  9. Hanna Instruments HI 775 User manual

Hanna Instruments HI 775 User manual

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 96821 User manual

Hanna Instruments

Hanna Instruments HI 96821 User manual

Hanna Instruments HI 746 User manual

Hanna Instruments

Hanna Instruments HI 746 User manual

Hanna Instruments IRIS HI801 User manual

Hanna Instruments

Hanna Instruments IRIS HI801 User manual

Hanna Instruments HI 9813-5 User manual

Hanna Instruments

Hanna Instruments HI 9813-5 User manual

Hanna Instruments HI 84429 User manual

Hanna Instruments

Hanna Instruments HI 84429 User manual

Hanna Instruments HI 84102 User manual

Hanna Instruments

Hanna Instruments HI 84102 User manual

Hanna Instruments HI97730 User manual

Hanna Instruments

Hanna Instruments HI97730 User manual

Hanna Instruments HI 705 User manual

Hanna Instruments

Hanna Instruments HI 705 User manual

Hanna Instruments HI 755 User manual

Hanna Instruments

Hanna Instruments HI 755 User manual

Hanna Instruments HI 2300 User manual

Hanna Instruments

Hanna Instruments HI 2300 User manual

Hanna Instruments HI 93703 User manual

Hanna Instruments

Hanna Instruments HI 93703 User manual

Hanna Instruments HI 771 User manual

Hanna Instruments

Hanna Instruments HI 771 User manual

Hanna Instruments HI 96734C User manual

Hanna Instruments

Hanna Instruments HI 96734C User manual

Hanna Instruments HI 9811-5 User manual

Hanna Instruments

Hanna Instruments HI 9811-5 User manual

Hanna Instruments HI96708C User manual

Hanna Instruments

Hanna Instruments HI96708C User manual

Hanna Instruments HI 255 User manual

Hanna Instruments

Hanna Instruments HI 255 User manual

Hanna Instruments HI 9810 User manual

Hanna Instruments

Hanna Instruments HI 9810 User manual

Hanna Instruments HI 4111 User manual

Hanna Instruments

Hanna Instruments HI 4111 User manual

Hanna Instruments HI 38018 User manual

Hanna Instruments

Hanna Instruments HI 38018 User manual

Hanna Instruments HI 709 User manual

Hanna Instruments

Hanna Instruments HI 709 User manual

Hanna Instruments HI 9126 User manual

Hanna Instruments

Hanna Instruments HI 9126 User manual

Hanna Instruments HI 120 User manual

Hanna Instruments

Hanna Instruments HI 120 User manual

Hanna Instruments HI 93751 User manual

Hanna Instruments

Hanna Instruments HI 93751 User manual

Hanna Instruments HI96752 User manual

Hanna Instruments

Hanna Instruments HI96752 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

IRIMO 984-6D-1 owner's manual

IRIMO

IRIMO 984-6D-1 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Alkalinity
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 775 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 1 bottle of HI 775 Alkalinity Liquid Reagent
• 1 mL syringe with tip
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 775
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking “0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
For more details about spare parts and accessories
see “Accessories”.
Technical specifications:
Range 0 to 500 ppm
Resolution 1 ppm
Accuracy ±5% of reading ±5 ppm @ 25°C
Typical EMC Dev. ±1 ppm
Light Source Light Emitting Diode @ 610 nm
Light Detector Silicon Photocell
Method Colorimetric method. The reaction causes a distinctive
range of colors from yellow to blue to develop. This
meter has been developed to work with fresh water
samples.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
www. .com information@itm.com1.800.561.8187
•Turn the meter on by pressing the button,
all segments will be displayed. When the
display shows ‘’Add’’, ‘’C.1’’ with ‘’Press’’
blinking, the meter is ready.
•Fill the cuvette to the 10 mL line on the
cuvette with unreacted sample and replace
the cap. Place the cuvette into the meter
and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
Note:Any chlorine present in the sample will
interfere with the reading. To remove the
chlorine interference add one drop of
HI 93755-53 Chlorine Remover to the
unreacted sample.
•Remove the cuvette, open it and using a
1 mL syringe carefully add exactly 1.00 mL
of Alkalinity Reagent to the sample. Replace
the cap and gently invert 5 times. Place the
cuvette back into the meter.
Note: Pay attention not to spill reagent otherwise
full color development may be inhibited.
•Press the button. The instrument directly
displays the concentration of alkalinity in
ppm of CaCO3. Alkalinity conversion:
1 ppm CaCO3= 0.02 meq/L
= 0.056 dkH
The meter automatically turns off after 10
minutes.
10 mL
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must bedry outside,
and completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI 731318
or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 775-26 Liquid Reagents for 25 Alkalinity tests, 1 syringe and 1 tip
OTHER ACCESSORIES
HI 775-11 Alkalinity Certified Standard Kit
HI 740142 1 mL graduated syringe with tip
HI 740028 1.5V AAA batteries (4 pcs)
HI 731318 Cloth for wiping cuvettes (4 pcs)
HI 731321 Glass cuvettes (4 pcs)
HI 731225 Cuvette cap for checker HC (4 pcs)
HI 93703-50 Cuvette cleaning solution (230 mL)
HI 93755-53 Chlorine Remover
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first
•Insert the battery cover and replace the screw with a screwdriver.
IST775 04 /12
Measurement procedure: Accessories:Tips for an accurate measurement