INSTRUMENT CONDITIONS
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CHANNEL 1
Units...........................................................................................pH
Resolution............................................................................. 0.01
Measurement..........................................................By stability
Criteria...........................................................................standard
Type of calibration ....................................Technical buffers
Frequency of calibration.....................................every 24 h.
CHANNEL 2
Units...........................................................................................ISE
Measurement..........................................................By stability
Criteria...........................................................................standard
Curve number...................................................................... 1, F-
Uniits measurement........................................................mol/l
Standard 1......................................................................0.00001
Standard 2........................................................................ 0.0001
Standard 3...........................................................................0.001
Standard 4............................................................................. 0.01
Standard 5.............................................................................. 0.1
SYSTEM
Stirrer number 2.................................................................. non
Codification ............................................................. automatic
Data send .........................................................................printer
Type of printeri.............................................................thermal
Type of report..............................................................reduced
Samples by report........................................................ several
Readjust temperature.........................................................NO
------------------------------------------------------------------------
ELECTRODE HISTORY Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
Installed : 01-03-07 15:39:32
In service: 06:30
Number of measurements: 216
Maximum value Minimum value
pH 13.75 1.15
mV -- --
T (˚C) 38.2 14.3
------------------------------------------------------------------------
MEASUREMENT DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
Sample pH ˚C Time St%
----------- ---------- --------- ---------- -------
15-03-07 14:36
38510 10.15 23.5 00:15 30
15-03-07 14:40
38511 9,85 23.2 00:22 30
15-03-07 14:46
38512 10.23 22.8 00:18 30
16-03-07 08:36
Pozo1 7.48 23.1 00:25 30
------------------------------------------------------------------------
pH CALIBRATION DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CALIBRATED 29-07-2010 15:42:51 22.5˚C
Buffers Slope Sens P. Asym. Time
Technical mV/pH % mV s
----------- ---------- ----- ---------- ----------
4.01,7.00 58.16 99.0 -1.5 18
7.00,9.21 57.91 98.5 -1.5
Stirring 30%
Calibration frequency, every 24 h
------------------------------------------------------------------------
INSTRUMENT CONDITIONS
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CHANNEL 1
Units...........................................................................................pH
Resolution............................................................................. 0.01
Measurement..........................................................By stability
Criteria...........................................................................standard
Type of calibration ....................................Technical buffers
Frequency of calibration.....................................every 24 h.
CHANNEL 2
Units...........................................................................................ISE
Measurement..........................................................By stability
Criteria...........................................................................standard
Curve number...................................................................... 1, F-
Uniits measurement........................................................mol/l
Standard 1......................................................................0.00001
Standard 2........................................................................ 0.0001
Standard 3...........................................................................0.001
Standard 4............................................................................. 0.01
Standard 5.............................................................................. 0.1
SYSTEM
Stirrer number 2.................................................................. non
Codification ............................................................. automatic
Data send .........................................................................printer
Type of printeri.............................................................thermal
Type of report..............................................................reduced
Samples by report........................................................ several
Readjust temperature.........................................................NO
------------------------------------------------------------------------
ELECTRODE HISTORY Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
Installed : 01-03-07 15:39:32
In service: 06:30
Number of measurements: 216
Maximum value Minimum value
pH 13.75 1.15
mV -- --
T (˚C) 38.2 14.3
------------------------------------------------------------------------
MEASUREMENT DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
Sample pH ˚C Time St%
----------- ---------- --------- ---------- -------
15-03-07 14:36
38510 10.15 23.5 00:15 30
15-03-07 14:40
38511 9,85 23.2 00:22 30
15-03-07 14:46
38512 10.23 22.8 00:18 30
16-03-07 08:36
Pozo1 7.48 23.1 00:25 30
------------------------------------------------------------------------
pH CALIBRATION DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CALIBRATED 29-07-2010 15:42:51 22.5˚C
Buffers Slope Sens P. Asym. Time
Technical mV/pH % mV s
----------- ---------- ----- ---------- ----------
4.01,7.00 58.16 99.0 -1.5 18
7.00,9.21 57.91 98.5 -1.5
Stirring 30%
Calibration frequency, every 24 h
------------------------------------------------------------------------
INSTRUMENT CONDITIONS
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CHANNEL 1
Units...........................................................................................pH
Resolution............................................................................. 0.01
Measurement..........................................................By stability
Criteria...........................................................................standard
Type of calibration ....................................Technical buffers
Frequency of calibration.....................................every 24 h.
CHANNEL 2
Units...........................................................................................ISE
Measurement..........................................................By stability
Criteria...........................................................................standard
Curve number...................................................................... 1, F-
Uniits measurement........................................................mol/l
Standard 1......................................................................0.00001
Standard 2........................................................................ 0.0001
Standard 3...........................................................................0.001
Standard 4............................................................................. 0.01
Standard 5.............................................................................. 0.1
SYSTEM
Stirrer number 2.................................................................. non
Codification ............................................................. automatic
Data send .........................................................................printer
Type of printeri.............................................................thermal
Type of report..............................................................reduced
Samples by report........................................................ several
Readjust temperature.........................................................NO
------------------------------------------------------------------------
ELECTRODE HISTORY Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
Installed : 01-03-07 15:39:32
In service: 06:30
Number of measurements: 216
Maximum value Minimum value
pH 13.75 1.15
mV -- --
T (˚C) 38.2 14.3
------------------------------------------------------------------------
MEASUREMENT DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
Sample pH ˚C Time St%
----------- ---------- --------- ---------- -------
15-03-07 14:36
38510 10.15 23.5 00:15 30
15-03-07 14:40
38511 9,85 23.2 00:22 30
15-03-07 14:46
38512 10.23 22.8 00:18 30
16-03-07 08:36
Pozo1 7.48 23.1 00:25 30
------------------------------------------------------------------------
pH CALIBRATION DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CALIBRATED 29-07-2010 15:42:51 22.5˚C
Buffers Slope Sens P. Asym. Time
Technical mV/pH % mV s
----------- ---------- ----- ---------- ----------
4.01,7.00 58.16 99.0 -1.5 18
7.00,9.21 57.91 98.5 -1.5
Stirring 30%
Calibration frequency, every 24 h
------------------------------------------------------------------------
INSTRUMENT CONDITIONS
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CHANNEL 1
Units...........................................................................................pH
Resolution............................................................................. 0.01
Measurement..........................................................By stability
Criteria...........................................................................standard
Type of calibration ....................................Technical buffers
Frequency of calibration.....................................every 24 h.
CHANNEL 2
Units...........................................................................................ISE
Measurement..........................................................By stability
Criteria...........................................................................standard
Curve number...................................................................... 1, F-
Uniits measurement........................................................mol/l
Standard 1......................................................................0.00001
Standard 2........................................................................ 0.0001
Standard 3...........................................................................0.001
Standard 4............................................................................. 0.01
Standard 5.............................................................................. 0.1
SYSTEM
Stirrer number 2.................................................................. non
Codification ............................................................. automatic
Data send .........................................................................printer
Type of printeri.............................................................thermal
Type of report..............................................................reduced
Samples by report........................................................ several
Readjust temperature.........................................................NO
------------------------------------------------------------------------
ELECTRODE HISTORY Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
Installed : 01-03-07 15:39:32
In service: 06:30
Number of measurements: 216
Maximum value Minimum value
pH 13.75 1.15
mV -- --
T (˚C) 38.2 14.3
------------------------------------------------------------------------
MEASUREMENT DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
Sample pH ˚C Time St%
----------- ---------- --------- ---------- -------
15-03-07 14:36
38510 10.15 23.5 00:15 30
15-03-07 14:40
38511 9,85 23.2 00:22 30
15-03-07 14:46
38512 10.23 22.8 00:18 30
16-03-07 08:36
Pozo1 7.48 23.1 00:25 30
------------------------------------------------------------------------
pH CALIBRATION DATA Channel 1
SN:70602 V1.0
Monday 30 July 2010 15:50
------------------------------------------------------------------------
CALIBRATED 29-07-2010 15:42:51 22.5˚C
Buffers Slope Sens P. Asym. Time
Technical mV/pH % mV s
----------- ---------- ----- ---------- ----------
4.01,7.00 58.16 99.0 -1.5 18
7.00,9.21 57.91 98.5 -1.5
Stirring 30%
Calibration frequency, every 24 h
------------------------------------------------------------------------
Next Prev.
DATA LOGGER
10-04-2010 17:15
Channel 1
Consult
Print
Erase
DATA LOGGER C1
Consult
Print
Reset
DATA LOGGER C1
Consult
Print
Erase
DATA LOGGER C1
DATA LOGGER C1
DATA LOGGER C1
Consult:
Measurement Data
Electrode Report
Calibration Data
Consult:
Measurement Data
Electrode Report
Calibration Data
Consult:
Measurement Data
Electrode Report
Calibration Data
DATA LOGGER C1
Print:
Measurement Data
Electrode Report
Calibration Data
Calibration Report
Instrument Conditions
DATA LOGGER C1
Reset:
Measurement Data
Electrode Report
DATA LOGGER C1
Warning!
All data will be
lost.
Erase
DATA LOGGER C1
Meas. 1/400
By stability
Water P1
pH 7.05 21.4˚C
1-2
DATA LOGGER C1
Electrode
16-04-2010 15:46:12
In service: 10:55
Measurement n.: 272
1-2
DATA LOGGER C1
pH caibration
16-04-2010 15:46:12
Technical buffers
4.00 7.02 9.25
1-3
DATA LOGGER C1
Exit
Exit
4. Operation
4.12. Data Logger
The
sensION™+MM340
automatically memorizes the last 400 measurements by each Channel, the current calibrations, the
sensor’s history and the two Channels measuring conditions.
All these data can be viewed at any moment on display, printed or sent to a PC.
Report examples
Report printing with configura-
tion “Printer data output”
Reset Data Logger
Display view
Standard and GLP report: If this type of
report is selected , there will be data
about the electrode, report header and
user’s name in the reports with calibration
data and electrode’s history.
First measurement
To see the rest of
the data about the
first measurement.
To see the rest of
the measurements
Electrode data
To see the rest
of the data.
Calibration data.
Last 9 calibration.
To see the rest
of the data.