manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hach
  6. •
  7. Transmitter
  8. •
  9. Hach si792 D series User manual

Hach si792 D series User manual

Other Hach Transmitter manuals

Hach CAx440EX User manual

Hach

Hach CAx440EX User manual

Hach si792 D-FF Series User manual

Hach

Hach si792 D-FF Series User manual

Hach si792 T User manual

Hach

Hach si792 T User manual

Hach si792 E User manual

Hach

Hach si792 E User manual

Hach si792 D-FF Series User manual

Hach

Hach si792 D-FF Series User manual

Hach CAx440EX User manual

Hach

Hach CAx440EX User manual

Hach si792 P series User manual

Hach

Hach si792 P series User manual

Popular Transmitter manuals by other brands

NIVELCO EasyTREK SCA-300 Series Programming manual

NIVELCO

NIVELCO EasyTREK SCA-300 Series Programming manual

Honeywell 5816WMBR installation instructions

Honeywell

Honeywell 5816WMBR installation instructions

Kamstrup READy MTU Installation and operation guide

Kamstrup

Kamstrup READy MTU Installation and operation guide

Omega ZW Series user guide

Omega

Omega ZW Series user guide

ABB Field 264H Instructions for installation

ABB

ABB Field 264H Instructions for installation

Alinco DJ-V17T Service manual

Alinco

Alinco DJ-V17T Service manual

Dejero EnGo 3x manual

Dejero

Dejero EnGo 3x manual

Rosemount 4600 Reference manual

Rosemount

Rosemount 4600 Reference manual

Speaka Professional 2342740 operating instructions

Speaka Professional

Speaka Professional 2342740 operating instructions

trubomat GAB 1000 instruction manual

trubomat

trubomat GAB 1000 instruction manual

Teledyne Analytical Instruments LXT-380 instructions

Teledyne Analytical Instruments

Teledyne Analytical Instruments LXT-380 instructions

Rondish UT-11 quick start guide

Rondish

Rondish UT-11 quick start guide

MG 10JRE instruction manual

MG

MG 10JRE instruction manual

Burkert 8177 operating instructions

Burkert

Burkert 8177 operating instructions

Firetide HotPort 3203 Hardware installation guide

Firetide

Firetide HotPort 3203 Hardware installation guide

Tele Radio T9-02 User instructions

Tele Radio

Tele Radio T9-02 User instructions

Rotronic HygroFlex HF3 Series user guide

Rotronic

Rotronic HygroFlex HF3 Series user guide

Jamara FCX6 Pro Tel Instruction

Jamara

Jamara FCX6 Pro Tel Instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

© Hach Company, 2007, 2014. All rights reserved. Printed in Germany. Edition 2 March 2014
For technical assistance or ordering information: visit www.hach.com, call 800-227-4224 or email [email protected]
Quick Start Guide
For si792 D series 2-wire transmitters DOC016.53.00797
Additional copies are available on www.hach.com
DANGER: Explosion Hazard. This Quick Start Guide is intended for setup of the si792
transmitter in a non-hazardous location. It is not a substitute for reading and following
the instructions in the user manual. Installation in a hazardous location must be
performed by qualified personnel per the applicable control drawing or ATEX certificate
in the si792 D on the provided documentation CD.
DANGER: Electric Shock Hazard. Disconnect power to all wiring before making
connection to the si792 D series transmitters.
1. Connect the power and communication (Table 1).
2. Connect the sensor wires (Table 2 and Table 3).
3. The transmitter is set by default for Oxysens and Oxyferm. Manually select the sensor type for the
Oxygold G (si792(x) D only. Refer to the user manual for si792x D-FF and si792x D-PA.)
a. Press CONF, ENTER.
bInput passcode: 1200. Press ENTER. The display will show “out.1MNU”. Press ENTER.
cThe display will show “tyPE A” and “o1 SnSR”. Use the ARROW keys to select “tyPE A” (Oxy-
ferm and Oxysens) or “tyPE B” (Oxygold). Press ENTER.
dPress CONF, ENTER to exit the configuration mode. The transmitter will remain in hold mode
for 20 seconds, then return to measurement mode.
4. Calibrate the sensor. The two most common calibration methods are described below. Refer to
the user manual for other details.
Calibration in air (recommended)
a. Press CAL, enter passcode: 1100, ENTER. The display will briefly show CAL and then the
entered (or default) humidity value.
bRemove the sensor from the process and place in air.
cEnter the actual relative humidity and press ENTER. The display will show the entered (or
default) barometric pressure.
dEnter the barometric pressure and press ENTER. A drift check will begin. The drift check may
take some time. When complete, the display will show the concentration in the units specified
in the configuration menu.
Note:To exit the drift check after more than 10 seconds, press CAL (accuracy will be reduced).
eAccept the calculated value for the concentration or adjust the value as necessary and then
press ENTER. The display will show the new slope and zero (relative to 25 °C and
1.013 mbar).
fPress ENTER to end the calibration. Return the sensor to the process. The measured value
will be shown in the display alternately with Hold.
gPress ENTER to return to the measure mode.
Calibration by Comparison
The sensor can be calibrated for DO (or gas volume concentration) by entering the sample
value from a measurement in an instrument such as a laboratory or portable meter. The sensor
does not need to be removed from the process during calibration. The sample temperature
should correspond to the measured process temperature for accurate results.
a. Press CAL, enter passcode: 1105, ENTER. The display will briefly show CAL PRD and then
StorE.
bCollect a grab sample for measurement in the lab. Press ENTER to store the sample value in
the transmitter.
cThe cal mode indicator will flash to indicate that calibration is not complete. The transmitter
will continue to measure and display the current sample values.
dMeasure the grab sample with a laboratory or portable meter.
ePress CAL, enter passcode: 1105, ENTER to access the product calibration once more. The
display will briefly show CAL PRD and then CALC.
fEdit the displayed value to match the value measured by the laboratory or portable meter and
press ENTER. The new slope and zero point will be shown.
gPress ENTER to end the calibration. Return the sensor to the process. The measured value
will be shown in the display alternately with Hold.
hPress ENTER to return to the measure mode.
Note:For comparison calibration via Foundation Fieldbus, see the user manual.
Table 1 Terminal assignments
Terminal No. 4–20 mA/HART Profibus PA Foundation Fieldbus
10 (+) (–) (–)
11 (–) (+) (+)
Table 2 Terminal assignments—si792(x) D (4–20 mA/HART)
Terminal No. Oxygold G/Oxyferm sensor wire color1Oxysens sensor wire color
1clear clear
5shield/red brown
6shield/yellow-green shield
7green yellow
8white blue
1With VP cable LZY079, LZY353 and LZY354
Table 3 Terminal assignments—si792x D-FF and si792x D-PA
Terminal No. Oxygold G/Oxyferm sensor wire color1
1With VP cable LZY079, LZY353 and LZY354
Oxysens sensor wire color
2clear clear
4shield/red brown
5shield/yellow-green shield
7green yellow
8white blue
© Hach Company, 2007, 2014. All rights reserved. Printed in Germany. Edition 2 March 2014
For technical assistance or ordering information: visit www.hach.com, call 800-227-4224 or email [email protected]
Quick Start Guide
For si792 D series 2-wire transmitters DOC016.53.00797
Additional Configuration
Note: Additional configuration is for 4-20mA /Hart version only. For Foundation Fieldbus or Profibus PA,
refer to the user manual for configuration instructions.
Refer to Table 4 for default passcode settings.
1. To enter the configuration mode, press CONF, configuration code: 1200, ENTER. There are 5
configuration menus:
• current output "out.1MNU"
• temperature compensation "tc MNU"
• corrections configuration "Corr MNU"
• Calibration setup "Cal MNU"
• alarm settings "ALrt MNU"
2. In configuration mode, press the RIGHT ARROW KEY to navigate to a specific configuration
menu. Press ENTER to access the menu options.
3. Use the ARROW KEYS to edit values. Press ENTER to save the settings. If a value is outside of
the acceptable range, “Err” will be displayed and the value will not be accepted.
4. To exit the configuration mode and return to measurement mode, press CONF, ENTER at
any time.
Refer to user manual for detailed configuration instructions and a complete list of menu options.
Table 4 Default passcode settings
Display Function Default setting
Cal INFO Calibration information 0000
CAL 0-CAL Calibration zero 1001
CAL CAL Calibration Saturation/Concentration 1100
CAL PROD Product calibration 1105
CAL RTD Temperature sensor adjustment 1015
CFG ERR Error information 0000
CFG CONF Configuration 1200
CFG SNSR MO Probe monitor 2222
CFG OUT Current source 5555
CFG SPCL ST Administrator passcode. Use ARROW keys to
select NO or YES 1989
NO SPCL ST Do not change
administrator passcode
Press ENTER to return to the default
passcode
Press CONF to exit the editor
YES SPCL ST Change administrator
passcode
Press ENTER to accept the new
passcode
Press CONF to exit the editor