EC MEASURE
New calibration
in 15 days
10-04-2010 17:15
EC MEASURE
Calibrate EC
10-04-2010 17:15
EC CALIBRATION
Calibrate EC
4. Operation
4.2. Start-up
Connect to the rear panel of the instrument:
- the conductivity cell
- the magnetic stirrer
- the plug-in power supply
Follow the next steps:
For the first time...
Successive times...
4.3. Factory configuration
The
sensION™+EC71
is an instrument for EC, salinity and TDS
measurements. The factory configuration is suitable for the
majority of applications.
Electrical conductivity (EC)
Temperature of reference (TRef) 25 ˚C
Temperature coefficient (CT) linear 2.00 %/ ºC.
Measurements: By stability.
In continuous pressing twice the key .
Calibration in 1, 2 or 3 points, with molar standards.
Calibration frequency, every 15 days.
Automatic sample coding.
Data sending, report type, etc. in accordance with the start up
configuration.
Salinity (NaCl)
Temperature of reference (TRef) 25 ˚C
Temperature coefficient (TC) linear 2.00 %/ ºC.
Measurements: By stability.
In continuous pressing twice the key .
Calibration in 1, 2 or 3 points, with molar standards.
Calibration frequency, every 15 days.
Automatic sample coding.
Data sending, report type, etc. in accordance with the start up
configuration.
Total dissolved solids (TDS)
Temperature of reference (TRef) 25 ˚C.
Temperature coefficient (TC) linear 2.00 %/ ºC.
Factor 0.64
Measurements: By stability.
In continuous pressing twice the key .
Calibration in 1, 2 or 3 points, with molar standards.
Calibration frequency, every 15 days.
Automatic sample coding.
Data sending, report type, etc. in accordance with the start up
configuration.
Resistivity ( )
Temperature of reference (TRef) 25 ˚C
Temperature coefficient (CT) linear 2.00 %/ ºC.
Measurements: By stability.
In continuous pressing twice the key .
Calibration in 1, 2 or 3 points, with EC molar standards.
Calibration frequency, every 15 days.
Automatic sample coding.
Data sending, report type, etc. in accordance with the start up
configuration.
Re-programming
To perform any modification in the default programming, for
example different measuring modes, calibration with other type
of buffers, etc. see corresponding chapters.
Press
Instrument switched off.
Press the keys to select the
language.
Press the keys to select the data
output. If a printer is available select the
corresponding options appearing on the
display. See page 17 (shady options)
Calibrate with standard solutions.
sensION+ EC71 HACH LANGE 7010.indd ANG:6sensION+ EC71 HACH LANGE 7010.indd ANG:6 8/7/10 16:14:478/7/10 16:14:47