manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hanna Instruments
  6. •
  7. Test Equipment
  8. •
  9. Hanna Instruments HI 739 User manual

Hanna Instruments HI 739 User manual

Other Hanna Instruments Test Equipment manuals

Hanna Instruments HI 38079 User manual

Hanna Instruments

Hanna Instruments HI 38079 User manual

Hanna Instruments Foodcare HI981036 User manual

Hanna Instruments

Hanna Instruments Foodcare HI981036 User manual

Hanna Instruments DiST 3 User manual

Hanna Instruments

Hanna Instruments DiST 3 User manual

Hanna Instruments HI 95707C User manual

Hanna Instruments

Hanna Instruments HI 95707C User manual

Hanna Instruments HI 93739 User manual

Hanna Instruments

Hanna Instruments HI 93739 User manual

Hanna Instruments HALO2 HI9810452 User manual

Hanna Instruments

Hanna Instruments HALO2 HI9810452 User manual

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments Gro Line HALO 2 User manual

Hanna Instruments

Hanna Instruments Gro Line HALO 2 User manual

Hanna Instruments HI98330 User manual

Hanna Instruments

Hanna Instruments HI98330 User manual

Hanna Instruments DiST 3 User manual

Hanna Instruments

Hanna Instruments DiST 3 User manual

Hanna Instruments Primo Series User manual

Hanna Instruments

Hanna Instruments Primo Series User manual

Hanna Instruments Pool Line HI9310014 User manual

Hanna Instruments

Hanna Instruments Pool Line HI9310014 User manual

Hanna Instruments HALO2 User manual

Hanna Instruments

Hanna Instruments HALO2 User manual

Hanna Instruments HALO 2 HI9810442 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810442 User manual

Hanna Instruments HI 38034 User manual

Hanna Instruments

Hanna Instruments HI 38034 User manual

Hanna Instruments HI 38076 User manual

Hanna Instruments

Hanna Instruments HI 38076 User manual

Hanna Instruments HI 902 Color User manual

Hanna Instruments

Hanna Instruments HI 902 Color User manual

Hanna Instruments Groline HI98118 User manual

Hanna Instruments

Hanna Instruments Groline HI98118 User manual

Hanna Instruments HI 38082 User manual

Hanna Instruments

Hanna Instruments HI 38082 User manual

Hanna Instruments HI 3830 User manual

Hanna Instruments

Hanna Instruments HI 3830 User manual

Hanna Instruments HI 98129 User manual

Hanna Instruments

Hanna Instruments HI 98129 User manual

Hanna Instruments FC2053 User manual

Hanna Instruments

Hanna Instruments FC2053 User manual

Hanna Instruments HI981037 User manual

Hanna Instruments

Hanna Instruments HI981037 User manual

Hanna Instruments HALO2 HI9810352 User manual

Hanna Instruments

Hanna Instruments HALO2 HI9810352 User manual

Popular Test Equipment manuals by other brands

Ambit P.F. Series instruction manual

Ambit

Ambit P.F. Series instruction manual

Megger BMM500 Series user guide

Megger

Megger BMM500 Series user guide

RJG Lynx Sensor Tester operating instructions

RJG

RJG Lynx Sensor Tester operating instructions

Mark-10 TT02 Series user guide

Mark-10

Mark-10 TT02 Series user guide

Martel 3001 Operator's manual

Martel

Martel 3001 Operator's manual

ElproSys DiagProg4 instruction manual

ElproSys

ElproSys DiagProg4 instruction manual

Redtech TRAILERteck T05 user manual

Redtech

Redtech TRAILERteck T05 user manual

Venmar AVS Constructo 1.0 HRV user guide

Venmar

Venmar AVS Constructo 1.0 HRV user guide

Test Instrument Solutions SafetyPAT operating manual

Test Instrument Solutions

Test Instrument Solutions SafetyPAT operating manual

Kistler 5495C Series instruction manual

Kistler

Kistler 5495C Series instruction manual

Waygate Technologies DM5E Basic quick start guide

Waygate Technologies

Waygate Technologies DM5E Basic quick start guide

StoneL DeviceNet CK464002A manual

StoneL

StoneL DeviceNet CK464002A manual

Seica RAPID 220 Site preparation guide

Seica

Seica RAPID 220 Site preparation guide

Kingfisher KI7400 Series Training manual

Kingfisher

Kingfisher KI7400 Series Training manual

Kurth Electronic CCTS-03 operating manual

Kurth Electronic

Kurth Electronic CCTS-03 operating manual

SMART KANAAD SBT XTREME 3G Series user manual

SMART

SMART KANAAD SBT XTREME 3G Series user manual

Agilent Technologies BERT Serial Getting started

Agilent Technologies

Agilent Technologies BERT Serial Getting started

Agilent Technologies N3280A user guide

Agilent Technologies

Agilent Technologies N3280A user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Fluoride High Range
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected].
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 739 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 1 bottle (30 mL) of HI 739AS Fluoride HR Reagent A
• 1 bottle (30 mL) of HI 739BS Fluoride HR Reagent B
• 2 x 1 mL syringe with tip
• 1 dropper
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 739
Technical specifications:
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking “0.00” indicates that the sample absorbs
more light than the zero reference. Check the procedure and make
sure you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Range 0.0 to 20.0 ppm
Resolution 0.1 ppm
Accuracy ±0.5 ppm ±5 % of reading @ 25 °C / 77 °F
Typical EMC Dev. ±0.1 ppm
Light Source Light Emitting Diode @ 575 nm
Light Detector Silicon Photocell
Method Adaptation of the
Standard Methods for the Examination
of Water and Wastewater, 18th edition
, SPADNS method.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
[email protected]
www.ponpe.com
•Turn the meter on by pressing the button,
all segments will be displayed. When the
display shows “Add”, “C.1” with “Press”
blinking, the meter is ready.
•Using supplied syringe add 2 mL (1 mL
twice) of HI 739AS reagent to the cuvette
and use the pipette to fill the cuvette up to
the mark with HI 739BS reagent.
•Replace the cap and invert several times to
mix.
•Place the cuvette into the meter and close
the meter’s cap.
•Press and hold the button until the timer is
displayed on the LCD. Alternatively wait for
one minute and press the button. When the
display shows “Add”, “C.2” with “Press”
blinking the meter is zeroed.
•Remove the cuvette from the meter and
unscrew the cap. Using the unused syringe,
add exactly 1.00 mL of sample.
•Replace the cap and invert several times to
mix.
•Place the cuvette back into the meter.
•Press and hold the button until the timer is
displayed on the LCD. Alternatively wait for
one minute and press the button.
•The instrument directly displays the
concentration of fluoride in ppm.
The meter automatically turns off after 10
minutes.
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must bedry outside,
and completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI 731318
or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 739-26 Liquid Reagent Set for 20 tests, 2 syringes and 2 tips
OTHER ACCESSORIES
HI 739-11 Fluoride HR Certified Standard Kit
HI 740142 1 mL graduated syringe
HI 740028 1.5V AAA batteries (4 pcs)
HI 731318 Cloth for wiping cuvettes (4 pcs)
HI 731321 Glass cuvettes (4 pcs)
HI 731225 Cuvette cap for checker HC (4 pcs)
HI 93703-50 Cuvette cleaning solution (230 mL).
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first.
•Insert the battery cover and replace the screw with a screwdriver.
IST739 05 /11
Measurement procedure: Accessories:Tips for an accurate measurement
10 mL
2x
www.ponpe.com
[email protected]
www.ponpe.com