INSTRUMENT CONDITIONS
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CHANNEL 1
Units...........................................................................................pH
Resolution............................................................................. 0.01
Measurement..........................................................By stability
Criteria...........................................................................standard
Type of calibration ....................................Technical buffers
Frequency of calibration.....................................every 24 h.
CHANNEL 2
Units...........................................................................................ISE
Measurement..........................................................By stability
Criteria...........................................................................standard
Curve number...................................................................... 1, F-
Uniits measurement........................................................mol/l
Standard 1......................................................................0.00001
Standard 2........................................................................ 0.0001
Standard 3...........................................................................0.001
Standard 4............................................................................. 0.01
Standard 5.............................................................................. 0.1
SYSTEM
Stirrer number 2.................................................................. non
Codification ............................................................. automatic
Data send .........................................................................printer
Type of printeri.............................................................thermal
Type of report..............................................................reduced
Samples by report........................................................ several
Readjust temperature.........................................................NO
------------------------------------------------------------------------
ELECTRODE HISTORY Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
Installed : 01-03-07 15:39:32
In service: 06:30
Number of measurements: 216
Maximum value Minimum value
pH 13.75 1.15
mV -- --
T (˚C) 38.2 14.3
------------------------------------------------------------------------
MEASUREMENT DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
Sample pH ˚C Time St%
----------- ---------- --------- ---------- -------
15-03-07 14:36
38510 10.15 23.5 00:15 30
15-03-07 14:40
38511 9,85 23.2 00:22 30
15-03-07 14:46
38512 10.23 22.8 00:18 30
16-03-07 08:36
Pozo1 7.48 23.1 00:25 30
------------------------------------------------------------------------
pH CALIBRATION DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CALIBRATED 29-07-2010 15:42:51 22.5˚C
Buffers Slope Sens P. Asym. Time
Technical mV/pH % mV s
----------- ---------- ----- ---------- ----------
4.01,7.00 58.16 99.0 -1.5 18
7.00,9.21 57.91 98.5 -1.5
Stirring 30%
Calibration frequency, every 24 h
------------------------------------------------------------------------
4. Operation
1. Channel 1 and Channel 2 can measure pH, mV or ISE with no difference.
2. The system configuration, sample codification, data output, type of report,
3. Data Logger. The measurements will be stored in the Data Logger of the
4.4. General diagram