GHT-117 HV Adapter Box
GHT-117
The GHT-117 High Voltage Adapter Box is an optional accessory of the GPT
series Electrical Safety Analyzer/Tester aimed at providing a convenient way to
connect the DUT’s AC line socket to the analyzer/tester for the AC withstanding,
DC withstanding and Insulation Resistance tests.
Consideration
The GPT series Electrical Safety Analyzers/Testers operate
at up to 5kVAC and 6kVDC. To prevent electrical shock, the
grounding wire of the adapter box (green/yellow wire)
must be connected to the ground connector of the GPT
series rear panel. Please also make sure that the GPT series
is properly grounded via its AC line input. Please refer to
the specific GPT series User Manual for more information
regarding safety and the earth ground connection.
Connection
POWER
START STOP
REMOTE
MANU/AUTO EDIT/SAVE UTILITY
ESC PAGE
GPT-9803 Tester
HIGH VOLTAGE
CAUTION 5.0 kVAC MAX.
6.0 kVDC MAX.
AC / DC Withstanding Voltage /
Insulation Resistance
RETURN
PASS FAIL READY TEST
DUT GPT-9000
Rear panel
GND
GHT-117
HV
Return
Ground