INSTRUCTION MANUAL
HI98304
EC Tester
Warranty
These meters are warranted for a period of one year against defects in
workmanship and materials when used for their intended purpose and
maintained according to instructions. Probes are warranted for a period
of six months. This warranty is limited to repair or replacement free of
charge. Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered. If service is required, contact your local Hanna
Instruments Office. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization (RGA) number from the Technical Service department
and then send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
Solutions
Code Description
HI70030P 12.88 mS/cm calibration solution,
20 mL sachets (25 pcs.)
HI7030M 12.88 mS/cm calibration solution,
230 mL bottle
HI7030L 12.88 mS/cm calibration solution,
500 mL bottle
Accessories
IST98304N 06/18
To obtain the highest accuracy for measurement it is important to follow
these tips:
• Calibration is only as good as the solution being used. The calibration
solution values change over time. Fresh solution should be used for each
calibration.
• The probe should be rinsed with purified water each time before placing
in calibration solution or sample to be tested.
Care and Maintenance
Error messages
During user calibration, if the reading is out of the accepted range, the meter
will display “---- Err“.i.e. 20.00 mS/cm blinking means the limit of the
measured value is out of range.
If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the
50.0 ºC or 0.0 ºC temperature value will blink on the LCD in measure mode.
1. Pour 2” (5 cm) of standard solution into a container.
2. Place the probe in calibration solution. The probe tip should be centered
in the solution and submersed 1.18” (3 cm). The meter will automatically
recognize the solution.
3. If the solution is not recognized or is out of the accepted range “---- Err“
is displayed.
4. If the calibration solution is recognized “REC“ is displayed until the
reading is stable and the calibration is accepted.
•After acceptance, the “Stor“ message is displayed and meter returns to
measurement mode.
Calibration Procedure
Measurement and Calibration
For better accuracy, frequent calibration of the meter is recommended. In
addition, the instrument must be recalibrated:
•When high accuracy is required.
•At least once a month.
Measurement
Press the ON/OFF button to turn the meter ON. Place the probe to the
desired solution. The conductivity value, automatically compensated for
temperature, will be shown on the LCD.
The measured temperature will be shown on the secondary LCD.
Note: Before taking any measurement make sure the meter has been
calibrated.
Clear Calibration
Place meter in calibration mode. Press ON/OFF and “Clr” is displayed. The
meter will now be at default calibration.
DiST 4
Battery Replacement
To change the CR2032 Li-ion battery, turn the battery cover located on the
back of the meter counterclockwise to unlock. Remove cover and replace with
new battery + side facing up.
Note: Batteries should only be replaced in a safe area using the
battery type specified in this instruction manual. Old batteries should
be disposed in accordance with local regulations.
The meter features a low battery indicator. When the battery is running low
(under 10%), the battery indicator will blink on the LCD. When the battery
is discharged “dEAd bAtt” will be displayed on the LCD for 2 seconds and
the meter will turn off.
0
0
4
4
Recommendations for Users
Before using Hanna Instruments products, make sure that they are entirely
suitable for your specific application and for the environment in which they
are used. Any variation introduced by the user to the supplied equipment
may degrade the meter' s performance. Avoid touching the electrode area.
To avoid damages or burns, do not put the meter in microwave oven. For
yours and the meter's safety do not use or store the meter in hazardous
environments.
•Press CAL button to enter calibration mode from measurement mode.
•The meter will enter the calibration mode, displaying “mS/cm 12.88
USE” with CAL tag blinking.
Liquid Crystal
Display
CAL button
Operation
Thank You
Thank you for choosing a Hanna Instruments product. Please read
this instruction manual carefully before using the instrument.
For more information about Hanna Instruments and our products,
visit www.hannainst.com or e-mail us at [email protected].
For technical support, contact your local Hanna Instruments
Office or e-mail us at tec[email protected]
Find your local Hanna Instruments Office on www.hannainst.com
Preliminary Examination
Remove the meter from the packing material and examine it carefully to
make sure that no damage has occurred during shipment. If noticeable
damage is evident, contact your local Hanna Instruments Office.
Each meter is supplied with:
•CR2032 battery (1 pc.)
•Storage / Protection sleeve
•12.88 mS/cm calibration solution sachet (4 pcs.)
•Instruction manual
•Quality Certificate
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original
packaging with the supplied accessories.
Preparation
The probe is shipped dry. Before using the electrode, remove the protective
cap. Then follow the calibration procedure.
•Turn the tester on by pressing ON/OFF button.
•Remove the protective cap and immerse the tip of the probe in the
sample to be tested.
•Stir gently and wait for a stability tag to disappear.
•The electrode automatically compensates for temperature variations.
•The reading on display is directly expressed in: mS/cm.
•For best results, recalibrate periodically.
•After use rinse the probe with water.
•Always replace the protective cap after each use.
DiST 4
CAL
mS/cm
Specifications
Calibration
Solutions
HI70030
(12.88 mS/cm)
Calibration automatic, single point (12.88 mS/cm)
Temperature Compensation automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF)
Battery type CR2032 Li-Ion 3V (1 pc.)
Battery life approximately 250 hours of continuous use
Environment 0 to 50 ºC (32 to 122 ºF);
RH 100% max
Weight
(without battery) 68 g (2.4 oz.)
ON/OFF button
Temperature
Sensor (inside)
Probe
Range 0.00 to 20.00 mS/cm
0.0 to 50.0 ˚C (32.0 to 122.0 ˚F)
Resolution 0.01 mS/cm
0.1 ˚C (0.1 ˚F)
Accuracy
(@25 °C/77 °F)
±2% F.S.
±0.5 °C (±1.0 °F)
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design,
construction, or appearance of its products without advance notice.
Operational Guide
To turn the meter ON and check the battery status
Press the ON/OFF button to turn the meter on. At start-up, all the LCD
segments are displayed for 1 second, then the percent indication of the
remaining battery life is displayed for another second. The meter then enters
the normal measuring mode.
Note: Keeping the ON/OFF button pressed while turning the meter on
will display all LCD segments as long as the button is pressed.
To enter calibration mode
Press the CAL button. “CAL“ message is displayed.
To enter setup mode
Remove the battery cover and press the Setup button located on the side
of the battery.
Meter Setup
While in measurement mode, remove the battery cover. Press
the Setup button located on the side of the battery in the battery
compartment. The meter will enter in setup mode. Press the
ON/OFF button to move through setup parameters. Change option by
pressing CAL button. The default settings are: “Set t” measure unit - °C,
“AOFF” - 8 min.
To select the temperature unit (°C/°F)
To select the temperature unit when “SET t” is displayed press the CAL
button to change between °C or °F.
To select the Auto-Off time
To select the AUTO OFF when “AOFF” is displayed press the CAL button to
change between 8 min, 60 min or --- (disabled).
To return to measurement mode
Press the ON/OFF button.
Setup
Button
Recommended
Immersion
Level

This manual suits for next models

1

Other Hanna Instruments Test Equipment manuals

Hanna Instruments HI 7101 User manual

Hanna Instruments

Hanna Instruments HI 7101 User manual

Hanna Instruments HI 38082 User manual

Hanna Instruments

Hanna Instruments HI 38082 User manual

Hanna Instruments 3815 User manual

Hanna Instruments

Hanna Instruments 3815 User manual

Hanna Instruments HI 95711C User manual

Hanna Instruments

Hanna Instruments HI 95711C User manual

Hanna Instruments HI7640-58 User manual

Hanna Instruments

Hanna Instruments HI7640-58 User manual

Hanna Instruments HI 98129 User manual

Hanna Instruments

Hanna Instruments HI 98129 User manual

Hanna Instruments HI 3812 User manual

Hanna Instruments

Hanna Instruments HI 3812 User manual

Hanna Instruments DiST 3 User manual

Hanna Instruments

Hanna Instruments DiST 3 User manual

Hanna Instruments HI 902 Color User manual

Hanna Instruments

Hanna Instruments HI 902 Color User manual

Hanna Instruments HI 38033 User manual

Hanna Instruments

Hanna Instruments HI 38033 User manual

Hanna Instruments Checker HI772 User manual

Hanna Instruments

Hanna Instruments Checker HI772 User manual

Hanna Instruments HI 38039 User manual

Hanna Instruments

Hanna Instruments HI 38039 User manual

Hanna Instruments HI 739 User manual

Hanna Instruments

Hanna Instruments HI 739 User manual

Hanna Instruments HI 98103 User manual

Hanna Instruments

Hanna Instruments HI 98103 User manual

Hanna Instruments HI9810342 User manual

Hanna Instruments

Hanna Instruments HI9810342 User manual

Hanna Instruments PICCOLO 2 HI 98112 User manual

Hanna Instruments

Hanna Instruments PICCOLO 2 HI 98112 User manual

Hanna Instruments Pool Line HI981204 User manual

Hanna Instruments

Hanna Instruments Pool Line HI981204 User manual

Hanna Instruments HI 98103 User manual

Hanna Instruments

Hanna Instruments HI 98103 User manual

Hanna Instruments PH GROCHEK User manual

Hanna Instruments

Hanna Instruments PH GROCHEK User manual

Hanna Instruments HI 38077 User manual

Hanna Instruments

Hanna Instruments HI 38077 User manual

Hanna Instruments HI 839800 User manual

Hanna Instruments

Hanna Instruments HI 839800 User manual

Hanna Instruments HI 98311 User manual

Hanna Instruments

Hanna Instruments HI 98311 User manual

Hanna Instruments Gro Line HALO 2 User manual

Hanna Instruments

Hanna Instruments Gro Line HALO 2 User manual

Hanna Instruments HI9810372 User manual

Hanna Instruments

Hanna Instruments HI9810372 User manual

Popular Test Equipment manuals by other brands

SPX Power Team HT50A operating instructions

SPX

SPX Power Team HT50A operating instructions

Tru-Test 702 user manual

Tru-Test

Tru-Test 702 user manual

VeEX LX100 E-manual

VeEX

VeEX LX100 E-manual

Würth 0715 53 090 user guide

Würth

Würth 0715 53 090 user guide

Bosch EPS 807 Repair instructions

Bosch

Bosch EPS 807 Repair instructions

Buckleys BathyCorrometer Calibration Kit instruction manual

Buckleys

Buckleys BathyCorrometer Calibration Kit instruction manual

Tektronix 2445A Service manual

Tektronix

Tektronix 2445A Service manual

YOKOGAWA CA150 user manual

YOKOGAWA

YOKOGAWA CA150 user manual

FUZRR ES1010 user manual

FUZRR

FUZRR ES1010 user manual

XtendLan DPA-MRMDS Setting manual

XtendLan

XtendLan DPA-MRMDS Setting manual

VECTECH 190 instruction manual

VECTECH

VECTECH 190 instruction manual

BOLDAN Miniroller D200 e manual

BOLDAN

BOLDAN Miniroller D200 e manual

Swagelok M200 User instruction

Swagelok

Swagelok M200 User instruction

Akku Vision TEST TOWER quick start guide

Akku Vision

Akku Vision TEST TOWER quick start guide

Fluke DTX-PCU6S user manual

Fluke

Fluke DTX-PCU6S user manual

Martel MC1000 manual

Martel

Martel MC1000 manual

Fluke Calibration 5560A Programmer's manual

Fluke Calibration

Fluke Calibration 5560A Programmer's manual

Keysight Technologies U3022AM06 User's and service guide

Keysight Technologies

Keysight Technologies U3022AM06 User's and service guide