HI739
Fluoride High Range
INSTRUCTION MANUAL
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Operation of
these instruments may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct such interferences. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance. To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in hazardous environments.
Recommendations for Users
Accessories
Reagent Sets
HI739-26 Reagents for 30 Fluoride High Range tests
Other Accessories
HI739-11 Fluoride HR Certified Standard Kit
HI731318 Cloth for wiping cuvettes (4 pcs.)
HI731321 Glass cuvettes and Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI731225 Cuvette Black Cap for Checker®HC Colorimeters (4 pcs.)
HI731353 Cuvette Seal Cap for Checker®HC Colorimeters (4 pcs.)
HI740028P 1.5V AAA batteries (12 pcs.)
HI740142P 1 mL graduated syringe (10 pcs.)
HI740144P Pipette tip for syringes (10 pcs.)
HI93703-50 Cuvette cleaning solution (230 mL)
IST739 05/18
HI739 is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
Warranty
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Tips for an Accurate Measurement
To save the battery, the instrument shuts down after 10 minutes of non-use and 2 minutes.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Battery Management
• Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “Add”, “C.1” with “Press” blinking, the
meter is ready.
• Using supplied syringe add 2 mL (1 mL twice)
of HI739AS reagent to the cuvette and use the
pipette to fill the cuvette up to the mark with
HI739BS reagent.
• Replace the cap and invert several times to mix.
• Place the cuvette into the meter and close the
meter’s cap.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 1 minute and then press
the button. When the display shows “Add”,
“C.2” with “Press” blinking the meter is zeroed.
• Remove the cuvette from the meter and unscrew
the cap. Using the unused syringe, add exactly
1 mL of sample.
• Replace the cap and invert several times to mix.
• Place the cuvette back into the meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 1 minute and then press
the button.
• The instrument directly displays the
concentration of fluoride in ppm. The meter
automatically turns off after 2 minutes.
Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.0” indicates that the sample absorbs
more light than the zero reference. Check the procedure and make
sure you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
Measurement Procedure
Hanna Instruments reserves the right to modify the design, construction or appearance of
its products without advance notice.
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
Functional Description
Errors and Warnings
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction manual
carefully before using the instrument.
For more information about Hanna Instruments and our products, visit www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected].
Find your local Hanna Instruments Office on www.hannainst.com.
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each HI739 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• One bottle Fluoride HR Reagent A
• One bottle Fluoride HR Reagent B
• 1 mL syringe wtih tip (2 pcs.)
• 1 plastic pipette
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Preliminary Examination
Range 0 to 20.0 ppm
Resolution 0.1 ppm
Accuracy ±0.5 ppm ±5% of reading @25 °C/77 °F
Light Source Light Emitting Diode @575 nm
Light Detector Silicon Photocell
Method Adaptation of the Standard Methods for the Examination of Water
and Wastewater,18th edition, SPADNS method.
Environment 0 to 50 °C (32 to 122 °F); max 95% RH non-condesing
Battery Type 1.5V AAA (1 pc.)
Auto-Shut off After 10 minutes of non-use and 2 minutes after reading
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 52 g (1.84 oz.)
Specifications

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 95734C User manual

Hanna Instruments

Hanna Instruments HI 95734C User manual

Hanna Instruments HI 221 User manual

Hanna Instruments

Hanna Instruments HI 221 User manual

Hanna Instruments HI747 User manual

Hanna Instruments

Hanna Instruments HI747 User manual

Hanna Instruments HI 83742 User manual

Hanna Instruments

Hanna Instruments HI 83742 User manual

Hanna Instruments HI 991002 User manual

Hanna Instruments

Hanna Instruments HI 991002 User manual

Hanna Instruments HI 723 User manual

Hanna Instruments

Hanna Instruments HI 723 User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments HI98198 OPDO User manual

Hanna Instruments

Hanna Instruments HI98198 OPDO User manual

Hanna Instruments Pronto HI 982401 User manual

Hanna Instruments

Hanna Instruments Pronto HI 982401 User manual

Hanna Instruments HI 99141 User manual

Hanna Instruments

Hanna Instruments HI 99141 User manual

Hanna Instruments HI 93754G-25 MR User manual

Hanna Instruments

Hanna Instruments HI 93754G-25 MR User manual

Hanna Instruments HI 993310 User manual

Hanna Instruments

Hanna Instruments HI 993310 User manual

Hanna Instruments HI 95759 User manual

Hanna Instruments

Hanna Instruments HI 95759 User manual

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HALO 2 HI9810392 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810392 User manual

Hanna Instruments HI 8424 User manual

Hanna Instruments

Hanna Instruments HI 8424 User manual

Hanna Instruments HI97710 User manual

Hanna Instruments

Hanna Instruments HI97710 User manual

Hanna Instruments HI781 User manual

Hanna Instruments

Hanna Instruments HI781 User manual

Hanna Instruments HI98165 User manual

Hanna Instruments

Hanna Instruments HI98165 User manual

Hanna Instruments HI 253 User manual

Hanna Instruments

Hanna Instruments HI 253 User manual

Hanna Instruments HI97747 User manual

Hanna Instruments

Hanna Instruments HI97747 User manual

Hanna Instruments HI 747 User manual

Hanna Instruments

Hanna Instruments HI 747 User manual

Hanna Instruments C22190 User manual

Hanna Instruments

Hanna Instruments C22190 User manual

Hanna Instruments HI 96745 User manual

Hanna Instruments

Hanna Instruments HI 96745 User manual

Popular Measuring Instrument manuals by other brands

Schweitzer Engineering Laboratories SEL-734 quick start guide

Schweitzer Engineering Laboratories

Schweitzer Engineering Laboratories SEL-734 quick start guide

Endress+Hauser Prothermo NMT81 operating instructions

Endress+Hauser

Endress+Hauser Prothermo NMT81 operating instructions

IMT Analytics FlowAnalyser Series user manual

IMT Analytics

IMT Analytics FlowAnalyser Series user manual

Larsen & Brusgaard ALTITRACK manual

Larsen & Brusgaard

Larsen & Brusgaard ALTITRACK manual

Novanta SYNRAD Pulstar p100 user manual

Novanta

Novanta SYNRAD Pulstar p100 user manual

LIMIT 1180-R operating manual

LIMIT

LIMIT 1180-R operating manual

Mahr MarCal 30 EWN operating instructions

Mahr

Mahr MarCal 30 EWN operating instructions

HOZO Design MEAZOR User instructions

HOZO Design

HOZO Design MEAZOR User instructions

Endress+Hauser StamoLys CA 70 operating instructions

Endress+Hauser

Endress+Hauser StamoLys CA 70 operating instructions

ITL MON-2697-COM Reference manual

ITL

ITL MON-2697-COM Reference manual

Ecolab EcoTrans II pH operating instructions

Ecolab

Ecolab EcoTrans II pH operating instructions

AEMC instruments MN185 user manual

AEMC instruments

AEMC instruments MN185 user manual

Amprobe VPC-10A user manual

Amprobe

Amprobe VPC-10A user manual

TFA 30.5033 instruction manual

TFA

TFA 30.5033 instruction manual

Testboy TV333 operating manual

Testboy

Testboy TV333 operating manual

ACL 395 Operation manual

ACL

ACL 395 Operation manual

AccuEnergy AcuRev 1300 user manual

AccuEnergy

AccuEnergy AcuRev 1300 user manual

Continental Refrigerator DTCO 1381 Process Guide

Continental Refrigerator

Continental Refrigerator DTCO 1381 Process Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.