Hanna Instruments Checker HI729 User manual

HI729
Fluoride Low Range
INSTRUCTION MANUAL
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI729-26 Reagents for 24 Fluoride Low Range tests
(4 x HI729S and a 1 mL syringe with tip)
Other Accessories
HI729-11 Fluoride LR Certified Standard Kit
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI740142P 1 mL graduated syringe (10 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST729 05/18
HI729 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 10 minutes of non-use and 2 minutes
after reading.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
• Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
• Using supplied syringe add 2 mL of
HI729S SPADNS reagent to both cuvettes.
• Fill the first cuvette to the line with 8 mL
of deionized water and the second cuvette
to the line with 8 mL of sample. Replace
the cap and invert several times to mix.
• Place the first cuvette (deionized water)
into the meter and close the cap.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 2 minutes and
press the button.
• When the display shows “Add”, “C.2”
with “Press” blinking the meter is zeroed.
• Remove the cuvette.
• Insert the second cuvette (sample) into the
instrument and press the button.
• The instrument directly displays the
concentration of fluoride in ppm. The meter
automatically turns off after 2 minutes.
8 mL 8 mL
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs
more light than the zero reference. Check the procedure and make sure
that the sample and zero cuvette are not inverted.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must
be replaced. Once this indication is displayed, normal operation of
the instrument will be interrupted. Change the battery and restart
the meter.
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI729 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagent for 6 tests
• 1 mL syringe with tip
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0.00 to 2.00 ppm
Resolution 0.01 ppm
Accuracy ±0.10 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @575 nm
Light Detector Silicon Photocell
Method Adaptation of the Standard Methods for the Examination of Water
and Wastewater, 18th edition, SPADNS method.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use and 2 minutes after reading
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI84530 User manual

Hanna Instruments

Hanna Instruments HI84530 User manual

Hanna Instruments HALO2 HI9810382 User manual

Hanna Instruments

Hanna Instruments HALO2 HI9810382 User manual

Hanna Instruments HI 2550 User manual

Hanna Instruments

Hanna Instruments HI 2550 User manual

Hanna Instruments HI97761 User manual

Hanna Instruments

Hanna Instruments HI97761 User manual

Hanna Instruments HI96746C User manual

Hanna Instruments

Hanna Instruments HI96746C User manual

Hanna Instruments HI 38020 User manual

Hanna Instruments

Hanna Instruments HI 38020 User manual

Hanna Instruments HI96704C User manual

Hanna Instruments

Hanna Instruments HI96704C User manual

Hanna Instruments HI97738 User manual

Hanna Instruments

Hanna Instruments HI97738 User manual

Hanna Instruments HI97762 User manual

Hanna Instruments

Hanna Instruments HI97762 User manual

Hanna Instruments HI 93712 User manual

Hanna Instruments

Hanna Instruments HI 93712 User manual

Hanna Instruments OHI 98140 User manual

Hanna Instruments

Hanna Instruments OHI 98140 User manual

Hanna Instruments HI 99163 User manual

Hanna Instruments

Hanna Instruments HI 99163 User manual

Hanna Instruments HI 96813 User manual

Hanna Instruments

Hanna Instruments HI 96813 User manual

Hanna Instruments HI 96770C User manual

Hanna Instruments

Hanna Instruments HI 96770C User manual

Hanna Instruments HI96739C User manual

Hanna Instruments

Hanna Instruments HI96739C User manual

Hanna Instruments HI 931100 User manual

Hanna Instruments

Hanna Instruments HI 931100 User manual

Hanna Instruments HI 93708 User manual

Hanna Instruments

Hanna Instruments HI 93708 User manual

Hanna Instruments HI 96769C User manual

Hanna Instruments

Hanna Instruments HI 96769C User manual

Hanna Instruments HI 93749 User manual

Hanna Instruments

Hanna Instruments HI 93749 User manual

Hanna Instruments IRIS HI801 User manual

Hanna Instruments

Hanna Instruments IRIS HI801 User manual

Hanna Instruments HI 22091 User manual

Hanna Instruments

Hanna Instruments HI 22091 User manual

Hanna Instruments HI 8732 User manual

Hanna Instruments

Hanna Instruments HI 8732 User manual

Hanna Instruments Chromium VI User manual

Hanna Instruments

Hanna Instruments Chromium VI User manual

Hanna Instruments HI 93701 User manual

Hanna Instruments

Hanna Instruments HI 93701 User manual

Popular Measuring Instrument manuals by other brands

QUARTA RADEX ONE user manual

QUARTA

QUARTA RADEX ONE user manual

Katronic KATflow 170 Operating and safety instructions

Katronic

Katronic KATflow 170 Operating and safety instructions

Lumel N30P user manual

Lumel

Lumel N30P user manual

Easy-Laser E980 manual

Easy-Laser

Easy-Laser E980 manual

Elcontrol STAR 3 user manual

Elcontrol

Elcontrol STAR 3 user manual

Agilent Technologies NFA Series Setup guide

Agilent Technologies

Agilent Technologies NFA Series Setup guide

Endress+Hauser nivotester FTC 420 Quick installation guide

Endress+Hauser

Endress+Hauser nivotester FTC 420 Quick installation guide

Flybox Omnia57-80 Series manual

Flybox

Flybox Omnia57-80 Series manual

bmcm AMS42-LAN16fx manual

bmcm

bmcm AMS42-LAN16fx manual

Schäfer RMS 442 quick start guide

Schäfer

Schäfer RMS 442 quick start guide

PCB Piezotronics 352A21/NC Installation and operating manual

PCB Piezotronics

PCB Piezotronics 352A21/NC Installation and operating manual

Schaller humimeter BMA user manual

Schaller

Schaller humimeter BMA user manual

THEIS TKL-7 operating instructions

THEIS

THEIS TKL-7 operating instructions

HYDROMETER IZAR CENTER user manual

HYDROMETER

HYDROMETER IZAR CENTER user manual

Omega HHF143 user guide

Omega

Omega HHF143 user guide

amc VIBRO AVM 1002D user manual

amc VIBRO

amc VIBRO AVM 1002D user manual

Mollet LF20 Series Instruction

Mollet

Mollet LF20 Series Instruction

DayTronic 3000PLUS instruction manual

DayTronic

DayTronic 3000PLUS instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.