Model H120
pH Meter with Silicon Chip Sensor
IQ Scientific Instruments
Calibration and Measurement Instructions
Parts
When finished, always rinse the probe and replace the protective cap.
C
Remove cap from probe
tip and turn on the meter.
Rinse the probe in tap
water and blot dry.
Apply one drop of 7.00
buffer to the probe tip. Be
sure to cover both the
reference junction and the
pH sensor.
Depress the CAL 1 button
with the tip of the clip.
The CAL and 7.0 will flash
until a stable reading is
obtained. When the CAL
disappears and 7.0 stops
blinking, calibration is
complete.
Rinse the probe in tap
water and blot dry.
To measure the pH of a
sample, go to Step 7.
Now measure the sample
pH by placing a single
drop on the probe tip or
dipping the probe tip
into solution.
1234567
6. Protective Cap 7. Clip
8. Bubble
9. POWER Button
10. CAL 1 Button 1. Probe Tip pH sensor and reference electrode.
2. pH Sensor Fast response silicon chip pH sensor with built-in temperature sensor.
3. Reference Electrode Replaceable saturated KCl Ag/AgCl electrode.
4. Reference Junction Measurement is made when the reference junction and the pH sensor are
immersed in the solution.
5. LCD Display CAL (blinking): Calibration in process. Measurement cannot be made.
CAL (off): Measurement can be made.
BAT: Low battery indicator. Replace batteries.
6. Protective Cap Always replace the cap after use.
7. Clip Use the tip of the clip to depress the CAL button.
8. Bubble The reference solution bubble indicates the life of the reference electrode.
9. POWER Button Press to turn the meter on or off.
10. CAL 1 Button Button for pH 7.0 calibration.
1. Probe Tip
2. pH Sensor
3. Reference Electrode
4. Reference Junction
5. LCD Display
OPERATING INSTRUCTIONS
pH Sensor
Reference Junction
DOC022.53.80059
mini_man_H120 9/8/09 9:26 AM Page 1
SPECIFICATIONS miniLab H120
Catalog Number H120
Model miniLab H120
Meter Pocket-sized waterproof pH meter
Sensor Silicon chip pH sensor
Calibration 1
Buffer Recognition Automatic pH 7.0
Temp. Compensation Automatic
Reference Replaceable KCl gel filled. Part No. RF01-01
pH Range: pH 2.0 to pH 12.0
Resolution: 0.1 pH
Accuracy ±0.1 pH
Operating Temp. Range 5 to 40 °C (40 °F to 105 °F)
Display LCD digital display
Power Auto Shut Off. Two 3v lithium batteries CR2032
Battery Life 150 hours continuous. 10 hour low battery warning.
Dimensions 152.4 x 28.57 x 16.38mm (6.0” x 1.125” x .645”) 53.94 g (1.03 oz)
Shipping Weight 1.0 lb. (.45 kg)
Troubleshooting Guide
If an error message is displayed first check the following:
No pH buffer on the pH sensor.
If the meter is reading 0.0 or 14.0, the sensor is dry
Air bubbles are trapped on the sensor surface.
pH sensor is dirty. See “Cleaning the Sensor”.
pH sensor and reference is not in solution.
BAT Message Displayed
Batteries are too low for reliable measurement. See “Replace Batteries”
(Battery Icon) Displayed
Approximately 10 hours of battery life remains. Replace batteries immediately.
Difficulty in Calibrating or in Obtaining a Stable Reading:
pH sensor is dirty. See “Cleaning the Sensor”.
The reference electrode has reached the end of its useful life. Replace reference electrode.
Interference from direct sunlight. Shade sensor from sunlight.
pH or temperature of sample is changing.
Sample has low ionic strength (tap water, distilled water).
Notes on calibration and measurement:
• Measurements cannot be made while CAL is displayed. Be sure to complete the calibration procedure before
taking pH measurements.
• Always begin a measuring session with a calibration.
• Be sure to use pH 7.0, buffer as a calibration standard.
• The use of other solutions to calibrate may make measurements inaccurate.
• Rinse the sensor with distilled or deionized water when testing the pH of tap water, rainwater, or clean water.
• Although the pH meter has automatic temperature compensation, always keep the pH buffers and the samples at
the same temperature.
• To measure solids such as soil, make a slurry of the sample in deionized or distilled water.
• Keep the sensor surface clean. See “Cleaning the Sensor”.
• A white powder or gel at the probe tip is KCl reference solution. Clean from sensor before use.
• To change C°or F° temperature unit, turn off meter. Press and hold CAL 1 button, then press the POWER button
to turn meter back on.
DO:
Soak the probe in pH 7.0 (neutral) pH buffer for 5 minutes if the probe is new or has not been in regular use.
Clean the probe regularly with soft cotton tipped swab.
STORE THE PROBE DRY with the protective cap covering the probe tip. No electrode storage solution is required.
For maximum accuracy always begin each measuring session with a calibration.
Calibrate at the same temperature as the sample solution. Although the meter has automatic temperature compensation,
best results will be achieved if the calibration buffers and sample are the same temperature.
Be sure the surface of the sensor in the probe is free from any deposits or films. See the cleaning instructions in this manual.
Always place the protective cap over the sensor tip when finished measuring.
DO NOT
DO NOT store the sensor in solution or use for long term pH measuring applications.
DO NOT use below 5 °C or above 40 °C (40 - 105 °F).
DO NOT leave the sensor uncapped for long periods of time.
DO NOT allow oil, fat, food particles, starch, protein, or other materials to remain on the pH sensor after use.
DO NOT use a sharp metal object (needle, pin, etc.) to clean the pH sensor surface.
DO NOT take readings in direct sunlight. Direct sunlight may cause unstable readings or difficulty in calibration.
DO NOT use in an environment that will damage pH sensor or meter: Organic solvents (acetone, toluene, thinner, oils), strong
acids (pH 0 - 2), strong alkalis (pH 12 - 14), abrasive samples, silicon etching compounds (hydrofluoric acid).
DO NOT press the POWER or CAL buttons with sharp objects.
DO NOT submerge the meter. The meter has water-resistant O-ring seals, it is not submersible.
DO NOT use in environments with static electricity. Electrostatic discharge may permanently damage the probe.
Error Codes:
CODE Description Solution
E06 pH sensor slope error Clean probe. If error persists, replace meter.
E07 pH sensor voltage error Clean probe. Be sure sample liquid covers both sensor and reference junction.
CODE Description Solution
E08 Too long to calibrate Clean probe. If error persists, replace reference. If error still persists,
replace meter.
E13 Temperature Sensor error Replace meter
Cleaning the Sensor:
1. Clean the sensor with soapy water and a toothbrush or
cotton tipped swab.
2. Rinse the sensor with water.
3. Recalibrate.
CAUTIONS:
Do not scratch the pH sensor.
Do not press the reference electrode junction.
White powder or gel on the pH sensor is KCl reference solution. Clean before using.
Do not clean the sensor with organic solvents such as acetone, methanol, or thinner.
Replacing the Batteries:
1. Wipe the pH meter dry.
2. Pull the meter case apart as shown to the right.
3. Gently pry out the batteries and replace with two 3v lithium CR2032 batteries.
The (+) side of the batteries should face up.
4. Be sure the O-ring is not damaged and is in the correct position.
5. Reassemble the meter.
6. Recalibrate.
CAUTIONS:
Do not open the meter case if the pH meter is wet.
Always replace both batteries at the same time.
Dispose of batteries properly.
O-Ring
NOTE:
The service life of the reference electrode will vary
according to the frequency of measurement, temperature,
and other measurement conditions.
CAUTIONS:
Do not remove the reference electrode if the meter is wet.
Be sure that the O-ring seals properly. If the O-ring does
not seal properly to the meter, liquid may enter and cause
permanent damage.
Replace the Reference When...
• Liquid in the reference electrode is
depleted. Shown to the right.
• Response time slows
• There is difficulty in obtaining a stable
reading.
Replacing the Reference.
1. Wipe the pH meter dry.
2. Pull the reference electrode out of the
pH meter as shown to the right.
3. Check to be sure that the watertight
O-ring is clean and properly seated on
the new reference.
4.
Insert the new reference into the pH
meter. Part No. RF01-01.
5. Recalibrate the pH meter.
!
!
!
!
Bubble
Large air bubble,
time to replace reference electrode.
O-Ring
mini_man_H120 9/8/09 9:26 AM Page 2

Other Hach Measuring Instrument manuals

Hach QbD1200+ User manual

Hach

Hach QbD1200+ User manual

Hach HQd Quick manual

Hach

Hach HQd Quick manual

Hach HQ440d Quick manual

Hach

Hach HQ440d Quick manual

Hach NA5600 sc Na+ Instruction Manual

Hach

Hach NA5600 sc Na+ Instruction Manual

Hach TL2350 Quick manual

Hach

Hach TL2350 Quick manual

Hach sensION3 User manual

Hach

Hach sensION3 User manual

Hach TU5300 sc Quick manual

Hach

Hach TU5300 sc Quick manual

Hach ORBISPHERE 3654 User manual

Hach

Hach ORBISPHERE 3654 User manual

Hach TPM870 User manual

Hach

Hach TPM870 User manual

Hach ORBISPHERE 410 User manual

Hach

Hach ORBISPHERE 410 User manual

Hach 1735 TRO User manual

Hach

Hach 1735 TRO User manual

Hach ISENA381 Series User manual

Hach

Hach ISENA381 Series User manual

Hach REFO 3 User manual

Hach

Hach REFO 3 User manual

Hach sc100 1720E User manual

Hach

Hach sc100 1720E User manual

Hach ORBISPHERE 3650 Atex Quick manual

Hach

Hach ORBISPHERE 3650 Atex Quick manual

Hach POLYMETRON 83 series User manual

Hach

Hach POLYMETRON 83 series User manual

Hach 9186 User manual

Hach

Hach 9186 User manual

Hach BODTrak II User manual

Hach

Hach BODTrak II User manual

Hach FT660 sc Quick manual

Hach

Hach FT660 sc Quick manual

Hach TU5200 Quick manual

Hach

Hach TU5200 Quick manual

Hach CA610 User manual

Hach

Hach CA610 User manual

Hach Polymetron 9610sc Operating instructions

Hach

Hach Polymetron 9610sc Operating instructions

Hach LDO101 User manual

Hach

Hach LDO101 User manual

Hach AF7000 Quick manual

Hach

Hach AF7000 Quick manual

Popular Measuring Instrument manuals by other brands

Emerson Rosemount 2410 manual

Emerson

Emerson Rosemount 2410 manual

EzCAP 293P quick start guide

EzCAP

EzCAP 293P quick start guide

Opti-Med JM022-001 user manual

Opti-Med

Opti-Med JM022-001 user manual

HBM C16i series Mounting instructions

HBM

HBM C16i series Mounting instructions

Anritsu MT8802A Service manual

Anritsu

Anritsu MT8802A Service manual

Endress+Hauser Proline Prosonic Flow P 500 Functional safety manual

Endress+Hauser

Endress+Hauser Proline Prosonic Flow P 500 Functional safety manual

Pyle PSPL25 manual

Pyle

Pyle PSPL25 manual

Kimans HYDRASONIC S8 user manual

Kimans

Kimans HYDRASONIC S8 user manual

V&A VA8070 user manual

V&A

V&A VA8070 user manual

Kanomax OMX-ADM Operation manual

Kanomax

Kanomax OMX-ADM Operation manual

Emerson Micro Motion G-Series installation manual

Emerson

Emerson Micro Motion G-Series installation manual

Love TID Series Specifications-installation and operating instructions

Love

Love TID Series Specifications-installation and operating instructions

SKF TMRT 1 Instructions for use

SKF

SKF TMRT 1 Instructions for use

Oval ALTI mass Series instructions

Oval

Oval ALTI mass Series instructions

Bosch GLM Professional 80 Original instructions

Bosch

Bosch GLM Professional 80 Original instructions

SilverNet SIL ECHO M5 user guide

SilverNet

SilverNet SIL ECHO M5 user guide

Holman DRG001 user guide

Holman

Holman DRG001 user guide

Maxtec Handi+N2 Operating manual & instructions for use

Maxtec

Maxtec Handi+N2 Operating manual & instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.