Measurement
Procedure
Turn the meter on by pressing the button.
All tags will be displayed. The display will
show “INSERT REF” with “PRESS” blinking
to indicate the meter is ready to calibrate.
Open the cover and insert the
sealed glycerol reference cuvette
into the meter. Close the cover.
Press the button to calibrate the meter.
The maple leaf icon will blink to show
that calibration is taking place.
The display will then show “INSERT
SAMPLE” with “PRESS” blinking when the
meter has successfully been calibrated.
Fill a clean, dry cuvette with 10 mL of
maple syrup sample and replace the cap.
Open the cover and insert the cuvette
into the meter. Close the cover.
Press the button to take the
measurement. The maple leaf
icon will blink to show that a
measurement is taking place.
The meter directly displays the grade of
maple syrup. To turn o, press and hold
the button for 3 seconds. The meter
automaticallyturns o after 10 minutes.
Recommendations for Users
Before using Hanna products, make sure that they are entirely suitable for your
specic application and for the environment in which they are used. Operation
of these instruments may cause unacceptable interferences to other electronic
equipment, thus requiring the operator to take all necessary steps to correct such
interferences. Any variation introduced by the user to the supplied equipment
may degrade the instrument’s EMC performance. To avoid damages or burns,
do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in hazardous environments.
Code Description
HI759-11 glycerol standard cuvettes (2 pcs)
HI93703-57 glycerol bottles, 30 mL (4 pcs)
HI731318 cloth for wiping cuvettes (4 pcs)
HI731321 glass cuvettes (4 pcs)
HI731225 black cuvette caps for Checker®HC (4 pcs)
HI740028 1.5V AAA batteries (4 pcs)
HI93703-50 cuvette cleaning solution (230 mL)
Accessories
HI759 INSTMAN 1.3 09/15
Hanna Instruments reserves the right to
modify the design, construction, or appearance
of its products without advance notice.
Warranty
HI759 is warranted for a period of one year after date of purchase against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. This warranty is limited
to repair or replacement free of charge. Damage due to accidents, misuse,
tampering or lack of prescribed maintenance is not covered. If service is
required, contact your local Hanna oce. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If
the repair is not covered by the warranty, you will be notied of the charges
incurred. If the instrument is to be returned to Hanna Instruments, rst
obtain a Return Goods Authorization number from the Technical Service
department and then send it with shipping costs prepaid. When shipping
any instrument, make sure it is properly packed for complete protection.
Battery Management
To conserve battery, the instrument shuts down after
10 minutes of non-use. To replace the battery, turn the
instrument o by holding the button until the meter shuts
o. Turn the instrument upside down and remove the
battery cover with a screwdriver. Remove the battery from
its location and replace it with a new one, inserting the
negative end rst. Replace the battery cover and screw.
Tips for an Accurate
Measurement
When taking measurements, the outside of the cuvette
must be dry and free of syrup, ngerprints, oil, or dirt.
Wipe it thoroughly with a lint-free cloth prior to insertion.
Make sure cuvettes and syringes are completely
dry before using. Excess water will leave streaks
in the syrup sample which may aect results.
Dispense or pour the sample slowly. When transferring
the syrup into the cuvette, be careful not to generate
bubbles as they may cause inaccurate readings.
Grade A Color
Classes Taste Light Transmittance
Grade A Golden Delicate ≥75.0
Grade A Amber Rich 50 to 74.9
Grade A Dark Robust 25 to 49.9
Grade A Very Dark Strong < 25
USDA Standards