HI713
Phosphate Low Range
INSTRUCTION MANUAL
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI713-25 Reagents for 25 Phosphate Low Range tests
Other Accessories
HI713-11 Phosphate LR Certified Standard Kit
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST713 05/18
HI713 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 3 minutes of non-use and 2 minutes
after reading.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
All manuals and user guides at all-guides.com
• Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “Add”, “C.1” with “Press” blinking,
the meter is ready.
• Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
• Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
• Remove the cuvette from the meter and
unscrew the cap. Add the content of one
packet of HI713-0 reagent. Replace the cap
and shake gently for 2 minutes until the
powder is completely dissolved. Place the
cuvette back into the meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 3 minutes and press the
button.
• The instrument directly displays the
concentration of phosphate in ppm. The
meter automatically turns off after 2 minutes.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use
the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI713 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0.00 to 2.50 ppm
Resolution 0.01 ppm
Accuracy ±0.04 ppm ±4% of reading @25 °C/77 °F
Light Source Light Emitting Diode @525 nm
Light Detector Silicon Photocell
Method
Adaptation of the Standard Methods for the Examination of Water
and Wastewater, 20th edition, Ascorbic Acid method. The reaction
between phosphate and the reagent causes a blue tint in the sample.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 3 minutes of non-use and 2 minutes after reading
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications
All manuals and user guides at all-guides.com

Other manuals for HI 713

5

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI96733C User manual

Hanna Instruments

Hanna Instruments HI96733C User manual

Hanna Instruments HI749 User manual

Hanna Instruments

Hanna Instruments HI749 User manual

Hanna Instruments HI 9813-5 User manual

Hanna Instruments

Hanna Instruments HI 9813-5 User manual

Hanna Instruments HI 88713 User manual

Hanna Instruments

Hanna Instruments HI 88713 User manual

Hanna Instruments Checktemp 1 User manual

Hanna Instruments

Hanna Instruments Checktemp 1 User manual

Hanna Instruments HI 5315 User manual

Hanna Instruments

Hanna Instruments HI 5315 User manual

Hanna Instruments HI 9210N User manual

Hanna Instruments

Hanna Instruments HI 9210N User manual

Hanna Instruments HI 4110 User manual

Hanna Instruments

Hanna Instruments HI 4110 User manual

Hanna Instruments HI 4101 User manual

Hanna Instruments

Hanna Instruments HI 4101 User manual

Hanna Instruments HI 96771C User manual

Hanna Instruments

Hanna Instruments HI 96771C User manual

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments HI 96706C User manual

Hanna Instruments

Hanna Instruments HI 96706C User manual

Hanna Instruments HI 96729C User manual

Hanna Instruments

Hanna Instruments HI 96729C User manual

Hanna Instruments HI97710 User manual

Hanna Instruments

Hanna Instruments HI97710 User manual

Hanna Instruments HI 83741 User manual

Hanna Instruments

Hanna Instruments HI 83741 User manual

Hanna Instruments HUALIX HI 715 User manual

Hanna Instruments

Hanna Instruments HUALIX HI 715 User manual

Hanna Instruments HI 60503 User manual

Hanna Instruments

Hanna Instruments HI 60503 User manual

Hanna Instruments HI 2214 User manual

Hanna Instruments

Hanna Instruments HI 2214 User manual

Hanna Instruments HI 95759 User manual

Hanna Instruments

Hanna Instruments HI 95759 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HI98494 User manual

Hanna Instruments

Hanna Instruments HI98494 User manual

Hanna Instruments HI 93740 User manual

Hanna Instruments

Hanna Instruments HI 93740 User manual

Hanna Instruments HI 3221 User manual

Hanna Instruments

Hanna Instruments HI 3221 User manual

Hanna Instruments HI 8010 User manual

Hanna Instruments

Hanna Instruments HI 8010 User manual

Popular Measuring Instrument manuals by other brands

Laser Quantum Ventus with smd12 operating manual

Laser Quantum

Laser Quantum Ventus with smd12 operating manual

RADWAG PUE CY10 operating manual

RADWAG

RADWAG PUE CY10 operating manual

EMIS EMIS-VIHR 200 Operation manual

EMIS

EMIS EMIS-VIHR 200 Operation manual

X-Rite iCPlate2 user guide

X-Rite

X-Rite iCPlate2 user guide

Chauvin Arnoux DECADES BOXES user manual

Chauvin Arnoux

Chauvin Arnoux DECADES BOXES user manual

Topcon GM-100 Series Quick manual

Topcon

Topcon GM-100 Series Quick manual

Kobold MAN-SD series operating instructions

Kobold

Kobold MAN-SD series operating instructions

Mitsubishi Electric EcoMonitorPro EMU2-RD3-F instruction manual

Mitsubishi Electric

Mitsubishi Electric EcoMonitorPro EMU2-RD3-F instruction manual

sparklike Handheld instruction manual

sparklike

sparklike Handheld instruction manual

DMO Wirewalker user guide

DMO

DMO Wirewalker user guide

Sealey AK9635D.V2 instructions

Sealey

Sealey AK9635D.V2 instructions

Silva Map Measurer Plus instruction manual

Silva

Silva Map Measurer Plus instruction manual

Orbis CONTAX D-6331 S0 user manual

Orbis

Orbis CONTAX D-6331 S0 user manual

ETS-Lindgren 5225 Series manual

ETS-Lindgren

ETS-Lindgren 5225 Series manual

PCE Instruments PCE-CL 10 user manual

PCE Instruments

PCE Instruments PCE-CL 10 user manual

Omega FMA 4000 user guide

Omega

Omega FMA 4000 user guide

Anritsu VNA Master MS2024B Maintenance manual

Anritsu

Anritsu VNA Master MS2024B Maintenance manual

GE moisture.IQ user manual

GE

GE moisture.IQ user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.