INSTRUCTION MANUAL
HI774
Phosphate Ultra Low Range
Reagent Sets
HI774-25 Reagents for 25 Phosphate Ultra Low Range tests
Other Accessories
HI774-11 Phosphate Ultra Low Range Certified Standard Kit
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731315 Glass Cuvettes and Caps for Checker®HC Colorimeters (2 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
Accessories
IST774 10/18
HI774 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Before using this product, make sure it is entirely suitable for your specific application and
for the environment in which it is used.
Any variation introduced by the user to the supplied equipment may degrade the meter’s
performance.
For yours and the meter’s safety do not use or store the meter in hazardous environments.
Recommendations for Users
Warranty
To save the battery, the instrument shuts down after 7 minutes of non-use and 2 minutes
after reading.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
Measurement Procedure
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI774 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 10 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction manual
carefully before using the instrument.
For more information about Hanna Instruments and our products, visit www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected]
Preliminary Examination
Specifications
Range 0.00 to 0.90 ppm
Resolution 0.01 ppm
Accuracy ±0.02 ppm ±5% of reading @25 °C/ 77 °F
Light Source Light Emitting Diode @525 nm
Light Detector Silicon Photocell
Method
Adaptation of Standard Methods for the Examination of Water and
Wastewater, 20th edition, Ascorbic Acid method. The reaction be-
tween phosphate and the reagent causes a blue tint in the sample.
Environment 0 to 50 °C (32 to 122 °F); RH max 95% non-condensing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 7 minutes of non-use and 2 minutes after reading
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use
the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
Functional Description
Errors and Warnings
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
• Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “Add”, “C.1” with “Press” blinking,
the meter is ready.
• Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
• Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
• Remove the cuvette from the meter and unscrew
the cap. Add the content of one packet of
HI774-0 reagent. Replace the cap and shake
gently for 2 minutes until the powder is
completely dissolved. Place the cuvette back
into the meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 3 minutes and press the
button.
• The instrument directly displays the
concentration of phosphate in ppm. The
meter automatically turns off after 2 minutes.
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.

Other manuals for HI774

2

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HI97720 User manual

Hanna Instruments

Hanna Instruments HI97720 User manual

Hanna Instruments HI97711 User manual

Hanna Instruments

Hanna Instruments HI97711 User manual

Hanna Instruments HI 93703 User manual

Hanna Instruments

Hanna Instruments HI 93703 User manual

Hanna Instruments HI 717 User manual

Hanna Instruments

Hanna Instruments HI 717 User manual

Hanna Instruments HI96700C User manual

Hanna Instruments

Hanna Instruments HI96700C User manual

Hanna Instruments HI 2221 User manual

Hanna Instruments

Hanna Instruments HI 2221 User manual

Hanna Instruments HI 98127 User manual

Hanna Instruments

Hanna Instruments HI 98127 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments HI 9813-5 User manual

Hanna Instruments

Hanna Instruments HI 9813-5 User manual

Hanna Instruments HI 96107 User manual

Hanna Instruments

Hanna Instruments HI 96107 User manual

Hanna Instruments Pronto HI 982401 User manual

Hanna Instruments

Hanna Instruments Pronto HI 982401 User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments HI 93707 User manual

Hanna Instruments

Hanna Instruments HI 93707 User manual

Hanna Instruments HI 98180 User manual

Hanna Instruments

Hanna Instruments HI 98180 User manual

Hanna Instruments HI 96706C User manual

Hanna Instruments

Hanna Instruments HI 96706C User manual

Hanna Instruments HI 4321 User manual

Hanna Instruments

Hanna Instruments HI 4321 User manual

Hanna Instruments HI 4105 User manual

Hanna Instruments

Hanna Instruments HI 4105 User manual

Hanna Instruments HI 8014 User manual

Hanna Instruments

Hanna Instruments HI 8014 User manual

Hanna Instruments HI6221 User manual

Hanna Instruments

Hanna Instruments HI6221 User manual

Hanna Instruments HI97713 User manual

Hanna Instruments

Hanna Instruments HI97713 User manual

Hanna Instruments HI97736 User manual

Hanna Instruments

Hanna Instruments HI97736 User manual

Hanna Instruments HI 96701C User manual

Hanna Instruments

Hanna Instruments HI 96701C User manual

Hanna Instruments edge DO User manual

Hanna Instruments

Hanna Instruments edge DO User manual

Popular Measuring Instrument manuals by other brands

Micronics Ultraflo 2000 operating manual

Micronics

Micronics Ultraflo 2000 operating manual

Pulsar FlowPulse Handheld instruction manual

Pulsar

Pulsar FlowPulse Handheld instruction manual

Fortinet FortiAnalyzer 1000B quick start guide

Fortinet

Fortinet FortiAnalyzer 1000B quick start guide

HP 431B CALIBRATION PROCEDURE

HP

HP 431B CALIBRATION PROCEDURE

Agilent Technologies M9210A Startup guide

Agilent Technologies

Agilent Technologies M9210A Startup guide

Farmcomp Wile 27 operating instructions

Farmcomp

Farmcomp Wile 27 operating instructions

Textron GREENLEE CMI-2000 instruction manual

Textron

Textron GREENLEE CMI-2000 instruction manual

Hengstler Tico 731.4 operating instructions

Hengstler

Hengstler Tico 731.4 operating instructions

Dräger DrugTest 5000 Technical manual

Dräger

Dräger DrugTest 5000 Technical manual

Perel EVM125 quick start guide

Perel

Perel EVM125 quick start guide

KROHNE BW25 Handbook

KROHNE

KROHNE BW25 Handbook

Gossen MetraWatt SECULIFE ITBASE operating instructions

Gossen MetraWatt

Gossen MetraWatt SECULIFE ITBASE operating instructions

Ludlum Measurements 334A Operation manual

Ludlum Measurements

Ludlum Measurements 334A Operation manual

Bosch GLM 100 C Professional Original instructions

Bosch

Bosch GLM 100 C Professional Original instructions

Vacom BARION atm II manual

Vacom

Vacom BARION atm II manual

Lincoln SSVDL user manual

Lincoln

Lincoln SSVDL user manual

Envertech EVB300 User's operating manual

Envertech

Envertech EVB300 User's operating manual

PCB Piezotronics IMI SENSORS LB608A11/-0002 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS LB608A11/-0002 Installation and operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.