Hanna Instruments Checker HI729 User manual

HI729
Fluoride Low Range
INSTRUCTION MANUAL
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI729-26 Reagents for 24 Fluoride Low Range tests
(4 x HI729S and a 1 mL syringe with tip)
Other Accessories
HI729-11 Fluoride LR Certified Standard Kit
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI740142P 1 mL graduated syringe (10 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST729 05/18
HI729 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 10 minutes of non-use and 2 minutes
after reading.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
• Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
• Using supplied syringe add 2 mL of
HI729S SPADNS reagent to both cuvettes.
• Fill the first cuvette to the line with 8 mL
of deionized water and the second cuvette
to the line with 8 mL of sample. Replace
the cap and invert several times to mix.
• Place the first cuvette (deionized water)
into the meter and close the cap.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 2 minutes and
press the button.
• When the display shows “Add”, “C.2”
with “Press” blinking the meter is zeroed.
• Remove the cuvette.
• Insert the second cuvette (sample) into the
instrument and press the button.
• The instrument directly displays the
concentration of fluoride in ppm. The meter
automatically turns off after 2 minutes.
8 mL 8 mL
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs
more light than the zero reference. Check the procedure and make sure
that the sample and zero cuvette are not inverted.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must
be replaced. Once this indication is displayed, normal operation of
the instrument will be interrupted. Change the battery and restart
the meter.
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction
manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI729 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagent for 6 tests
• 1 mL syringe with tip
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0.00 to 2.00 ppm
Resolution 0.01 ppm
Accuracy ±0.10 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @575 nm
Light Detector Silicon Photocell
Method Adaptation of the Standard Methods for the Examination of Water
and Wastewater, 18th edition, SPADNS method.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use and 2 minutes after reading
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 91610 User manual

Hanna Instruments

Hanna Instruments HI 91610 User manual

Hanna Instruments HI96802 User manual

Hanna Instruments

Hanna Instruments HI96802 User manual

Hanna Instruments Combo Grochek HI 991404 User manual

Hanna Instruments

Hanna Instruments Combo Grochek HI 991404 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments HI 216 User manual

Hanna Instruments

Hanna Instruments HI 216 User manual

Hanna Instruments HI 83540 User manual

Hanna Instruments

Hanna Instruments HI 83540 User manual

Hanna Instruments HI709 User manual

Hanna Instruments

Hanna Instruments HI709 User manual

Hanna Instruments HI 726 User manual

Hanna Instruments

Hanna Instruments HI 726 User manual

Hanna Instruments HI 8731 User manual

Hanna Instruments

Hanna Instruments HI 8731 User manual

Hanna Instruments HI 120 User manual

Hanna Instruments

Hanna Instruments HI 120 User manual

Hanna Instruments Color of Water HI 727 User manual

Hanna Instruments

Hanna Instruments Color of Water HI 727 User manual

Hanna Instruments HI83303 User manual

Hanna Instruments

Hanna Instruments HI83303 User manual

Hanna Instruments HI96752 User manual

Hanna Instruments

Hanna Instruments HI96752 User manual

Hanna Instruments HI97747 User manual

Hanna Instruments

Hanna Instruments HI97747 User manual

Hanna Instruments HI 9829 User manual

Hanna Instruments

Hanna Instruments HI 9829 User manual

Hanna Instruments HI99165 User manual

Hanna Instruments

Hanna Instruments HI99165 User manual

Hanna Instruments HI 93714 User manual

Hanna Instruments

Hanna Instruments HI 93714 User manual

Hanna Instruments HI 8014 User manual

Hanna Instruments

Hanna Instruments HI 8014 User manual

Hanna Instruments HI 96721C User manual

Hanna Instruments

Hanna Instruments HI 96721C User manual

Hanna Instruments HI98164 User manual

Hanna Instruments

Hanna Instruments HI98164 User manual

Hanna Instruments HI 983302 User manual

Hanna Instruments

Hanna Instruments HI 983302 User manual

Hanna Instruments HI 931101 User manual

Hanna Instruments

Hanna Instruments HI 931101 User manual

Hanna Instruments HI 96811 User manual

Hanna Instruments

Hanna Instruments HI 96811 User manual

Hanna Instruments pH 209 User manual

Hanna Instruments

Hanna Instruments pH 209 User manual

Popular Measuring Instrument manuals by other brands

AEMC MN353 user manual

AEMC

AEMC MN353 user manual

Eickemeyer LIFEVET O2 user manual

Eickemeyer

Eickemeyer LIFEVET O2 user manual

Giropes GI620 COMPACT user manual

Giropes

Giropes GI620 COMPACT user manual

Megger DCM340 user guide

Megger

Megger DCM340 user guide

TINRS Edgecutter manual

TINRS

TINRS Edgecutter manual

Bender PEM353 quick start guide

Bender

Bender PEM353 quick start guide

Knick SensoGate WA 131 H user manual

Knick

Knick SensoGate WA 131 H user manual

3M The Edge quick start guide

3M

3M The Edge quick start guide

Campbell LI200R instruction manual

Campbell

Campbell LI200R instruction manual

GE SimpliNano Original instructions

GE

GE SimpliNano Original instructions

Balmac 205ce Operation manual

Balmac

Balmac 205ce Operation manual

Shanghai Grandway Telecom Tech. Co., Ltd. FHP12 user guide

Shanghai Grandway Telecom Tech. Co., Ltd.

Shanghai Grandway Telecom Tech. Co., Ltd. FHP12 user guide

Nikon Nivo Series instruction manual

Nikon

Nikon Nivo Series instruction manual

Elma Instruments RT606 A user manual

Elma Instruments

Elma Instruments RT606 A user manual

Minolta SPOTMETER F - IR REMOTE CONTRO LRC-3 manual

Minolta

Minolta SPOTMETER F - IR REMOTE CONTRO LRC-3 manual

Megger DLRO-10 user guide

Megger

Megger DLRO-10 user guide

iSTEK K100-TB instruction manual

iSTEK

iSTEK K100-TB instruction manual

AITE ATC1006A Operation manual

AITE

AITE ATC1006A Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.