Hanna Instruments HI 717 User manual

Phosphate High Range
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected].
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 717 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• One bottle (5 mL) of HI 717AS Phosphate HR Reagent A
• Twenty packets of HI 717B-25 Phosphate HR Reagent B
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 717
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking “0.0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Cuvette cap.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Technical specifications:
Range 0.0 to 30.0 ppm
Resolution 0.1 ppm
Accuracy ±1.0 ppm ±5 % of reading @ 25 °C / 77 °F
Typical EMC Dev. ±0.2 ppm
Light Source Light Emitting Diode @ 525 nm
Light Detector Silicon Photocell
Method Adaptation of the
Standard Methods for the Examination
of Water and Wastewater, 18Th edition
, Heteropoly-
molybdenum Blue method. The reaction between
orthophosphate (reactive phosphorus) and the reagent
causes a blue tint in the sample.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
•Turn the meter on by pressing the button.
all segments will be displayed. When the
display shows ‘’Add’’, ‘’C.1’’ with ‘’Press’’
blinking, the meter is ready.
•Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
•Remove the cuvette from the meter and
unscrew the cap. Add 10 drops of HI 717AS
reagent and the content of one packet of
HI 717B-25 reagent. Replace the cap and
shake gently until the reagent is completely
dissolved. Place the cuvette back into the
meter.
•Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively , wait for 5 minutes and press
the button.
•The instrument displays the results in mg/L of
phosphate. The meter automatically turns
off after 10 minutes.
INTERFERENCE
Interference may be caused by:
Sulfide (S2-)
Fluoride (F-)
Arsenate (AsO4
3-)
Ferrous iron (Fe2+) above 100 mg/L
10 mL
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must bedry outside,
and completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI 731318
or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 717-25 Reagent set for 40 tests
OTHER ACCESSORIES
HI 717-11 Phosphate HR Certified Standard Kit
HI 740028 1.5V AAA batteries (4 pcs)
HI 731318 Cloth for wiping cuvettes (4 pcs)
HI 731321 Glass cuvettes (4 pcs)
HI 731225 Cuvette cap for checker HC (5 pcs)
HI 93703-50 Cuvette cleaning solution (230 mL).
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first
•Insert the battery cover and replace the screw with a screwdriver.
IST717 05/11
Measurement procedure: Accessories:Tips for an accurate measurement

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI83303 User manual

Hanna Instruments

Hanna Instruments HI83303 User manual

Hanna Instruments Chromium VI User manual

Hanna Instruments

Hanna Instruments Chromium VI User manual

Hanna Instruments HI 8734 User manual

Hanna Instruments

Hanna Instruments HI 8734 User manual

Hanna Instruments HI 993310 User manual

Hanna Instruments

Hanna Instruments HI 993310 User manual

Hanna Instruments C 221 User manual

Hanna Instruments

Hanna Instruments C 221 User manual

Hanna Instruments HI 96729C User manual

Hanna Instruments

Hanna Instruments HI 96729C User manual

Hanna Instruments Checker HI 721 User manual

Hanna Instruments

Hanna Instruments Checker HI 721 User manual

Hanna Instruments HI 9812-0 User manual

Hanna Instruments

Hanna Instruments HI 9812-0 User manual

Hanna Instruments HI 96750 User manual

Hanna Instruments

Hanna Instruments HI 96750 User manual

Hanna Instruments HI 83746 User manual

Hanna Instruments

Hanna Instruments HI 83746 User manual

Hanna Instruments HI96786 User manual

Hanna Instruments

Hanna Instruments HI96786 User manual

Hanna Instruments HI 96710C User manual

Hanna Instruments

Hanna Instruments HI 96710C User manual

Hanna Instruments HI 96714 User manual

Hanna Instruments

Hanna Instruments HI 96714 User manual

Hanna Instruments Champ HI 98106 User manual

Hanna Instruments

Hanna Instruments Champ HI 98106 User manual

Hanna Instruments HI 2211 User manual

Hanna Instruments

Hanna Instruments HI 2211 User manual

Hanna Instruments HI 758 User manual

Hanna Instruments

Hanna Instruments HI 758 User manual

Hanna Instruments HI 93723 User manual

Hanna Instruments

Hanna Instruments HI 93723 User manual

Hanna Instruments HI 983301N User manual

Hanna Instruments

Hanna Instruments HI 983301N User manual

Hanna Instruments Pool Line HI981914 User manual

Hanna Instruments

Hanna Instruments Pool Line HI981914 User manual

Hanna Instruments HI 91610 User manual

Hanna Instruments

Hanna Instruments HI 91610 User manual

Hanna Instruments HI 221 User manual

Hanna Instruments

Hanna Instruments HI 221 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments HI96000 Series User manual

Hanna Instruments

Hanna Instruments HI96000 Series User manual

Hanna Instruments HI 96721 User manual

Hanna Instruments

Hanna Instruments HI 96721 User manual

Popular Measuring Instrument manuals by other brands

BAHAG Wisent manual

BAHAG

BAHAG Wisent manual

biochrom WPA Lightwave 3 user manual

biochrom

biochrom WPA Lightwave 3 user manual

YOKOGAWA PX8000 user manual

YOKOGAWA

YOKOGAWA PX8000 user manual

biochrom Ultrospec 50 user manual

biochrom

biochrom Ultrospec 50 user manual

Omron EQUO Series user manual

Omron

Omron EQUO Series user manual

Ludlum Measurements 3001 manual

Ludlum Measurements

Ludlum Measurements 3001 manual

Bently Nevada 3300/40 Operation manual

Bently Nevada

Bently Nevada 3300/40 Operation manual

LB Altimeters ALFA QUICK START & ROAD MAP

LB Altimeters

LB Altimeters ALFA QUICK START & ROAD MAP

Sper scientific 800020 instruction manual

Sper scientific

Sper scientific 800020 instruction manual

Absolute Process Instruments API 4008 G quick start guide

Absolute Process Instruments

Absolute Process Instruments API 4008 G quick start guide

Sensitech TempTale Ultra Instructions for use

Sensitech

Sensitech TempTale Ultra Instructions for use

Kaiterra Sensedge Mini installation guide

Kaiterra

Kaiterra Sensedge Mini installation guide

Velleman CAMCOLBUL4 quick start guide

Velleman

Velleman CAMCOLBUL4 quick start guide

Stoneridge SE5000 quick guide

Stoneridge

Stoneridge SE5000 quick guide

JPT YDFLP-150-M7-L1-X user manual

JPT

JPT YDFLP-150-M7-L1-X user manual

TOOLCRAFT LDM 50 U operating instructions

TOOLCRAFT

TOOLCRAFT LDM 50 U operating instructions

TDE Instruments Digalox DPM72-AVP instruction manual

TDE Instruments

TDE Instruments Digalox DPM72-AVP instruction manual

Industrial Data Systems IDS 430 Installation/Calibration/Operation

Industrial Data Systems

Industrial Data Systems IDS 430 Installation/Calibration/Operation

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.