor, after that they will be transmit-
ted to the device for further processing.
By this factor you can adapt the device to your transmitter or select a pre-scaling factor for large values.
Example 1: Your flow rate transmitter supplies 165 pulses per litre. When setting a pre-scaling factor
of 165 every 165th pulse (so 1 pulse per litre) will be used for further processing.
SENS
OVTP
D,V
ED6E

E44.X.00.6C-03 Connection and operating manual GIR 300 page 18 of 28
Example 2: Your transmitter is supplying about 5 000 000 pulses during the measurement, which ex-
ceeds the limit of the device.
But when setting a pre-scaling factor of 1000 only every 1000th pulse is used for further
processing. So you only got a value 5000 which won’t exceed the limit of the device.
- Press button 1 to validate your selection. The display shows “diu“ again.
- Press button 1 again. The display shows “Co.Hi“ (counter high = upper counting range limit).
- Use button 2 and button 3 to select the maximum pulse-count (after pre-scaling factor) for the count-
ing process.
Example: Your flow rate transmitter is supplying 1800 pulses per litre, you selected a pre-scaling factor
of 100 and you are expecting a maximum flow rate of 300 litres during the measurement.
With a pre-scaling factor of 100 selected, you will get 18 pulses per litre.
With a maximum flow rate of 300 litres you will be getting a pulse count of 18 * 300 = 5400.
- Press button 1 to validate your selection. The display shows “Co.Hi“ again.
- When pressing button 1 again, the device will be displaying “dP“ (decimal point).
- Use button 2 and button 3 to select the desired decimal point position.
- Press button 1 to validate your selected decimal point position. The display shows “dP“ again.
- Press button 1 again. The display shows “di.Hi“ (display high = upper display range limit).
- Use button 2 and button 3 to set the value to be displayed when the maximum pulse (setting of co.Hi)
count is reached.
Example: Your flow rate transmitter is supplying 1800 pulses per litre and you are expecting a maxi-
mum flow rate of 300 litres. You selected a pre-scaling factor of 100 and a counter range
limit of 5400.
When wanting a resolution of 0.1 litres shown in the display of the device you would have to
set the decimal point position to ---.- and a display range limit of 300.0.
- Press button 1 to validate your selection. The display shows “di.Hi“ again.
- Press button 1. The display will show “Li“ (Limit = measuring range limit).
- Use button 2 and button 3 to select the desired measuring range limit (counter range limit.
Display
Measuring range limit
Notes
off
Inactive
Exceeding of the counter range is tolerable
until you reach the maximum measuring
range limit.
on.Er
active, (error indicator)
The measuring range is exactly bounded by
the selected counter-range-limit. When ex-
ceeding or shortfalling of the limit the device
will display an error message.
on.rG
active, (measuring range limit)
The measuring range is exactly bounded by
the selected counter-range-limit. When ex-
ceeding or shortfalling of the limit the device
will display the upper counter-range-limit or 0.
Hint: The lower counter-range-limit (for configured downwards counter) is fixed to 0.
- Press button 1 to validate your selection. The display shows “Li“ again.
Now your device is adjusted to your signal source. The only thing left to do is to adjust the outputs of the device.
- When pressing button 1 again, the display will show “outP“. (Output)
For configuring the outputs of the device, please follow the instructions shown in chapter 5.8.
OVTP
DP
(O.K,
D,.K,
L,

E44.X.00.6C-03 Connection and operating manual GIR 300 page 19 of 28
5.7 Interface mode
When the device is in the interface mode it won’t make any measurements by itself.
The value shown in the device’s display is sent via serial interface.
But the switching and alarm functions of the displayed value are still available.
The EASYBus-Address of the device needed for the communication can be set manually with the device itself or with
the help of an EASYBus-software (like EASYBus-Configurator).
Please note, when carrying out an EASYBus-system-initialisation the device’s address will be reset automatically.
This chapter describes how to configure the device as an EASYBus-display.
This instruction demands that you selected “SEri“ as your desired input type like it is explained in chapter 5.1.
The device has to display “InP“.
- When pressing button 1 again, the device will display “Adr“ (address).
- Use button 2 and button 3 to select the desired address [0 ... 239] of the device.
- Press button 1 to validate the selected device address. The display shows “Adr“ again.
You don’t need any further configuration but the outputs.
- When pressing button 1 again, the device will be displaying “outP“ (output).
For configuring the outputs please follow the instructions given in chapter 5.8.
5.8 Selection of the output function
- After configuration of the input (chapter 5.2 –5.7) you have to select the output function.
The display shows “outP“ (output).
- Use button 2 and button 3 (middle or right button) to select the desired output-function.
Description
Function
To select as
output
For switching point
setting please
refer to chapter
Output 1
Output 2
No output, device is used
as display unit
inactive
(contact is open)
inactive
(contact is open)
no
--
2-point-controller
switching function 1
(active = contact closed)
switching function 1
(active = contact is open)
2P
6.1
3- point-controller
switching function 1
(active = contact closed)
switching function 2
*1
(active = contact closed)
3P
6.1
2- point-controller with
min-/max-alarm
switching function 1
(active = contact closed)
min-/max-alarm
(alarm = contact closed)
2P.AL
6.2
min-/max-alarm,
common
min-/max-alarm
(alarm = contact is open)
min-/max-alarm
(alarm = contact closed)
AL.F1
6.3
min-/max-alarm,
individual*2
max-alarm
(alarm = contact closed)
min-alarm
(alarm = contact closed)
AL.F2
6.3
*1 = Please take notice, that output 2 is a breaking contact of the relay, which means, that the contact will be
closed if the device have no power supply!
2= Note: please take notice, that output 1 and 2 has different contact types.
Through this the device will have different contact states by max- and min-alarm in case of no
power supply is present! (Max alarm = contact is open, min alarm = contact is closed)
ADR
OVTP

E44.X.00.6C-03 Connection and operating manual GIR 300 page 20 of 28
- Press button 1 to validate the selected output function. The display shows “outP“ again.
Depending on your output function setting, it may be possible that one or more settings described below won’t be
available.
- When pressing button 1 again, the device will display “1.dEL“ (delay of output 1).
- Use button 2 and button 3 to set the desired value for the switching-delay of output 1.
Hint: The selected value [0.01 ... 2.00] will be in seconds.
- Press button 1 to validate the selection. The display shows “1.dEL“ again.
- When pressing button 1 again, the device will display “1.Err“ (preferred state of output 1).
- Use button 2 and button 3 (middle or right button) to set the desired initial state in case of an error.
Display
Preferred state of the output
Notes
off
Inactive in case of an error
(contact is open)
on
Active in case of an error
(contact ist closed)
- Press button 1 to validate the selection. The display shows “1.Err“ again.
- In case you selected a 3-point-controller you have to make the following settings similar to the settings
you already made for output 1:
“2.dEL“ (delay of output 2) and
“2.Err“ (preferred state of output 2).
Depending on the selected output function you have to make the settings for switching and alarm points.
See description in chapter 6 „Switching points and alarm-boundaries“ for further information.
Hint: The settings for the switching and alarm points can be made later in an extra menu (see chapter 6)
1.DEL
1.ERR
2.DEL
2.ERR
Table of contents
Other GREISINGER Controllers manuals

GREISINGER
GREISINGER GEWAS 191 User manual

GREISINGER
GREISINGER EB 3000 Instruction manual

GREISINGER
GREISINGER gir 2000Pt Technical document

GREISINGER
GREISINGER GRA 0420 VO User manual

GREISINGER
GREISINGER gir 2000Pt Instruction manual

GREISINGER
GREISINGER GIR 230 Pt1000 / DIF Instruction manual

GREISINGER
GREISINGER GIR 230 Pt User manual