INSTRUCTIONS FOR USE
Second observer
tube
4. Edition / 2021 – 08

2
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
INSTRUCTIONS FOR USE
Second observer tube
4. Edition / 2021 – 08
Preface
Thank you for choosing a Haag-Streit device. Provided you comply carefully with the
regulations in these instructions for use, we can guarantee reliable and trouble-free
use of our product.
WARNING!
Read the instruction manual carefully before commissioning this product.
It contains important information regarding the safety of the user and
patient.
NOTE!
For USA only: Federal law restricts this device to sale by or on the order
of a physician or licensed practitioner.
Contents
¥ 1 Safety...................................................................................................................................................3
◦ 1.1 Comments on these instructions for use........................................................................................3
◦ 1.2 Ambient conditions.........................................................................................................................3
◦ 1.3 Shipment and unpacking ...............................................................................................................3
◦ 1.4 Installation warnings ......................................................................................................................3
◦ 1.5 Operation, environment .................................................................................................................3
◦ 1.6 Disinfection ....................................................................................................................................4
◦ 1.7 Warranty and product liability ........................................................................................................4
◦ 1.8 Reporting obligation .......................................................................................................................4
◦ 1.9 Description of symbols...................................................................................................................4
¥ 2 Purpose of use...............................................................................................................................4
¥ 3 Introduction .....................................................................................................................................5
◦ 3.1 Overview ........................................................................................................................................5
¥ 4 Device assembly / installation .............................................................................................5
¥ 5 Commissioning .............................................................................................................................6
¥ 6 Operation...........................................................................................................................................6
◦ 6.1 Setting the viewing height (Second observer tube long)................................................................6
◦ 6.2 Adjusting the image upright position (Second observer tube long)................................................7
¥ 7 Decommissioning........................................................................................................................7
¥ 8 Technical data ............................................................................................................................... 7
◦ 8.1 Second observer tube short ..........................................................................................................7
◦ 8.2 Second observer tube long............................................................................................................ 7
¥ 9 Maintenance.................................................................................................................................... 7
◦ 9.1 Device inspection ..........................................................................................................................7
◦ 9.2 Servicing........................................................................................................................................ 7
◦ 9.3 Cleaning and disinfection ..............................................................................................................8
¥ 10 Appendix ........................................................................................................................................ 8
◦ 10.1 Accessories / functionals parts / detachable parts / consumables .............................................. 8
◦ 10.2 Disposal....................................................................................................................................... 8

3
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
DEUTSCHENGLISHFRANÇAISITALIANOESPAÑOLNEDERLANDS PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
1Safety
DANGER!
Failure to comply with these instructions may result in material damage
or pose a danger to patients or users.
WARNING!
These warnings must absolutely be complied with to guarantee safe
operation of the product and to avoid any danger to users and to
patients.
NOTE!
Important information, please read carefully.
1.1 Comments on these instructions for use
NOTE!
In these instructions for use the point is used as decimal separator.
1.2 Ambient conditions
Temperature −40 °C ... +70 °C
Air pressure 500 hPa ... 1060 hPa
Transport
Relative humidity 10 % ... 95 %
Temperature −10 °C ... +55 °C
Air pressure 700 hPa ... 1060 hPa
Storage
Relative humidity 10 % ... 95 %
Temperature +10 °C ... +35 °C
Air pressure 800 hPa ... 1060 hPa
Use
Relative humidity 30 % ... 90 %
1.3 Shipment and unpacking
• Before unpacking the device, check whether the packaging shows traces of
improper handling or damage. If this is the case, notify the transport company
that delivered the goods to you.
• Unpack the device together with a representative of the transport company.
Make a report of any damaged parts. This report must be signed by you and by
the representative of the transport company.
• Leave the device in the packaging for a few hours before unpacking it
(condensation).
• Check the device for damage after it is unpacked.
• Return defective devices in the appropriate packaging.
• Store packaging material carefully so that it can be used for potential returns or
when moving.
• Check that the connection parts of the accessories are in the correct position
(screw connections, quick-release fasteners).
1.4 Installation warnings
WARNING!
• Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained
specialists.
• Only original Haag-Streit spare parts may be used.
1.5 Operation, environment
WARNING!
• Due to the risk of injury, it must not be possible for passing persons
to bump against the second observer tube during an examination.
• The second observer tube must be mounted where its contours
which protrude over the table top will not impede passing persons.
• Avoid moisture.

4
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
NOTE!
• This device must only be operated by qualified personnel. The
owner is responsible for their training.
• This device may only be used in accordance with the instructions in
the 'Purpose of use' chapter.
1.6 Disinfection
NOTE!
The device can, but does not need to be disinfected. For more
information, please refer to the 'Maintenance' chapter.
1.7 Warranty and product liability
• Haag-Streit products must be used only for the purposes and in the manner
described in the documents distributed with the product.
• The product must be treated as described in the ‘Safety’ chapter. Improper
handling can damage the product. This would void all guarantee claims.
• Continued use of a product damaged by incorrect handling may lead to personal
injury. In such a case, the manufacturer will not accept any liability.
• Haag-Streit does not grant any warranties, either expressed or implied, including
implied warranties of merchantability or fitness for a particular use.
• Haag-Streit expressly disclaims liability for incidental or consequential damage
resulting from the use of the product.
• This product is covered by a limited warranty granted by your seller.
• For USA only: This product is covered by a limited warranty, which may be
reviewed at www.haag-streit-usa.com.
1.8 Reporting obligation
NOTE!
Any serious incident that has occurred in relation to the device must be
reported to Haag-Streit and the competent authority of the Member
State in your country.
1.9 Description of symbols
General warning, read the
accompanying documentation
European certificate of
conformity
Trademark of the manufacturer Haag-
Streit AG Manufacturer
Haag-Streit reference number Serial number
2Purpose of use
The second observer tube was developed exclusively for the Haag-Streit slit lamp
BQ 900. It enables additional persons to take part in the examination.

5
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
DEUTSCHENGLISHFRANÇAISITALIANOESPAÑOLNEDERLANDS PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
3Introduction
The observation in the second observer tube is monocular and takes place in the
right or left-hand side of the beam path of the stereo-microscope depending on the
slit lamp connection.
3.1 Overview
Required accessories for the slit lamp BQ 900:
1. Beam splitter 50/50 Haag-Streit
part number 1400308
2. Second observer tube short
(1400301)
3. Eyepiece
4. CE label / Haag-Streit part number
(back) type designation
5. Tonometer stop
6. Union nut
7. Second observer tube long
(1400034)
8. Twist grip for image upright
position
9. Release button
10. Selectable relief springs
11. Bayonet joint
12. Joy stick base locking screw
13. Microscope arm locking screw
14. Cover
4 Device assembly / installation
NOTE!
• If the tube is mounted on the same side as the tonometer, the
tonometer stop (5) must be fixed on the end of the tube.
• If the tube is left on the slit lamp permanently, the selectable relief
springs (10) must be reset (See instruction manual for the BQ 900
slit lamp)
• Tighten the control knob (12) on the cross slide to block the microscope arm and
the screw to block (13) horizontal movements.
• Loosening the bayonet joint (11) on the microscope (caution, hold the
microscope tight!) enables the stereo-microscope to be dismantled and the
beam splitter to be mounted.
• Carefully mount the stereo-microscope on the beam splitter with the bayonet
joint.
• Loosen the lid mounted on the side of the beam splitter (14).
• Hold the second observer tube in accordance with the figure. Align the groove
on the end of the tube vertically to the wedges of the beam splitter connecting
sleeve.
• Push the tube into the stop and fix it in place with the union nut (6).

6
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
5 Commissioning
Following installation the device will be immediately ready for use; no further
measures are required.
6Operation
NOTE!
The eyepiece (3) must be set according to the refraction of the co-
observer.
6.1 Setting the viewing height (Second observer tube long)
The second observer tube can be turned around the horizontal axis coming laterally
out of the beam splitter. This enables the viewing height in the eyepiece to be varied
so that the second observer can look into the eyepiece at a freely adjustable
inclination angle.
FORBIDDEN!
The viewing height should not be adjusted during the examination, as
the patient or person carrying out the examination could be injured.
To adjust the viewing height, the release button (9) must be released first of all. To
do this, the elbow at the side of the beam splitter must be held so that the thumb is
positioned on top of the release button. To stabilize the slit lamp, the other hand
should be used to support the tube. Press and hold the release button and turn the
tube about the horizontal axis on the beam splitter. When the desired inclination
angle is reached, release the release button and turn the tube carefully until the next
engaging point is reached.

7
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
DEUTSCHENGLISHFRANÇAISITALIANOESPAÑOLNEDERLANDS PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
6.2 Adjusting the image upright position (Second observer
tube long)
The image visible through the eyepiece is turned about the visual axis by adjusting
the viewing height of the second observer. The image is uprighted by adjusting the
twist grip (8).
7 Decommissioning
Decommissioning is not required. However, we recommend dismounting the second
observer tube (long) after use in order to prevent accidental impacts.
8 Technical data
8.1 Second observer tube short
Focal length f 132 mm
Image size and frame Are identical in the binocular tube and second observer
tube
Length of the tube ~150 mm
Stop for tonometer Can be mounted / dismantled as required
8.2 Second observer tube long
Focal length f 132 mm
Image size and frame Are identical in the binocular tube and second observer
tube
Length of the tube ~410 mm
Stop for tonometer Can be mounted / dismantled as required
9 Maintenance
WARNING!
• Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained
specialists.
• Contact your Haag-Streit representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
• Only original Haag-Streit spare parts may be used.
9.1 Device inspection
The following procedure should be used to check that the second observer tubes
are working correctly:
• Insert the test rod into the radial movement bearing, whilst at the same time
aligning the surface to the microscope at a right angle.
• Set the slit length to 8 or 14 mm.
• Set the illumination intensity to 50%.
• Set the magnification in the microscope to max.
• Set the eyepieces in such a way that the test rod is in sharp focus. In doing so,
turn the ocular from the (+) to the (-) side.
• For all magnifications, the structure of the test rod must be shown in sharp
focus.
• In the second observer tubes, the same area must also be in sharp focus at all
magnifications
9.2 Servicing
To guarantee a long service life, the device must be cleaned weekly as described
and protected with the dust cover when not in use. We recommend having the
device inspected once a year by an authorized service technician.

8
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
9.3 Cleaning and disinfection
The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped
down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friendly
disinfectants (containing aldehyde or aldehyde-free) are also permitted, such as
Kohrsolin FF.
WARNING!
• The preparation instructions provided do not apply to tonometer
measuring prisms.
• Tonometer measuring prisms must be prepared in accordance with
a different manual.
• Too strong or aggressive disinfectants or cleaning liquids e.g.
hydrogen peroxide will damage the finish and coating of the device.
• Do not use sprays.
• Observe the manufacturer's safety instructions.
• Do not use any cloths that drip.
• Wring out any soaked cloths before use when necessary.
• Ensure that no liquid penetrates the device.
• Comply with the stipulated exposure time.
• Clean optical surfaces after disinfection with a very soft cloth.
NOTE!
IP code: IPX0 (device is not protected against liquids)
10 Appendix
10.1 Accessories / functionals parts / detachable parts /
consumables
Components REF
Beam splitter 50/50 1400308
Eyepiece 12.5 × with dioptre adjustment 1400303
10.2 Disposal
For correct disposal, please contact your Haag-Streit representative.
This will guarantee that no hazardous substances enter the
environment and that valuable raw materials are recycled.

9
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS SVENSKA
DEUTSCHENGLISHFRANÇAISITALIANOESPAÑOLNEDERLANDS PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08
Should you have any further questions, please contact your Haag-Streit representative at: