manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hach
  6. •
  7. Transmitter
  8. •
  9. Hach si792 P series User manual

Hach si792 P series User manual

© Hach Company, 2007. All rights reserved. Printed in Germany. For technical assistance or ordering information: visit www.hach.com, call 800-227-4224 or email [email protected] Edition 1 December 2007
Quick Start Guide
For si792 P series 2-wire transmitters DOC016.53.00794
Additional copies are available on www.hach.com
DANGER: Explosion Hazard. This Quick Start Guide is intended for setup of the si792
transmitter in a non-hazardous location. It is not a substitute for reading and following
the instructions in the user manual. Installation in a hazardous location must be
performed by qualified personnel per the applicable control drawing or ATEX certificate
on the prived documentation CD.
DANGER: Electric Shock Hazard. Disconnect power to all wiring before making
connection to the si792 P series transmitters.
1. Connect the 24 VDC power and communication (Table 1).
2. Connect the sensor wires (Table 2).
Basic transmitter configuration—si792(x) P (4–20 mA/HART) only
Note: Refer to the manual for si792x P-FF (Foundation Fieldbus) and si792x P-PA (Profibus PA)
configuration instructions.
1. To enter the configuration mode, press CONF, configuration code: 1200, ENTER. There are 4
configuration menus:
• output "out.1MNU"
• temperature compensation "tc MNU"
• calibration setup "CAL MNU"
• alarm settings "ALrt MNU"
2. In configuration mode, press the RIGHT ARROW KEY to navigate to a specific configuration menu.
Press ENTER to access the menu options.
3. Use the ARROW KEYS to edit values. Press ENTER to save the settings. If a value is outside of
the acceptable range, “Err” will be displayed and the value will not be accepted.
To exit the configuration mode and return to measurement mode, press CONF, ENTER at any time.
Configure the transmitter for ORP sensors
The transmitter has default settings for use with pH sensors. For use with an ORP sensor:
1. Select the "out.1MNU" configuration menu.
2. Change the sensor type from pH to ORP.
a. Press CONF, ENTER. Input passcode 1200 and press ENTER.
bUse the RIGHT ARROW KEY to change from pH to ORP under the "01.UniT" display code.
Press ENTER.
cPress CONF, ENTER to exit configuration mode.
Configure the pH buffer
For accurate pH calibration, configure the correct manufacturer of pH buffers. Calibration is rarely
performed at exactly 25 °C, and different pH buffers can exhibit different temperature dependencies.
3. Press CONF, ENTER. Input passcode 1200 and press ENTER.
4. Select "CAL MNU".
5. Select the appropriate code from Table 3.
Note:If the buffer is not listed in Table 3, select manual entry (MAN). Enter the pH values when
prompted during calibration.
6. Press ENTER. Press CONF, ENTER to exit configuration mode.
Table 1 Terminal assignments
Terminal No. si792(x) P si792x P-FF si792x P-PA
14 (–) (–) (–)
15 (+) (+) (+)
Table 2 Terminal assignments—differential sensors
Terminal No. Assignment Hach wire color
1Measure electrode (pHD is Reference) red
4Reference electrode (pHD is Measure) green
5
Cable shield connections (2 shield wires for
Hach pHD, 1 shield wire for Hach LCP)
Note:
Do not connect to earth ground.
shield
8RTD (resistive temperature device) yellow
17 –6 V white
18 GND black
Table 3 Calibration mode setup menu—CAL mnu
Select pH buffer set
–01–BUF Knick/Mettler-Toledo
–02–BUF Merck Titrisols, Riedel Fixanals
–03–BUF Ciba (94)
–04–BUF NIST technical buffers
–05–BUF NIST standard buffers
–06–BUF HACH buffers
–07–BUF WTW technical buffers
–08–BUF Hamilton Duracal
MAN (manual entry, for buffers not listed in this table)
DAT—Entry of offset and slope of premeasured electrodes
© Hach Company, 2007. All rights reserved. Printed in Germany. For technical assistance or ordering information: visit www.hach.com, call 800-227-4224 or email [email protected] Edition 1 December 2007
Quick Start Guide
For si792 P series 2-wire transmitters DOC016.53.00794
7. Calibrate the sensor. The two most common calibration methods are described below. Refer to
the user manual for other details. Refer to Table 4 for default passcode settings.
Calibration with automatic buffer recognition
Complete the following steps to calibrate the pH probe:
a. Press CAL, enter passcode: 1100, ENTER. The display will show CAL1.
bRemove the pH and temperature probes from the process. Clean the probes and immerse
in the first buffer solution (any order). Briefly stir the solution with the probes.
cPress ENTER. Automatic recognition of the buffer solution will begin. The display will first
show the hourglass and no buffer value, then the recognized buffer value followed by the
measured millivolt value. When the measurement is complete, the display will show CAL2.
dWhen CAL2 is displayed, remove the probes from the first buffer solution and rinse thor-
oughly.
eComplete a one-point calibration or start a two-point calibration:
•One-point calibration: press CAL to end the calibration. The slope (%) and offset (mV) of the
probe (based on 25 °C) will be displayed.
•Two-point calibration: immerse the probes in the second buffer solution. Press ENTER. The
buffer detection will repeat as for the first buffer solution. When complete, the slope (%) and off-
set (mV) of the probe (based on 25 °C) will be displayed.
fPress ENTER, ENTER to return to the measurement mode. The HOLD mode will remain
active for 20 seconds.
gRemove the probes from the buffer, rinse and return to the process.
Calibration by Comparison
The probe can be calibrated by entering the sample value from measurement in an instru-
ment such as a laboratory or portable meter (one-point calibration). The probe does not need
to be removed from the process during calibration. The sample temperature should corre-
spond to the measured process temperature for accurate results.
a. Press CAL, enter passcode: 1105, ENTER. The display will briefly show CAL PRD and
then StorE.
bCollect a grab sample for measurement in the lab. Press ENTER to store the sample value
in the transmitter.
cThe cal mode indicator will flash to indicate that calibration is not complete. The transmit-
ter will continue to measure and display the current sample values.
dMeasure the grab sample with a laboratory or portable meter.
ePress CAL, enter passcode: 1105, ENTER to access the product calibration once more.
The display will briefly show CAL PRD and then CALC.
fEdit the displayed value to match the value measured by the laboratory or portable meter
and press ENTER. The slope and offset will be shown.
gPress ENTER to end the calibration. The display will show the measured value alternating
with Hold.
hPress ENTER to return to the measuring mode. The outputs will remain in the hold mode
for approximately 20 seconds.
Note:For comparison calibration via Foundation Fieldbus, see the pH user manual,
DOC026.53.00794. Comparison calibration via Profibus is not available.
Table 4 Default passcode settings
Display Function Default setting
Cal INFO Calibration information 0000
CAL 0-CAL Calibration zero 1001 (not for
si792x P-PA)
CAL CAL Calibration 1100
CAL PROD Product calibration 1105 (not for
si792x P-PA)
CAL RTD Temperature probe adjustment 1015
CFG ERR Error information 0000
CFG CONF Configuration 1200
CFG SNSR MO Sensor monitor 2222
CFG OUT Current source 5555
CFG SPCL ST Administrator passcode. Use ARROW keys to
select NO or YES 1989
NO SPCL ST Do not change
administrator passcode
Press ENTER to return to the default
passcode
Press CONF to exit the editor
YES SPCL ST Change administrator
passcode
Press ENTER to accept the new
passcode
Press CONF to exit the editor

Other Hach Transmitter manuals

Hach CAx440EX User manual

Hach

Hach CAx440EX User manual

Hach si792 D-FF Series User manual

Hach

Hach si792 D-FF Series User manual

Hach si792 D series User manual

Hach

Hach si792 D series User manual

Hach si792 T User manual

Hach

Hach si792 T User manual

Hach si792 D-FF Series User manual

Hach

Hach si792 D-FF Series User manual

Hach si792 E User manual

Hach

Hach si792 E User manual

Hach CAx440EX User manual

Hach

Hach CAx440EX User manual

Popular Transmitter manuals by other brands

AVGear AVG-HDWP70HD user manual

AVGear

AVGear AVG-HDWP70HD user manual

Endress+Hauser iTEMP TMT82 operating instructions

Endress+Hauser

Endress+Hauser iTEMP TMT82 operating instructions

Danfoss MBS 4201 instructions

Danfoss

Danfoss MBS 4201 instructions

Sennheiser SI 10-92 manual

Sennheiser

Sennheiser SI 10-92 manual

Rotronic HygroFlex 1 instruction manual

Rotronic

Rotronic HygroFlex 1 instruction manual

Jamara Compa X3 Evo Instruction

Jamara

Jamara Compa X3 Evo Instruction

Quick RRCH02 Manual of installation and use

Quick

Quick RRCH02 Manual of installation and use

Magnetrol Eclipse 706-512 manual

Magnetrol

Magnetrol Eclipse 706-512 manual

Tekon DUOS inCO2 installation guide

Tekon

Tekon DUOS inCO2 installation guide

Shure P4T user guide

Shure

Shure P4T user guide

Eaton MTL83*C Modbus Software manual

Eaton

Eaton MTL83*C Modbus Software manual

CMOSTEK NextGenRF CMT2189B user guide

CMOSTEK

CMOSTEK NextGenRF CMT2189B user guide

Nautel GV10 installation manual

Nautel

Nautel GV10 installation manual

Erone 295W user manual

Erone

Erone 295W user manual

Soundcast VGtx user manual

Soundcast

Soundcast VGtx user manual

Ness RR2 Installation notes

Ness

Ness RR2 Installation notes

Miller Edge MWR13 installation instructions

Miller Edge

Miller Edge MWR13 installation instructions

Monacor JTS TG-98RA /5 instruction manual

Monacor

Monacor JTS TG-98RA /5 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.