Hanna Instruments HI706 User manual

HI706
Phosphorus High Range
INSTRUCTION MANUAL
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI706-25 Reagents for 40 Phosphorus High Range tests
Other Accessories
HI706-11 Phosphorus HR Certified Standard Kit
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST706 05/18
HI706 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 10 minutes of non-use and 2 minutes
after reading.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
• Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
• Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the
cuvette into the meter and close the meter’s
cap.
• Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
• Remove the cuvette from the meter and
unscrew the cap. Add 10 drops of HI706AS
reagent and the content of one packet of
HI706B-0 reagent. Replace the cap and
shake gently until the reagent is completely
dissolved. Place the cuvette back into the
meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 5 minutes and
press the button.
• The instrument displays the results in
mg/L (ppm) of phosphorus (P). The meter
automatically turns off after 2 minutes.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use
the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI706 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 20 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0.0 to 15.0 ppm
Resolution 0.1 ppm
Accuracy ±0.3 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @525 nm
Light Detector Silicon Photocell
Method
Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, Heteropolymolybdenum Blue method. The
reaction between reactive phosphorus and the reagent causes a blue
tint in the sample.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use and 2 minutes after reding
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 2209 User manual

Hanna Instruments

Hanna Instruments HI 2209 User manual

Hanna Instruments HI 9813-6 User manual

Hanna Instruments

Hanna Instruments HI 9813-6 User manual

Hanna Instruments HI96719C User manual

Hanna Instruments

Hanna Instruments HI96719C User manual

Hanna Instruments HI 9026 User manual

Hanna Instruments

Hanna Instruments HI 9026 User manual

Hanna Instruments HI97708 User manual

Hanna Instruments

Hanna Instruments HI97708 User manual

Hanna Instruments HI 8314 User manual

Hanna Instruments

Hanna Instruments HI 8314 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments Pool Line HI981914 User manual

Hanna Instruments

Hanna Instruments Pool Line HI981914 User manual

Hanna Instruments HI 4101 User manual

Hanna Instruments

Hanna Instruments HI 4101 User manual

Hanna Instruments HI 8424 User manual

Hanna Instruments

Hanna Instruments HI 8424 User manual

Hanna Instruments HI 84102 User manual

Hanna Instruments

Hanna Instruments HI 84102 User manual

Hanna Instruments HI 721 User manual

Hanna Instruments

Hanna Instruments HI 721 User manual

Hanna Instruments HI 96700 User manual

Hanna Instruments

Hanna Instruments HI 96700 User manual

Hanna Instruments HI97739 User manual

Hanna Instruments

Hanna Instruments HI97739 User manual

Hanna Instruments HI 93102 User manual

Hanna Instruments

Hanna Instruments HI 93102 User manual

Hanna Instruments HI 93715 User manual

Hanna Instruments

Hanna Instruments HI 93715 User manual

Hanna Instruments HI 96770 User manual

Hanna Instruments

Hanna Instruments HI 96770 User manual

Hanna Instruments HI 2210 User manual

Hanna Instruments

Hanna Instruments HI 2210 User manual

Hanna Instruments HI 60503 User manual

Hanna Instruments

Hanna Instruments HI 60503 User manual

Hanna Instruments HI 96770C User manual

Hanna Instruments

Hanna Instruments HI 96770C User manual

Hanna Instruments HI98162 User manual

Hanna Instruments

Hanna Instruments HI98162 User manual

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments HI96718C User manual

Hanna Instruments

Hanna Instruments HI96718C User manual

Hanna Instruments HI96719C User manual

Hanna Instruments

Hanna Instruments HI96719C User manual

Popular Measuring Instrument manuals by other brands

Waters ionKey/MS Guide

Waters

Waters ionKey/MS Guide

Powerex MH-C9000 instructions

Powerex

Powerex MH-C9000 instructions

Master Meter eLinx quick start guide

Master Meter

Master Meter eLinx quick start guide

R&S RT-ZC20 user manual

R&S

R&S RT-ZC20 user manual

BEKA A504G-SS-PM manual

BEKA

BEKA A504G-SS-PM manual

Emerson Daniel PT Series Installation and maintenance instructions

Emerson

Emerson Daniel PT Series Installation and maintenance instructions

Denver Instrument IR-200 Operation manual

Denver Instrument

Denver Instrument IR-200 Operation manual

Omega APH2 instruction sheet

Omega

Omega APH2 instruction sheet

Kaelus iVA0727A operating manual

Kaelus

Kaelus iVA0727A operating manual

ATAGO PAL-26S instruction manual

ATAGO

ATAGO PAL-26S instruction manual

ICP DAS USA iWSN-110X-RCT Series quick start

ICP DAS USA

ICP DAS USA iWSN-110X-RCT Series quick start

Rohde & Schwarz AMN6500 user manual

Rohde & Schwarz

Rohde & Schwarz AMN6500 user manual

Agilent Technologies E4980A Operation manual

Agilent Technologies

Agilent Technologies E4980A Operation manual

Dwyer Instruments FLMG Series instructions

Dwyer Instruments

Dwyer Instruments FLMG Series instructions

Spectris PARTICLE MEASURING SYSTEMS AirSentry II Operation manual

Spectris

Spectris PARTICLE MEASURING SYSTEMS AirSentry II Operation manual

National Instruments NI 5911 user manual

National Instruments

National Instruments NI 5911 user manual

Apator Powogaz JS operating instructions

Apator Powogaz

Apator Powogaz JS operating instructions

FLOWMAX 221 quick start guide

FLOWMAX

FLOWMAX 221 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.